  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 4
    }
  },
  "results": [
    {
      "cfres_id": "144222",
      "product_res_number": "Z-1329-2016",
      "event_date_initiated": "2015-12-22",
      "event_date_posted": "2016-04-05",
      "recall_status": "Terminated",
      "event_date_terminated": "2016-10-03",
      "res_event_number": "73456",
      "product_code": "MCH",
      "product_description": "CA500 Epix Universal Clip Applier\n\nProduct Usage: \nThe Epix disposable laparoscopic clip applier is indicated for ligation of tubular structures or vessels in laparoscopic and general surgical procedures.",
      "code_info": "Lot #: 1231562",
      "firm_fei_number": "2027111",
      "recalling_firm": "Applied Medical Resources Corp",
      "address_1": "22872 Avenida Empresa",
      "city": "Rancho Santa Margarita",
      "state": "CA",
      "postal_code": "92688-2650",
      "additional_info_contact": "949-713-8000",
      "reason_for_recall": "Applied Medical Resources Corporation is recalling CA500 Epix Universal Clip Applier, kit models STD GK258 AME Standard Kit 44, STD KIT GK213, AME Standard Kit 15, and STD KIT GK260, Standard Set 360 Z, which contain the affected CA500 lot, because of a potential loading mechanism nonconformance.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "Customer notification letters were sent on December 22, 2015 for the U.S. and on January 6, 2016 for those in Europe.\n\nThe customer notification letter requested that the following actions be taken:\n- Check inventory for recalled product.\n- Complete the attached Recall Notification Confirmation Form to acknowledge the\nrecall, indicate if facility is returning or has already used this lot, and provide a no-charge PO number if replacements requested.\n- Return the Recall Notification Confirmation Form to Applied Medical by emailing to\nrecall60684384@appliedmedical.com or faxing to (949) 713-8936.\n- Return affected product and a copy of the Recall Notification Confirmation Form to Applied Medical \n\nFor product return questions contact Karen Mitchell, Sales Operations Supervisor, RGA Dept. at (949) 713-8622 or by email at kmitchell@appliedmedical.com.\n\nFor regulatory questions contact Lauren Contursi, at (949) 713-8767 or by email at lcontursi@appliedmedical.com.",
      "product_quantity": "1771 units",
      "distribution_pattern": "U.S .Nationwide Distribution in the states of GA, AL, LA, TX, MN, AK, CA, ID, OR, UT, WY, NM, and AZ.\n\nForeign distirbution to the following; Austria, Belgium, Denmarl, Canada, Japan, France, Germany, Ireland, Italy, Greece, Netherlands, Russia, Spain, Switzerland, Turkey, Great Britain",
      "openfda": {
        "k_number": [
          "K896998",
          "K040148",
          "K972745"
        ],
        "registration_number": [
          "2431166",
          "8010273",
          "3004215117",
          "3015895045",
          "3014279513",
          "3008770252"
        ],
        "fei_number": [
          "2523865",
          "3004215117",
          "4486",
          "3015895045",
          "3014279513",
          "3008770252"
        ],
        "device_name": "Clip, Hemostatic",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4300",
        "device_class": "2"
      }
    }
  ]
}