  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "87632",
      "product_res_number": "Z-0719-2010",
      "event_date_initiated": "2009-11-12",
      "event_date_posted": "2010-01-29",
      "recall_status": "Terminated",
      "event_date_terminated": "2010-10-25",
      "res_event_number": "54149",
      "product_code": "cac",
      "k_numbers": [
        "K952224"
      ],
      "product_description": "Stryker CBCII ConstaVac Blood conservation system,  sterile,  Stryker Instruments, Kalamazoo, MI, REF 225-28E.  \n\nBlood recovery system used post-operatively to collect, filter and allow for reinfusion of autologous blood.",
      "code_info": "serial numbers 07303012 through 09173032",
      "firm_fei_number": "1811755",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E. Milham Ave.",
      "city": "Kalamazoo",
      "state": "MI",
      "postal_code": "49001",
      "additional_info_contact": "Jennifer Mars\n269-323-7700",
      "reason_for_recall": "The product may not be sterile due to package not being intact.",
      "root_cause_description": "Packaging process control",
      "action": "Stryker issued a \"Medical Device Recall Notification\" via email  to foreign consignees dated November 12, 2009.  Stryker further issued recall letters via overnight mail to domestic consginees beginning November 20, 2009.  These notifications informed consignees of the affected devices instructing them to destroy all affected product and return a completed Business Reply Form to the firm.\n\nFor further information, contact Stryker Instruments at 1-800-800-4236 extension 3808.",
      "product_quantity": "79,617 for all products",
      "distribution_pattern": "Worldwide Distribution.",
      "openfda": {}
    }
  ]
}