  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "93123",
      "product_res_number": "Z-0069-2011",
      "event_date_initiated": "2007-10-05",
      "event_date_posted": "2010-10-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2010-10-19",
      "res_event_number": "56278",
      "product_code": "LNX",
      "product_description": "Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1.\n\nSCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760.  \n\nManufacture Dates:\n4.0.2.38.1 06/15/07  \n4.02.39.2 06/29/07, \n4.02.40.1 07/19/07, \n4.02.41.1 08/20/07.  \n\nThe Result Reporting Interface is a highly versatile interface designed to electronically convey clinical laboratory orders and results for the SCC applications to other recipient information systems.",
      "code_info": "Versions 4.0.2.38.1, 4.0.2.39.2, 4.02.40.1, and 4.0.2.41.1.",
      "firm_fei_number": "1000307651",
      "recalling_firm": "SCC Soft Computer",
      "address_1": "5400 Tech Data Drive",
      "city": "Clearwater",
      "state": "FL",
      "postal_code": "33760",
      "additional_info_contact": "727-789-0100",
      "reason_for_recall": "SCC Soft Computer, Clearwater, FL issued a Correction for Hospital Information System Results Reporting Interface Versions 4.0.2.38.1, 4.0.2.39.2, 4.0.2.40.1, and 4.0.2.41.1 on 10/05/2007.  A software anomaly in the result reporting interface caused result comments entered after an empty line  not to be sent to the HIS system.  This included interpretative comments, results, and normal range c",
      "root_cause_description": "Software design",
      "action": "SCC Soft Computer sent a RISK-TO-HEALTH NOTIFICATION TASK VERBIAGE dated October 5, 2007, through their Task Management System to all affected clients on October 17, 2007.  The notice identified the product, the problem and the action clients should take to correct the problem.  \nClients were to acknowledge receipt of the task notice and to identify if the Workflow or functionality is not used and if they Agree to use alternative working solutions.  Each client was asked to grant permission to load the required software to correct the issue.\nFor any questions regarding this recall call (727) 789-0100.",
      "product_quantity": "35",
      "distribution_pattern": "Worldwide Distribution - USA, including AL, CO, FL, GA, IN, KS, KY, MA, MO, NC, NH, NJ, NY, OH, PA, TN, and WA and the country of Canada",
      "openfda": {
        "k_number": [
          "K921846"
        ]
      }
    }
  ]
}