  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 6
    }
  },
  "results": [
    {
      "cfres_id": "214014",
      "product_res_number": "Z-2023-2025",
      "event_date_initiated": "2025-05-20",
      "event_date_posted": "2025-06-25",
      "recall_status": "Open, Classified",
      "res_event_number": "96910",
      "product_code": "MJA",
      "k_numbers": [
        "K021582"
      ],
      "product_description": "BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA).",
      "code_info": "Catalog No. 245128; UDI-DI: (01) 0038290245128; Lot No. 4150998, 4177908, 4178500, 4262031, 4262036, 4262037, 4262039, 4262040, 4262044, 4284438, 4284441, 4284443, 4284445, 4284449, 4304789, 4304796, 4347228, 4347230, 4352110, 4362418, 5007456, 5020153, 5050670, 5050671, 5050672, 5050675, 5050678.",
      "firm_fei_number": "1119779",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "city": "Sparks",
      "state": "MD",
      "postal_code": "21152-9212",
      "reason_for_recall": "BD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis isolates.",
      "root_cause_description": "Under Investigation by firm",
      "action": "***UPDATE 11/13/2025***\nBD Issued an URGENT: Medical Device Correction letters were sent to customers. BD is pleased to announce that we have resumed production of a modified version of the BD BACTEC MGIT 960 PZA Kit which includes modified inoculation methods and reduced shelf-life. BD has conducted a thorough review of the performance from all supported inoculum sources using the synthetic raw material and the following change has been implemented for the BD BACTEC MGIT 960 PZA Kit. 1. Only inoculum prepared from MGIT tubes 3-5 days post-positivity can be currently supported. 2. Inoculum prepared from MGIT tubes 1-2 days past instrument positivity are currently not supported. 3. Inoculum prepared from solid media are currently not supported.\nAdditionally, the shelf life of the product has been adjusted from 18 months to 13 months to reflect the latest internal supporting data. Clinical users should refer to the updated Instructions for Use (IFU) available at https://www.bd.qarad.eifu.online/hcp. Additional languages will be added as translations are completed.\n\nAn URGENT: Medical Device Correction notice dated 5/20/25 was sent to consignees via mail. The notification instructs consignees to inspect their inventory for recalled units and destroy them per their institution's process of destruction. The recall notification is to be shared at consignee locations with those who utilize the device and to customers who received distributed product. Consignees are to return a completed Customer Response Form to BD. Consignees with any questions are to email BDRC12@bd.com. Consignees are to report any complaints by phone to 1-844-823-5433 or by email at productcomplaints@bd.com.",
      "product_quantity": "12,256 eaches",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AR, AU, BD, BE, BN, BR, CA, CN, CR, DO, GT, HK, HT, ID, IN, JP, KH, KR, MO, MX, MY, NP, NZ, PE, PH, PK, SG, SV, TH, TW, VN.",
      "openfda": {
        "k_number": [
          "K003062",
          "K021582",
          "K972772",
          "K014123",
          "K052323"
        ],
        "device_name": "Susceptibility Test Powders, Antimycobacterial",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.1640",
        "device_class": "2",
        "registration_number": [
          "1119779",
          "1111096"
        ],
        "fei_number": [
          "1119779",
          "1111096"
        ]
      }
    }
  ]
}