  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "93117",
      "product_res_number": "Z-0067-2011",
      "event_date_initiated": "2007-03-28",
      "event_date_posted": "2010-10-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2010-10-19",
      "res_event_number": "56275",
      "product_code": "LNX",
      "product_description": "SoftPath GUI Releases 3.17 through 4.1.0.\n\nSCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760.\n\nManufacture/Distribution Dates:\nVersion 3.17.5 10/14/2003.  \nVersion 3.17.6 03/28/2003.  \nVersion 3.17.7 07/02/2003.  \nVersion 3.17.8 03/03/2004.  \nVersion 4.1.0 07/03/2003.  \nVersion 4.1.1 10/03/2003.  \n\nIntended for use in the Pathology Lab for receiving and accessioning specimens, documenting tissue processing, producing material labels, generating final diagnostic reports, and creating billing and management reports for surgical, GYN, NON-GYN, and Autopsy cases.",
      "code_info": "Versions 3.17.5,  3.17.6, 3.17.7, 3.17.8, 4.1.0, and 4.1.1.",
      "firm_fei_number": "1000307651",
      "recalling_firm": "SCC Soft Computer",
      "address_1": "5400 Tech Data Drive",
      "city": "Clearwater",
      "state": "FL",
      "postal_code": "33760",
      "additional_info_contact": "727-789-0100",
      "reason_for_recall": "SCC Soft Computer, Clearwater, FL issued a Correction for SoftPath GUI Releases 3.17 and 4.1 in March 2007 for clients with high volume output of final reports who use the Batch Print for Final Reports in the Scheduler, there is a potential for some reports not printing.  This issue was corrected for the one affected user in October 2003.",
      "root_cause_description": "Software design",
      "action": "SCC Soft Computer sent a SAFETY NOTIFICATION TASK VERBIAGE dated January 5, 2004, to one client.  The Notice identified the product, the problem, and the action to be taken by the client.  \nSCC recommended the client set the Search Date Range to a small enough range to assure that the number of cases scanned during the job would be less than 10,000.\nThe client was asked to acknowledge receipt of the task and to indicate if the Workflow or functionality was used/not used and Agree to use the alternative working solution.   \nFor any questions regarding this recall call (727) 780-0100, ext. 4235.",
      "product_quantity": "84",
      "distribution_pattern": "Nationwide - USA, including the states of CA, CO, CT, FL, GA, IL, IN, LA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, PA, SD, TX, WI, and WV",
      "openfda": {
        "k_number": [
          "K921846"
        ]
      }
    }
  ]
}