  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "34727",
      "product_res_number": "Z-1487-04",
      "event_date_initiated": "2004-04-22",
      "event_date_posted": "2004-09-29",
      "recall_status": "Terminated",
      "event_date_terminated": "2005-01-12",
      "res_event_number": "29886",
      "product_code": "FWN",
      "product_description": "Alveolus Tracheobronchial Stent Technology System (TB-STS), Manufactured for: Alveolus Inc., 401 North Tryon Street, Charlotte, North Carolina USA 28202",
      "code_info": "Size 12x40, Part #90129-20, Lot #AC0406; Size 14x40, Part #90129-21, Lot #AC0409; Size 16x40, Part #90129-22, Lot #AA0411; Size 16x60, Part #90129-23, Lot #A0413; Size 12x20, Part #90129-24, Lot #AA0404; Size 14x40, Part #90129-25, Lot 3AA0409; Size 14x40, Part #90129-26, Lot #AC0409; Size 16x60, Part #90129-27, Lot #AA0413",
      "firm_fei_number": "3004039296",
      "recalling_firm": "Alveolus, Inc.",
      "address_1": "401 N Tryon St Fl 10th",
      "city": "Charlotte",
      "state": "NC",
      "postal_code": "28202-2108",
      "additional_info_contact": "David R. Hamilton\n704-998-6014",
      "reason_for_recall": "The tracheal stent may collapse resulting in blockage of the air passage because the stents were not properly coated with the polyurethane material to the very end of the stent.",
      "root_cause_description": "Other",
      "action": "Consignees were notified by telephone on 4/22/2004.",
      "distribution_pattern": "FL, SC, TX",
      "openfda": {
        "k_number": [
          "K842287",
          "K962693",
          "K896667",
          "K000154",
          "K853726",
          "K841347",
          "K091135",
          "K884810",
          "K864936",
          "K842800"
        ],
        "device_name": "Prosthesis, Larynx (Stents And Keels)",
        "medical_specialty_description": "Ear, Nose, Throat",
        "regulation_number": "874.3620",
        "device_class": "2",
        "registration_number": [
          "1219795",
          "1032347",
          "1220850"
        ],
        "fei_number": [
          "1219795",
          "1032347",
          "1220850"
        ]
      }
    }
  ]
}