  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "84239",
      "product_res_number": "Z-1145-2010",
      "event_date_initiated": "2009-07-14",
      "event_date_posted": "2010-03-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-03-18",
      "res_event_number": "52889",
      "product_code": "DHC",
      "k_numbers": [
        "K032868"
      ],
      "product_description": "BIO-RAD 425-2020 Autoimmune EIA Anti-Cardiolipin IgM, 96 Tests...2010 02 16...Lot CL-863.  Positive Control included in lot is identified as \"REF 425-2025\" and \"LOT CPM-591\". in vitro diagnostic.",
      "code_info": "Kit identified on label as \"REF 425-2020\" and \"LOT CL-863\".  Positive Control included in lot is identified as \"REF 425-2025\" and \"LOT CPM-591\".",
      "firm_fei_number": "2915274",
      "recalling_firm": "Bio-Rad Laboratories Inc",
      "address_1": "4000 Alfred Nobel Dr",
      "city": "Hercules",
      "state": "CA",
      "postal_code": "94547-1803",
      "additional_info_contact": "510-724-7000",
      "reason_for_recall": "The firm received customer complaints of microbial contamination in Positive Controls in its Anti-Cardiolipin IgM Test Kit Lot CL-863.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "Bio-Rad initiated its field correction on July 14, 2009.  The firm notified domestic consignees by phone followed by fax, and each subsidiary will be given an urgent notice, subsidiary response form, customer medical device correction notification, acknowledgement form and a listing of consignees.  The subsidiaries must contact customers as instructed.",
      "product_quantity": "213 kits",
      "distribution_pattern": "Worldwide distribution:  USA,  Australia, Canada, France, Italy, Korea, Mexico, Panama, Singapore, and Trinidad.",
      "openfda": {
        "k_number": [
          "K800605",
          "K872910",
          "K001398",
          "K821385",
          "K001352",
          "K810981",
          "K790106",
          "K914801",
          "K810226",
          "K885048",
          "K790257",
          "K782129",
          "K800991",
          "K930247",
          "K930421",
          "K800992"
        ],
        "registration_number": [
          "2242436",
          "2029372",
          "1832216",
          "1641328",
          "1721937"
        ],
        "fei_number": [
          "2242436",
          "1832216",
          "1641328",
          "1000117401",
          "1000306617"
        ],
        "device_name": "System, Test, Systemic Lupus Erythematosus",
        "medical_specialty_description": "Immunology",
        "regulation_number": "866.5820",
        "device_class": "2"
      }
    }
  ]
}