{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 8
    }
  },
  "results": [
    {
      "cfres_id": "113066",
      "product_res_number": "Z-0071-2013",
      "event_date_initiated": "2009-10-02",
      "event_date_posted": "2012-10-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2013-01-29",
      "res_event_number": "63225",
      "product_code": "KRI",
      "product_description": "Terumo Backplate Replacement CXFVR 1500 Holder\r\n\r\nThe Flexible venous Reservoir is mounted on the Backplate/Holder and secures the flexible reservoir by strategically placed mounting arms. Additionally, there are three port supports the user incorporates to position the lines that enter and exit the venous bag. The Backplate/Holder is designed to allow the user to move the port support location to accommodate different tubing circuit designs to permit mounting flexibility.",
      "code_info": "No lot/serial number; Catalog number: PN 816280",
      "firm_fei_number": "1828100",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Road",
      "city": "Ann Arbor",
      "state": "MI",
      "postal_code": "48103-9586",
      "additional_info_contact": "734-741-6173",
      "reason_for_recall": "Backplate reservoir holders were manufactured with an incorrect configuration of connector supports. The devices were manufactured and configured with two 3/8 inch and one  inch connector supports instead of one 3/8 inch and two  inch connector supports (See location of supports as # 7 and # 8 on backplate attachment). The backplates were manufactured as described in manufacturing instructions a",
      "root_cause_description": "Labeling design",
      "action": "The firm, Terumo, interviewed the Sr. Product Manager from the Marketing Department of the company on August 29, 2012.  The interview included the description of the product, problem and action taken. Replacement connectors were provided to users and all units were corrected.\r\n\r\nIf you have any questions, call 734-663-4145 or Manager, Regulatory Affairs at 734-741-5576.",
      "product_quantity": "33 units",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) including states of: CA, MI, NC, NJ, PA, and WI;and country of: EUROPE.",
      "openfda": {
        "k_number": [
          "K992635",
          "K802542",
          "K852455",
          "K822097",
          "K790192",
          "K934782",
          "K860298",
          "K811368",
          "K830167",
          "K892856",
          "K780984",
          "K874605",
          "K874925",
          "K926233",
          "K802906",
          "K781861",
          "K841296",
          "K862510",
          "K802541",
          "K841812",
          "K990512",
          "K864890",
          "K851077",
          "K862549",
          "K770681",
          "K802539",
          "K851289",
          "K841032",
          "K810079",
          "K820406",
          "K884640",
          "K884893",
          "K922404",
          "K792151",
          "K811348",
          "K020365",
          "K860672",
          "K832704",
          "K813271",
          "K791829",
          "K854113",
          "K832411"
        ],
        "registration_number": [
          "3010402691",
          "8010026",
          "3003661577",
          "1124841",
          "3045892899",
          "3006073153",
          "3011468686",
          "1646747",
          "3007899424",
          "1055236",
          "3004605321",
          "1043214",
          "3006849754",
          "3011373951",
          "2916596",
          "1218444",
          "2244478",
          "1643817",
          "3003306248",
          "3011216300",
          "1718850",
          "8020347",
          "8010697",
          "3009507273",
          "9680841",
          "2183924",
          "1054241",
          "3013267343",
          "1422634",
          "1625519",
          "3009810261",
          "3003905499",
          "9611109",
          "3021599320",
          "3010163695",
          "3019807891",
          "3017540705",
          "2432320",
          "1828100",
          "3003752502",
          "3010009632",
          "3021562887",
          "3004215117",
          "8010762",
          "9681834",
          "3016746283"
        ],
        "fei_number": [
          "3010402691",
          "3003661577",
          "1000160162",
          "3045892899",
          "3006073153",
          "3011468686",
          "3007899424",
          "1000116132",
          "3004605321",
          "1043214",
          "3006849754",
          "3011373951",
          "2916596",
          "1218444",
          "1643817",
          "3003306248",
          "3011216300",
          "1000514656",
          "1718850",
          "1000112877",
          "3009507273",
          "3013267343",
          "1000112968",
          "1625519",
          "2828",
          "3009810261",
          "3002808327",
          "3003905499",
          "3005426550",
          "1000521377",
          "3021599320",
          "3009649876",
          "3010163695",
          "3019807891",
          "3000204839",
          "3017540705",
          "3005612924",
          "3004053982",
          "1828100",
          "3003752502",
          "3010009632",
          "3021562887",
          "3004215117",
          "1000136269",
          "3016746283",
          "3002808267"
        ],
        "device_name": "Accessory Equipment, Cardiopulmonary Bypass",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.4200",
        "device_class": "1"
      }
    }
  ]
}