{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 43
    }
  },
  "results": [
    {
      "cfres_id": "130325",
      "product_res_number": "Z-0086-2015",
      "event_date_initiated": "2014-09-18",
      "event_date_posted": "2014-10-17",
      "recall_status": "Terminated",
      "event_date_terminated": "2015-02-24",
      "res_event_number": "69371",
      "product_code": "MUE",
      "k_numbers": [
        "K122286"
      ],
      "product_description": "Siemens Mammomat Inspiration mammography systems.\r\n\r\n The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.",
      "code_info": "material #10140000, with serial numbers: 5018 3849 5042",
      "firm_fei_number": "2240869",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "city": "Malvern",
      "state": "PA",
      "postal_code": "19355",
      "additional_info_contact": "Customer Support\n610-219-6300",
      "reason_for_recall": "The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i",
      "root_cause_description": "Process control",
      "action": "A safety advisory notice, dated September 18, 2014, was sent to end users that identified the product, problem, and actions to be taken.",
      "product_quantity": "3",
      "distribution_pattern": "US Distribution in states of: NY, VA, and MA.",
      "openfda": {
        "k_number": [
          "K193166",
          "K130704",
          "K190809",
          "K163252",
          "K254090",
          "K210151",
          "K162268",
          "K202902",
          "K163302",
          "K133972",
          "K111361",
          "K193334",
          "K110729",
          "K110798",
          "K123520",
          "K151645",
          "K211215",
          "K121674",
          "K260087",
          "K131885",
          "K211720",
          "K202822",
          "K170930",
          "K120246",
          "K173576",
          "K123995",
          "K110572",
          "K220073",
          "K103485",
          "K110717",
          "K200424",
          "K173132",
          "K120255",
          "K212873",
          "K211725",
          "K243420",
          "K241113",
          "K233539",
          "K121963",
          "K110025",
          "K172027",
          "K243849",
          "K173408",
          "K254092",
          "K123873",
          "K162670",
          "K192317",
          "K111324",
          "K182951",
          "K111434",
          "K163328",
          "K253678",
          "K122510",
          "K122286",
          "K172404",
          "K111646"
        ],
        "pma_number": [
          "P030010",
          "P990066",
          "P010017",
          "P010025",
          "P080018",
          "P050014"
        ],
        "registration_number": [
          "3008833258",
          "3003775006",
          "3023003446",
          "3004485675",
          "1317307",
          "3009196578",
          "3005406110",
          "3017585294",
          "3004977335",
          "3008787461",
          "3009868110",
          "9680027",
          "3005877899",
          "3017375271",
          "3035950510",
          "3001722928",
          "3003768251",
          "3012127758",
          "3009717300",
          "3000126629",
          "3008355361",
          "3025442726",
          "3015276088",
          "3021059265",
          "3001556265",
          "3005496266",
          "3003769120",
          "9611343",
          "3005172103",
          "3009048384",
          "3009963139",
          "3010398867",
          "9613152",
          "3027339477",
          "8043728",
          "1423253",
          "3010703925",
          "3005843418",
          "3031503890",
          "3008496839",
          "3006013411",
          "3005810333",
          "2433615"
        ],
        "fei_number": [
          "3008833258",
          "3003775006",
          "3023003446",
          "3004485675",
          "1317307",
          "3009196578",
          "3005406110",
          "3017585294",
          "3004977335",
          "3008787461",
          "3009868110",
          "1000284313",
          "3005877899",
          "3017375271",
          "3035950510",
          "3001722928",
          "3003768251",
          "3012127758",
          "3009717300",
          "3000126629",
          "3008355361",
          "3025442726",
          "3015276088",
          "3021059265",
          "1000514697",
          "3002807903",
          "3002807317",
          "3001556265",
          "3003772616",
          "3005496266",
          "3003769120",
          "3005172103",
          "3009048384",
          "3009963139",
          "3010398867",
          "3027339477",
          "3010703925",
          "3005843418",
          "3031503890",
          "3008496839",
          "3006013411",
          "3005810333",
          "2433615"
        ],
        "device_name": "Full Field Digital, System, X-Ray, Mammographic",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1715",
        "device_class": "2"
      }
    }
  ]
}