  "results": [
    {
      "cfres_id": "87753",
      "product_res_number": "Z-0763-2010",
      "event_date_initiated": "2009-10-12",
      "event_date_posted": "2010-04-12",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-09-06",
      "res_event_number": "54169",
      "product_code": "KWQ",
      "k_numbers": [
        "K062310"
      ],
      "product_description": "Stryker Reflex Hybrid Screw Extractor Inner Shaft;\nNot Sterile;\nManufactured by: Stryker Spine SAS, Cestas, France;\nDistributed in the US by: Stryker Spine, 2 Pearl Court, Allendale, NJ.\n\nThe threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.",
      "code_info": "Ref # 48511905B, all lots.",
      "firm_fei_number": "3004024955",
      "recalling_firm": "Stryker Spine",
      "address_1": "2 Pearl Ct",
      "city": "Allendale",
      "state": "NJ",
      "postal_code": "07401-1611",
      "additional_info_contact": "Ms. Michelle Barry\n201-760-8150",
      "reason_for_recall": "The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.",
      "root_cause_description": "Employee error",
      "action": "An \"Urgent Product Recall\" letter was sent to Stryker Branches/Agencies on October 12, 2009 and to surgeons who have used the device on October 26, 2009.  The recall letter described the affected products, issue and potential hazards.  Consignees were instructed to examine their inventory and hospital locations for the affected products and return to Stryker Spine as indicated in the recall letter.  Customers were also instructed to complete the enclosed Customer Response form and fax to 1-201-760-8370.\n\nDirect questions about the recall to Stryker Spine by calling 1-201-760-8206.",
      "product_quantity": "983",
      "distribution_pattern": "Worldwide Distribution.",
      "openfda": {
        "k_number": [
          "K170342",
          "K905023",
          "K143626",
          "K173867",
          "K020244",
          "K220275",
          "K981008",
          "K132994",
          "K002592",
          "K133911",
          "K790404",
          "K231460",
          "K192049",
          "K070922",
          "K132653",
          "K122216",
          "K200170",
          "K152015",
          "K213820",
          "K240935",
          "K070273",
          "K180537",
          "K022997",
          "K231593",
          "K081548",
          "K030595",
          "K062371",
          "K020650",
          "K070994",
          "K040397",
          "K232256",
          "K993066",
          "K050117",
          "K152239",
          "K070775",
          "K100265",
          "K102334",
          "K982875",
          "K024326",
          "K071905",
          "K233215",
          "K041447",
          "K030866",
          "K091134",
          "K130868",
          "K103505",
          "K121049",
          "K200956",
          "K990603",
          "K051764",
          "K182489",
          "K043548",
          "K091071",
          "K991092",
          "K070148",
          "K010003",
          "K191786",
          "K021461",
          "K964623",
          "K032102",
          "K182473",
          "K121323",
          "K223627",
          "K020620",
          "K190227",
          "K213219",
          "K120724",
          "K200883",
          "K261309",
          "K221572",
          "K150983",
          "K190425",
          "K243369",
          "K131521",
          "K030500",
          "K000009",
          "K062831",
          "K971883",
          "K201521",
          "K181543",
          "K001864",
          "K200885",
          "K173935",
          "K143706",
          "K013877",
          "K042235",
          "K222628",
          "K080773",
          "K220441",
          "K831982",
          "K112861",
          "K181549",
          "K192494",
          "K192678",
          "K042922",
          "K031276",
          "K002858",
          "K132012",
          "K043151",
          "K063158",
          "K203065",
          "K021761",
          "K100061",
          "K030667",
          "K121801",
          "K213456",
          "K113170",
          "K032752",
          "K181244",
          "K012916",
          "K183060",
          "K041989",
          "K963780",
          "K973060",
          "K974885",
          "K112809",
          "K022965",
          "K213901",
          "K001272",
          "K953978",
          "K021564",
          "K973923",
          "K133518",
          "K180090",
          "K081038",
          "K802222",
          "K223753",
          "K211903",
          "K022605",
          "K971248",
          "K141500",
          "K042317",
          "K172086",
          "K130640",
          "K220379",
          "K040197",
          "K143065",
          "K233640",
          "K051971",
          "K220334",
          "K052072",
          "K121791",
          "K140611",
          "K201079",
          "K183214",
          "K041460",
          "K063430",
          "K160362",
          "K101353",
          "K182418",
          "K041794",
          "K061274",
          "K111135",
          "K183073",
          "K040130",
          "K160702",
          "K121514",
          "K992168",
          "K212139",
          "K032368",
          "K031100",
          "K000486",
          "K170224",
          "K081380",
          "K051815",
          "K172585",
          "K964797",
          "K171797",
          "K022791",
          "K172104",
          "K072407",
          "K041205",
          "K043082",
          "K092765",
          "K181846",
          "K001847",
          "K930807",
          "K082037",
          "K231766",
          "K090861",
          "K112724",
          "K923242",
          "K083663",
          "K080646",
          "K043181",
          "K092360",
          "K162638",
          "K873826",
          "K130030",
          "K983622",
          "K904446",
          "K040003",
          "K922855",
          "K150036",
          "K073708",
          "K112748",
          "K943523",
          "K120515",
          "K953720",
          "K022344",
          "K121815",
          "K241175",
          "K080191",
          "K190170",
          "K061385",
          "K993855",
          "K072650",
          "K012184",
          "K062836",
          "K141993",
          "K082821",
          "K223719",
          "K240690",
          "K052763",
          "K910681",
          "K061482",
          "K103423",
          "K991678",
          "K171112",
          "K181562",
          "K041217",
          "K211932",
          "K201136",
          "K150700",
          "K050965",
          "K260015",
          "K091044",
          "K001794",
          "K061229",
          "K073349",
          "K050774",
          "K073530",
          "K140417",
          "K993030",
          "K230721",
          "K053581",
          "K050813",
          "K031214",
          "K122497",
          "K143688",
          "K070285",
          "K162824",
          "K192076",
          "K993181",
          "K212446",
          "K072339",
          "K150449",
          "K082187",
          "K242361",
          "K142237",
          "K072703",
          "K260570",
          "K071151",
          "K153526",
          "K123578",
          "K143576",
          "K051531",
          "K112533",
          "K031782",
          "K974706",
          "K945700",
          "K060491",
          "K082504",
          "K974435",
          "K043066",
          "K200523",
          "K240484",
          "K233360",
          "K993822",
          "K221728",
          "K052552",
          "K180755",
          "K103137",
          "K212405",
          "K021039",
          "K160522",
          "K223321",
          "K222493",
          "K983899",
          "K040844",
          "K061789",
          "K133475",
          "K211408",
          "K210424",
          "K113509",
          "K152455",
          "K040962",
          "K130202",
          "K201979",
          "K040017",
          "K220147",
          "K943725",
          "K092531",
          "K792352",
          "K202309",
          "K111796",
          "K100070",
          "K171439",
          "K163602",
          "K100614",
          "K260697",
          "K970635",
          "K994347",
          "K140260",
          "K250486",
          "K150666",
          "K131533",
          "K021226",
          "K042544",
          "K062310",
          "K091396",
          "K001844",
          "K121060",
          "K111410",
          "K230301",
          "K202624",
          "K192314",
          "K182885",
          "K210325",
          "K163366",
          "K180166",
          "K202192",
          "K041625",
          "K943819",
          "K040401",
          "K141332",
          "K934756",
          "K970955",
          "K061353",
          "K230424",
          "K102265",
          "K142060",
          "K102406",
          "K121837",
          "K140234",
          "K083341",
          "K111866",
          "K202064",
          "K040721",
          "K190053",
          "K923703",
          "K072546",
          "K011556",
          "K060379",
          "K000634",
          "K092474",
          "K202878",
          "K182808",
          "K173375",
          "K122081",
          "K253201",
          "K992784",
          "K190666",
          "K013665",
          "K982443",
          "K240515",
          "K000536",
          "K233735",
          "K090415",
          "K050390",
          "K173091",
          "K963655",
          "K063764",
          "K970806",
          "K994239",
          "K023983",
          "K090751",
          "K193494",
          "K950709",
          "K162250",
          "K984251",
          "K220182",
          "K992110",
          "K133245",
          "K062407",
          "K080429",
          "K013558",
          "K071416",
          "K162194",
          "K041469",
          "K243904",
          "K172131",
          "K163104",
          "K093776",
          "K103838",
          "K163625",
          "K080020",
          "K111528",
          "K251436",
          "K061002",
          "K231145",
          "K003061",
          "K080302",
          "K131025",
          "K050588",
          "K013979",
          "K023133",
          "K982798",
          "K081567",
          "K970462",
          "K790941",
          "K243015",
          "K152281",
          "K081418",
          "K910874",
          "K926453",
          "K073437",
          "K012305",
          "K060442",
          "K143279",
          "K051039",
          "K010466",
          "K173521",
          "K132310",
          "K974757",
          "K092659",
          "K171538",
          "K190565",
          "K965193",
          "K972718",
          "K110990",
          "K990959",
          "K992344",
          "K132366",
          "K990005",
          "K973854",
          "K130427",
          "K062643",
          "K101255",
          "K092070",
          "K143028",
          "K970048",
          "K191584",
          "K923446",
          "K841189",
          "K254182",
          "K202227",
          "K941367",
          "K040261",
          "K982006",
          "K150834",
          "K032059",
          "K173306",
          "K123167",
          "K240592",
          "K173099",
          "K112206",
          "K210286",
          "K251965",
          "K033965",
          "K121078",
          "K062590",
          "K063544",
          "K943694",
          "K001357",
          "K221053",
          "K071667",
          "K180022",
          "K133194",
          "K060025",
          "K142212",
          "K002824",
          "K010115",
          "K042859",
          "K081391",
          "K140742",
          "K071990",
          "K081568",
          "K201587",
          "K042798",
          "K922087",
          "K083020",
          "K062686",
          "K160982",
          "K101848",
          "K040426",
          "K220324",
          "K203154",
          "K964416",
          "K062110",
          "K050804",
          "K070681",
          "K092124",
          "K230993",
          "K162664",
          "K091926",
          "K953730",
          "K161641",
          "K070891",
          "K002371",
          "K050892",
          "K080690",
          "K980228",
          "K050451",
          "K925351",
          "K252611",
          "K093804",
          "K020600",
          "K955730",
          "K200927",
          "K122910",
          "K190584",
          "K231839",
          "K073514",
          "K992792",
          "K033844",
          "K231090",
          "K182055",
          "K103750",
          "K113329",
          "K073500",
          "K042499",
          "K210182",
          "K103393",
          "K984342",
          "K042025",
          "K122733",
          "K181063",
          "K000513",
          "K052292",
          "K163061",
          "K092146",
          "K131374",
          "K091253",
          "K073479",
          "K103491",
          "K152282",
          "K090222",
          "K202972",
          "K103068",
          "K120092",
          "K101151",
          "K162211",
          "K132589",
          "K012881",
          "K010911",
          "K901507",
          "K082070",
          "K092108",
          "K131512",
          "K251940",
          "K851414",
          "K173885",
          "K020649",
          "K041105",
          "K181190",
          "K153042",
          "K211205",
          "K002980",
          "K180772",
          "K071329",
          "K103070",
          "K013439",
          "K073643",
          "K131520",
          "K081132",
          "K072231",
          "K083338",
          "K231715",
          "K212937",
          "K250602",
          "K994122",
          "K240423",
          "K132324",
          "K141632",
          "K231251",
          "K170530",
          "K213443",
          "K082273",
          "K121862",
          "K211718",
          "K183409",
          "K914362",
          "K010639",
          "K151553",
          "K200224",
          "K201029",
          "K222572",
          "K012045",
          "K853033",
          "K040655",
          "K103819",
          "K092695",
          "K130825",
          "K152623",
          "K090316",
          "K950011",
          "K220843",
          "K191500",
          "K061147",
          "K913730",
          "K073275",
          "K031855",
          "K091936",
          "K093200",
          "K200995",
          "K091159",
          "K063100",
          "K201078",
          "K200846",
          "K931653",
          "K111132",
          "K152193",
          "K022798",
          "K021979",
          "K922543",
          "K033951",
          "K936269",
          "K100243",
          "K992012",
          "K023377",
          "K031702",
          "K080815",
          "K142699",
          "K102820",
          "K934429",
          "K962486",
          "K051665",
          "K100215",
          "K252776",
          "K190016",
          "K032265",
          "K820034",
          "K203253",
          "K042073",
          "K022419",
          "K033148",
          "K072040",
          "K022399",
          "K014107",
          "K994383",
          "K212527",
          "K782061",
          "K140361",
          "K974525",
          "K081770",
          "K121658",
          "K053053",
          "K150759",
          "K940744",
          "K192792",
          "K993513",
          "K072943",
          "K083562",
          "K211712",
          "K131200",
          "K030327",
          "K183056",
          "K950209",
          "K153097",
          "K070335",
          "K073232",
          "K133567",
          "K160604",
          "K250446",
          "K024125",
          "K261112",
          "K143271",
          "K183084",
          "K232318",
          "K052546",
          "K161524",
          "K032815",
          "K971730",
          "K100708",
          "K000742",
          "K192481"
        ],
        "registration_number": [
          "",
          "3020703621",
          "3008583793",
          "3006791286",
          "3011577940",
          "3000203391",
          "3009941480",
          "3004464325",
          "3013055499",
          "2532027",
          "3007125392",
          "3010097171",
          "3010032903",
          "3010560653",
          "3010120104",
          "3011015572",
          "3004638532",
          "3010370554",
          "3014257776",
          "3003639920",
          "3009887475",
          "1225838",
          "3007441485",
          "3011301313",
          "3010120148",
          "1423662",
          "2133928",
          "3009882675",
          "3006128100",
          "3009106092",
          "9681465",
          "3009973699",
          "3037407500",
          "3008791302",
          "3009189869",
          "3008455034",
          "2031093",
          "3005677016",
          "9611813",
          "9615021",
          "3014725904",
          "1417485",
          "3010120339",
          "3005180920",
          "3005031160",
          "3004788213",
          "9611390",
          "3009962553",
          "3009165919",
          "3005823819",
          "3010057495",
          "1835831",
          "3017521423",
          "3009417901",
          "3008850074",
          "3015941638",
          "3010440433",
          "3014937101",
          "3017528621",
          "3018094310",
          "3016676864",
          "3013194153",
          "1527105",
          "3010531060",
          "3010123256",
          "2134285",
          "3023138345",
          "1319429",
          "1835296",
          "3010178296",
          "3010363503",
          "3012764194",
          "3015869493",
          "3016438694",
          "3006493760",
          "3015426608",
          "3004941535",
          "2134470",
          "2531477",
          "2183449",
          "1824199",
          "3003597504",
          "3005596514",
          "3006404019",
          "3003855635",
          "3012491394",
          "3004717358",
          "3014678447",
          "1030489",
          "1722511",
          "3003387384",
          "3015399803",
          "1824313",
          "1526439",
          "1530530",
          "1649379",
          "3008812560",
          "3019878714",
          "3005977257",
          "1226544",
          "3010068573",
          "3001239363",
          "3015398319",
          "3006755626",
          "3005144609",
          "3005497913",
          "1221053",
          "2648623",
          "1935627",
          "3010509633",
          "3004024955",
          "1825034",
          "3009971621",
          "3014655115",
          "3010197239",
          "1225675",
          "3021945817",
          "3009756327",
          "3005061536",
          "3012428435",
          "3009169996",
          "3007014520",
          "1720929",
          "3009050824",
          "3014273424",
          "1450662",
          "1223434",
          "3012711663",
          "3031261833",
          "3043543260",
          "3009760038",
          "3010866843",
          "3012599928",
          "3009144915",
          "3002976036",
          "3006356043",
          "1000432246",
          "3005874553",
          "3007993775",
          "3011390763",
          "3005819474",
          "3010120326",
          "3012120772",
          "3009504230",
          "1000461041",
          "3033509898",
          "2029275",
          "3007728266",
          "3014680795",
          "3011795235",
          "3019837678",
          "3008114965",
          "2029005",
          "3006639944",
          "3010160527",
          "3003477135",
          "3010041993",
          "3009554293",
          "3011277306",
          "3010041430",
          "3008868758",
          "3003935342",
          "1422634",
          "3009241418",
          "3016084569",
          "3010863450",
          "3010047454",
          "1056553",
          "3003761012",
          "3015440604",
          "3009998573",
          "3008754074",
          "3010143171",
          "2031966",
          "3013422236",
          "3020282716",
          "2032112",
          "1000517406",
          "3006110997",
          "3013422238",
          "2027467",
          "3031240334",
          "3009173317",
          "3005739886",
          "3004176895",
          "1420032",
          "3011181154",
          "3010197224",
          "1220477",
          "3006783837",
          "3012329926",
          "3000270450",
          "1836357",
          "3006801265",
          "3010665091",
          "3017936978",
          "3009217531",
          "3009790163",
          "3009380063",
          "3011489618",
          "2531195",
          "3004774118",
          "3011127597",
          "3006215390",
          "9612420",
          "3009826895",
          "3009959868",
          "3010663372",
          "3005360851",
          "3009888740",
          "3036756245",
          "3005076207",
          "1531174",
          "3017597868",
          "3015991317",
          "3004748528",
          "3007922509",
          "3009973336",
          "3010041499",
          "3015831087",
          "1528646",
          "3008285983",
          "9617297",
          "3011221483",
          "1835444",
          "2530808",
          "3010462278",
          "2134947",
          "3025603301",
          "8010468",
          "3008262872",
          "1526534",
          "3006378738",
          "3005325790",
          "1057425",
          "3009446038",
          "1835251",
          "3003120897",
          "3016443334",
          "3012447612",
          "3023852420",
          "1649518",
          "2647346",
          "3011187779",
          "3010370500",
          "2087234",
          "3010287685",
          "3012994232",
          "2028632",
          "1224360",
          "3008110533",
          "3017435639",
          "3007289746",
          "3010009693",
          "3005706667",
          "3011530718",
          "9617544",
          "3010049501",
          "3004153896",
          "2183946",
          "3016050940",
          "3016237080",
          "1000200989",
          "3015179588",
          "1828288",
          "3009957608",
          "3004049923",
          "2032098",
          "2032093",
          "1833986",
          "3014938863",
          "3010162973",
          "3013548554",
          "3016678045",
          "3007485169",
          "3010365473",
          "3004638600",
          "3009051471",
          "3005226664",
          "3002807112",
          "3015869492",
          "3009704301",
          "3006846753",
          "3007539489",
          "3003526896",
          "3002807561",
          "3016496998",
          "3013491327",
          "3027807601",
          "3010370880",
          "9681642",
          "9610612",
          "3022159082",
          "3008992889",
          "2245304",
          "1424263",
          "3005641619",
          "3016112537",
          "3015869121",
          "2024024",
          "3015877656",
          "3015709929",
          "3010326971",
          "3014252644",
          "3034676720",
          "3010123206",
          "3007675554",
          "3010120135",
          "3007210720",
          "3006017180",
          "3008534770",
          "3003525643",
          "3012131184",
          "3015876273",
          "3004961578",
          "1424434",
          "1644312",
          "3019767615",
          "3030412764",
          "1043214",
          "3006404071",
          "1834331",
          "3004635447",
          "9615800",
          "1833824",
          "1833920",
          "3011302058",
          "1833506",
          "3003541440",
          "3004142400",
          "3004719693",
          "3006272282",
          "9614209",
          "3004499989",
          "8043792",
          "9615128",
          "3015212339",
          "3009973505",
          "3009049161",
          "1045254",
          "3003898943",
          "3004188066",
          "3007923096",
          "3000170817",
          "1221763",
          "9612277",
          "1058584",
          "3009169872",
          "2031917",
          "3004893332",
          "3018210489",
          "3006217744",
          "3006082533",
          "3011874600",
          "1828464",
          "3011513267",
          "3011556898",
          "3008526708",
          "1834379",
          "3013011598"
        ],
        "fei_number": [
          "",
          "3020703621",
          "3008583793",
          "3006791286",
          "3011577940",
          "3000203391",
          "3009941480",
          "3004464325",
          "3008647423",
          "3013055499",
          "3007125392",
          "3010097171",
          "3010032903",
          "3010560653",
          "1000120314",
          "3010120104",
          "3011015572",
          "3004638532",
          "3010370554",
          "3014257776",
          "3003051987",
          "3003639920",
          "3009887475",
          "3007441485",
          "3011301313",
          "3001814066",
          "3010120148",
          "3003110073",
          "3006128100",
          "3009106092",
          "3009973699",
          "3037407500",
          "3008791302",
          "3009189869",
          "3008455034",
          "3005677016",
          "3002806577",
          "3014725904",
          "3002807113",
          "3003128516",
          "1417485",
          "3003076831",
          "3010120339",
          "3005180920",
          "3005031160",
          "3004788213",
          "3009962553",
          "3009165919",
          "1000615957",
          "3010057495",
          "3000718985",
          "3017521423",
          "3009417901",
          "3008850074",
          "3010440433",
          "3014937101",
          "3017528621",
          "3018094310",
          "3016676864",
          "3013194153",
          "3010531060",
          "3010123256",
          "3023138345",
          "3010363503",
          "3012764194",
          "3015869493",
          "3016438694",
          "3006493760",
          "3015426608",
          "3004941535",
          "2183449",
          "3002806725",
          "1824199",
          "3000207350",
          "3003597504",
          "3005596514",
          "1000306303",
          "3003855635",
          "3001236277",
          "3012491394",
          "3004717358",
          "3014678447",
          "1000221916",
          "3001642380",
          "3003387384",
          "3015399803",
          "1824313",
          "1526439",
          "3003292403",
          "3001551180",
          "3008812560",
          "3019878714",
          "3005977257",
          "3002980729",
          "3010068573",
          "3003909293",
          "3003200961",
          "1000395711",
          "3001239363",
          "1000160570",
          "1000125025",
          "3006755626",
          "3005144609",
          "3007932279",
          "3005497913",
          "1221053",
          "2648623",
          "1935627",
          "3010509633",
          "3004024955",
          "1825034",
          "1000477302",
          "3009971621",
          "1000143502",
          "3009256700",
          "3002808269",
          "3014655115",
          "3010197239",
          "3021945817",
          "3005061536",
          "3012428435",
          "3009169996",
          "3007014520",
          "1720929",
          "3009050824",
          "3002806685",
          "3014273424",
          "1450662",
          "3012711663",
          "3005532635",
          "3000215273",
          "3031261833",
          "3043543260",
          "3009760038",
          "3010866843",
          "3012599928",
          "3009144915",
          "3002976036",
          "3006356043",
          "1000432246",
          "3005874553",
          "3007993775",
          "1000132370",
          "3011390763",
          "3005819474",
          "3010120326",
          "3012120772",
          "3009504230",
          "1000461041",
          "3033509898",
          "3007728266",
          "1000117748",
          "3014680795",
          "3011795235",
          "3019837678",
          "3008114965",
          "3006639944",
          "3003477135",
          "3010041993",
          "3009554293",
          "3011277306",
          "1000136451",
          "3010041430",
          "3008868758",
          "3003935342",
          "3009241418",
          "3016084569",
          "3010863450",
          "3010047454",
          "3003761012",
          "3015440604",
          "3009998573",
          "3000717687",
          "3008754074",
          "3010143171",
          "3002807040",
          "3002929183",
          "3013422236",
          "3020282716",
          "1000517406",
          "3006110997",
          "3013422238",
          "2027467",
          "3031240334",
          "3009173317",
          "3005739886",
          "3004176895",
          "1420032",
          "3011181154",
          "3010197224",
          "1220477",
          "3012329926",
          "3000270450",
          "3011764740",
          "1000306431",
          "3001236549",
          "3002462840",
          "3006801265",
          "3010665091",
          "3017936978",
          "3009217531",
          "3002807041",
          "3009790163",
          "3009380063",
          "3011489618",
          "1000115331",
          "3011127597",
          "3006215390",
          "3009826895",
          "3009959868",
          "3010663372",
          "3005360851",
          "3009888740",
          "3036756245",
          "3000718490",
          "3005076207",
          "3017597868",
          "3015991317",
          "3004748528",
          "3001451451",
          "3007922509",
          "3009973336",
          "3015831087",
          "3008285983",
          "3011221483",
          "3010462278",
          "3025603301",
          "3008262872",
          "1000514656",
          "1526534",
          "3006378738",
          "3005325790",
          "3002795782",
          "3008599177",
          "3009446038",
          "3001644153",
          "3003120897",
          "3016443334",
          "3012447612",
          "3023852420",
          "1000137563",
          "2647346",
          "3009322142",
          "3011187779",
          "3010370500",
          "1000116116",
          "3010287685",
          "3012994232",
          "3000205591",
          "1000522098",
          "3008820440",
          "3003015552",
          "3003058448",
          "3002806315",
          "3002807038",
          "3008110533",
          "3017435639",
          "3007289746",
          "1000514086",
          "3010009693",
          "3005706667",
          "3011530718",
          "3002807328",
          "3010049501",
          "3004153896",
          "3016050940",
          "3016237080",
          "1000200989",
          "3015179588",
          "1000139987",
          "1828288",
          "3009957608",
          "3004049923",
          "3014938863",
          "3002807156",
          "3013548554",
          "3016678045",
          "3007485169",
          "3010365473",
          "3004638600",
          "3009051471",
          "3005226664",
          "3002807112",
          "3015869492",
          "3009704301",
          "3006846753",
          "3007539489",
          "3003526896",
          "3002807561",
          "3016496998",
          "3013491327",
          "3027807601",
          "3010370880",
          "3002993297",
          "1000115656",
          "3022159082",
          "3002287318",
          "3008992889",
          "2245304",
          "3005641619",
          "3016112537",
          "3015869121",
          "3000210784",
          "2024024",
          "3015877656",
          "3015709929",
          "3010326971",
          "3014252644",
          "3034676720",
          "3010123206",
          "1000219637",
          "3007675554",
          "3010120135",
          "3007210720",
          "3006017180",
          "3008534770",
          "3003525643",
          "3012131184",
          "3015876273",
          "3004961578",
          "1000306340",
          "3019767615",
          "3030412764",
          "1043214",
          "3006404071",
          "3004635447",
          "3003112432",
          "1000069183",
          "3011302058",
          "3003541440",
          "3004142400",
          "3004719693",
          "3006272282",
          "3003064967",
          "3003058119",
          "3004499989",
          "3010938185",
          "3015212339",
          "3009973505",
          "3009049161",
          "1045254",
          "3003898943",
          "3004188066",
          "3007923096",
          "1832656",
          "3000170817",
          "1221763",
          "3009169872",
          "3000717902",
          "3004893332",
          "3018210489",
          "3006217744",
          "3011874600",
          "1828464",
          "3006082533",
          "3011513267",
          "3002806773",
          "3011556898",
          "3008526708",
          "3013011598"
        ],
        "device_name": "Appliance, Fixation, Spinal Intervertebral Body",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3060",
        "device_class": "2"
      }
    }
  ]
}