  "results": [
    {
      "cfres_id": "95276",
      "product_res_number": "Z-0389-2011",
      "event_date_initiated": "2010-10-07",
      "event_date_posted": "2010-11-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2010-12-17",
      "res_event_number": "57052",
      "product_code": "DWF",
      "k_numbers": [
        "K881330"
      ],
      "product_description": "Sorin OptiPack, Perfusion Tubing Set, Sterile EO, REF 084504300, Manufactured by Sorin Group USA, Inc. 14401 W. 65th Way Arvada, CO 80004\n\nIndicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours.",
      "code_info": "Lot No. 1021400039",
      "firm_fei_number": "1718850",
      "recalling_firm": "Sorin Group USA, Inc.",
      "address_1": "14401 W 65th Way",
      "city": "Arvada",
      "state": "CO",
      "postal_code": "80004",
      "additional_info_contact": "303-467-6306",
      "reason_for_recall": "Defective product packaging may compromise product sterility.",
      "root_cause_description": "Packaging process control",
      "action": "Sorin Group sent an \"URGENT: FIELD SAFETY NOTICE\" letter dated October 7, 2010, to all customers.  The letter described the product, problem and action to be taken by customer.  The customers were instructed to check all affected inventory to exam for holes.  Any damaged products were to be returned to Sorin.  If no damage was found, product could be used.  The customers were also instructed to contact Sorin Cardiopulmonary Customer Service at 1-800-650-2623 to return product and receive replacement; they were to share this information with all personnel who may use this product; and to complete and return the attached Customer Response Form via the self-addressed prepaid envelope provided or fax to (303) 467-6502 as soon as possible.\n\nFor further information, contact 1-800-650-2623.",
      "product_quantity": "18",
      "distribution_pattern": "Nationwide distribution: CA, FL, IL, IN, KS, MD, MO, NE, NJ, NM, TN, TX, and VA.",
      "openfda": {
        "k_number": [
          "K250937",
          "K880779",
          "K881352",
          "K971023",
          "K955121",
          "K883412",
          "K140165",
          "K834134",
          "K152161",
          "K901074",
          "K880421",
          "K110980",
          "K171979",
          "K970914",
          "K802261",
          "K890024",
          "K020693",
          "K081820",
          "K033264",
          "K801397",
          "K990772",
          "K880180",
          "K840914",
          "K090878",
          "K023880",
          "K131269",
          "K854640",
          "K001371",
          "K010737",
          "K062597",
          "K844140",
          "K043179",
          "K801396",
          "K964195",
          "K832124",
          "K895352",
          "K030696",
          "K014303",
          "K930684",
          "K791832",
          "K896777",
          "K022272",
          "K052081",
          "K910021",
          "K972261",
          "K180979",
          "K801770",
          "K151523",
          "K991170",
          "K865079",
          "K905638",
          "K133445",
          "K971021",
          "K253671",
          "K993677",
          "K872043",
          "K964199",
          "K162500",
          "K831332",
          "K203067",
          "K801403",
          "K821316",
          "K884393",
          "K813037",
          "K872788",
          "K854487",
          "K081560",
          "K801392",
          "K881330",
          "K861310",
          "K842375",
          "K861096",
          "K092509",
          "K905312",
          "K883221",
          "K863342",
          "K040173",
          "K141951",
          "K132451",
          "K896636",
          "K110983",
          "K964194",
          "K031037",
          "K020515",
          "K201100",
          "K931586",
          "K123762",
          "K983791",
          "K001961",
          "K934127",
          "K955818",
          "K133293",
          "K203057",
          "K123303",
          "K260043",
          "K981009",
          "K861496",
          "K033416",
          "K845045",
          "K831130",
          "K983270",
          "K200289",
          "K871844",
          "K162214",
          "K120612",
          "K872762",
          "K034058",
          "K971291",
          "K873212",
          "K923483",
          "K810820",
          "K822317",
          "K113459",
          "K961245",
          "K964205",
          "K791947",
          "K897026",
          "K060101",
          "K833830",
          "K123298",
          "K870421",
          "K790565",
          "K991541",
          "K874896",
          "K180456",
          "K890980",
          "K153598",
          "K801359",
          "K860988",
          "K001509",
          "K012538",
          "K091304",
          "K964374",
          "K897022",
          "K102420",
          "K841482",
          "K850925",
          "K113869",
          "K844139",
          "K212386",
          "K994243",
          "K031518",
          "K120988",
          "K941166",
          "K834039",
          "K953945",
          "K990396",
          "K013013",
          "K031827",
          "K791858",
          "K964193",
          "K012617",
          "K833112",
          "K882034",
          "K843542",
          "K852984",
          "K831759",
          "K132995",
          "K991066",
          "K024069",
          "K123370",
          "K925442",
          "K994171",
          "K023622",
          "K052524",
          "K874853",
          "K833047",
          "K964391",
          "K042984",
          "K920902",
          "K881691",
          "K951088",
          "K981995",
          "K894684",
          "K850917",
          "K961927",
          "K140569",
          "K870819",
          "K032977",
          "K821149",
          "K974736",
          "K891576",
          "K063618",
          "K002857",
          "K994209",
          "K881072",
          "K200612",
          "K100274",
          "K061254",
          "K082344",
          "K052570",
          "K974259",
          "K253106",
          "K002578",
          "K032041",
          "K845046",
          "K091542",
          "K131666",
          "K182302",
          "K020784",
          "K860136",
          "K072172",
          "K860497",
          "K781990",
          "K884310",
          "K013593",
          "K922471",
          "K881733",
          "K132811",
          "K230960",
          "K860743",
          "K000830",
          "K950814",
          "K122913",
          "K844137",
          "K022174",
          "K770431",
          "K961717",
          "K790246",
          "K000776",
          "K033464",
          "K901231",
          "K213540",
          "K231206",
          "K861116",
          "K930593",
          "K822210",
          "K792070",
          "K003371",
          "K032632",
          "K896338",
          "K932208",
          "K964373",
          "K082425",
          "K910004",
          "K874899",
          "K143083",
          "K862190",
          "K152190",
          "K120987",
          "K031776",
          "K925049",
          "K943934",
          "K972665",
          "K861322",
          "K861103",
          "K052372",
          "K875048",
          "K971928",
          "K802378",
          "K964248",
          "K141465",
          "K102532",
          "K792202",
          "K002273",
          "K972993",
          "K882771",
          "K142593",
          "K844138",
          "K870109",
          "K902277",
          "K233895",
          "K831591",
          "K861790",
          "K834025",
          "K924529",
          "K190557",
          "K964200",
          "K250610",
          "K140192",
          "K915268",
          "K022042",
          "K022321",
          "K033463",
          "K081881",
          "K964198",
          "K010974",
          "K860149",
          "K013944",
          "K111147",
          "K240193",
          "K880103",
          "K012774",
          "K090869",
          "K081933",
          "K915193",
          "K200499",
          "K971041",
          "K002019",
          "K810548",
          "K150536",
          "K113411",
          "K961594",
          "K972503",
          "K922992",
          "K171308",
          "K832475",
          "K022280",
          "K190594",
          "K020768",
          "K140999",
          "K770068",
          "K150422",
          "K850385",
          "K860742",
          "K143107",
          "K863740",
          "K031151",
          "K801401",
          "K992932",
          "K981601",
          "K875353",
          "K900753",
          "K041650",
          "K260195",
          "K902394",
          "K771499",
          "K810515",
          "K142673",
          "K961180",
          "K162215",
          "K802543",
          "K961270",
          "K924642",
          "K770429",
          "K831769",
          "K120780",
          "K093730",
          "K000274",
          "K023251",
          "K250283",
          "K160257",
          "K930620",
          "K791972",
          "K022022",
          "K133151",
          "K120543",
          "K874834",
          "K241248",
          "K881734",
          "K964393",
          "K801398",
          "K840002",
          "K042156",
          "K002366",
          "K881074",
          "K182586",
          "K851323",
          "K810415",
          "K880887",
          "K001565",
          "K030231",
          "K771692",
          "K941916",
          "K020983",
          "K123187",
          "K130639",
          "K842889",
          "K905224",
          "K251258",
          "K024058",
          "K170488",
          "K870507",
          "K770430",
          "K872033",
          "K862634",
          "K792152",
          "K880777",
          "K221353",
          "K843444",
          "K863558",
          "K971019",
          "K831330",
          "K850356",
          "K140208",
          "K840001",
          "K872159",
          "K924643",
          "K883981",
          "K120072",
          "K112565",
          "K023056",
          "K853128",
          "K083301",
          "K253998",
          "K180453",
          "K885188",
          "K133236",
          "K830728",
          "K090533",
          "K821445",
          "K874852",
          "K150567",
          "K993818",
          "K200358",
          "K181150",
          "K896778",
          "K080034",
          "K962835",
          "K253203",
          "K061832"
        ],
        "registration_number": [
          "",
          "9611109",
          "3009493875",
          "3005173255",
          "3010163695",
          "3009211636",
          "2244478",
          "1222928",
          "1450662",
          "2030598",
          "1054241",
          "1724474",
          "3009380063",
          "3011427673",
          "3010531069",
          "1722746",
          "2528981",
          "2320762",
          "3018503755",
          "3002807090",
          "3011468686",
          "1928237",
          "1720929",
          "9680841",
          "1319211",
          "1221051",
          "3030125051",
          "1721676",
          "1836324",
          "1218444",
          "2024311",
          "2011171",
          "3017059666",
          "1723241",
          "3031571797",
          "1921846",
          "2032228",
          "3004080795",
          "1124841",
          "1000393132",
          "1423662",
          "3015173212",
          "2184009",
          "3007829657",
          "3008500478",
          "3009420598",
          "8020347",
          "3021632375",
          "1519227",
          "3003955307",
          "3019807891",
          "2648988",
          "1417592",
          "1123137",
          "3008998256",
          "3003418325",
          "8010762",
          "1929756",
          "1000523114",
          "1423537",
          "1649914",
          "3005448067",
          "3044156918",
          "1643116",
          "1220948",
          "1043214",
          "1643817",
          "3017987980",
          "2916596",
          "3017540705",
          "3007681502",
          "3011554160",
          "3003752502",
          "1450019",
          "3014479313",
          "1718850",
          "1647149",
          "2183744",
          "1047843",
          "3013267343",
          "3016746283",
          "1424263",
          "9617601",
          "2029275",
          "3010155661",
          "2020394",
          "1055236",
          "3006367242",
          "1649139",
          "8040510",
          "3005941719",
          "1061927",
          "3007421149",
          "3014990448",
          "3038776570",
          "1319639",
          "1422634",
          "9680721",
          "1831948",
          "3010017366",
          "9611665",
          "1220648",
          "1820334",
          "1625519"
        ],
        "fei_number": [
          "",
          "3009493875",
          "3005173255",
          "3003015552",
          "3010163695",
          "3009211636",
          "1450662",
          "1000518731",
          "3009380063",
          "3011427673",
          "1000306431",
          "3010531069",
          "2528981",
          "3002807095",
          "3018503755",
          "3011468686",
          "1928237",
          "1000116158",
          "1720929",
          "1319211",
          "1221051",
          "3002534070",
          "3030125051",
          "1000138054",
          "3005426550",
          "1721676",
          "1218444",
          "2024311",
          "2011171",
          "3017059666",
          "3003166981",
          "3031571797",
          "1000514656",
          "1921846",
          "3004080795",
          "1000393132",
          "3015173212",
          "3007829657",
          "3003040625",
          "3000203000",
          "3009420598",
          "3000215273",
          "3021632375",
          "3002808267",
          "1519227",
          "3003955307",
          "3019807891",
          "1417592",
          "1000112877",
          "3003094851",
          "3008998256",
          "3003418325",
          "3006003617",
          "1929756",
          "1000523114",
          "1000306139",
          "1000306216",
          "3005448067",
          "3044156918",
          "3001236905",
          "1643116",
          "1220948",
          "1043214",
          "1643817",
          "3005612924",
          "3017987980",
          "2916596",
          "3017540705",
          "1000517719",
          "3001665800",
          "3001319476",
          "3007681502",
          "3011554160",
          "3003752502",
          "3002998595",
          "1450019",
          "3014479313",
          "1718850",
          "2183744",
          "1047843",
          "3013267343",
          "3016746283",
          "3000204839",
          "1000120613",
          "3002406063",
          "3010155661",
          "2020394",
          "3000204775",
          "3006367242",
          "1000123180",
          "1000112968",
          "1000244733",
          "3007421149",
          "3009649876",
          "3002807111",
          "3014990448",
          "3038776570",
          "1000136269",
          "1831948",
          "1713910",
          "3010017366",
          "1220648",
          "1820334",
          "1625519"
        ],
        "device_name": "Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.4210",
        "device_class": "2"
      }
    }
  ]
}