  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "153355",
      "product_res_number": "Z-1394-2017",
      "event_date_initiated": "2015-08-27",
      "event_date_posted": "2017-02-27",
      "recall_status": "Terminated",
      "event_date_terminated": "2017-03-06",
      "res_event_number": "76492",
      "product_code": "MLA",
      "k_numbers": [
        "K931374"
      ],
      "product_description": "BD  AffirM VPIII Microbial Identification Test",
      "code_info": "Catalog Number\tBatch Number  446252\t                     5126980                       \t 5135673 \t                     5138774 \t                     5140625 \t                     5142767 \t                     5153674 \t                     5153675 \t                     5224882  446257\t                     5141655 \t                     5148857 \t                     5156873 \t                     5173648",
      "firm_fei_number": "1119779",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir BD Diagnostic",
      "city": "Sparks",
      "state": "MD",
      "postal_code": "21152-9212",
      "additional_info_contact": "BD\n410-316-4000",
      "reason_for_recall": "BD has confirmed that a portion of tests associated with the affected lots( 446252 and 446257 ) of product may exhibit blue speckles or fibers on the beads and cards, as shown below. This issue would result in an invalid test, as described in the package insert.",
      "root_cause_description": "Employee error",
      "action": "Becton Dickinson sent an Urgent Product Recall Letter dated August 2015.  The distributors were provided with a written communication via email and customers were provided with a written communication via UPS. Customers were not asked to send product back.\n\nIf further assistance is needed regarding replacements, please contact BD Customer Service Department at 1-800-675-0908. For all other inquiries, please contact BD Technical Services Department  at 1-800-638-8663.\n\nFor further questions please call (410) 316-4428.",
      "product_quantity": "Qty Distributed to Field - 4,502",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and  WY\nand to the countries of : AU, BE, IN, KR, SG,",
      "openfda": {
        "registration_number": [
          "1119779",
          "2647876",
          "3043226252",
          "3012494290"
        ],
        "fei_number": [
          "1119779",
          "2647876",
          "3043226252",
          "3012494290"
        ],
        "device_name": "Dna Probe, Yeast",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.2660",
        "device_class": "1",
        "k_number": [
          "K931151",
          "K931374"
        ]
      }
    }
  ]
}