  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "36840",
      "product_res_number": "Z-0633-05",
      "event_date_initiated": "2004-10-20",
      "event_date_posted": "2005-03-18",
      "recall_status": "Terminated",
      "event_date_terminated": "2006-07-22",
      "res_event_number": "30930",
      "product_code": "LAZ",
      "product_description": "ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycoplasma pneumoniae in human serum.  Test kit for 192 determinations.  Catalog No. 4700.  For In Vitro Diagnostic use.  manufactured for DiaSorin Inc., 1950 Northwestern Avenue P.O. Box 285, Stillwater, MN  55082-0285",
      "code_info": "lot # E052660D, E052790D",
      "firm_fei_number": "2182595",
      "recalling_firm": "Diasorin Inc.",
      "address_1": "1951 Northwestern Ave S",
      "city": "Stillwater",
      "state": "MN",
      "postal_code": "55082-7536",
      "additional_info_contact": "DiaSorin Technical Services\n800-328-1482",
      "reason_for_recall": "2 lots of ETI-MP IgM assay kits showed a decline in Optical Density (OD) reading of the positive control and calibrators, P10 and P50.  The decline in OD of the positive control and calibrators (P10 & P50) would cause the assay validation criteria to fail.",
      "root_cause_description": "Other",
      "action": "Customers were sent notification letters. Recommended to detroy the remaining kits and requested a form be returned which indicates the number of kits destroyed so they can arrange replacement.",
      "distribution_pattern": "MA, OH & IL",
      "openfda": {
        "k_number": [
          "K811461",
          "K211667",
          "K823127",
          "K851474"
        ],
        "device_name": "N-Acetylprocainamide Control Materials",
        "medical_specialty_description": "Clinical Toxicology",
        "regulation_number": "862.3280",
        "device_class": "1"
      }
    }
  ]
}