  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "33484",
      "product_res_number": "Z-1039-04",
      "event_date_initiated": "2004-05-28",
      "event_date_posted": "2004-07-20",
      "recall_status": "Terminated",
      "event_date_terminated": "2004-11-23",
      "res_event_number": "29292",
      "product_code": "LSO",
      "product_description": "Digene Hybrid Capture System CMV DNA Test (version 2.0), packaged in a cardboard box containing test reagents for performing 60 tests. Digene Catalog #4403-2003.",
      "code_info": "Lot number 6234, EXP 10/20/04",
      "firm_fei_number": "3003572099",
      "recalling_firm": "Digene Corp",
      "address_1": "1201 Clopper Rd",
      "city": "Gaithersburg",
      "state": "MD",
      "postal_code": "20878-4000",
      "additional_info_contact": "Donna Sowers\n301-944-7007",
      "reason_for_recall": "In vitro diagnostic test kit may produce false negative clinical results.",
      "root_cause_description": "Other",
      "action": "The recalling firm notified consignees by phone 5/28/04 and by faxed letter 6/2/04. The letter advised of possible false negative test results and that such results should be retested immediately with replacement kits provided by the firm. The notification further advised that positive results should be considered valid if the assay meets all verification and calibration criteria. Consignees were requested to return  product and response form.",
      "product_quantity": "206 kits",
      "distribution_pattern": "Product was sold to medical facility clinical laboratories nationwide and a distributor in the United Kingdom.",
      "openfda": {
        "device_name": "Dna-Reagents, Cytomegalovirus",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.3175",
        "device_class": "2"
      }
    }
  ]
}