  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "94758",
      "product_res_number": "Z-0360-2011",
      "event_date_initiated": "2009-08-28",
      "event_date_posted": "2010-11-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-11-16",
      "res_event_number": "56873",
      "product_code": "LNX",
      "product_description": "SoftID.NET versions 2.0.0.6 - 2.0.0.19.  \nSCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760.  \n\nSoftID.NET is to be used in a medical or clinical laboratory setting by knowledgeable, trained, and experienced professionals. SoftID.NET supports bedside collection of patient specimens used in diagnostic testing. This product is integrated with Soft Computer's SoftLab, SoftScape and can also interface with all other SCC laboratory systems and Hospital Information Systems (HIS).",
      "code_info": "Versions: SoftID.NET 2.0.0.6 - 4/3/2007,       SoftID.NET 2.0.0.7 - 6/21/2007,    SoftID.NET 2.0.0.8 - 8/29/2007,      SoftID.NET 2.0.0.9 - 10/01/2007,    SoftID.NET 2.0.0.10 - 10/25/2007,    SoftID.NET 2.0.0.11 - 11/07/2007,    SoftID.NET 2.0.0.12 - 2/08/2008,    SoftID.NET 2.0.0.13 - 5/13/2008,    SoftID.NET 2.0.0.14 - 6/12/2008,    SoftID.NET 2.0.0.15 - 8/15/2008,    SoftID.NET 2.0.0.16 - 10/15/2008,    SoftID.NET 2.0.0.17 - 12/15/2008,    SoftID.NET 2.0.0.18 - 2/02/2009,    SoftID.NET 2.0.0.19 -  5/18/2009.",
      "firm_fei_number": "1000307651",
      "recalling_firm": "SCC Soft Computer",
      "address_1": "5400 Tech Data Drive",
      "city": "Clearwater",
      "state": "FL",
      "postal_code": "33760",
      "additional_info_contact": "727-789-0100",
      "reason_for_recall": "SCC Soft Computer issued a Correction for SoftID.NET versions 2.0.0.6 - 2.0.0.19.  When using the Cancel functionality, if the collection list refreshes during the transaction and changes the position of the patient on the list, the wrong patient order can be cancelled.  All affected clients have been notified.",
      "root_cause_description": "Software design",
      "action": "SCC Soft Computer notified Customers on August 28, 2009, via the proprietary communication software entitled \"Risk-to-Health Notification Task Verbiage\"  dated August 17, 2009.  The notification described the product, problem and action to be taken by customer. The customers were instructed to acknowledge receipt of the notification; document a request to load the software, if they wish to receive the correction; document that the Alternative Working Solution will be used by the customers facility until the correction has been taken live.\n\nIf you have any questions, contact (727) 789-0100.",
      "product_quantity": "41",
      "distribution_pattern": "Nationwide distribution: AR, CA, FL, IL, IN, KS,LA, MA, MI, MN, MO, NC, ND,NH, NJ, NY, OH, PA, RI, TN, TX, VA and WI.",
      "openfda": {
        "k_number": [
          "K921846"
        ]
      }
    }
  ]
}