  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "82561",
      "product_res_number": "Z-1582-2009",
      "event_date_initiated": "2009-04-09",
      "event_date_posted": "2009-07-07",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-08-27",
      "res_event_number": "52223",
      "product_code": "DRN",
      "k_numbers": [
        "K050172"
      ],
      "product_description": "Cardiomedics CardiAssist External Counterpulsation System, circulatory assist device, Series 4000.",
      "code_info": "CardioAssist Model 4000 Serial 4018",
      "firm_fei_number": "2025845",
      "recalling_firm": "Cardiomedics Incorporated",
      "address_1": "7 Whatney",
      "address_2": "Suite B",
      "city": "Irvine",
      "state": "CA",
      "postal_code": "92618-2839",
      "additional_info_contact": "949-863-2500",
      "reason_for_recall": "Cuff may underinflate, or device may stop working during use. Cardioassist software version  4.3.1 utilized a signal from the plethysmograph that was found not to be sensitive enough due to the shape of the waveform in certain circumstances where there was excessive motion artifact.",
      "root_cause_description": "Software design",
      "action": "Service visit to customer in May 2009 to upgrade and validate CardioAssist 4000 software to version 4.3.5.",
      "product_quantity": "1",
      "distribution_pattern": "Florida",
      "openfda": {
        "k_number": [
          "K010261",
          "K033617",
          "K022107",
          "K971026",
          "K023701",
          "K020857",
          "K152115",
          "K940264",
          "K173483",
          "K023016",
          "K081359",
          "K792430",
          "K030587",
          "K052611",
          "K022078",
          "K882401",
          "K792354",
          "K250756",
          "K202108",
          "K012141",
          "K050172",
          "K980937",
          "K030532",
          "K082134",
          "K033657",
          "K003469",
          "K872034",
          "K023427",
          "K190683",
          "K833627",
          "K130439",
          "K191955",
          "K021340",
          "K042413"
        ],
        "registration_number": [
          "3012470925",
          "3002806999",
          "2435300",
          "1319639",
          "3028606342",
          "3010564728",
          "3010135889",
          "3010145926"
        ],
        "fei_number": [
          "1000138054",
          "3012470925",
          "3002806999",
          "2435300",
          "3028606342",
          "3010564728",
          "3010135889",
          "3010145926"
        ],
        "device_name": "Device, Counter-Pulsating, External",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.5225",
        "device_class": "2"
      }
    }
  ]
}