  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "71854",
      "product_res_number": "Z-2200-2008",
      "event_date_initiated": "2008-06-16",
      "event_date_posted": "2008-09-11",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-06-23",
      "res_event_number": "48679",
      "product_code": "HFN",
      "k_numbers": [
        "K905563",
        "K955443",
        "K961424"
      ],
      "product_description": "Intran Plus, Disposable Intrauterine Pressure Catheter System, Part numbers IUP-400, IUP-400S, IUP-450, IUP-450S, IUP-500, IUP-550, IUP-600, IUP-600S, IUP-650, IUP-650S, IUP-700, IUP-750, Sterile, Single Use, Utah Medical Products Inc., 7043 South 300 West, Midvale, Utah 84047.",
      "code_info": "Lot numbers less than 1072755. Also 1080086, 1080087, 1080088, 1080089, 1080091, 1080093, 1080095, 1080245, 1080246, 1080247, 1080248, 1080249, 1080250, 1080266, 1080296, 1080297, 1080298, 1080299, 1080301, 1080302, 1080393, 1080394, 1080395, 1080432, 1080610, 1080669, and 1080721.",
      "firm_fei_number": "1718873",
      "recalling_firm": "Utah Medical Products, Inc",
      "address_1": "7043 South 300 West",
      "city": "Midvale",
      "state": "UT",
      "postal_code": "84047",
      "additional_info_contact": "801-566-1200",
      "reason_for_recall": "Possibility of compromised package sterility prior to the expiration date.",
      "root_cause_description": "Component change control",
      "action": "All consignees were notified by a faxed Recall Notice letter beginning on 6/16/2008. Customers were asked to return all unused product for immediate replacement with units that have a protective barrier within the tray packaging.  The firm cautioned customers to visually inspect packaging before use.  Follow-up phone calls will be made the following week.",
      "product_quantity": "300,000 units.",
      "distribution_pattern": "Worldwide Distribution --- USA including states of CA, KS, MO, NC, NV, TX, VA, and WA and country of Japan.",
      "openfda": {
        "k_number": [
          "K882224",
          "K250897",
          "K893503",
          "K954955",
          "K955443",
          "K920375",
          "K961424",
          "K974389",
          "K973746"
        ],
        "registration_number": [
          "3003971136",
          "3009077524",
          "9681377",
          "1722684",
          "2032098",
          "9610816",
          "1718873",
          "1720929",
          "3014541700",
          "3000126629",
          "1066270",
          "1218950"
        ],
        "fei_number": [
          "3003971136",
          "3009077524",
          "3003058448",
          "1000220650",
          "1000300598",
          "1718873",
          "1720929",
          "3014541700",
          "3000126629",
          "3000957797",
          "3002807167",
          "1218950"
        ],
        "device_name": "Transducer, Pressure, Intrauterine",
        "medical_specialty_description": "Obstetrics/Gynecology",
        "regulation_number": "884.2700",
        "device_class": "2"
      }
    }
  ]
}