  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "170632",
      "product_res_number": "Z-0972-2020",
      "event_date_initiated": "2019-01-04",
      "event_date_created": "2020-02-04",
      "recall_status": "Terminated",
      "event_date_terminated": "2020-05-29",
      "res_event_number": "82026",
      "product_code": "GBL",
      "product_description": "Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822\nG14823",
      "code_info": "8011030 8033568 8063430 8063431 8150491 8417631 8417632 8587162 8587163 8590532 8590533 7990830 8003084 8033571 8063428 8590520 8590522 8629475 NS8629473",
      "firm_fei_number": "1820334",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "city": "Bloomington",
      "state": "IN",
      "postal_code": "47404-9120",
      "additional_info_contact": "Cook Medical Customer Relations Department\n800-457-4500",
      "reason_for_recall": "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.",
      "root_cause_description": "Process control",
      "action": "On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.",
      "product_quantity": "13388 total",
      "distribution_pattern": "Domestic distribution to nationwide US.\n\nForeign distribution to Australia, Belgium,\nBolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic,\nDenmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India,\nIreland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia,\nNetherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto\nRico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan,\nTrinidad and Tobago, United Arab Emirates, and the United Kingdom.",
      "openfda": {
        "device_name": "Catheter, Ureteral, General & Plastic Surgery",
        "medical_specialty_description": "Gastroenterology, Urology",
        "regulation_number": "876.5130",
        "device_class": "N",
        "registration_number": [
          "3005987240"
        ],
        "fei_number": [
          "3005987240"
        ]
      }
    }
  ]
}