  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "149421",
      "product_res_number": "Z-2793-2016",
      "event_date_initiated": "2016-08-02",
      "event_date_posted": "2016-09-14",
      "recall_status": "Terminated",
      "event_date_terminated": "2017-03-20",
      "res_event_number": "75125",
      "product_code": "LLG",
      "k_numbers": [
        "K093612"
      ],
      "product_description": "Syngo Plaza picture archiving and communication system (PAC) (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.",
      "code_info": "Model Numbers: 10863171, 10863172, 10863173, 10592457  Serial Numbers:  100262,100389,100198,100258,100259,100538,100135,100133,100213,100214,100031,100032,100484,100362,100233,100140,100141,100142,100143,100144,100416,100105,100106,100109,100110,100111,100112,100113,100114,100115,100116,100117,100119,100120,100461,100457,100245,100447,100364,100314,100298,100299,100386,200277,100139,100027,100028,100300,100559,100440,100441,100197,100170,100025,100026,100084,100186,100189,100190,100191,100192,100195,100196,100550,100033,100034,100306,100307,100232,100449,100450,100185,100596,100256,100199,100188,100193,100194,100257,100226,100228,100425,100029,100030,100253,100203,100204,100132,100247,100248,100434,100215,100491,100495,100230,100221,100212,100539,100206",
      "firm_fei_number": "2240869",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "city": "Malvern",
      "state": "PA",
      "postal_code": "19355-1418",
      "additional_info_contact": "Anastasia Sokolova\n610-219-4834",
      "reason_for_recall": "Siemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values.  This suffix stands for on projection and indicates that a shown measurement value may be different from the real value.",
      "root_cause_description": "Error in labeling",
      "action": "Siemens mailed to customers on August 04, 2016 a Customer Advisory Notice that provided interpretations about the meaning of measurement values like area, distance and perimeter shown as \"o.p.\" by the syngo.plaza.",
      "product_quantity": "101 Systems",
      "distribution_pattern": "Nationwide Distribution",
      "openfda": {
        "k_number": [
          "K841465",
          "K840257"
        ],
        "device_name": "Kit, Igg, Platelet Associated",
        "medical_specialty_description": "Hematology",
        "regulation_number": "864.5425",
        "device_class": "2"
      }
    }
  ]
}