  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "94639",
      "product_res_number": "Z-0088-2011",
      "event_date_initiated": "2010-07-19",
      "event_date_posted": "2010-10-19",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-07-16",
      "res_event_number": "56851",
      "product_code": "LNX",
      "product_description": "SoftMic Versions 4.0.4.7, 4.5.1.3.  \nVersion 4.0.4.7 Manufactured 02/03/2010.  \nVersion 4.5.1.3 Manufactured 03/18/2010.  \nSCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760.",
      "code_info": "Versions 4.0.4.7, 4.5.1.3.",
      "firm_fei_number": "1000307651",
      "recalling_firm": "SCC Soft Computer",
      "address_1": "5400 Tech Data Drive",
      "city": "Clearwater",
      "state": "FL",
      "postal_code": "33760",
      "additional_info_contact": "Kathryn Branca\n727-789-0100",
      "reason_for_recall": "Softmic system may delay or omit reporting of clinically significant results including organisms and drug sensitivities.  Incomplete culture results may appear complete. SCC Soft Computer is issuing a Correction for SoftMic Versions 4.0.4.7-4.5.1.3 software due to this potential hazard.",
      "root_cause_description": "Software design",
      "action": "SCC Soft Computer issued Risk-to-Health notifications to clients on July 19, 2010 via proprietary communication software.  Corrective actions will be via programming correction via hot fix (software modification) or such so that Global variable is not used.  Clients are asked to acknowledge receipt of this task and update the task with permission to load the required correction.",
      "product_quantity": "35",
      "distribution_pattern": "Worldwide Distribution: USA, including the states of CA, CT, DE, FL, IA, KY, MA, MA, MI, MN, NJ, NY, OH, PA, TX, VA, and WI, and the country of Canada.",
      "openfda": {
        "k_number": [
          "K921846"
        ]
      }
    }
  ]
}