  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "213679",
      "product_res_number": "Z-2220-2025",
      "event_date_initiated": "2025-03-13",
      "event_date_posted": "2025-07-30",
      "recall_status": "Open, Classified",
      "res_event_number": "96793",
      "product_code": "MSB",
      "k_numbers": [
        "K991594"
      ],
      "product_description": "Brand Name: Celsior Cold Storage Solution\nProduct Name: Celsior Cold Storage Solution\nModel/Catalog Number: 533682\nProduct Description: In US Celsior is a class II medical device. Celsior is a clear to slightly yellow, sterile, non pyrogenic solution for\nhypothermic flushing and storage of hearts. Celsior can be used for hypothermic cardiac flushing and storage in\npreparation for transportation and eventual transplantation of the heart into the recipient. Celsior is packaged in\none liter bag with 4 units per carton box. Celsior shall be stored between 36 to 46F.\nIn Europe Celsior is a drug and can be used for thoracic organs (heart & lung) and abdominal organs (liver, kidney, pancreas).",
      "code_info": "Lot Code: UDI carton box: 23700911900123 & UDI bag: 13700911900126 (for US reference).\nThe product reference recalled in France is different and there is no UDI on this one (not required for medicine\nin France). Batches concerned by the recall are the following:\nLot SL240261-1\nLot SL240266-1\nLot SL240272-1\nLot SL240280-2\nThese batches have been sold only in France",
      "firm_fei_number": "3024820181",
      "recalling_firm": "INSTITUT GEORGES LOPEZ",
      "address_1": "6 ROUTE DE SAINT BONNET",
      "city": "ST LAURENT DE CHAMOUSSET",
      "country": "France",
      "reason_for_recall": "The potential for risk of microbiological contamination of products due to inability to ensure sterility assurance throughout aseptic manufacturing.",
      "root_cause_description": "Under Investigation by firm",
      "action": "On March 13, 2025 notification emails were sent to OUS customers informing them of the recall and instructing them to check their inventory for affected product and to destroy it. If they did not wish to destroy, to arrange for the product to be returned to the recalling firm. Customers were instructed to acknowledge receipt of the email and provide proof of destruction, if performed.",
      "product_quantity": "1,532 units",
      "distribution_pattern": "International distribution to the country of France.",
      "openfda": {
        "k_number": [
          "K123326",
          "K180194",
          "K143074",
          "K133432",
          "K032794",
          "K261913",
          "K222535",
          "K182735",
          "K190063",
          "K022730",
          "K991594",
          "K211314",
          "K000881",
          "K081997"
        ],
        "registration_number": [
          "3027669473",
          "3010765841",
          "3017223202",
          "3007685782",
          "3010034260",
          "3011175555",
          "1220477",
          "3007583504",
          "3003708554",
          "3024820181",
          "9615922",
          "1047843",
          "3031884833",
          "2123774",
          "3003152463",
          "1420032"
        ],
        "fei_number": [
          "3027669473",
          "3010765841",
          "3017223202",
          "3007685782",
          "3010034260",
          "3001640191",
          "3011175555",
          "1220477",
          "3007583504",
          "3003708554",
          "3024820181",
          "1047843",
          "3031884833",
          "2123774",
          "3003152463",
          "1420032"
        ],
        "device_name": "System And Accessories, Isolated Heart, Transport And Preservation",
        "medical_specialty_description": "Gastroenterology, Urology",
        "regulation_number": "876.5880",
        "device_class": "2"
      }
    }
  ]
}