  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "157691",
      "product_res_number": "Z-3072-2017",
      "event_date_initiated": "2017-06-14",
      "event_date_created": "2017-08-25",
      "recall_status": "Terminated",
      "event_date_terminated": "2021-08-24",
      "res_event_number": "77706",
      "product_code": "GBT",
      "k_numbers": [
        "K120088",
        "K143031"
      ],
      "product_description": "mcompass 2-channel Balloon Catheters\n\nThe mcompass Biofeedback Anorectal Manometry System is for use on patients requiring anorectal pressure studies. mcompass Anorectal Manometry System (RMD-001-001) which includes a FOB component connected to the balloon catheter to inflate and deflate and a medical grade tablet PC to run the software. The software is not included in the recall, only the balloon catheter component.",
      "code_info": "Part # RMD-002-004 Lot No #'s 160627-10 and 160627-11.",
      "firm_fei_number": "3008414621",
      "recalling_firm": "Medspira, Llc",
      "address_1": "2718 Summer St NE",
      "city": "Minneapolis",
      "state": "MN",
      "postal_code": "55413-2820",
      "additional_info_contact": "Paul Bradik\n612-789-0013",
      "reason_for_recall": "Potential failure in the balloon bond in the inner catheter stem, leading to the separation of the balloon from the inner stem, leaving it in the rectal cavity.",
      "root_cause_description": "Employee error",
      "action": "The firm Medspira plans to contact all foreign and domestic consignees in regards to the recall. International consignees are to destroy recalled product by cutting the catheter shaft and balloon with scissors. It is also requested that their own customers be notified of the recall. Medspira plans to provide product replacement. Domestic customers are asked to return affected catheters to the recalling firm. Medspira indicates that they will cove shipping costs and replacements.  All consignees are asked to complete and return the attached recall response form.  For further questions please call, (612) 789-0013.",
      "product_quantity": "355 catheters (255 US - 100 foreign. )",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the states of :CA, FL, IA, ID, and TX., and to the countries of : Austria, Italy, Malaysia, Philippines, Sweden, and United Kingdom",
      "openfda": {
        "device_name": "Catheter, Rectal",
        "medical_specialty_description": "Gastroenterology, Urology",
        "regulation_number": "876.5980",
        "device_class": "2",
        "k_number": [
          "K924549",
          "K890466",
          "K941391"
        ],
        "registration_number": [
          "3014656749"
        ],
        "fei_number": [
          "3014656749"
        ]
      }
    }
  ]
}