  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "95738",
      "product_res_number": "Z-0445-2011",
      "event_date_initiated": "2010-11-03",
      "event_date_posted": "2010-11-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-12-07",
      "res_event_number": "57282",
      "product_code": "OGX",
      "k_numbers": [
        "K071529"
      ],
      "product_description": "Lumen Biomedical Xtract Aspiration Catheter system, intravascular catheter,  consisting of One Catheter, Two 30 mL Syringes, One Extension Tube with Stopcock, and One 40 um Cell Strainer.  Single-user, 0.014\" guidewire compatible intravascular extraction and aspiration catheter, Part number 1450-0001 size 4.2F, and part number 1451-003 size 5.8F,  distributed by Volcano Corporation.",
      "code_info": "Part number 1450-001 size 4.2F and 1451-003 size 5.8F, all lots.",
      "firm_fei_number": "2939520",
      "recalling_firm": "Volcano Corporation",
      "address_1": "2870 Kilgore Rd",
      "city": "Rancho Cordova",
      "state": "CA",
      "postal_code": "95670-6133",
      "additional_info_contact": "Nancy Lelicoff\n916-638-8008 Ext. 4503",
      "reason_for_recall": "Potential for product failure. Packaging/Storage issue results in compromised integrity and potential product failure during use, leading to potential patient injury.",
      "root_cause_description": "Device Design",
      "action": "Firm initiated recall on 11/03/2010, and sent notifications to its consignees via letter.  The firm has requested that consignees provide the notification to all personnel within the organization and letters will be sent directly to known customers on 11/04/2010 by Volcano.  Customers have been instructed to cease use and contact customer service to arrange returns, and credit will be issued.",
      "product_quantity": "4724 units",
      "distribution_pattern": "Worldwide distribution: USA, Greece, Egypt, Kuwait, Netherlands, Saudi Arabia, Italy, France, Sweden, Portugal, Great Britain, Austria, Switzerland, Germany, Romania, Poland, and Bulgaria",
      "openfda": {}
    }
  ]
}