  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "28970",
      "product_res_number": "Z-0745-04",
      "event_date_initiated": "2002-06-01",
      "event_date_posted": "2004-07-20",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-05-03",
      "res_event_number": "26998",
      "product_code": "MOT",
      "k_numbers": [
        "K974210"
      ],
      "product_description": "RAD SOURCE RS 3000 Blood Irradiator Three Bag Option",
      "code_info": "Serial numbers: 1002 through 1005, 1007 through 1009, 1011 through 1023.",
      "firm_fei_number": "3002459090",
      "recalling_firm": "Rad-Source Technologies, Inc.",
      "address_1": "20283 State Rd. 7",
      "address_2": "Suite 107",
      "city": "Boca Raton",
      "state": "FL",
      "postal_code": "33498",
      "additional_info_contact": "Randy Kirk\n888-396-3339",
      "reason_for_recall": "The firm completed a cooling system retrofit to preclude overheating and failure.",
      "root_cause_description": "Other",
      "action": "The firm e-mailed a notification on May 9, 2002 to its consignees giving them instructions on using additional radiation indictors to assure that the blood units were correctly irradiated.  They also initiated a cooling tube retrofit of all units on that date and this was completed on 8/13/2002.",
      "product_quantity": "20",
      "distribution_pattern": "Units were distributed to 20 blood banks and transfusion services in WI, KS, IN, MI, GA, NC, WA, NE, FL, IN, TX, MD, Washington DC, SC, and NM.",
      "openfda": {
        "k_number": [
          "K161324",
          "K974210",
          "K172087",
          "K201177",
          "K032684",
          "K865027",
          "K837346",
          "K952291",
          "K051065",
          "K181737",
          "K082921",
          "K963497",
          "K915767",
          "K915766",
          "K050963",
          "K851828"
        ],
        "registration_number": [
          "3006946288",
          "3002459090",
          "3016367050",
          "8043725"
        ],
        "fei_number": [
          "3009631751",
          "3006946288",
          "3002459090",
          "3016367050"
        ],
        "device_name": "Irradiator, Blood To Prevent Graft Versus Host Disease",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "U"
      }
    }
  ]
}