  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "95558",
      "product_res_number": "Z-1943-2011",
      "event_date_initiated": "2010-11-01",
      "event_date_posted": "2011-04-11",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-04-17",
      "res_event_number": "57189",
      "product_code": "DTY",
      "k_numbers": [
        "K902654"
      ],
      "product_description": "Terumo Cardiovascular Systems Corporations, CDI 101 Hematocrit/Oxygen Saturation Monitoring System.  In-line  blood-gas sensor used in cardiopulmonary bypass.",
      "code_info": "CDI 101  1070-1697",
      "firm_fei_number": "1828100",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Road",
      "city": "Ann Arbor",
      "state": "MI",
      "postal_code": "48103-9586",
      "additional_info_contact": "Tracy Bellanca\n734-741-6173",
      "reason_for_recall": "Inaccurate result.  Possibility for displayed values for potassium, CO2, hematocrit, O2 saturation, pH, hemoglobin and other measured blood parameters to be inaccurate.",
      "root_cause_description": "Device Design",
      "action": "On 11/1/2010, Terumo sent \"URGENT MEDICAL DEVICE CORRECTION\" notices to Chief Perfusionist, Department of Cardiovascular Surgery or Director of Operating Room Services.  The letter informed the consignees about the incidents of the CDI 100 and 101 displaying inaccurate values.  The letter states that when a displayed value is significantly different from expectations based on the clinical situation, the user should verify the accuracy by independent means before treating. The letter included  \"Caution\" stickers that were to be applied to the CDI system operators manual.",
      "product_quantity": "468 monitors",
      "distribution_pattern": "US, India, Colombia, Taiwan, Venezuela, Vietnam, Puerto Rico, Malaysia, Sri Lanka, Brazil, Indonesia, Pakistan, Chile, Germany, Hong Kong, Japan, AE, Singapore, Austria, Belgium and  Thailand.",
      "openfda": {
        "k_number": [
          "K900860",
          "K954501",
          "K902654",
          "K963698",
          "K093650",
          "K840749",
          "K864703",
          "K822000"
        ],
        "registration_number": [
          "1828100",
          "1422634",
          "1928237",
          "2184009",
          "2011171"
        ],
        "fei_number": [
          "1000116158",
          "1000514656",
          "1828100",
          "1928237",
          "2011171"
        ],
        "device_name": "Sensor, Blood-Gas, In-Line, Cardiopulmonary Bypass",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.4410",
        "device_class": "2"
      }
    }
  ]
}