  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 12
    }
  },
  "results": [
    {
      "cfres_id": "94358",
      "product_res_number": "Z-0068-2011",
      "event_date_initiated": "2010-09-07",
      "event_date_posted": "2010-10-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-05-13",
      "res_event_number": "56719",
      "product_code": "MLM",
      "k_numbers": [
        "K070820"
      ],
      "product_description": "Architect Tacrolimus Reagent Kit; an in-vitro diagnostic chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of tacrolimus in human whole blood; Abbott Laboratories, Abbott Diagnostics Division, Abbott Park, IL 60064 USA;   list 1L77-25\n\nThe Architect Tacrolimus assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of tacrolimus in human whole blood as an aid in the management of liver and kidney allograft patients receiving tacrolimus therapy.",
      "code_info": "list 1L77-25, lot number 86599M500, expiration date September 01, 2011",
      "firm_fei_number": "1415939",
      "recalling_firm": "Abbott Laboratories",
      "address_1": "100 Abbott Park Rd",
      "city": "Abbott Park",
      "state": "IL",
      "postal_code": "60064-3502",
      "additional_info_contact": "Customer Service\n877-422-2688",
      "reason_for_recall": "There are reagent barcode read errors when using reagent lot 86599M500 due to the print quality of the barcode.",
      "root_cause_description": "Other",
      "action": "Abbott Laboratories sent a \"Product Recall Immediate Action Required\" letter dated September 7, 2010, to all customers. The letter described product, problem and action to be taken by customers.  The customers were instructed to determine if they are currently using and/or have inventory of lot 86599M500; discontinue use of and destroy any remaining inventory; inform and provide copies of letter to other laboratories if appropriate; retain the letter for laboratory records and complete and return the Customer Reply Immediate Action Required form via fax at 1-800-777-0051 or email QAGCO@abbott.com.  \n\nIMPORTANT NOTE: If you do not currently have replacement material than you may continue to use this reagent lot until your laboratory receives replacement material. Upon receipt of the replacement material, discontinue use of and discard lot 86599M500 and switch to the replacement material. Contact Abbott Customer service for replacement material (1-877-4ABBOTT, Option 1).\n\nIf you or any of the health care providers you serve have any questions regarding this information, U.S. customers should call Customers Service at 1-877-4ABBOTT. Customers outside of the U.S., please contact your local Customer Service representative.",
      "product_quantity": "1,995 kits",
      "distribution_pattern": "Worldwide distribution: USA and countries including Puerto Rico, Argentina, Australia, Austria, Belarus, Belgium, Bosnia, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Egypt, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Lithuania, Montenegro, Netherlands, New Zealand, Norway, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom and Vietnam.",
      "openfda": {
        "k_number": [
          "K150168",
          "K173857",
          "K123343",
          "K060385",
          "K203833",
          "K060502",
          "K063868",
          "K070820",
          "K050206"
        ],
        "pma_number": [
          "P970007",
          "P970025"
        ],
        "registration_number": [
          "3003951180",
          "2521625",
          "1415939",
          "3008344661",
          "9616673",
          "9610126",
          "3005333358",
          "1218959",
          "3026848661",
          "2517506",
          "2432235",
          "1643621",
          "3010939897"
        ],
        "fei_number": [
          "3003179344",
          "3003951180",
          "2521625",
          "3002806559",
          "1415939",
          "3008344661",
          "3005333358",
          "1218959",
          "3026848661",
          "2517506",
          "2432235",
          "1643621",
          "3010939897"
        ],
        "device_name": "Enzyme Immunoassay, Tracrolimus",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1678",
        "device_class": "2"
      }
    }
  ]
}