  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "26005",
      "product_res_number": "Z-0727-03",
      "event_date_initiated": "2003-02-10",
      "event_date_posted": "2003-04-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2005-01-19",
      "res_event_number": "25541",
      "product_code": "JWM",
      "k_numbers": [
        "K903315",
        "K914350"
      ],
      "product_description": "ICON 9.5 Software used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203",
      "code_info": "All Diacam, Multispect 2, Multispect 3 and e.cam camera systems with ICON workstations with ICON software version 9.5 and the optional Whole Body SPECT software.",
      "firm_fei_number": "1000514697",
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "address_1": "2501 N. Barrington Road",
      "city": "Hoffman Estates",
      "state": "IL",
      "postal_code": "60195-5203",
      "additional_info_contact": "Mr. Ron Nolte\n847-304-7700",
      "reason_for_recall": "ICON 9.5 software does not apply flood corrections during whole body SPECT studies with multiple bed positions.",
      "root_cause_description": "Other",
      "action": "Siemens sent Customer Advisory letters dated 2/10/03 to all ICON 9.5 customers with Whole Body SPECT functionality on integrated camera systems including Diacam, Multispect 2, Multispect 3 and e.cam camera systems aquired on an ICON workstation.  The accounts were informed of the problem related to applying the flood correction matrix to Whole Body SPECT studies in ICON version 9.5 software, and were provided the follwoing workaround:\nTo ensure that the flood correction is applied, the user should acquire multiple single bed positions with the General SPECT predefine, indexing the bed manually between each SPECT acquisition.  The data can then be combined into a Whole Body SPECT in the Whole Body SPECT Display program for review.  In addition, users were requested to routinely perform standard quality control procedures at the facility to ensure proper image quality.  No software upgrade is planned.",
      "product_quantity": "600 units",
      "distribution_pattern": "Nationwide, United Arab Emirates, Argentina, Austria, Australia, Barbados, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Equador, Egypt, France, Germany,  Great Britain, Georgia, Greece, Hong Kong, Israel, India, Italy, Jordan, Japan, Korea, Latvia, Mexico, Niger, Norway, New Zealand, Philippines, Pakistan, Poland, Russia, Saudi Arabia, SIngapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine and the Netherlands.",
      "openfda": {
        "k_number": [
          "K873075",
          "K874473",
          "K851686",
          "K832737",
          "K780368",
          "K851685",
          "K810873",
          "K850885",
          "K914350",
          "K841905",
          "K904599",
          "K851687",
          "K812966",
          "K780369",
          "K901545",
          "K801917",
          "K903315"
        ],
        "device_name": "System, Tomographic, Nuclear",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1310",
        "device_class": "2"
      }
    }
  ]
}