  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-12",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "73479",
      "product_res_number": "Z-0030-2009",
      "event_date_initiated": "2008-08-25",
      "event_date_posted": "2008-10-02",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-12-10",
      "res_event_number": "49462",
      "product_code": "DQX",
      "k_numbers": [
        "K935997",
        "K945379"
      ],
      "product_description": "Platinum Plus\" LT , REF 46-601, UPN  M001466010, LT/0.018/180, Sterilized with ethylene oxide gas, Rx only, Legal Manufacturer Boston Scientific Corporation, One Boston scientific Place, Natick, MA 01760-1537, Made in USA: 8600 NW 41st Street, Miami, FL 33166-6202.  The Platinum Plus\" Guidewire facilitates placement of a catheter during diagnostic or interventional intravascular procedures. The wire can be torqued to facilitate navigation through tortuous arteries and/or avoid unwanted side branches.",
      "code_info": "Lot # 11574626",
      "firm_fei_number": "3002095335",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "1 Scimed Pl",
      "city": "Maple Grove",
      "state": "MN",
      "postal_code": "55311-1565",
      "additional_info_contact": "763-494-1700",
      "reason_for_recall": "Boston Scientific is initiating a voluntary recall of the Platinum Plus Guidewire (peripheral).  BSC has received reports that product may be labeled as a 260 CM long guidewire when the actual packaged device is a 180 CM long guidewire and the product may be labeled as a 180 CM long guidewire when the actual packaged device is a 260 CM long guidewire.  There is no expected potential for injury to",
      "root_cause_description": "Labeling mix-ups",
      "action": "A \"Boston Scientific Urgent  Voluntary Medical Device Recall or Correction - Immediate Action Required\" letter dated 8/25/08  was sent to each affected account (To The Hospital Field Action Contact).  The letter includes the customer steps for recall, a complete listing of all product batches/lots being recalled, the Reply Verification Tracking Form(s), and a return shipping label.  Contact Boston Scientific Corporation at 1-763-494-1700 for assistance.",
      "product_quantity": "47",
      "distribution_pattern": "USA:  \nCA\nFL\nGA\nIL\nIN\nLA\nMN\nMO\nNJ\nNV\nNY\nOH\nPA\nTN\nTX\nUT\nWA\nWI and\nWV\n\nOUS: France, Germany, Great Britain, Italy, and the Netherlands.",
      "openfda": {
        "k_number": [
          "K022762",
          "K171513",
          "K160785",
          "K003084",
          "K232570",
          "K130588",
          "K063819",
          "K141831",
          "K874776",
          "K220075",
          "K181353",
          "K890807",
          "K905639",
          "K122590",
          "K011906",
          "K902153",
          "K923696",
          "K241784",
          "K902755",
          "K141295",
          "K122468",
          "K955637",
          "K950535",
          "K895903",
          "K182457",
          "K900386",
          "K890505",
          "K780493",
          "K915114",
          "K925947",
          "K862567",
          "K935997",
          "K121776",
          "K913353",
          "K181647",
          "K841071",
          "K060551",
          "K180128",
          "K200120",
          "K962765",
          "K070410",
          "K153397",
          "K143587",
          "K950352",
          "K943896",
          "K992169",
          "K930622",
          "K970244",
          "K011968",
          "K911547",
          "K171897",
          "K003398",
          "K051374",
          "K981381",
          "K212268",
          "K142051",
          "K925074",
          "K950156",
          "K963320",
          "K201595",
          "K033321",
          "K040592",
          "K002346",
          "K864585",
          "K971281",
          "K240997",
          "K981936",
          "K935577",
          "K912293",
          "K032031",
          "K153443",
          "K942066",
          "K152991",
          "K063372",
          "K931327",
          "K040037",
          "K133230",
          "K182985",
          "K901347",
          "K904477",
          "K100351",
          "K861765",
          "K861606",
          "K972944",
          "K882321",
          "K931514",
          "K022813",
          "K905432",
          "K770977",
          "K101985",
          "K971086",
          "K073082",
          "K093397",
          "K113454",
          "K193194",
          "K833574",
          "K110241",
          "K120644",
          "K020821",
          "K122573",
          "K914032",
          "K181001",
          "K202543",
          "K973945",
          "K001198",
          "K945129",
          "K002907",
          "K182553",
          "K880722",
          "K223432",
          "K232574",
          "K970690",
          "K973494",
          "K990823",
          "K946240",
          "K101986",
          "K180795",
          "K080144",
          "K964352",
          "K941710",
          "K890207",
          "K855089",
          "K062192",
          "K192599",
          "K954552",
          "K150802",
          "K080563",
          "K961271",
          "K072552",
          "K173277",
          "K974773",
          "K083904",
          "K150465",
          "K934359",
          "K930091",
          "K760944",
          "K944677",
          "K193548",
          "K945379",
          "K993000",
          "K945934",
          "K211898",
          "K140485",
          "K211741",
          "K934474",
          "K112381",
          "K032615",
          "K240387",
          "K121745",
          "K970528",
          "K051062",
          "K922706",
          "K043422",
          "K083094",
          "K140695",
          "K905581",
          "K103057",
          "K912798",
          "K162842",
          "K934602",
          "K901899",
          "K083713",
          "K950216",
          "K133396",
          "K933334",
          "K974560",
          "K971254",
          "K152709",
          "K920884",
          "K213949",
          "K253262",
          "K213854",
          "K874202",
          "K931584",
          "K040825",
          "K922674",
          "K210235",
          "K260168",
          "K935998",
          "K924202",
          "K171912",
          "K972762",
          "K132545",
          "K030019",
          "K101648",
          "K040100",
          "K952860",
          "K871882",
          "K110009",
          "K170915",
          "K153685",
          "K031652",
          "K253146",
          "K220137",
          "K871492",
          "K912156",
          "K180830",
          "K172073",
          "K970376",
          "K170417",
          "K081021",
          "K032183",
          "K882804",
          "K922912",
          "K960755",
          "K123024",
          "K873942",
          "K914194",
          "K903923",
          "K120298",
          "K952562",
          "K873543",
          "K982876",
          "K964551",
          "K183467",
          "K092303",
          "K030549",
          "K952430",
          "K160018",
          "K171764",
          "K920891",
          "K061453",
          "K183062",
          "K925342",
          "K914581",
          "K042373",
          "K122659",
          "K053265",
          "K182315",
          "K140536",
          "K864787",
          "K921702",
          "K864058",
          "K151234",
          "K193653",
          "K781561",
          "K964611",
          "K860258",
          "K990271",
          "K192886",
          "K083706",
          "K955304",
          "K060449",
          "K213834",
          "K203529",
          "K261640",
          "K932801",
          "K083732",
          "K854511",
          "K170636",
          "K060454",
          "K893626",
          "K895899",
          "K915723",
          "K974241",
          "K971395",
          "K193571",
          "K240859",
          "K021243",
          "K241521",
          "K920824",
          "K161887",
          "K943522",
          "K012861",
          "K991646",
          "K002206",
          "K191464",
          "K904582",
          "K031678",
          "K081337",
          "K994358",
          "K141218",
          "K942382",
          "K953533",
          "K933517",
          "K955801",
          "K201690",
          "K950316",
          "K063210",
          "K914828",
          "K961445",
          "K163426",
          "K124007",
          "K914531",
          "K213868",
          "K141054",
          "K254137",
          "K250203",
          "K063695",
          "K201720",
          "K850827",
          "K961777",
          "K001033",
          "K222631",
          "K862360",
          "K780438",
          "K142415",
          "K173795",
          "K943358",
          "K101116",
          "K052347",
          "K980119",
          "K201834",
          "K120881",
          "K191502",
          "K970494",
          "K901224",
          "K163575",
          "K821190",
          "K822597",
          "K955551",
          "K260544",
          "K880332",
          "K935232",
          "K973845",
          "K810675",
          "K003394",
          "K150445",
          "K981326",
          "K922536",
          "K980360",
          "K221351",
          "K022357",
          "K852354",
          "K050964",
          "K020283",
          "K213315",
          "K250075",
          "K905502",
          "K914031",
          "K925381",
          "K955175",
          "K950069",
          "K871137",
          "K883579",
          "K946029",
          "K171997",
          "K251181",
          "K052783",
          "K951109",
          "K884490",
          "K834237",
          "K933466",
          "K781393",
          "K032050",
          "K253667",
          "K013024",
          "K772021",
          "K983346",
          "K790882",
          "K113498",
          "K874542",
          "K032411",
          "K032899",
          "K094062",
          "K020340",
          "K043398",
          "K002212",
          "K953023",
          "K163444",
          "K911460",
          "K131710",
          "K122089",
          "K173052",
          "K240818",
          "K863178",
          "K942333",
          "K050873",
          "K033708",
          "K122506",
          "K970938",
          "K963171",
          "K965140",
          "K943340",
          "K143316",
          "K831839",
          "K905548",
          "K862048",
          "K173235",
          "K172453",
          "K170684",
          "K860840",
          "K081187",
          "K950534",
          "K936032",
          "K925979",
          "K113162",
          "K923134",
          "K961107",
          "K860430",
          "K173423",
          "K880762",
          "K112674",
          "K880761",
          "K072431",
          "K882507",
          "K073162",
          "K101011",
          "K103101",
          "K221390",
          "K041740",
          "K033758",
          "K122469",
          "K010497",
          "K041531",
          "K954642",
          "K061670",
          "K061171",
          "K032129",
          "K243733",
          "K961917",
          "K043457",
          "K040140",
          "K851574",
          "K133155",
          "K894633",
          "K140608",
          "K213898",
          "K100470",
          "K950141",
          "K971719",
          "K200144",
          "K061984",
          "K820699",
          "K170933",
          "K884198",
          "K040867",
          "K080364",
          "K883880",
          "K982559",
          "K963702",
          "K882928",
          "K072705",
          "K982543",
          "K040345",
          "K053251",
          "K993257",
          "K242824",
          "K884895",
          "K915496",
          "K102211",
          "K083146",
          "K862409",
          "K191907",
          "K923240",
          "K910280",
          "K914863",
          "K163004",
          "K062186",
          "K935905",
          "K091825",
          "K212615",
          "K972142",
          "K934122",
          "K914986",
          "K251596",
          "K001736",
          "K093647",
          "K894940",
          "K951307",
          "K913631",
          "K031916",
          "K111485",
          "K011287",
          "K042996",
          "K221575",
          "K181351",
          "K152404",
          "K021285",
          "K250031",
          "K123792",
          "K082922",
          "K021228",
          "K870664",
          "K080765",
          "K962932",
          "K861574",
          "K031442",
          "K912878",
          "K121398",
          "K152231",
          "K003483",
          "K180784",
          "K112957",
          "K230418",
          "K171445",
          "K881787",
          "K852123",
          "K024021",
          "K093681",
          "K011986",
          "K090554",
          "K192783",
          "K910969",
          "K122856",
          "K960657",
          "K841926",
          "K070212",
          "K873972",
          "K984394",
          "K021990",
          "K171948",
          "K991152",
          "K091385",
          "K031864",
          "K950113",
          "K112745",
          "K854912",
          "K033929",
          "K982592",
          "K080637",
          "K950266",
          "K120863",
          "K161702",
          "K173532",
          "K071029",
          "K926457",
          "K771230",
          "K021638",
          "K943390",
          "K832607",
          "K081775",
          "K201760",
          "K112631",
          "K152740",
          "K041973",
          "K100093",
          "K130104",
          "K890404",
          "K850828",
          "K021219",
          "K013833",
          "K241702",
          "K021363",
          "K924009",
          "K972290",
          "K092965",
          "K930961",
          "K000011",
          "K141782",
          "K913328",
          "K925852",
          "K053187",
          "K950752",
          "K210548",
          "K930787",
          "K100710",
          "K915115",
          "K980294",
          "K052022",
          "K896379",
          "K221917",
          "K012534",
          "K041624",
          "K160643",
          "K131288",
          "K140673",
          "K110553",
          "K093515",
          "K042611",
          "K902152",
          "K894318",
          "K042338",
          "K030617",
          "K914030",
          "K070150",
          "K001251",
          "K033742",
          "K913178",
          "K922410",
          "K010511",
          "K884183",
          "K101339",
          "K050915",
          "K082873",
          "K032146",
          "K122795",
          "K002289",
          "K980977",
          "K042854",
          "K960563",
          "K955135",
          "K023300",
          "K240864",
          "K924558",
          "K123609",
          "K931577",
          "K090193",
          "K934339",
          "K972793",
          "K170532",
          "K230637",
          "K050756",
          "K130798",
          "K822586",
          "K924385",
          "K881897",
          "K983033",
          "K042264",
          "K253746",
          "K853999",
          "K945159",
          "K011905",
          "K771231",
          "K991898",
          "K865017",
          "K863138",
          "K123707",
          "K120119",
          "K203533",
          "K241510",
          "K943022",
          "K193126",
          "K082304",
          "K913074",
          "K031277",
          "K083145",
          "K983506",
          "K982272",
          "K991194",
          "K963214",
          "K802473",
          "K192454",
          "K073655",
          "K001454",
          "K111498",
          "K974713",
          "K080863",
          "K132623",
          "K961232",
          "K942453",
          "K100578",
          "K962572",
          "K914138",
          "K791276",
          "K941895",
          "K103377",
          "K875323",
          "K014075",
          "K252674",
          "K931171",
          "K992237",
          "K143252",
          "K133865",
          "K861115",
          "K090274",
          "K884647",
          "K141707",
          "K013092",
          "K850677",
          "K971815",
          "K884262",
          "K033542",
          "K840756",
          "K130766",
          "K961015",
          "K250552",
          "K854830",
          "K180726",
          "K261281",
          "K862983",
          "K171933",
          "K925804",
          "K071721",
          "K152497",
          "K922787",
          "K902612",
          "K042370",
          "K232578",
          "K143135",
          "K962266",
          "K965023",
          "K982083",
          "K971688",
          "K083738",
          "K954622",
          "K952067",
          "K093822",
          "K241962",
          "K142598",
          "K924987",
          "K172187",
          "K253847",
          "K120137",
          "K113716",
          "K890959",
          "K941485",
          "K946356",
          "K223791",
          "K041762",
          "K954913",
          "K111395",
          "K151244",
          "K880287",
          "K040786",
          "K802260",
          "K100425",
          "K151840",
          "K141339",
          "K880927",
          "K872220",
          "K240460",
          "K212167",
          "K902207",
          "K897152",
          "K251385",
          "K973708",
          "K230377",
          "K023700",
          "K840391",
          "K002962",
          "K091582",
          "K111488",
          "K142393",
          "K160594",
          "K232662",
          "K860533",
          "K052339",
          "K860841",
          "K905423",
          "K940984",
          "K083727",
          "K990639",
          "K790258",
          "K834223",
          "K941229",
          "K864316",
          "K935170",
          "K190176",
          "K070945",
          "K946180",
          "K950835",
          "K082094",
          "K922060",
          "K053268",
          "K110479",
          "K191687",
          "K915554",
          "K082935",
          "K935022",
          "K110163",
          "K102818",
          "K011084",
          "K161584",
          "K072460",
          "K871850",
          "K242597",
          "K111288",
          "K123067",
          "K021323",
          "K960718",
          "K880288",
          "K922302",
          "K871517",
          "K153106",
          "K170700",
          "K953760",
          "K002707",
          "K843012",
          "K100930",
          "K142254",
          "K884274",
          "K873403",
          "K983771",
          "K930982",
          "K971322",
          "K101729",
          "K862937",
          "K925833"
        ],
        "registration_number": [
          "1000523114",
          "3002808486",
          "3030447506",
          "3027260597",
          "",
          "3010273872",
          "9614279",
          "2184149",
          "3007146453",
          "3010163695",
          "9611446",
          "2248146",
          "2183870",
          "3006950086",
          "3050539341",
          "3003898360",
          "9616684",
          "2011171",
          "9616088",
          "3004091615",
          "3003768277",
          "8010026",
          "9614758",
          "3008114965",
          "3011706110",
          "3015124694",
          "3001124136",
          "2017865",
          "1423537",
          "3010432890",
          "1055236",
          "1644312",
          "3006283319",
          "3010532612",
          "3015910259",
          "1028232",
          "3020283264",
          "3011137372",
          "3012931345",
          "1043214",
          "2024024",
          "3011642792",
          "3007830707",
          "1061124",
          "3015173212",
          "3006010712",
          "3009546466",
          "2134285",
          "3014246639",
          "3035642068",
          "3009600098",
          "3018784828",
          "3013556777",
          "3018070217",
          "9680794",
          "3008403546",
          "3013666336",
          "3008264254",
          "1118880",
          "3009121749",
          "3011469355",
          "9681477",
          "9616662",
          "9617592",
          "3004111573",
          "3012536737",
          "2133928",
          "2022435",
          "3005941719",
          "3016119728",
          "2953200",
          "3010882065",
          "3013595987",
          "3014656749",
          "2527072",
          "3017636737",
          "3009564766",
          "2528981",
          "3024682301",
          "2029214",
          "2939520",
          "1047429",
          "3005737652",
          "1824619",
          "3004859241",
          "1721676",
          "2518902",
          "3014162263",
          "3002917664",
          "2025587",
          "3010041511",
          "3008061490",
          "3015859709",
          "3007738736",
          "3009211636",
          "3031768596",
          "3006425876",
          "3042175841",
          "3015489752",
          "3030446844",
          "2020394",
          "1000563940",
          "3003289691",
          "3008853977",
          "3009526575",
          "3016678045",
          "2029275",
          "3003775027",
          "3012050423",
          "3004605321",
          "3004519921",
          "1123137",
          "1319639",
          "9612812",
          "3009756153",
          "9680001",
          "3014579161",
          "1724474",
          "3013943846",
          "3002244583",
          "3017961114",
          "3006082230",
          "2134812",
          "1625425",
          "2024311",
          "1225687",
          "1221806",
          "3039642752",
          "3003915875",
          "1000121056",
          "3015531723",
          "3015225571",
          "3006891611",
          "3008988055",
          "2246552",
          "3009092139",
          "3007695959",
          "2939561",
          "3006946276",
          "3039020025",
          "3012154226",
          "3014279513",
          "3007284006",
          "2126666",
          "3002807314",
          "3008837339",
          "3015453963",
          "3013656790",
          "3013750125",
          "2134244",
          "1016427",
          "3011564431",
          "2521402",
          "1018233",
          "2030624",
          "1064858",
          "3009888344",
          "3009039068",
          "3020968442",
          "1820334",
          "2124215",
          "3017210488",
          "3007138831",
          "3007635982",
          "9681834",
          "3000247873",
          "9610139",
          "2024168",
          "3009380063",
          "3009957947",
          "9681260",
          "9611665",
          "3012223769",
          "3010155661",
          "3009018440",
          "3041276591",
          "3005334138",
          "3004365956",
          "3010665433",
          "1220477",
          "1223004",
          "3007064586",
          "1282497",
          "3013557562",
          "3008439199",
          "3030452144",
          "3035372913",
          "3008307705",
          "1721504",
          "1048735",
          "3008002401",
          "8023054",
          "1319211",
          "2134265",
          "3012497308",
          "1450662",
          "3026317086",
          "3000126629",
          "1417485",
          "1649395",
          "3013162291",
          "2135147",
          "3030733800",
          "3003084417",
          "3008363989",
          "3013023134",
          "3033589330",
          "3030574705",
          "9610849",
          "3000268902",
          "1030451",
          "3011050570",
          "1018470",
          "3007361214",
          "1720929",
          "3014716786",
          "3014499739",
          "3010419931",
          "3003780911",
          "3004748528",
          "1928237",
          "1226348",
          "3005012805",
          "3007628272",
          "3008627763",
          "2219920",
          "3015309643",
          "3013288201",
          "3036802499",
          "2182208",
          "3013764800",
          "2183744",
          "1422634",
          "3014977413",
          "3004784537",
          "2245270",
          "1313046",
          "3024743081",
          "3020778",
          "3015522130",
          "1721686",
          "3009490946",
          "1293141",
          "3014590708",
          "1220452",
          "3019751610",
          "3005580113",
          "3014461810",
          "2523676",
          "1000393132",
          "3007113487",
          "3010131137",
          "8043892",
          "3013611763",
          "1054811",
          "3015550451",
          "1054241",
          "3012159165",
          "1649869",
          "3010291427",
          "3014222368",
          "1222301",
          "3020954746"
        ],
        "fei_number": [
          "1000523114",
          "3002808486",
          "3001451463",
          "3030447506",
          "3027260597",
          "3010273872",
          "",
          "2828",
          "3007146453",
          "3010163695",
          "3003767881",
          "3008397795",
          "3002806565",
          "2183870",
          "3006950086",
          "3050539341",
          "3003898360",
          "3002807725",
          "2011171",
          "3001632632",
          "3004091615",
          "3002807095",
          "3000206435",
          "3008114965",
          "3011706110",
          "3015124694",
          "3001124136",
          "2017865",
          "3010432890",
          "3002827704",
          "3006283319",
          "3010532612",
          "3015910259",
          "1028232",
          "3020283264",
          "3012931345",
          "1043214",
          "2024024",
          "3011642792",
          "3007830707",
          "3015173212",
          "3006010712",
          "1000117748",
          "3009546466",
          "3014246639",
          "3035642068",
          "3009600098",
          "3018784828",
          "3013556777",
          "3018070217",
          "3002902735",
          "3006097687",
          "3008403546",
          "3003094851",
          "3003737899",
          "3013666336",
          "3008264254",
          "1118880",
          "3009121749",
          "3011469355",
          "3002710305",
          "3004111573",
          "3012536737",
          "2022435",
          "3016119728",
          "3013595987",
          "3014656749",
          "3001452571",
          "2527072",
          "3017636737",
          "3009564766",
          "2528981",
          "3024682301",
          "2939520",
          "1047429",
          "3005737652",
          "3004635528",
          "1824619",
          "3004859241",
          "1721676",
          "2518902",
          "3014162263",
          "3004216185",
          "3002917664",
          "3002808327",
          "2025587",
          "3010041511",
          "3008061490",
          "3015859709",
          "3007738736",
          "3009211636",
          "3002095335",
          "3031768596",
          "3006425876",
          "3001236905",
          "3015489752",
          "3030446844",
          "3042175841",
          "1000306431",
          "2020394",
          "1000563940",
          "1000139754",
          "1000514656",
          "3003289691",
          "3008853977",
          "3009526575",
          "3016678045",
          "3003775027",
          "3004605321",
          "3012050423",
          "3000717902",
          "3014579161",
          "3017961114",
          "3009756153",
          "3002244583",
          "1000121071",
          "3013943846",
          "3006082230",
          "1625425",
          "2024311",
          "1221806",
          "3002468371",
          "3039642752",
          "3003915875",
          "3002806500",
          "1000121056",
          "3015531723",
          "3015225571",
          "3006891611",
          "3008988055",
          "3002807391",
          "2246552",
          "3009092139",
          "3005747797",
          "3007695959",
          "3006946276",
          "3039020025",
          "3012154226",
          "3014279513",
          "3007284006",
          "2126666",
          "3000718472",
          "3002807314",
          "3008837339",
          "3015453963",
          "3013656790",
          "3013750125",
          "1016427",
          "3011564431",
          "2521402",
          "1018233",
          "3002947440",
          "3009888344",
          "3009039068",
          "1000114185",
          "3020968442",
          "1820334",
          "2124215",
          "3017210488",
          "1000123180",
          "1000112877",
          "1000120555",
          "3007138831",
          "3007635982",
          "3000247873",
          "2024168",
          "3002808050",
          "3009380063",
          "3009957947",
          "3004365956",
          "3010155661",
          "3009018440",
          "3041276591",
          "3005334138",
          "3010665433",
          "1220477",
          "3007064586",
          "1282497",
          "3013557562",
          "3008439199",
          "3030452144",
          "3035372913",
          "3008307705",
          "1000222374",
          "1048735",
          "1721504",
          "3008002401",
          "1319211",
          "3002807439",
          "1000112968",
          "3012497308",
          "1450662",
          "3026317086",
          "3000126629",
          "1417485",
          "3002808445",
          "3013162291",
          "1000512168",
          "3030733800",
          "3003084417",
          "3005846326",
          "3002287318",
          "3008363989",
          "3013023134",
          "1000220862",
          "3030574705",
          "3033589330",
          "3000268902",
          "3002807111",
          "1030451",
          "3011050570",
          "1018470",
          "3007361214",
          "1720929",
          "3014716786",
          "3014499739",
          "3010419931",
          "3003780911",
          "3004748528",
          "1928237",
          "3005012805",
          "3007628272",
          "3008627763",
          "2219920",
          "3015309643",
          "3013288201",
          "3036802499",
          "2182208",
          "3013764800",
          "2183744",
          "3014977413",
          "3004784537",
          "2245270",
          "1313046",
          "3024743081",
          "3020778",
          "3015522130",
          "1721686",
          "3009490946",
          "3002807880",
          "1293141",
          "3014590708",
          "1220452",
          "3019751610",
          "3005580113",
          "1000138054",
          "3014461810",
          "3008411636",
          "2523676",
          "3002913962",
          "1000393132",
          "3007113487",
          "3010131137",
          "3013611763",
          "3000335778",
          "3003407487",
          "3015550451",
          "3012159165",
          "3010291427",
          "3014222368",
          "3002808331",
          "3020954746"
        ],
        "device_name": "Wire, Guide, Catheter",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1330",
        "device_class": "2"
      }
    }
  ]
}