  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "96195",
      "product_res_number": "Z-1397-2011",
      "event_date_initiated": "2010-12-20",
      "event_date_posted": "2011-02-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-11-19",
      "res_event_number": "57431",
      "product_code": "LSK",
      "k_numbers": [
        "K092704"
      ],
      "product_description": "Abbott RealTime CT/NG Assay, List 8L07, in vitro diagnostic; Abbott Molecular Inc., Des Plaines, IL 60018 USA\nThe Abbott RealTime CT/NG assay is an in vitro polymerase chain reaction (PCR) assay for the direct, qualitative detection of the plasmid DNA for C. trachomatis and the genomic DNA of N. gonorrhoeae in female endocervical or vaginal swab specimens, male urethral swab specimens, or in male and female urine specimens.",
      "code_info": "All in date lots",
      "firm_fei_number": "3005248192",
      "recalling_firm": "Abbott Molecular",
      "address_1": "1300 E Touhy Ave",
      "city": "Des Plaines",
      "state": "IL",
      "postal_code": "60018-3315",
      "additional_info_contact": "224-361-7133",
      "reason_for_recall": "Internal  testing identified the potential of under filled and over filled enzyme reagent vials which may result in obtaining less than the total number of tests per kit as described on the package insert.",
      "root_cause_description": "Other",
      "action": "Abbott Molecular sent a Field Advisory Notice letter to all RealTime HIV-1, CT/NG and HBV assay customers on December 20, 2010, informing them that internal testing identified the potential of under filled and over filled enzyme vials which may result in obtaining less than the total number of tests per kit as described in the package insert.  The customers were told that specimens that meet acceptance criteria, per the product labeling are not affected by the situation and are considered valid results, and that the issue has no impact on results for those runs that generate valid results.  Any questions were directed to Abbott Technical Support at 1-800-553-7042, Option 2.",
      "distribution_pattern": "Nationwide Distribution",
      "openfda": {
        "k_number": [
          "K010891",
          "K874878",
          "K920378",
          "K010892",
          "K871878",
          "K920302",
          "K990023",
          "K092704"
        ],
        "device_name": "Dna-Reagents, Chlamydia",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.3120",
        "device_class": "1",
        "registration_number": [
          "1122376"
        ],
        "fei_number": [
          "3003572099"
        ]
      }
    }
  ]
}