  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 5
    }
  },
  "results": [
    {
      "cfres_id": "163588",
      "product_res_number": "Z-2037-2018",
      "event_date_initiated": "2018-04-20",
      "event_date_created": "2018-05-31",
      "recall_status": "Terminated",
      "event_date_terminated": "2020-11-04",
      "res_event_number": "79822",
      "product_code": "LEI",
      "pma_numbers": [
        "P100049"
      ],
      "product_description": "LINX Reflux Management System, Implant Suture, 12-17 Bead, 0.7T, Product Code LS-xx, sterile, Rx only.  Labeled as:\n\na. Product Code LS-12;  \nb. Product Code LS-13; \nc. Product Code LS-14;  \nd. Product Code LS-15;  \ne. Product Code LS-16;  \nf. Product Code LS-17;  \n\n Laparoscopic, fundic-sparing anti-reflux procedure indicated for patients diagnosed with Gastroesophageal Reflux Disease (GERD) as defined by abnormal pH testing, and who are seeking an alternative to continuous acid suppression therapy (i.e. proton pump inhibitors or equivalent) in the management of their GERD.",
      "code_info": "a. Product Code LS-12, Part No. 2170-6, GTIN 00855106005011;   b. Product Code LS-13, Part No. 2170-7, GTIN 00855106005028;  c. Product Code LS-14, Part No. 2170-8, GTIN 00855106005035;   d. Product Code LS-15, Part No. 2170-9, GTIN 00855106005042;   e. Product Code LS-16, Part No. 2170-10, GTIN 00855106005059;   f. Product Code LS-17, Part No. 2170-11, GTIN 00855106005066;    Lot No. 6100 through 14055, 15286, 15287, 15314, 15315, 15317, and 16557",
      "firm_fei_number": "3008766073",
      "recalling_firm": "Torax Medical, Inc.",
      "address_1": "4188 Lexington Ave N",
      "city": "Shoreview",
      "state": "MN",
      "postal_code": "55126-6106",
      "additional_info_contact": "877-384-4266",
      "reason_for_recall": "Possible out of specification condition which may allow a bead component to separate from an adjacent wire link. This condition may result in a discontinuous or open LINX device.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "The firm, Consignees Torax Medical Inc., sent an \"URGENT: FIELD SAFETY NOTICE (REMOVAL)\" letter mailed on approximately 04/20/2018 (US consignees and physicians) and 04/23/2018 (OUS consignees and physicians).  The letter described the product, problem and actions to be taken. The customers were instructed to examine inventory for affected product, quarantine any affected product, communicate the issue to all relevant operating room or materials management personnel, forward the recall information to facilities where the product may have been forwarded to,  complete the Business Reply Form confirming receipt of this notice and return it to Torax Medical, Inc. within three (3) business days via email at brf@toraxmedical.com; ( return the BRF even if you do not have the product lot subject to this recall.), and arrange the return of any affected product on hand.  \n\nAdditionally, a recall notification will be sent to sent to physicians outlining the reason for recall and to advise them to monitor their patients.\n\nQuestion regarding the recall can be directed to the Customer Support Cetner at 1-877-384-4266, which is open Monday through Friday, 7:30 AM to 6:30 PM ET.",
      "product_quantity": "9,131 devices in total",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) in states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY; and in countries of: Austria, France, Germany, Italy, Scotland, Spain, Switzerland, United Kingdom.",
      "openfda": {
        "device_name": "Implant, Anti-Gastroesophageal Reflux",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3",
        "pma_number": [
          "P790006",
          "P100049",
          "P250018"
        ],
        "registration_number": [
          "",
          "3006638824",
          "3008766073",
          "1450662",
          "3015217460",
          "3013445086"
        ],
        "fei_number": [
          "",
          "3006638824",
          "3008766073",
          "1450662",
          "3015217460",
          "3013445086"
        ]
      }
    }
  ]
}