{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 8
    }
  },
  "results": [
    {
      "cfres_id": "122496",
      "product_res_number": "Z-0525-2014",
      "event_date_initiated": "2013-10-09",
      "event_date_posted": "2013-12-17",
      "recall_status": "Terminated",
      "event_date_terminated": "2018-05-17",
      "res_event_number": "66496",
      "product_code": "LTK",
      "k_numbers": [
        "K983875"
      ],
      "product_description": "VITROS CA 125 II Calibrators\n\nFor use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, VITROS 3600 Immunodiagnostic System and VITROS 5600 Integrated System for the quantitative measurement of OC 125 defined antigen in human serum and plasma (EDTA or heparin).",
      "code_info": "product code: 680 0034; Lot numbers 1220, 1230, 1240, 1250",
      "firm_fei_number": "1000136573",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "city": "Rochester",
      "state": "NY",
      "postal_code": "14626-5101",
      "reason_for_recall": "Ortho Clinical Diagnostics issued a Product Correction Notification for VITROS Immunodiagnostic Products CA 125 II Reagent Pack and Calibrators due to a positive shift in patient results.",
      "root_cause_description": "Under Investigation by firm",
      "action": "The firm, Ortho Clinical Diagnostics, sent a \"IMPORTANT PRODUCT CORRECTION NOTIFICATION\" letter dated October 9, 2013  to their clients/customers informing them of a shift in results that may be observed when performing patient sample correlation tests between VITROS CA 125 II Calibrators and Reagent Packs. The letter described the product, problem and actions to be taken.  The customers were informed that it is acceptable to continue using any remaining inventory until its expiration date providing that quality control results are within acceptable limits (Note: Lots 1220 and 1230 are expired). The customers were also instructed to consider the need to review patient results using the affected lot(s). If you suspect that a previously reported result may have been affected, provide this information to your Laboratory Medical Director and the requesting physician or health care provider so that appropriate actions may be taken for the patients; Complete and return the attached Confirmation of Receipt form no later than October 18, 2013 via FAX TO: Ortho Clinical Diagnostics at 1-888-557-3759 or 1-585-453-4110, and Forward the information in this notification if you have distributed this product outside of your facility.\n\nIf you have any additional questions, please contact Customer Technical Services at 1-800-421-3311.",
      "product_quantity": "601 units",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of: AK, CA, CT, CO, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV, WY, and Puerto Rico, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.",
      "openfda": {
        "k_number": [
          "K964020",
          "K200199",
          "K143534",
          "K983715",
          "K964098",
          "K240479",
          "K964185",
          "K980240",
          "K142895",
          "K023891",
          "K221355",
          "K213510",
          "K983391",
          "K031297",
          "K042731",
          "K072794",
          "K080561",
          "K983875",
          "K080469",
          "K972162",
          "K020828",
          "K990431",
          "K023597",
          "K981297",
          "K100321",
          "K022329"
        ],
        "pma_number": [
          "P850009"
        ],
        "registration_number": [
          "3000308930",
          "8020888",
          "2517506",
          "3005622096",
          "3002806944",
          "2032839",
          "1219913",
          "2250051",
          "1036362",
          "3007111389",
          "3014643041",
          "3004192065",
          "3002809144",
          "2432235",
          "8031673",
          "3011989923",
          "2122870",
          "3006198300",
          "9610126",
          "3005333358",
          "3005529799",
          "2020726",
          "1643621",
          "2521625",
          "3002642396",
          "9612316",
          "3012963943",
          "1319681",
          "3005360469",
          "3002895169"
        ],
        "fei_number": [
          "3000308930",
          "2517506",
          "3005622096",
          "3002806944",
          "1219913",
          "2250051",
          "1036362",
          "3002808328",
          "3007111389",
          "3014643041",
          "3004185580",
          "3002806559",
          "3004192065",
          "3002809144",
          "2432235",
          "3011989923",
          "2122870",
          "3007019572",
          "3006198300",
          "2967",
          "3005333358",
          "3005529799",
          "2020726",
          "1643621",
          "2521625",
          "3002642396",
          "1000136573",
          "3012963943",
          "3005670738",
          "3002895169"
        ],
        "device_name": "Test, Epithelial Ovarian Tumor-Associated Antigen (Ca125)",
        "medical_specialty_description": "Immunology",
        "regulation_number": "866.6010",
        "device_class": "2"
      }
    }
  ]
}