  "results": [
    {
      "cfres_id": "74685",
      "product_res_number": "Z-1021-2009",
      "event_date_initiated": "2008-10-10",
      "event_date_posted": "2009-03-10",
      "recall_status": "Terminated",
      "event_date_terminated": "2010-10-27",
      "res_event_number": "50084",
      "product_code": "KIJ",
      "k_numbers": [
        "K943323",
        "K972367"
      ],
      "product_description": "Stryker Cordless Driver 3 Handpiece.  Stryker Instruments, Kalamazoo, MI; REF 4300.  \r\n\r\n\r\nThe device is intended for use in surgical procedures involving drilling, reaming, driving wire or pins, and cutting bone and hard tissue.",
      "code_info": "Serial number 723505023 through 804326503.",
      "firm_fei_number": "1811755",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "city": "Portage",
      "state": "MI",
      "postal_code": "49002",
      "additional_info_contact": "888-311-4521",
      "reason_for_recall": "The handpieces may continue to run even when the trigger is not depressed, which could result in the unintended cutting of bone or soft tissue damage.",
      "root_cause_description": "Component design/selection",
      "action": "Consignees were notified of the problem via letter dated October 10, 2008, provided interim instructions to reduce the risk, and informed that Stryker instruments will replace the trigger on all handpieces.  Consignees were also instructed on the following:  1) Review the letter and distribute to all affected departments, 2) Complete and sign the enclosed Business Reply Form and fax to 336-499-8603 and 3) Call Stryker Instruments Service Department at 888-308-1983 to schedule service for the handpieces.",
      "product_quantity": "25,312 of all units",
      "distribution_pattern": "Worldwide Distribution including USA and countries of Australia, Brazil, Canada, Chile, Colombia, Costa Rica, England, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Romania, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan and Turkey.",
      "openfda": {
        "k_number": [
          "K800311",
          "K780645",
          "K933337",
          "K941285",
          "K844163",
          "K960711",
          "K163308",
          "K884839",
          "K771218",
          "K800313",
          "K951450",
          "K944488",
          "K970378",
          "K873766",
          "K820367",
          "K112599",
          "K932307",
          "K931395",
          "K800312",
          "K101563",
          "K933101",
          "K844162",
          "K800314",
          "K972367",
          "K822790",
          "K864137",
          "K850789",
          "K790063",
          "K800315"
        ],
        "registration_number": [
          "1526534",
          "3013422236",
          "1834331",
          "3015151147",
          "3015967359",
          "2020601",
          "3017196117",
          "3043620689",
          "1835296",
          "3014747863",
          "3034669683",
          "3008709738",
          "3037012186",
          "1818910",
          "3004049923",
          "3004832480",
          "1219602",
          "3009599228",
          "1834379",
          "3030427687",
          "1835251",
          "3004719693",
          "2135156",
          "3007997036",
          "3015516266",
          "3020155054",
          "3016577622",
          "3025436467",
          "3014967969",
          "3010045785",
          "3016112537",
          "3004217743",
          "3007125392",
          "3010373263",
          "3029826185",
          "3030966822",
          "3006460162",
          "3025141",
          "3007123908",
          "1649518",
          "1531174",
          "2032098",
          "1000517406",
          "3010703925",
          "3004060780",
          "3024082778",
          "3008590661",
          "3008793310",
          "1720929",
          "1450662",
          "3007084590",
          "1423662",
          "3003604053",
          "3036756245",
          "1043653",
          "2027754",
          "2029275",
          "1221934",
          "3012764194",
          "9616671",
          "3007344102",
          "3019932374",
          "3009971621",
          "3013756169",
          "2132111",
          "3025603301",
          "3034932459",
          "1220477",
          "9610905",
          "1422572",
          "3008951116",
          "2028632",
          "1643264",
          "3009973699",
          "3006017180",
          "3007958831",
          "1057946",
          "3009106092",
          "3017181414",
          "3007441485",
          "3026311463",
          "3011530718",
          "2916714",
          "3007539489",
          "3033509898",
          "1226544"
        ],
        "fei_number": [
          "1526534",
          "3013422236",
          "1000306303",
          "3015151147",
          "3015967359",
          "1000125930",
          "2020601",
          "3017196117",
          "3043620689",
          "3014747863",
          "3034669683",
          "3008709738",
          "3037012186",
          "1818910",
          "3004049923",
          "3004832480",
          "1219602",
          "3009599228",
          "3030427687",
          "3004719693",
          "1000220336",
          "3007997036",
          "3015516266",
          "3020155054",
          "3001236277",
          "3016577622",
          "3025436467",
          "3014967969",
          "3010045785",
          "3016112537",
          "3004217743",
          "3007125392",
          "3010373263",
          "3029826185",
          "3030966822",
          "3006460162",
          "3007123908",
          "3000215273",
          "1000517406",
          "3000210784",
          "3010703925",
          "1000306340",
          "1000306431",
          "3003179719",
          "3004060780",
          "1000525271",
          "3024082778",
          "3008590661",
          "3008793310",
          "1720929",
          "1450662",
          "3007084590",
          "3002993297",
          "3003604053",
          "3036756245",
          "1043653",
          "2027754",
          "1221934",
          "3012764194",
          "3007344102",
          "3019932374",
          "3002854251",
          "3009971621",
          "3013756169",
          "3025603301",
          "3034932459",
          "1220477",
          "1000125025",
          "3008951116",
          "1643264",
          "3009973699",
          "3006017180",
          "3007958831",
          "3003128516",
          "3000205591",
          "3009106092",
          "1000140026",
          "3003058448",
          "3017181414",
          "3007441485",
          "3026311463",
          "3011530718",
          "2916714",
          "3007539489",
          "3033509898",
          "3002807464"
        ],
        "device_name": "Instrument, Surgical, Orthopedic, Dc-Powered Motor And Accessory/Attachment",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4820",
        "device_class": "1"
      }
    }
  ]
}