  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "144067",
      "product_res_number": "Z-1188-2016",
      "event_date_initiated": "2016-02-29",
      "event_date_created": "2019-07-09",
      "recall_status": "Terminated",
      "event_date_terminated": "2019-07-03",
      "res_event_number": "73348",
      "product_code": "FWE",
      "product_description": "The Vesa 75/100 adaptation with rotation is an accessory to a surgical camera intended to be used to record operative procedures.",
      "code_info": "FLEXION-PORT 400 (68cm) with Product number 32116068 and the following affected serial numbers: 001X001107, 001X001108, 001X001109, 001X001110, 001X001111, 001X001142, 001X001143, 001X001144, 001X001145, 001X001146, 001X001147, 001X001148, 001X001149, 001X001150, 001X001151, 001X001168, 001X001169, 001X001170, 001X001171, 001X001172, 001X001216, 001X001217, 001X001218, 001X001219, 001X001220, 001X001221, 001X001222, 001X001223, 001X001224, 001X001225, 001X001226, 001X001227, 001X001228, 001X001230, 001X001256, 001X001257, 001X001258, 001X001259, 001X001260, 001X001261, 001X001262, 001X001263, 001X001264, 001X001265, 001X001376, 001X001377, 001X001378, 001X001379, 001X001380, 001X001381, 001X001382, 001X001383, 001X001384, 001X001385  and 001X001229.   FLEXION-PORT 300+350 (93cm) with Product number 32116093 and the following affected serial numbers: 001X001173, 001X001174, 001X001175, 001X001176, 001X001177, 001X001235, 001X001237, 001X001239, 001X001299, 001X001309, 001X001310, 001X001313, 001X001314, 001X001315, 001X001366, 001X001367, 001X001368, 001X001369, 001X001370, 001X001371, 001X001372, 001X001373, 001X001374, 001X001375, 001X001402, 001X001403, 001X001406, 001X001231, 001X001232, 001X001233, 001X001234, 001X001236, 001X001238, 001X001240, 001X001296, 001X001297, 001X001298, 001X001300, 001X001301, 001X001302, 001X001303, 001X001304, 001X001305, 001X001306, 001X001307, 001X001308, 001X001311, 001X001312, 001X001404, 001X001405, 001X001439, 001X001440, 001X001441, 001X001442, 001X001443, 001X001444, 001X001445, 001X001446, 001X001447 and 001X001448.   FLEXION-PORT 400+350 (103cm) with Product number 32116103 and the following affected serial numbers: 001X001241, 001X001242, 001X001243, 001X001244, 001X001245, 001X001246, 001X001247, 001X001248, 001X001249, 001X001250, 001X001321, 001X001323, 001X001324, 001X001325, 001X001336, 001X001337, 001X001338, 001X001339, 001X001340, 001X001341, 001X001342, 001X001343, 001X001344, 001X001345, 001X001316, 001X001317, 001X001318, 001X001319, 001X001320 and 001X001322.",
      "firm_fei_number": "1418479",
      "recalling_firm": "Richard Wolf Medical Instruments Corp.",
      "address_1": "353 Corporate Woods Pkwy",
      "city": "Vernon Hills",
      "state": "IL",
      "postal_code": "60061-3110",
      "additional_info_contact": "Lisa Williams\n847-913-1113",
      "reason_for_recall": "Supplier of the \"VESA 75 / 100 adaptation with rotation\", which is part of their monitor mounts and support arms may be missing riveting.",
      "root_cause_description": "Process control",
      "action": "RICHARD WOLF sent an Urgent Safety Information letter dated March 2, 2016, to all affected consignees. The letter described the nature of the recall regarding the monitor holding plate (VESA 75/100 mount) in our Flexion-Port 300, 350 and 400 swivel arm product series.\n\nAccording to the firm insufficient riveting of the rotation elements the rivets may loosen during continuous use (e.g. when adjusting the tilt angle) and in an extremely unlikely event the monitor holding plate (VESA 75/100 mount) may loosen.\nAccording to Richard Wolf this can result in a potential danger for the patient, user and others which we want to prevent with this measure.\n\nThe firm is asking all end users to take the following step:\n\nDiscontinue use of the affected devices until the conversion kits have been installed by a qualified service technician.\n\nMake sure that all users of affected medical devices are  informed in their organization of the urgent safety information. If the end user has passed on products to third parties they should pass on a copy of this information and inform the recall contact person at Richard Wolf.\n\nPlease make sure that the information is observed at their organization until all necessary internal measures have been completed.\n\nThe end user should provide confirmation of receipt of this urgent safety information and answer the enclosed questionaire form which the firm asks to send to Richard Wolf at the following e-mail, marek.rast@richard-wolf.com by February 22, 2016. \n\nThe firm asking that recipients to acknowledge receipt of the safety information in order to avoid receiving more reminders from Richard Wolf.\n\nThe firm will send new monitor holding plate(s}, (643301264 conversion kit for monitor holding plate) once the form is  received.\n\nThe firm asks that qualified service technician install the new monitor holding plate (643301264 Conversion kit for monitor holding plate) in accordance with the enclosed assembly instructions and discard",
      "product_quantity": "108 units",
      "distribution_pattern": "Nationwide Distribution including Virginia, Pennsylvania, Florida, California, Texas, North Carolina and Illinois.",
      "openfda": {
        "device_name": "Camera, Television, Microsurgical, With Audio",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4160",
        "device_class": "1"
      }
    }
  ]
}