{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 5
    }
  },
  "results": [
    {
      "cfres_id": "104063",
      "product_res_number": "Z-0077-2012",
      "event_date_initiated": "2011-09-09",
      "event_date_posted": "2011-10-20",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-10-11",
      "res_event_number": "59948",
      "product_code": "DXS",
      "k_numbers": [
        "K852232"
      ],
      "product_description": "Medtronic Pressure Display Box, Models 66000, 64000, 63000, 60000.  It is capable of measuring pressure between -100 mmHg and +500 mmHg with an accuracy of +/-5 mmHg from -100 mmHg to +300 mmHg and +/-10 mmHg from +301 to +500 mmHg.  The Pressure Display Box has an upper and lower alarm setting. The alarm will be activated when the pressure is equal to or beyond the set value.    Manufacturer:  Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN  55431-5604.\n\nProduct  Usage:  This product is intended for use in displaying line pressures during cardiopulmonary bypass surgery.",
      "code_info": "serial numbers:  301010, 301012, 301013, 301015, 301016, 301017, 301018, 301019, 301020, 301023, 301024, 301025, 301026, 301027, 301029 thru 301041, 301043 thru 301050, 301052, 301053, 301054, 301055, 301056, 301057, 301058, 301060, 301061, 301062, 301063, 301068 thru 301085, 301087, 301088, 301090, 301091, 301092, 301093, 301094, 301300, 301301, 301302, 301303, 301371 thru 301383, 301385, 301387, 301388, 301389, 301390, 301394, 301396, 301397, 301399 thru 301407, 301409, 301410, 301431 thru 301450, 301460, 301461, 301462, 301464, 301467 thru 301485, 301487 thru 301507, 301509 thru 301526, 301560 thru 301663, 301665 thru 301704, 301707 thru 301751, 301753, 301754, 301755, 301756, 301758 thru 301817, 301819 thru 301842, 301844 thru 302148, 302150 thru 302161,   PDBR400000 thru PDBR400127, PDBR400129 thru PDBR400835, PDBR400847 thru PDBR401104, PDBR401107 thru PDBR401178.",
      "firm_fei_number": "1000116158",
      "recalling_firm": "Medtronic Perfusion Systems",
      "address_1": "7611 Northland Dr N",
      "city": "Brooklyn Park",
      "state": "MN",
      "postal_code": "55428-1088",
      "additional_info_contact": "Medtronic Cardiovascular Tech Support\n877-526-7890",
      "reason_for_recall": "The DFU for the Medtronic Pressure Display Box, model 66000, states that, for display readings of -100 to +300 mmHg there will be an accuracy of +/-5 mmHg, and for display readings of +301 to +500 mmHg there will be an accuracy of +/-10 mmHg.  Through reports from customers, Medtronic has become aware that the Pressure Display Box may not always accurately display readings.  During use in the -40",
      "root_cause_description": "Device Design",
      "action": "Medtronic Perfusion Systems, sent an \"Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers.   The letter identified the affected devices, described the issues, and the actions needed to be taken.  The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers.  Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form.  The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification.  For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.",
      "product_quantity": "1694 units",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of:   Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela, and  Yemen.",
      "openfda": {
        "k_number": [
          "K822822",
          "K771764",
          "K944764",
          "K781890",
          "K842194",
          "K852232",
          "K900797",
          "K873802",
          "K862836",
          "K820912",
          "K912045",
          "K831750",
          "K113235"
        ],
        "registration_number": [
          "3003955307",
          "1928237",
          "2184009",
          "1649914"
        ],
        "fei_number": [
          "3001665800",
          "3003955307",
          "1928237",
          "1000116158"
        ],
        "device_name": "Gauge, Pressure, Coronary, Cardiopulmonary Bypass",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.4310",
        "device_class": "2"
      }
    }
  ]
}