  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 3
    }
  },
  "results": [
    {
      "cfres_id": "33829",
      "product_res_number": "Z-1238-04",
      "event_date_initiated": "2004-06-29",
      "event_date_posted": "2004-08-04",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-05-03",
      "res_event_number": "29487",
      "product_code": "HQP",
      "product_description": "Battery-operated Cautery\nLow Temperature, fine/elongated Tip",
      "code_info": "Sterile lot numbers: 0204B 0204D 0204E 0304A 0304C 0304D 0404A 0404E 0404G 0404H 0404J 0504A 0504C 0504E 0504F. EXP DATES 2/1/08-5/1/04",
      "firm_fei_number": "3007593903",
      "recalling_firm": "Aaron Medical Industries, Inc.",
      "address_1": "7100 30th Ave N",
      "city": "St Petersburg",
      "state": "FL",
      "postal_code": "33710-2902",
      "additional_info_contact": "Vera MacElroy\n727-803-8513",
      "reason_for_recall": "The cover caps on some of these cauteries are not properly engaged to prevent inadvertent activation.",
      "root_cause_description": "Other",
      "action": "Each consignee will be notified by letter of the recall with instruction for the disposal of the product determined to be hazardous. Each letter will have a questionnaire attached with a 800 fax number for replying to the firm.",
      "product_quantity": "16700",
      "distribution_pattern": "Distributed to: 291 accounts in the USA, 8 in Canada, 18 in Europe, 1 in the Middle East, 1 in Asia, 4 in the Pacific Rim 1 in Africa and 5 US goverment/Military.",
      "openfda": {
        "k_number": [
          "K840425",
          "K841326",
          "K023506",
          "K875260",
          "K945763",
          "K810016",
          "K933079",
          "K840426",
          "K121441",
          "K952784",
          "K820157",
          "K002547",
          "K083428",
          "K810013",
          "K945761",
          "K771189",
          "K810012",
          "K062659",
          "K945762",
          "K994019",
          "K945493",
          "K801078",
          "K990728",
          "K841325"
        ],
        "device_name": "Unit, Cautery, Thermal, Battery-Powered",
        "medical_specialty_description": "Ophthalmic",
        "regulation_number": "886.4115",
        "device_class": "2",
        "registration_number": [
          "3005528784",
          "3013421458",
          "3011050570",
          "8030607",
          "3041276591",
          "1061927",
          "1319639",
          "1211998",
          "3007593903",
          "3015136750",
          "3003072803",
          "3013466775",
          "1219619",
          "3009103001",
          "3007748817",
          "1928237",
          "1836161",
          "1220477",
          "1643817",
          "1319660"
        ],
        "fei_number": [
          "3005528784",
          "3013421458",
          "3011050570",
          "3041276591",
          "1211998",
          "3007593903",
          "3015136750",
          "3003072803",
          "1000136478",
          "3013466775",
          "1219619",
          "3009103001",
          "3002750084",
          "3000203000",
          "3007748817",
          "1928237",
          "1220477",
          "3006509795",
          "1643817",
          "1000138054"
        ]
      }
    }
  ]
}