  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "150817",
      "product_res_number": "Z-0673-2017",
      "event_date_initiated": "2016-10-31",
      "event_date_created": "2016-11-29",
      "recall_status": "Terminated",
      "event_date_terminated": "2017-09-18",
      "res_event_number": "75569",
      "product_code": "CDS",
      "k_numbers": [
        "K032311"
      ],
      "product_description": "MSS CASSETTE GLU/LAC/UREA (BUN) as a part of the\nRoche OMNI S Analyzer: cobas b 221<6>Roche OMNI S6 system\nThe Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate",
      "code_info": "Lot Number(s): All lots within shelf life. Expiration Date(s): All",
      "firm_fei_number": "1823260",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "city": "Indianapolis",
      "state": "IN",
      "postal_code": "46256-1025",
      "additional_info_contact": "800-526-2272",
      "reason_for_recall": "Incorrect Urea results may be generated sporadically on cobas b 221<6>Roche OMNI S6 system and cannot be detected by QC. QC measurement results can be affected as well. This issue is only detectable if comparison result of a second measurement is available. The cause for this issue is an error in the software measurement procedure.\nIn children, decreased urea results can point at inborn disorders or urea cycle. Therefore, false normal urea values could lead to delayed/incorrect diagnosis of such a disorder.\nBesides the patients most at risk, the risk for the overall patient population with falsely low or high values is considered remote. No severe adverse health effects are expected; but, unnecessary diagnostic measures may be initiated.\nFalsely normal results may result in further necessary examination not being carried out, with the possibility of a delay in the diagnosis and treatment. A medical risk cannot entirely be excluded.",
      "root_cause_description": "Software design",
      "action": "Roche sent an Urgent Medical Correction letter dated October 31, 2016, to all affected customers.   The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.  \n\nActions Required\n\" Use the workaround outlined in this Urgent Medical Device Correction (UMDC) for Urea\nmeasurement until the updated software version is available.\n\" If your facility has distributed the affected product to another site, please ensure this UMDC is\nprovided to that site.\n\" Complete the attached fax form and fax it to 1-877-766-7471.\n\" File this UMDC for future reference.\n\nQuestions\nPlease contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272 if you have questions about the information contained in this UMDC.",
      "product_quantity": "97",
      "distribution_pattern": "Nationwide Distribution including ALCA CO CT FL IL IN KS MD MI MO NC NE NJ NM NY OH OK PA PR SC TX VA and WI",
      "openfda": {
        "device_name": "Electrode, Ion Specific, Urea Nitrogen",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1770",
        "device_class": "2",
        "k_number": [
          "K800071",
          "K033056",
          "K791018",
          "K811193",
          "K040973",
          "K760430",
          "K812308",
          "K860543",
          "K855091",
          "K800590",
          "K771060",
          "K171247",
          "K800616",
          "K925550",
          "K761061"
        ],
        "registration_number": [
          "2245578",
          "2050010",
          "3006198300",
          "3003795116",
          "3023162659",
          "3012817228",
          "9610126",
          "1219029",
          "3003030793",
          "3005643513",
          "3004959793",
          "3004102403",
          "2050012"
        ],
        "fei_number": [
          "2245578",
          "2050010",
          "3006198300",
          "3003795116",
          "3023162659",
          "3012817228",
          "3002806559",
          "1219029",
          "3003030793",
          "3005643513",
          "3004959793",
          "3004102403",
          "2050012"
        ]
      }
    }
  ]
}