{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "86620",
      "product_res_number": "Z-0769-2010",
      "event_date_initiated": "2009-11-13",
      "event_date_posted": "2010-03-04",
      "recall_status": "Terminated",
      "event_date_terminated": "2014-10-24",
      "res_event_number": "53827",
      "product_code": "MKJ",
      "k_numbers": [
        "K031987",
        "K040438",
        "K052161"
      ],
      "product_description": "Cardiac Science Powerheart  9300P  automated external defibrillator.  This semi-automatic model requires the user to press a shock button in order for the device to deliver defibrillation energy (when appropriate).\n\nIntended for use in either in-hospital or out-of-hospital settings for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing.",
      "code_info": "SERIAL NUMBERS:  391776, 393102, 402793, 404926, 900002, 900003, 900004, 900005, 900006, 900007, 900008, 900009, 900010, 900011, 900012, 900013, 900014, 900015, 900016, 900017, 900018, 900019, 900020, 900021, 900022, 900023, 900024, 900025, 900026, 900027, 900028, 900029, 900030, 900035, 900036, 900037, 900038, 900039, 900040, 900041, 900042, 900044, 900045, 900046, 900047, 900048, 900049, 900050, 900051, 900052, 900053, 900054, 900055, 900056, 900057, 900058, 900059, 900060, 900062, 900063, 900064, 900065, 900066, 900067, 900068, 900069, 900070, 900071, 900072, 900073, 900075, 900076, 900077, 900078, 900079, 900080, 900081, 900082, 900083, 900084, 900085, 900086, 900087, 900088, 900089, 900090, 900091, 900092, 900093, 900094, 900096, 900097, 900099, 900100, 900102, 900103, 900104, 900105, 900106, 900107, 900108, 900109, 900121, 900123, 900124, 900126, 900127, 900128, 900129, 900131, 900132, 900133, 900134, 900135, 900136, 900138, 900139, 900140, 900142, 900143, 900144, 900145, 900146, 900147, 900148, 900149, 900150, 900151, 900152, 900153, 900154, 900156, 900158, 900159, 900160, 900161, 900162, 900165, 900166, 900168, 900169, 900170, 900172, 900173, 900174, 900176, 900177, 900178, 900179, 900180, 900181, 900182, 900183, 900184, 900185, 900186, 900187, 900188, 900189, 900190, 900191, 900192, 900193, 900194, 900195, 900196, 900197, 900199, 900200, 900201, 900202, 900203, 900205, 900206, 900207, 900211, 900212, 900214, 900216, 900217, 900218, 900219, 900220, 900222, 900223, 900224, 900226, 900227, 900228, 900229, 900230, 900231, 900232, 900233, 900234, 900235, 900237, 900238, 900241, 900242, 900243, 900244, 900245, 900246, 900247, 900248, 900249, 900250, 900251, 900252, 900254, 900256, 900257, 900259, 900260, 900261, 900262, 900263, 900264, 900265, 900266, 900268, 900269, 900270, 900271, 900272, 900273, 900274, 900276, 900277, 900278, 900279, 900280, 900281, 900283, 900284, 900288, 900289, 900290, 900291, 900294, 900295, 900297, 900298, 900299, 900300, 900301, 900302, 900305, 900307, 900308, 900309, 900310, 900311, 900312, 900313, 900314, 900315, 900316, 900317, 900318, 900319, 900321, 900322, 900323, 900324, 900325, 900326, 900327, 900328, 900329, 900330, 900331, 900333, 900334, 900335, 900336, 900337, 900338, 900341, 900342, 900343, 900344, 900345, 900346, 900347, 900348, 900349, 900350, 900351, 900352, 900353, 900354, 900356, 900357, 900358, 900360, 900361, 900362, 900363, 900364, 900365, 900366, 900367, 900368, 900369, 900370, 900371, 900372, 900373, 900374, 900375, 900376, 900377, 900378, 900380, 900381, 900382, 900383, 900384, 900388, 900390, 900391, 900394, 900395, 900396, 900397, 900399, 900401, 900402, 900403, 900404, 900405, 900406, 900407, 900409, 900412, 900414, 900418, 900420, 900421, 900422, 900423, 900425, 900426, 900427, 900428, 900429, 900430, 900433, 900434, 900435, 900437, 900439, 900440, 900441, 900444, 900447, 900448, 900449, 900450, 900452, 900453, 900455, 900456, 900457, 900458, 900464, 900467, 900468, 900471, 900473, 900474, 900475, 900480, 900481, 900485, 900488, 900491, 900493, 900495, 900497, 900498, 900501, 900502, 900504, 900506, 900507, 900508, 900510, 900512, 900513, 900514, 900518, 900519, 900520, 900521, 900522, 900523, 900524, 900525, 900527, 900528, 900531, 900532, 900533, 900537, 900538, 900539, 900540, 900541, 900542, 900545, 900546, 900548, 900549, 900550, 900551, 900552, 900553, 900555, 900556, 900557, 900559, 900560, 900561, 900562, 900563, 900565, 900568, 900569, 900570, 900571, 900572, 900573, 900574, 900575, 900576, 900578, 900579, 900580, 900581, 900582, 900584, 900585, 900586, 900587, 900589, 900592, 900593, 900594, 900595, 900596, 900598, 900599, 900600, 900601, 900603, 900604, 900605, 900606, 900607, 900608, 900609, 900611, 900615, 900616, 900617, 900620, 900621, 900622, 900623, 900625, 900628, 900629, 900630, 900631, 900632, 900633, 900634, 900635, 900639, 900640, 900641, 900642, 900643, 900645, 900646, 900700, 900701, 900702, 900703, 900706, 900707, 900708, 900710, 900711, 900712, 900713, 900714, 900715, 900717, 900718, 900719, 900720, 900721, 900722, 900723, 900724, 900725, 900726, 900727, 900728, 900729, 900730, 900732, 900733, 900734, 900736, 900737, 900738, 900739, 900740, 900741, 900742, 900743, 900745, 900747, 900748, 900750, 900751, 900753, 900754, 900757, 900758, 900759, 900760, 900761, 900762, 900764, 900765, 900767, 900768, 900769, 900771, 900772, 900774, 900776, 900777, 900779, 900861, 900862, 900863, 900864, 900865, 900868, 900869, 900870, 900874, 900876, 900877, 900880, 900881, 900882, 900883, 900885, 900886, 900890, 900892, 900893, 900894, 900895, 900897, 900898, 900899, 900900, 900902, 900903, 900905, 900907, 900908, 900909, 900910, 900911, 900912, 900914, 900915, 900916, 900918, 900919, 900920, 900921, 900923, 900924, 900925, 900927, 900928, 900929, 900930, 900931, 900933, 900934, 900935, 900936, 900938, 900939, 900950, 900951, 900952, 900953, 900954, 900955, 900956, 900958, 900959, 900961, 900962, 900963, 900964, 900965, 900966, 900967, 900969, 900970, 900971, 900972, 900974, 901036, 901037, 901038, 901039, 901040, 901041, 901042, 901043, 901044, 901045, 901046, 901047, 901048, 901049, 901050, 901051, 901052, 901053, 901054, 901055, 901056, 901057, 901058, 901059, 901060, 901061, 901063, 901067, 901068, 901069, 901070, 901071, 901072, 901073, 901074, 901075, 901076, 901077, 901079, 901080, 901081, 901082, 901083, 901084, 901085, 901136, 901137, 901138, 901139, 901140, 901141, 901142, 901143, 901144, 901145, 901146, 901147, 901148, 901149, 901150, 901152, 901154, 901155, 901156, 901157, 901158, 901159, 901160, 901161, 901162, 901163, 901164, 901165, 901166, 901167, 901168, 901169, 901170, 901171, 901173, 901175, 901176, 901177, 901179, 901181, 901183, 901184, 901236, 901237, 901238, 901239, 901240, 901241, 901242, 901243, 901244, 901245, 901246, 901247, 901248, 901250, 901251, 901252, 901253, 901254, 901255, 901256, 901257, 901258, 901259, 901260, 901261, 901262, 901263, 901264, 901265, 901267, 901268, 901269, 901270, 901271, 901272, 901273, 901274, 901275, 901276, 901277, 901278, 901280, 901281, 901282, 901283, 901284, 901285, 901286, 901287, 901288, 901289, 901290, 901291, 901293, 901294, 901295, 901346, 901347, 901348, 901349, 901350, 901351, 901352, 901353, 901354, 901355, 901357, 901358, 901359, 901360, 901361, 901362, 901363, 901364, 901365, 901366, 901367, 901368, 901369, 901370, 901371, 901372, 901373, 901374, 901375, 901376, 901377, 901378, 901379, 901380, 901381, 901382, 901383, 901384, 901385, 901386, 901387, 901388, 901389, 901390, 901392, 901393, 901394, 901396, 901397, 901398, 901399, 901400, 901401, 901402, 901403, 901404, 901405, 901406, 901407, 901408, 901409, 901411, 901412, 901413, 901414, 901415, 901416, 901418, 901419, 901420, 901422, 901423, 901424, 901425, 901426, 901427, 901428, 901429, 901430, 901431, 901432, 901433, 901434, 901436, 901437, 901438, 901439, 901440, 901441, 901442, 901443, 901444, 901446, 901448, 901449, 901450, 901451, 901452, 901453, 901455, 901456, 901457, 901458, 901459, 901460, 901461, 901463, 901464, 901466, 901467, 901469, 901470, 901471, 901472, 901473, 901475, 901476, 901479, 901480, 901481, 901482, 901483, 901485, 901486, 901487, 901489, 901491, 901492, 901493, 901495, 901496, 901497, 901498, 901499, 901501, 901502, 901503, 901504, 901505, 901506, 901507, 901508, 901509, 901510, 901511, 901512, 901513, 901514, 901515, 901516, 901517, 901518, 901519, 901520, 901521, 901522, 901523, 901524, 901525, 901526, 901527, 901528, 901529, 901530, 901531, 901532, 901533, 901534, 901535, 901536, 901537, 901538, 901539, 901540, 901541, 901542, 901543, 901544, 901545, 901546, 901547, 901549, 901550, 901551, 901552, 901553, 901554, 901555, 901556, 901557, 901558, 901559, 901560, 901561, 901562, 901563, 901564, 901565, 901566, 901567, 901568, 901570, 901571, 901572, 901573, 901574, 901575, 901576, 901577, 901578, 901580, 901581, 901582, 901583, 901584, 901585, 901586, 901587, 901588, 901589, 901590, 901591, 901592, 901593, 901594, 901595, 901596, 901597, 901598, 901599, 901600, 901601, 901602, 901603, 901604, 901605, 901606, 901608, 901609, 901610, 901611, 901612, 901613, 901614, 901615, 901616, 901617, 901618, 901619, 901620, 901621, 901623, 901624, 901625, 901626, 901627, 901630, 901631, 901632, 901633, 901634, 901635, 901636, 901637, 901638, 901640, 901641, 901642, 901643, 901644, 901645, 901646, 901647, 901648, 901649, 901650, 901651, 901652, 901653, 901654, 901655, 901656, 901657, 901658, 901659, 901660, 901661, 901662, 901663, 901664, 901665, 901666, 901667, 901668, 901669, 901670, 901671, 901673, 901674, 901675, 901676, 901677, 901678, 901679, 901680, 901681, 901683, 901684, 901685, 901686, 901687, 901688, 901689, 901690, 901691, 901693, 901694, 901695, 901696, 901697, 901698, 901699, 901700, 901701, 901702, 901703, 901704, 901705, 901706, 901707, 901708, 901709, 901711, 901712, 901713, 901714, 901715, 901716, 901717, 901718, 901719, 901720, 901721, 901722, 901723, 901724, 901725, 901726, 901727, 901728, 901729, 901730, 901731, 901732, 901733, 901734, 901735, 901736, 901737, 901738, 901739, 901740, 901741, 901742, 901743, 901744, 901745, 901747, 901748, 901749, 901750, 901751, 901752, 901753, 901754, 901756, 901757, 901758, 901759, 901760, 901761, 901762, 901763, 901764, 901765, 901766, 901767, 901808, 901809, 901810, 901811, 901812, 901813, 901814, 901815, 901816, 901817, 901818, 901819, 901820, 901822, 901823, 901824, 901825, 901826, 901827, 901828, 901829, 901830, 901831, 901832, 901833, 901834, 901835, 901836, 901837, 901838, 901839, 901840, 901841, 901842, 901843, 901844, 901845, 901846, 901849, 901850, 901852, 901853, 901854, 901855, 901856, 901857, 901859, 901860, 901861, 901862, 901863, 901864, 901865, 901866, 901867, 901868, 901869, 901870, 901871, 901872, 901873, 901874, 901875, 901877, 901878, 901879, 901880, 901881, 901882, 901883, 901885, 901886, 901887, 901888, 901889, 901890, 901891, 901892, 901893, 901894, 901895, 901896, 901897, 901898, 901899, 901900, 901901, 901902, 901903, 901904, 901905, 901906, 901907, 901908, 901909, 901910, 901911, 901913, 901915, 901916, 901917, 901918, 901919, 901920, 901921, 901922, 901923, 901924, 901925, 901926, 901927, 901929, 901930, 901931, 901932, 901933, 901934, 901935, 901936, 901937, 901938, 901939, 901940, 901941, 901943, 901944, 901945, 901946, 901947, 901948, 901949, 901950, 901951, 901952, 901954, 901955, 901956, 901957, 901958, 901959, 901960, 901961, 901962, 901963, 901964, 901965, 901966, 901967, 901968, 901969, 901970, 901971, 901972, 901973, 901977, 901978, 901979, 901980, 901981, 901982, 901983, 901984, 901985, 901986, 901987, 902028, 902029, 902030, 902031, 902032, 902033, 902034, 902035, 902036, 902037, 902038, 902039, 902040, 902043, 902044, 902045, 902046, 902049, 902050, 902051, 902052, 902053, 902054, 902056, 902057, 902058, 902059, 902061, 902062, 902063, 902064, 902065, 902066, 902068, 902069, 902071, 902072, 902073, 902075, 902076, 902077, 902078, 902080, 902081, 902082, 902083, 902085, 902086, 902087, 902088, 902089, 902090, 902091, 902092, 902093, 902094, 902096, 902097, 902098, 902099, 902100, 902101, 902102, 902103, 902104, 902105, 902107, 902108, 902109, 902110, 902111, 902112, 902113, 902114, 902115, 902117, 902118, 902119, 902120, 902121, 902122, 902123, 902124, 902125, 902126, 902127, 902128, 902129, 902130, 902132, 902133, 902134, 902135, 902136, 902137, 902138, 902140, 902141, 902142, 902143, 902144, 902145, 902146, 902147, 902148, 902150, 902151, 902152, 902153, 902154, 902155, 902156, 902157, 902158, 902159, 902160, 902161, 902162, 902163, 902164, 902165, 902166, 902168, 902169, 902170, 902208, 902209, 902210, 902211, 902212, 902213, 902214, 902215, 902216, 902217, 902218, 902219, 902220, 902221, 902222, 902223, 902224, 902225, 902226, 902227, 902228, 902229, 902230, 902231, 902232, 902233, 902234, 902235, 902236, 902237, 902268, 902269, 902270, 902271, 902272, 902273, 902274, 902275, 902276, 902277, 902278, 902279, 902281, 902282, 902283, 902284, 902285, 902286, 902287, 902288, 902289, 902290, 902291, 902292, 902293, 902294, 902296, 902297, 902298, 902299, 902300, 902301, 902303, 902304, 902305, 902306, 902308, 902311, 902312, 902313, 902315, 902316, 902317, 902318, 902319, 902322, 902323, 902324, 902325, 902327, 902328, 902329, 902331, 902332, 902337, 902339, 902341, 902342, 902343, 902344, 902346, 902347, 902349, 902351, 902352, 902353, 902354, 902355, 902356, 902357, 902388, 902389, 902390, 902391, 902392, 902393, 902394, 902395, 902396, 902397, 902398, 902399, 902400, 902401, 902402, 902404, 902405, 902406, 902407, 902408, 902409, 902410, 902411, 902412, 902413, 902414, 902415, 902416, 902417, 902418, 902419, 902420, 902421, 902422, 902423, 902424, 902425, 902426, 902427, 902428, 902429, 902430, 902431, 902432, 902433, 902434, 902435, 902436, 902437, 902438, 902439, 902440, 902441, 902442, 902443, 902444, 902445, 902446, 902447, 902568, 902569, 902571, 902572, 902573, 902574, 902575, 902576, 902577, 902578, 902579, 902580, 902581, 902582, 902583, 902584, 902585, 902586, 902587, 902588, 902589, 902590, 902591, 902592, 902593, 902594, 902595, 902596, 902597, 902598, 902599, 902600, 902601, 902602, 902603, 902604, 902605, 902606, 902607, 902608, 902609, 902610, 902611, 902612, 902613, 902614, 902615, 902616, 902617, 902618, 902619, 902620, 902621, 902622, 902623, 902624, 902625, 902626, 902627, 902628, 902629, 902630, 902631, 902632, 902633, 902634, 902635, 902636, 902637, 902638, 902639, 902640, 902641, 902642, 902693, 902694, 902695, 902696, 902697, 902698, 902699, 902700, 902701, 902702, 902703, 902704, 902705, 902706, 902707, 902708, 902709, 902710, 902711, 902712, 902713, 902715, 902716, 902717, 902718, 902719, 902720, 902721, 902722, 902723, 902724, 902725, 902726, 902727, 902728, 902729, 902730, 902731, 902732, 902733, 902734, 902735, 902736, 902737, 902738, 902739, 902740, 902741, 902742, 902743, 902744, 902745, 902746, 902747, 902748, 902749, 902750, 902751, 902752, 902753, 902754, 902755, 902756, 902757, 902758, 902759, 902760, 902761, 902762, 902763, 902764, 902765, 902766, 902767, 902768, 902769, 902770, 902771, 902772, 902773, 902774, 902775, 902776, 902777, 902778, 902779, 902780, 902781, 902782, 902783, 902784, 902785, 902786, 902787, 902788, 902789, 902790, 902791, 902792, 902793, 902794, 902795, 902796, 902798, 902799, 902800, 902801, 902802, 902803, 902804, 902805, 902806, 902807, 902808, 902809, 902810, 902811, 902812, 902813, 902814, 902815, 902816, 902818, 902819, 902820, 902821, 902822, 902823, 902824, 902825, 902826, 902827, 902828, 902829, 902830, 902831, 902832, 902833, 902834, 902836, 902837, 902838, 902839, 902840, 902841, 902842, 902843, 902844, 902845, 902846, 902847, 902848, 902849, 902850, 902851, 902852, 902853, 902854, 902856, 902857, 902858, 902859, 902860, 902861, 902862, 902864, 902865, 902866, 902868, 902869, 902870, 902871, 902872, 902873, 902874, 902875, 902876, 902877, 902878, 902879, 902880, 902881, 902882, 902883, 902884, 902885, 902886, 902887, 902888, 902889, 902890, 902891, 902893, 902894, 902895, 902896, 902897, 902898, 902899, 902900, 902901, 902902, 902903, 902904, 902905, 902906, 902907, 902908, 902909, 902910, 902911, 902912, 902913, 902914, 902916, 902917, 902918, 902919, 902921, 902922, 902923, 902924, 902925, 902926, 902927, 902928, 902929, 902930, 902931, 902932, 902933, 902934, 902935, 902936, 902937, 902938, 902939, 902940, 902941, 902942, 902943, 902944, 902945, 902946, 902947, 902948, 902949, 902950, 902951, 902952, 902953, 902954, 902955, 902956, 902957, 902958, 902959, 902960, 902961, 902962, 902963, 902964, 902965, 902966, 902967, 902968, 902969, 902970, 902971, 902972, 902973, 902974, 902975, 902976, 902977, 902979, 902980, 902981, 902982, 902983, 902984, 902985, 902986, 902987, 902988, 902989, 902990, 902991, 902992, 902993, 902994, 902995, 902996, 902997, 902998, 902999, 903000, 903001, 903002, 903003, 903004, 903005, 903006, 903007, 903008, 903009, 903010, 903011, 903012, 903014, 903015, 903016, 903017, 903018, 903019, 903020, 903021, 903022, 903023, 903024, 903025, 903026, 903027, 903028, 903029, 903030, 903031, 903032, 903033, 903034, 903035, 903036, 903037, 903039, 903040, 903041, 903042, 903043, 903044, 903045, 903046, 903047, 903048, 903049, 903050, 903051, 903052, 903193, 903194, 903195, 903196, 903197, 903198, 903199, 903200, 903201, 903202, 903203, 903204, 903206, 903207, 903208, 903209, 903210, 903211, 903212, 903214, 903215, 903216, 903217, 903219, 903220, 903221, 903222, 903223, 903224, 903225, 903226, 903227, 903228, 903230, 903231, 903232, 903233, 903234, 903235, 903236, 903237, 903238, 903239, 903240, 903241, 903242, 903243, 903244, 903245, 903246, 903247, 903248, 903249, 903250, 903251, 903252, 4000101, 4000102, 4000103, 4000104, 4000106, 4000107, 4000108, 4000109, 4000110, 4000111, 4000113, 4000114, 4000115, 4000116, 4000117, 4000118, 4000119, 4000120, 4000121, 4000122, 4000123, 4000124, 4000125, 4000126, 4000127, 4000128, 4000129, 4000130, 4000131, 4000132, 4005683, 4005684, 4005685, 4005686, 4005687, 4005688, 4005689, 4005690, 4005691, 4005692, 4005693, 4005694, 4005695, 4005696, 4005697, 4005698, 4005699, 4005700, 4005701, 4005702, 4005703, 4005704, 4005705, 4005706, 4005707, 4005708, 4005709, 4005710, 4005711, 4005712, 4005713, 4005714, 4005715, 4005716, 4005718, 4005719, 4005720, 4005721, 4005722, 4005723, 4005724, 4005725, 4005726, 4005727, 4005728, 4005729, 4005730, 4005731, 4005732, 4005733, 4005734, 4005735, 4005736, 4005737, 4005738, 4005739, 4005740, 4005741, 4005742, 4005743, 4005745, 4005746, 4005747, 4005748, 4005749, 4005750, 4005751, 4005752, 4005753, 4005754, 4005755, 4005756, 4005757, 4005758, 4005759, 4005760, 4005761, 4005762, 4005763, 4005764, 4005765, 4005766, 4005767, 4005768, 4005769, 4005770, 4005771, 4005772, 4005773, 4005774, 4005775, 4005776, 4005777, 4005778, 4005779, 4005780, 4005781, 4005782, 4006758, 4006759, 4006760, 4006761, 4006762, 4006763, 4006764, 4006765, 4006766, 4006767, 4006768, 4006769, 4006770, 4006772, 4006774, 4006775, 4006776, 4006777, 4006778, 4006779, 4006780, 4006781, 4006782, 4006783, 4006784, 4006785, 4006786, 4006787, 4006788, 4006789, 4006790, 4006791, 4006792, 4006794, 4006795, 4006796, 4006797, 4006798, 4006799, 4006800, 4006801, 4006802, 4006803, 4006805, 4006806, 4006807, 4006808, 4006809, 4006810, 4006811, 4006812, 4006813, 4006814, 4006815, 4006816, 4006817, 4006818, 4006819, 4006820, 4006821, 4006822, 4006823, 4006824, 4006825, 4006826, 4006827, 4006828, 4006829, 4006830, 4006831, 4006832, 4006833, 4006834, 4006835, 4006836, 4006837, 4006838, 4006839, 4006840, 4006841, 4006842, 4006843, 4006844, 4006845, 4006846, 4006847, 4006848, 4006849, 4006850, 4006851, 4006852, 4006853, 4006854, 4006855, 4006857, 4008703, 4008704, 4008705, 4008706, 4008707, 4008708, 4008709, 4008710, 4008711, 4008712, 4008713, 4008714, 4008715, 4008716, 4008717, 4008718, 4008719, 4008720, 4008721, 4008722, 4008723, 4008724, 4008725, 4008726, 4008727, 4008728, 4008729, 4008730, 4008731, 4008732, 4008733, 4008734, 4008735, 4008736, 4008737, 4008738, 4008739, 4008740, 4008741, 4008742, 4008743, 4008744, 4008745, 4008747, 4008748, 4008749, 4008750, 4008751, 4008752, 4013834, 4013835, 4013836, 4013837, 4013838, 4013839, 4013840, 4013841, 4013842, 4013843, 4013844, 4013845, 4013846, 4013847, 4013848, 4013849, 4013850, 4013851, 4013852, 4013853, 4013854, 4013855, 4013856, 4013857, 4013858, 4013859, 4013860, 4013861, 4013862, 4013863, 4013864, 4013865, 4013866, 4013867, 4013868, 4013870, 4013871, 4013872, 4013873, 4013874, 4013875, 4013876, 4013877, 4013879, 4013880, 4013882, 4013883, 4013884, 4013885, 4013886, 4013887, 4013888, 4013889, 4013890, 4013891, 4013892, 4013893, 4013894, 4013895, 4013896, 4013897, 4013898, 4013899, 4013900, 4013901, 4013902, 4013903, 4013904, 4013905, 4013906, 4013907, 4013908, 4013909, 4013910, 4013911, 4013912, 4013913, 4013914, 4013915, 4013916, 4013917, 4013918, 4013919, 4013920, 4013921, 4013922, 4013924, 4013925, 4013926, 4013927, 4013928, 4013929, 4013930, 4013931, 4013932, 4013933, 4014936, 4014937, 4014938, 4014939, 4014940, 4014941, 4014942, 4014943, 4014944, 4014945, 4014946, 4014947, 4014948, 4014949, 4014950, 4014952, 4014953, 4014954, 4014955, 4014956, 4014957, 4014958, 4014959, 4014960, 4014961, 4014962, 4014963, 4014964, 4014965, 4014966, 4014967, 4014968, 4014969, 4014970, 4014971, 4014972, 4014973, 4014974, 4014975, 4014976, 4014977, 4014978, 4014979, 4014980, 4014981, 4014982, 4014983, 4014984, 4016136, 4016137, 4016138, 4016140, 4016141, 4016142, 4016143, 4016144, 4016145, 4016146, 4016147, 4016148, 4016150, 4016151, 4016152, 4016153, 4016154, 4016155, 4016156, 4016157, 4016158, 4016159, 4016160, 4016161, 4016162, 4016163, 4016164, 4016165, 4016166, 4016167, 4016168, 4016169, 4016170, 4016171, 4016174, 4016175, 4016176, 4016177, 4016178, 4016179, 4016181, 4016182, 4016183, 4016184, 4016185, 4016186, 4016187, 4016188, 4016189, 4016190, 4016191, 4016192, 4016193, 4016194, 4016195, 4016196, 4016197, 4016198, 4016199, 4016200, 4016201, 4016203, 4016204, 4016205, 4016206, 4016207, 4016208, 4016209, 4016210, 4016211, 4016212, 4016213, 4016214, 4016215, 4016216, 4016217, 4016218, 4016219, 4016220, 4016221, 4016222, 4016223, 4016225, 4016226, 4016227, 4016228, 4016229, 4016230, 4016231, 4016232, 4016233, 4016234, 4016235, 4018998, 4018999, 4019000, 4019001, 4019002, 4019003, 4019004, 4019005, 4019006, 4019007, 4019008, 4019009, 4019010, 4019011, 4019012, 4019013, 4019014, 4019015, 4019016, 4019017, 4019018, 4019019, 4019020, 4019021, 4019022, 4019023, 4019024, 4019025, 4019026, 4019027, 4019028, 4019029, 4019030, 4019031, 4019032, 4019034, 4019035, 4019036, 4019037, 4019038, 4019039, 4019040, 4019041, 4019042, 4019043, 4019044, 4019045, 4019046, 4019047, 4019049, 4019050, 4019051, 4019052, 4019053, 4019054, 4019055, 4019056, 4019057, 4019058, 4019059, 4019060, 4019062, 4019063, 4019064, 4019065, 4019066, 4019067, 4019068, 4019069, 4019070, 4019071, 4019072, 4019073, 4019074, 4019075, 4019076, 4019077, 4019078, 4019079, 4019081, 4019082, 4019083, 4019084, 4019085, 4019086, 4019088, 4019089, 4019090, 4019091, 4019092, 4019093, 4019094, 4019095, 4019096, 4019097, 4027934, 4027935, 4027936, 4027937, 4027938, 4027939, 4027940, 4027941, 4027942, 4027943, 4027944, 4027945, 4027946, 4027947, 4027948, 4027949, 4027950, 4027951, 4027952, 4027953, 4027954, 4027955, 4027956, 4027957, 4027958, 4027959, 4027960, 4027961, 4027962, 4027963, 4027964, 4027965, 4027966, 4027967, 4027968, 4027969, 4027970, 4027971, 4027972, 4027973, 4027974, 4027975, 4027977, 4027978, 4027979, 4027980, 4027982, 4027983, 4027984, 4027985, 4027986, 4027987, 4027989, 4027990, 4027991, 4027992, 4027993, 4027994, 4027995, 4027996, 4027997, 4027999, 4028000, 4028001, 4028003, 4028004, 4028005, 4028006, 4028007, 4028008, 4028010, 4028011, 4028012, 4028014, 4028015, 4028016, 4028017, 4028018, 4028019, 4028020, 4028021, 4028022, 4028023, 4028024, 4028025, 4028026, 4028027, 4028028, 4028030, 4028033, 4028034, 4028036, 4028037, 4028038, 4028039, 4028040, 4028041, 4028043, 4028044, 4028045, 4028046, 4028047, 4028048, 4028050, 4028051, 4028056, 4028057, 4028058, 4028059, 4028060, 4028062, 4028063, 4028064, 4028065, 4028066, 4028067, 4028068, 4028069, 4028070, 4028071, 4028076, 4028077, 4028078, 4028079, 4028080, 4028081, 4028082, 4028083, 4028084, 4028085, 4028086, 4028087, 4028088, 4028089, 4028090, 4028091, 4028092, 4028093, 4028094, 4028095, 4028096, 4028097, 4028098, 4028099, 4028100, 4028101, 4028102, 4028103, 4028104, 4028105, 4028106, 4028107, 4028108, 4028109, 4028110, 4028111, 4028112, 4028113, 4028114, 4028115, 4028116, 4028117, 4028119, 4028120, 4028121, 4028122, 4028123, 4028124, 4028125, 4028126, 4028127, 4028128, 4028129, 4028130, 4028131, 4028132, 4028133, 4030859, 4030860, 4030861, 4030862, 4030863, 4030864, 4030865, 4030866, 4030867, 4030868, 4030869, 4030871, 4030873, 4030874, 4030875, 4030876, 4030877, 4030878, 4030879, 4030880, 4030881, 4030882, 4030883, 4030884, 4030886, 4030887, 4030888, 4030889, 4030890, 4030891, 4030892, 4030893, 4030894, 4030895, 4030896, 4030897, 4030899, 4030900, 4030901, 4030902, 4030903, 4030904, 4030905, 4030906, 4030907, 4030908, 4033088, 4033089, 4033090, 4033091, 4033092, 4033093, 4033094, 4033095, 4033096, 4033097, 4033098, 4033099, 4033100, 4033101, 4033102, 4033103, 4033104, 4033105, 4033106, 4033107, 4033108, 4033109, 4033110, 4033111, 4033112, 4033113, 4033114, 4033115, 4033116, 4033117, 4033118, 4033119, 4033120, 4033121, 4033122, 4033123, 4033124, 4033125, 4033126, 4033127, 4033128, 4033129, 4033130, 4033131, 4033132, 4033133, 4033134, 4033135, 4033136, 4033137, 4033138, 4033139, 4033140, 4033141, 4033142, 4033143, 4033144, 4033145, 4033146, 4033147, 4033148, 4033149, 4033150, 4033151, 4033152, 4033153, 4033154, 4033155, 4033156, 4033157, 4033159, 4033160, 4033161, 4033162, 4033163, 4033164, 4033165, 4033166, 4033167, 4033168, 4033169, 4033170, 4033171, 4033172, 4033173, 4033174, 4033175, 4033176, 4033177, 4033178, 4033179, 4033180, 4033181, 4033182, 4033183, 4033184, 4033185, 4033186, 4033187, 4033188, 4033189, 4033190, 4033191, 4033192, 4033193, 4033194, 4033195, 4033196, 4033197, 4033198, 4033199, 4033200, 4033201, 4033202, 4033203, 4033204, 4033205, 4033206, 4033207, 4033208, 4033209, 4033210, 4033211, 4033212, 4033213, 4033214, 4033215, 4033216, 4033217, 4033218, 4033219, 4033220, 4033221, 4033222, 4033223, 4033224, 4033226, 4033227, 4033228, 4033229, 4033230, 4033231, 4033232, 4033233, 4033234, 4033235, 4033236, 4033237, 4033238, 4033239, 4033240, 4033241, 4033242, 4033243, 4033244, 4033245, 4033246, 4033247, 4033248, 4033249, 4033250, 4033251, 4033252, 4033253, 4033254, 4033255, 4033256, 4033257, 4033258, 4033259, 4033260, 4033261, 4033262, 4033263, 4033264, 4033265, 4033266, 4033267, 4033268, 4033269, 4033270, 4033271, 4033272, 4033273, 4033274, 4033275, 4033276, 4033277, 4033278, 4033279, 4033280, 4033281, 4033282, 4033283, 4033284, 4033285, 4033286, 4033287, 4039833, 4039834, 4039835, 4039836, 4039837, 4039838, 4039839, 4039840, 4039841, 4039842, 4039843, 4039844, 4039845, 4039846, 4039847, 4039848, 4039849, 4039850, 4039851, 4039852, 4039853, 4039854, 4039855, 4039856, 4039857, 4039858, 4039859, 4039860, 4039861, 4039862, 4039863, 4039864, 4039865, 4039866, 4039867, 4039868, 4039869, 4039870, 4039871, 4039872, 4039873, 4039874, 4039875, 4039876, 4039877, 4039878, 4039879, 4039880, 4039881, 4039882, 4039883, 4039884, 4039885, 4039886, 4039887, 4039888, 4039889, 4039890, 4039891, 4039892, 4039893, 4039894, 4039895, 4039896, 4039897, 4039898, 4039899, 4039900, 4039901, 4039902, 4039903, 4039904, 4039905, 4039906, 4039907, 4039908, 4039909, 4039910, 4039911, 4039912, 4039913, 4039914, 4039915, 4039916, 4039917, 4039918, 4039919, 4039920, 4039921, 4039922, 4039923, 4039924, 4039925, 4039926, 4039927, 4039928, 4039929, 4039930, 4039931, 4039932, 4039933, 4039934, 4039935, 4039936, 4039937, 4039938, 4039939, 4039940, 4039941, 4039942, 4039943, 4039944, 4039945, 4039946, 4039947, 4039948, 4039949, 4039950, 4039951, 4039952, 4039953, 4039954, 4039955, 4039956, 4039957, 4039958, 4039959, 4039960, 4039961, 4039962, 4044139, 4044140, 4044141, 4044142, 4044143, 4044144, 4044145, 4044146, 4044147, 4044148, 4044149, 4044150, 4044151, 4044152, 4044153, 4044154, 4044155, 4044156, 4044157, 4044158, 4044159, 4044160, 4044161, 4044162, 4044163, 4044164, 4044165, 4044166, 4044167, 4044168, 4044169, 4044170, 4044171, 4044172, 4044173, 4044174, 4044175, 4044176, 4044177, 4044178, 4044179, 4044180, 4044181, 4044182, 4044183, 4044184, 4044185, 4044186, 4044187, 4044188, 4044189, 4044190, 4044191, 4044192, 4044193, 4044194, 4044195, 4044196, 4044197, 4044198, 4044199, 4044200, 4044201, 4044202, 4044203, 4044204, 4044205, 4044206, 4044207, 4044208, 4044209, 4044210, 4044211, 4044212, 4044213, 4044214, 4044215, 4044216, 4044217, 4044218, 4044219, 4044220, 4044221, 4044222, 4044223, 4044224, 4044225, 4044226, 4044227, 4044228, 4044229, 4044230, 4044231, 4044232, 4044233, 4044234, 4044235, 4044236, 4044237, 4044238, 4044239, 4044240, 4044241, 4044242, 4044243, 4044244, 4044245, 4044246, 4044247, 4044248, 4044249, 4044250, 4044251, 4044252, 4044253, 4044254, 4044255, 4044256, 4044257, 4044258, 4044259, 4044260, 4044261, 4044262, 4044263, 4044264, 4044265, 4044266, 4044267, 4044268, 4044269, 4044270, 4044271, 4044272, 4044273, 4044274, 4044275, 4044276, 4044277, 4044278, 4044279, 4044280, 4044281, 4044282, 4044283, 4044284, 4044285, 4044286, 4044287, 4044288, 4047839, 4047840, 4047841, 4047842, 4047843, 4047844, 4047845, 4047846, 4047847, 4047848, 4047849, 4047850, 4047851, 4047852, 4047853, 4047854, 4047855, 4047856, 4047857, 4047858, 4047859, 4047860, 4047861, 4047862, 4047863, 4047864, 4047865, 4047866, 4047867, 4047868, 4047869, 4047870, 4047871, 4047872, 4047873, 4047874, 4047875, 4047876, 4047877, 4047878, 4047879, 4047880, 4047881, 4047882, 4047883, 4047884, 4047885, 4047886, 4047887, 4047888, 4047889, 4047890, 4047891, 4047892, 4047893, 4047894, 4047895, 4047896, 4047897, 4047898, 4047899, 4047900, 4047901, 4047902, 4047903, 4047904, 4047905, 4047906, 4047907, 4047908, 4047909, 4047910, 4047911, 4047912, 4047913, 4047914, 4047915, 4047916, 4047917, 4047918, 4047919, 4047920, 4047921, 4047922, 4047923, 4047924, 4047925, 4047926, 4047927, 4047928, 4047929, 4047930, 4047931, 4047932, 4047933, 4047934, 4047935, 4047936, 4047937, 4047938, 4047939, 4047940, 4047942, 4047943, 4047944, 4047945, 4047946, 4047947, 4047948, 4048050, 4048051, 4048052, 4048053, 4048054, 4048055, 4048056, 4048057, 4048058, 4048059, 4048060, 4048061, 4048062, 4048063, 4048064, 4048065, 4048066, 4048067, 4048068, 4048069, 4048070, 4048071, 4048072, 4048073, 4048074, 4048075, 4048076, 4048077, 4048078, 4048079, 4048080, 4048081, 4048083, 4048084, 4048085, 4048086, 4048087, 4048088, 4048089, 4048090, 4048091, 4048092, 4048093, 4048094, 4048095, 4048096, 4048097, 4048098, 4048099, 4048100, 4048101, 4048102, 4048103, 4048104, 4048105, 4048106, 4048107, 4048108, 4048109, 4048110, 4048111, 4048112, 4048113, 4048114, 4048115, 4048116, 4048117, 4048118, 4048119, 4048120, 4048121, 4048122, 4048123, 4048124, 4048125, 4048126, 4048127, 4048128, 4048129, 4048130, 4048131, 4048132, 4048133, 4048134, 4048135, 4048136, 4048137, 4048138, 4048139, 4048140, 4048141, 4048142, 4048143, 4048144, 4048145, 4048146, 4048147, 4048148, 4048159, 4048160, 4048161, 4048162, 4048163, 4048174, 4048175, 4048176, 4048177, 4048178, 4048179, 4048180, 4048181, 4048182, 4048183, 4048184, 4048185, 4048186, 4048187, 4048188, 4048189, 4048190, 4048191, 4048192, 4048193, 4048264, 4048265, 4048266, 4048267, 4048268, 4048269, 4048270, 4048271, 4048272, 4048273, 4048274, 4048275, 4048276, 4048277, 4048278, 4048279, 4048280, 4048281, 4048282, 4048283, 4048284, 4048285, 4048286, 4048287, 4048288, 4048289, 4048290, 4048291, 4048292, 4048293, 4048294, 4048295, 4048296, 4048297, 4048298, 4048299, 4048300, 4048301, 4048302, 4048303, 4048304, 4048305, 4048306, 4048307, 4048308, 4048309, 4048310, 4048311, 4048312, 4048313, 4048314, 4048315, 4048316, 4048317, 4048318, 4048323, 4048324, 4048325, 4048326, 4048327, 4048328, 4048329, 4048330, 4048331, 4048332, 4048333, 4048334, 4048335, 4048336, 4048337, 4048338, 4048339, 4048340, 4048341, 4048342, 4048343, 4048344, 4048345, 4048346, 4048347, 4048348, 4048349, 4048350, 4048351, 4048352, 4048353, 4048354, 4048355, 4048356, 4048357, 4048358, 4048359, 4048360, 4048361, 4048362, 4048363, 4048364, 4048365, 4048366, 4048367, 4048368, 4048369, 4048370, 4048371, 4048372, 4048373, 4048374, 4048375, 4048376, 4048377, 4048378, 4048379, 4048380, 4048381, 4048382, 4048383, 4048384, 4048385, 4048386, 4048387, 4048388, 4048389, 4048390, 4048391, 4048392, 4048393, 4048394, 4048395, 4048396, 4048397, 4048398, 4048399, 4048400, 4048401, 4048402, 4048403, 4048404, 4048405, 4048406, 4048407, 4048408, 4048409, 4048410, 4048412, 4048413, 4048414, 4048415, 4048416, 4048417, 4048418, 4048419, 4048420, 4048421, 4048422, 4048423, 4048424, 4048425, 4048426, 4048427, 4048428, 4048429, 4048430, 4048432, 4048433, 4048434, 4048435, 4048436, 4048437, 4048438, 4048439, 4048440, 4048441, 4048442, 4048443, 4048444, 4048445, 4048446, 4048447, 4048448, 4048449, 4048450, 4048451, 4048452, 4048453, 4048454, 4048455, 4048456, 4048457, 4048458, 4048459, 4048460, 4048461, 4048462, 4048463, 4048464, 4048465, 4048466, 4048468, 4048470, 4048471, 4048472, 4048523, 4048524, 4048525, 4048526, 4048527, 4048528, 4048529, 4048530, 4048531, 4048532, 4048533, 4048534, 4048535, 4048536, 4048537, 4048538, 4048539, 4048540, 4048541, 4048542, 4048543, 4048544, 4048545, 4048546, 4048547, 4048548, 4048549, 4048550, 4048551, 4048552, 4048553, 4048554, 4048555, 4048556, 4048557, 4048558, 4048559, 4048560, 4048561, 4048562, 4048563, 4048564, 4048565, 4048566, 4048568, 4048569, 4048570, 4048571, 4048572, 4048573, 4048574, 4048575, 4048576, 4048577, 4048578, 4048579, 4048580, 4048581, 4048582, 4048583, 4048584, 4048585, 4048586, 4048587, 4048588, 4048589, 4048590, 4048591, 4048592, 4048593, 4048594, 4048595, 4048596, 4048597, 4048598, 4048599, 4048600, 4048601, 4048602, 4048603, 4048604, 4048606, 4048607, 4048608, 4048609, 4048610, 4048611, 4048612, 4048613, 4048614, 4048615, 4048616, 4048617, 4048618, 4048619, 4048620, 4048621, 4048622, 4048623, 4048624, 4048625, 4048626, 4048627, 4048628, 4048629, 4048630, 4048631, 4048632, 4048633, 4048634, 4048635, 4048636, 4048637, 4048638, 4048639, 4048640, 4048641, 4048642, 4048643, 4048644, 4048645, 4048646, 4048647, 4048648, 4048649, 4048650, 4048651, 4048652, 4048653, 4048654, 4048655, 4048656, 4048657, 4048658, 4048659, 4048660, 4048661, 4048662, 4048663, 4048664, 4048665, 4048666, 4048667, 4048668, 4048669, 4048670, 4048671, 4048672, 4048673, 4048674, 4048675, 4048676, 4048677, 4048678, 4048679, 4048680, 4048681, 4048682, 4048683, 4048684, 4048685, 4048686, 4048687, 4048688, 4048689, 4048690, 4048691, 4048692, 4048693, 4048694, 4048695, 4048696, 4048697, 4048698, 4048699, 4048700, 4048701, 4048702, 4048703, 4048704, 4048705, 4048706, 4048707, 4048708, 4048709, 4048710, 4048711, 4048712, 4048713, 4048714, 4048715, 4048716, 4048717, 4048718, 4048719, 4048720, 4048721, 4048722, 4048723, 4048724, 4048725, 4048726, 4048727, 4048728, 4048729, 4048730, 4048731, 4048732, 4048733, 4048734, 4048735, 4048736, 4048737, 4048738, 4048739, 4048740, 4048741, 4048742, 4048743, 4048744, 4048745, 4048746, 4048747, 4048748, 4048749, 4048750, 4048751, 4048752, 4048753, 4048754, 4048755, 4048756, 4048757, 4048758, 4048759, 4048760, 4048761, 4048762, 4048763, 4048764, 4048765, 4048766, 4048767, 4048768, 4048769, 4048770, 4048771, 4048772, 4048823, 4048824, 4048825, 4048826, 4048827, 4048828, 4048829, 4048830, 4048831, 4048832, 4048833, 4048834, 4048835, 4048836, 4048837, 4048838, 4048839, 4048840, 4048841, 4048842, 4048843, 4048844, 4048845, 4048846, 4048847, 4048848, 4048849, 4048850, 4048851, 4048852, 4048853, 4048854, 4048855, 4048856, 4048857, 4048858, 4048859, 4048860, 4048861, 4048862, 4048863, 4048864, 4048865, 4048866, 4048867, 4048868, 4048869, 4048870, 4048871, 4048872, 4048873, 4048874, 4048875, 4048876, 4048877, 4048878, 4048879, 4048880, 4048881, 4048882, 4048883, 4048885, 4048886, 4048887, 4048888, 4048889, 4048890, 4048891, 4048892, 4048893, 4048894, 4048895, 4048896, 4048897, 4048898, 4048899, 4048900, 4048901, 4048902, 4048903, 4048904, 4048905, 4048906, 4048907, 4048908, 4048909, 4048910, 4048911, 4048912, 4048913, 4048914, 4048915, 4048916, 4048917, 4048918, 4048919, 4048920, 4048921, 4048922, 4048923, 4048924, 4048925, 4048926, 4048927, 4048928, 4048929, 4048930, 4048931, 4048932, 4048933, 4048934, 4048935, 4048936, 4048937, 4048938, 4048939, 4048940, 4048941, 4048942, 4048943, 4048944, 4048945, 4048946, 4048947, 4048948, 4048949, 4048950, 4048951, 4048952, 4048953, 4048954, 4048955, 4048956, 4048957, 4048959, 4048960, 4048961, 4048962, 4048963, 4048964, 4048965, 4048966, 4048967, 4048968, 4048969, 4048970, 4048971, 4048972, 4049023, 4049024, 4049025, 4049026, 4049027, 4049028, 4049029, 4049030, 4049031, 4049032, 4049033, 4049034, 4049035, 4049036, 4049037, 4049038, 4049039, 4049040, 4049041, 4049042, 4049043, 4049044, 4049045, 4049046, 4049047, 4049048, 4049049, 4049050, 4049051, 4049052, 4049053, 4049054, 4049055, 4049056, 4049057, 4049058, 4049059, 4049060, 4049061, 4049062, 4049063, 4049064, 4049065, 4049066, 4049067, 4049068, 4049069, 4049070, 4049071, 4049072, 4049073, 4049074, 4049075, 4049076, 4049077, 4049078, 4049079, 4049080, 4049081, 4049082, 4049083, 4049084, 4049085, 4049086, 4049087, 4049088, 4049089, 4049090, 4049091, 4049092, 4049093, 4049094, 4049095, 4049096, 4049097, 4049098, 4049099, 4049100, 4049101, 4049102, 4049103, 4049104, 4049105, 4049106, 4049107, 4049108, 4049109, 4049110, 4049111, 4049112, 4049113, 4049114, 4049115, 4049116, 4049118, 4049119, 4049120, 4049121, 4049122, 4049123, 4049124, 4049125, 4049126, 4049127, 4049128, 4049129, 4049130, 4049131, 4049132, 4049133, 4049134, 4049135, 4049136, 4049137, 4049138, 4049139, 4049140, 4049141, 4049142, 4049143, 4049144, 4049145, 4049146, 4049147, 4049148, 4049149, 4049150, 4049151, 4049152, 4049153, 4049154, 4049155, 4049156, 4049157, 4049158, 4049159, 4049160, 4049161, 4049162, 4049163, 4049164, 4049165, 4049166, 4049167, 4049168, 4049169, 4049170, 4049171, 4049172, 4049173, 4049174, 4049175, 4049176, 4049177, 4049178, 4049179, 4049180, 4049181, 4049182, 4049183, 4049184, 4049185, 4049186, 4049187, 4049188, 4049189, 4049190, 4049191, 4049192, 4049193, 4049194, 4049195, 4049196, 4049197, 4049198, 4049199, 4049200, 4049201, 4049202, 4049203, 4049204, 4049205, 4049206, 4049207, 4049208, 4049209, 4049210, 4049211, 4049212, 4049213, 4049214, 4049215, 4049216, 4049217, 4049218, 4049219, 4049220, 4049221, 4049222, 4049223, 4049224, 4049225, 4049226, 4049227, 4049228, 4049229, 4049230, 4049231, 4049232, 4049233, 4049234, 4049235, 4049236, 4049237, 4049239, 4049240, 4049241, 4049242, 4049243, 4049244, 4049245, 4049246, 4049247, 4049248, 4049249, 4049250, 4049251, 4049252, 4049253, 4049254, 4049255, 4049256, 4049257, 4049258, 4049259, 4049260, 4049261, 4049262, 4049263, 4049264, 4049265, 4049266, 4049267, 4049268, 4049269, 4049270, 4049271, 4049272, 4146763, 4146764, 4146765, 4146766, 4146767, 4146768, 4146769, 4146770, 4146771, 4146772, 4146773, 4146774, 4146775, 4146776, 4146777, 4146778, 4146779, 4146780, 4146781, 4146782, 4146783, 4146784, 4146785, 4146786, 4146787, 4146788, 4146789, 4146790, 4146791, 4146792, 4146793, 4146794, 4146795, 4146796, 4146797, 4146798, 4146799, 4146800, 4146801, 4146802, 4146803, 4146804, 4146805, 4146806, 4146807, 4146808, 4146809, 4146810, 4146811, 4146812, 4146813, 4146814, 4146815, 4146816, 4146818, 4146819, 4146820, 4146821, 4146822, 4146823, 4146824, 4146825, 4146826, 4146827, 4146828, 4146829, 4146830, 4146831, 4146832, 4146833, 4146834, 4146835, 4146836, 4146837, 4146838, 4146839, 4146840, 4146841, 4146842, 4146843, 4146844, 4146845, 4146846, 4146847, 4146848, 4146849, 4146850, 4146851, 4146852, 4146853, 4146854, 4146855, 4146856, 4146857, 4146858, 4146859, 4146860, 4146861, 4146862, 4146863, 4146864, 4146865, 4146866, 4146867, 4146868, 4146869, 4146870, 4146871, 4146872, 4146873, 4146874, 4146875, 4146876, 4146877, 4146878, 4146879, 4146880, 4146881, 4146882, 4146883, 4146884, 4146885, 4146886, 4146887, 4146888, 4146889, 4146890, 4146891, 4146892, 4146893, 4146894, 4146895, 4146897, 4146898, 4146899, 4146900, 4146901, 4146902, 4146903, 4146904, 4146905, 4146906, 4146907, 4146908, 4146909, 4146910, 4146912, 4146913, 4146914, 4146915, 4146916, 4146918, 4146921, 4146922, 4146923, 4146924, 4146925, 4146926, 4146927, 4146928, 4146929, 4146930, 4146931, 4146932, 4146933, 4146934, 4146935, 4146936, 4146937, 4146938, 4146939, 4146940, 4146941, 4146942, 4146943, 4146944, 4146945, 4146946, 4146947, 4146948, 4146949, 4146950, 4146951, 4146952, 4146953, 4146954, 4146955, 4146956, 4146957, 4146958, 4146959, 4146960, 4146961, 4146962, 4151013, 4151014, 4151015, 4151016, 4151017, 4151018, 4151019, 4151020, 4151021, 4151022, 4151023, 4151024, 4151025, 4151026, 4151027.4151028, 4151029, 4151030, 4151031, 4151032, 4151033, 4151034, 4151035, 4151036, 4151037, 4151038, 4151039, 4151040, 4151041, 4151042, 4151043, 4151044, 4151045, 4151046, 4151047, 4151048, 4151049, 4151050, 4151051, 4151053, 4151054, 4151055, 4151056, 4151057, 4151058, 4151059, 4151060, 4151061, 4151063, 4151064, 4151065, 4151066, 4151067, 4151068, 4151069, 4151070, 4151071, 4151072, 4151073, 4151074, 4151075, 4151076, 4151077, 4151078, 4151079, 4151080, 4151081, 4151082, 4151083, 4151084, 4151085, 4151087, 4151088, 4151089, 4151090, 4151092, 4151093, 4151094, 4151095, 4151096, 4151097, 4151098, 4151099, 4151100, 4151101, 4151102, 4151103, 4151104, 4151105, 4151106, 4151107, 4151108, 4151109, 4151110, 4151111, 4151112, 4151113, 4151114, 4151115, 4151116, 4151117, 4151118, 4151119, 4151120, 4151121, 4151122, 4151123, 4151124, 4151125, 4151126, 4151127, 4151128, 4151129, 4151130, 4151131, 4151132, 4151133, 4151134, 4151135, 4151136, 4151137, 4151138, 4151139, 4151140, 4151141, 4151142, 4151143, 4151144, 4151145, 4151146, 4151147, 4151148, 4151149, 4151150, 4151151, 4151152, 4151153, 4151154, 4151155, 4151156, 4151157, 4151158, 4151159, 4151160, 4151161, 4151162, 4151163, 4151164, 4151165, 4151166, 4151167, 4151168, 4151169, 4151170, 4151171, 4151172, 4151173, 4151174, 4151175, 4151176, 4151177, 4151178, 4151179, 4151180, 4151181, 4151182, 4151183, 4151184, 4151185, 4151186, 4151187, 4151188, 4151189, 4151190, 4151191, 4151192, 4151193, 4151194, 4151195, 4151196, 4151197, 4151198, 4151199, 4151200, 4151201, 4151202, 4151203, 4151204, 4151205, 4151206, 4151207, 4151208, 4151209, 4151210, 4151211, 4151212, 4156566, 4156567, 4156568, 4156569, 4156570, 4156571, 4156572, 4156573, 4156574, 4156575, 4156576, 4156577, 4156578, 4156579, 4156580, 4156581, 4156582, 4156583, 4156584, 4156585, 4156586, 4156587, 4156588, 4156589, 4156590, 4156591, 4156592, 4156593, 4156594, 4156595, 4156596, 4156597, 4156598, 4156599, 4156600, 4156601, 4156602, 4156603, 4156604, 4156605, 4156606, 4156608, 4156610, 4156611, 4156612, 4156613, 4156614, 4156615, 4156616, 4156617, 4156618, 4156619, 4156620, 4156621, 4156622, 4156623, 4156624, 4156625, 4156626, 4156627, 4156628, 4156629, 4156630, 4156631, 4156632, 4156633, 4156634, 4156635, 4156636, 4156637, 4156638, 4156639, 4156640, 4156641, 4156642, 4156643, 4156644, 4156645, 4156646, 4156647, 4156648, 4156649, 4156650, 4156651, 4156652, 4156653, 4156654, 4156655, 4156656, 4156657, 4156658, 4156659, 4156660, 4156661, 4156662, 4156663, 4156664, 4156665, 4156666, 4156667, 4156668, 4156669, 4156670, 4156671, 4156672, 4156673, 4156674, 4156675, 4156676, 4156677, 4156678, 4156679, 4156680, 4156681, 4156682, 4156683, 4156684, 4156685, 4156686, 4156687, 4156688, 4156689, 4156690, 4156691, 4156692, 4156693, 4156694, 4156695, 4156696, 4156697, 4156698, 4156699, 4156700, 4156701, 4156702, 4156703, 4156704, 4156705, 4156706, 4156707, 4156708, 4156709, 4156710, 4156711, 4156712, 4156713, 4156714, 4156715, 4156716, 4156717, 4156718, 4156719, 4156720, 4156721, 4156722, 4156723, 4156724, 4156725, 4156726, 4156727, 4156728, 4156729, 4156730, 4156731, 4156732, 4156733, 4156734, 4156735, 4156736, 4156737, 4156738, 4156739, 4156740, 4156741, 4156742, 4156743, 4156744, 4156745, 4156746, 4156747, 4156748, 4156749, 4156750, 4156751, 4156754, 4156755, 4156756, 4156757, 4156758, 4156759, 4156760, 4156761, 4156762, 4156763, 4156764, 4156765, 4156766, 4156767, 4156768, 4156769, 4156770, 4156771, 4156772, 4156773, 4156774, 4156775, 4156776, 4156777, 4156778, 4156779, 4156780, 4156781, 4156782, 4156783, 4156784, 4156785, 4156786, 4156787, 4156788, 4156789, 4156790, 4156791, 4156792, 4156793, 4156794, 4156795, 4156796, 4156797, 4156798, 4156799, 4156800, 4156801, 4156802, 4156803, 4156804, 4156805, 4156806, 4156807, 4156808, 4156809, 4156810, 4156811, 4156812, 4156813, 4156814, 4156815, 4156816, 4156817, 4156818, 4156819, 4156820, 4156821, 4156822, 4156823, 4156824, 4156825, 4156826, 4156827, 4156828, 4156829, 4156830, 4156831, 4156832, 4156833, 4156834, 4156835, 4156836, 4156837, 4156838, 4156839, 4156840, 4156841, 4156842, 4156843, 4156844, 4156845, 4156846, 4156847, 4156848, 4156849, 4156850, 4156851, 4156852, 4156853, 4156854, 4156855, 4156856, 4156857, 4156858, 4156859, 4156860, 4156861, 4156862, 4156863, 4156864, 4156865, 4167526, 4167527, 4167528, 4167529, 4167530, 4167531, 4167532, 4167533, 4167534, 4167535, 4167536, 4167537, 4167538, 4167539, 4167540, 4167541, 4167542, 4167543, 4167544, 4167545, 4167546, 4167547, 4167548, 4167549, 4167550, 4167551, 4167552, 4167553, 4167554, 4167555, 4167556, 4167557, 4167558, 4167559, 4167560, 4167561, 4167562, 4167563, 4167564, 4167565, 4167566, 4167567, 4167568, 4167569, 4167570, 4167571, 4167572, 4167573, 4167574, 4167575, 4167576, 4167577, 4167578, 4167579, 4167580, 4167581, 4167582, 4167583, 4167584, 4167585, 4167586, 4167587, 4167588, 4167589, 4167590, 4167591, 4167592, 4167593, 4167594, 4167595, 4167596, 4167597, 4167599, 4167600, 4167601, 4167602, 4167603, 4167604, 4167605, 4167606, 4167607, 4167608, 4167609, 4167610, 4167611, 4167612, 4167613, 4167614, 4167615, 4167616, 4167617, 4167618, 4167619, 4167620, 4167621, 4167622, 4167623, 4167624, 4167625, 4174701, 4174702, 4174703, 4174704, 4174705, 4174706, 4174707, 4174708, 4174709, 4174710, 4174711, 4174712, 4174713, 4174714, 4174715, 4174716, 4174717, 4174718, 4174719, 4174720, 4174721, 4174722, 4174723, 4174724, 4174725, 4174726, 4174727, 4174728, 4174729, 4174730, 4174731, 4174732, 4174733, 4174734, 4174735, 4174736, 4174737, 4174738, 4174739, 4174740, 4174741, 4174742, 4174743, 4174744, 4174745, 4174746, 4174747, 4174748, 4174749, 4174750, 4174751, 4174752, 4174753, 4174754, 4174755, 4174756, 4174757, 4174758, 4174759, 4174760, 4174761, 4174762, 4174763, 4174764, 4174765, 4174766, 4174768, 4174769, 4174770, 4174771, 4174772, 4174773, 4174774, 4174775, 4174776, 4174777, 4174778, 4174779, 4174780, 4174781, 4174783, 4174784, 4174785, 4174786, 4174787, 4174788, 4174789, 4174791, 4174792, 4174793, 4174795, 4174796, 4174797, 4174798, 4174799, 4174800, 4174801, 4174802, 4174803, 4174804, 4174805, 4174806, 4174807, 4174808, 4174809, 4174810, 4174811, 4174812, 4174813, 4174814, 4174815, 4174816, 4174817, 4174818, 4174819, 4174820, 4174821, 4174822, 4174823, 4174824, 4174825, 4174826, 4174827, 4174828, 4174829, 4174830, 4174831, 4174832, 4174833, 4174834, 4174835, 4174836, 4174837, 4174838, 4174839, 4174840, 4174841, 4174842, 4174843, 4174844, 4174845, 4174846, 4174847, 4174848, 4174849, 4174850, 4174851, 4174852, 4174853, 4174854, 4174855, 4174856, 4174857, 4174858, 4174859, 4174860, 4174861, 4174862, 4174863, 4174864, 4174865, 4174866, 4174867, 4174868, 4174869, 4174870, 4174871, 4174872, 4174873, 4174874, 4174875, 4174876, 4174877, 4174878, 4174879, 4174880, 4174881, 4174882, 4174883, 4174884, 4174885, 4174886, 4174887, 4174888, 4174889, 4174890, 4174891, 4174892, 4174893, 4174894, 4174895, 4174896, 4174897, 4174898, 4174899, 4174900, 4177751, 4177752, 4177753, 4177754, 4177755, 4177756, 4177757, 4177758, 4177759, 4177760, 4177761, 4177762, 4177763, 4177764, 4177765, 4177766, 4177767, 4177768, 4177769, 4177770, 4177771, 4177772, 4177773, 4177774, 4177775, 4177776, 4177778, 4177779, 4177780, 4177781, 4177782, 4177783, 4177784, 4177785, 4177786, 4177787, 4177788, 4177789, 4177790, 4177791, 4177792, 4177793, 4177794, 4177795, 4177796, 4177797, 4177798, 4177799, 4177801, 4177802, 4177803, 4177804, 4177805, 4177806, 4177807, 4177808, 4177809, 4177810, 4177811, 4177812, 4177813, 4177814, 4177815, 4177816, 4177817, 4177818, 4177819, 4177820, 4177821, 4177822, 4177823, 4177824, 4177825, 4177826, 4177827, 4177828, 4177829, 4177830, 4177831, 4177832, 4177833, 4177834, 4177835, 4177836, 4177837, 4177838, 4177839, 4177840, 4177841, 4177842, 4177843, 4177844, 4177845, 4177846, 4177847, 4177848, 4177849, 4177850, 4177851, 4177852, 4177853, 4177854, 4177855, 4177856, 4177857, 4177858, 4177859, 4177860, 4177861, 4177862, 4177863, 4177864, 4177865, 4177866, 4177867, 4177868, 4177869, 4177870, 4177871, 4177872, 4177874, 4177875, 4177876, 4177877, 4177878, 4177879, 4177880, 4177881, 4177882, 4177883, 4177884, 4177885, 4177886, 4177887, 4177888, 4177889, 4177890, 4177891, 4177892, 4177893, 4177894, 4177895, 4177896, 4177897, 4177898, 4177899, 4177900, 4177901, 4177902, 4177903, 4177904, 4177905, 4177906, 4177907, 4177908, 4177909, 4177910, 4177911, 4177912, 4177913, 4177914, 4177915, 4177916, 4177917, 4177918, 4177919, 4177920, 4177921, 4177922, 4177923, 4177924, 4177925, 4177926, 4177927, 4177928, 4177929, 4177930, 4177931, 4177932, 4177933, 4177934, 4177935, 4177936, 4177937, 4177938, 4177939, 4177941, 4177942, 4177943, 4177944, 4177945, 4177946, 4177948, 4177949, 4177950, 4187100, 4187101, 4187102, 4187103, 4187104, 4187105, 4187106, 4187107, 4187108, 4187109, 4187110, 4187111, 4187112, 4187113, 4187114, 4187115, 4187116, 4187117, 4187118, 4187119, 4187120, 4187121, 4187122, 4187123, 4187124, 4187125, 4187126, 4187127, 4187128, 4187129, 4187130, 4187131, 4187132, 4187133, 4187134, 4187135, 4187136, 4187137, 4187138, 4187139, 4187140, 4187141, 4187142, 4187143, 4187144, 4187145, 4187146, 4187147, 4187148, 4187149, 4187150, 4187151, 4187152, 4187153, 4187154, 4187155, 4187156, 4187157, 4187158, 4187159, 4187160, 4187161, 4187162, 4187163, 4187164, 4187165, 4187166, 4187167, 4187168, 4187169, 4187170, 4187171, 4187172, 4187173, 4187174, 4187175, 4187176, 4187177, 4187178, 4187179, 4187180, 4187182, 4187183, 4187184, 4187185, 4187186, 4187188, 4187189, 4187190, 4187191, 4187192, 4187193, 4187194, 4187195, 4187196, 4187197, 4187198, 4187199, 4187200, 4187201, 4187202, 4187203, 4187204, 4187205, 4187206, 4187207, 4187208, 4187209, 4187210, 4187211, 4187212, 4187213, 4187214, 4187215, 4187216, 4187217, 4187218, 4187219, 4187220, 4187221, 4187222, 4187223, 4187224, 4187225, 4187226, 4187227, 4187228, 4187229, 4187230, 4187231, 4187232, 4187233, 4187234, 4187235, 4187236, 4187237, 4187238, 4187239, 4187240, 4187241, 4187242, 4187243, 4187244, 4187245, 4187246, 4187247, 4187248, 4187249, 4187250, 4187251, 4187252, 4187253, 4187254, 4187255, 4187256, 4187257, 4187258, 4187259, 4187260, 4187261, 4187262, 4187263, 4187264, 4187265, 4187266, 4187267, 4187268, 4187269, 4187270, 4187271, 4187272, 4187273, 4187274, 4187275, 4187276, 4187277, 4187278, 4187279, 4187280, 4187281, 4187282, 4187283, 4187284, 4187285, 4187286, 4187287, 4187288, 4187289, 4187290, 4187291, 4187292, 4187293, 4187294, 4187295, 4187296, 4187297, 4187298, 4187299, 4187300, 4187301, 4187302, 4187303, 4187304, 4187305, 4187306, 4187307, 4187308, 4187309, 4187310, 4187311, 4187312, 4187313, 4187314, 4187315, 4187316, 4187317, 4187318, 4187319, 4187320, 4187321, 4187322, 4187323, 4187324, 4187325, 4187326, 4193327, 4193328, 4193329, 4193330, 4193331, 4193332, 4193333, 4193334, 4193335, 4193336, 4193337, 4193338, 4193339, 4193340, 4193341, 4193342, 4193343, 4193344, 4193345, 4193346, 4193347, 4193348, 4193349, 4193350, 4193351, 4193352, 4193353, 4193354, 4193355, 4193356, 4193357, 4193358, 4193359, 4193360, 4193361, 4193362, 4193363, 4193364, 4193365, 4193366, 4193367, 4193368, 4193369, 4193370, 4193371, 4193372, 4193373, 4193374, 4193375, 4193376, 4193377, 4193378, 4193379, 4193380, 4193381, 4193382, 4193383, 4193384, 4193385, 4193386, 4193387, 4193388, 4193389, 4193390, 4193391, 4193392, 4193393, 4193394, 4193395, 4193396, 4193397, 4193398, 4193399, 4193400, 4193401, 4193402, 4193403, 4193404, 4193405, 4193406, 4193407, 4193408, 4193409, 4193410, 4193411, 4193412, 4193413, 4193414, 4193415, 4193416, 4193417, 4193418, 4193419, 4193420, 4193421, 4193422, 4193423, 4193424, 4193425, 4193426, 4200128, 4200129, 4200130, 4200131, 4200132, 4200133, 4200134, 4200135, 4200136, 4200137, 4200138, 4200139, 4200140, 4200141, 4200142, 4200143, 4200144, 4200145, 4200146, 4200147, 4200148, 4200149, 4200150, 4200151, 4200152, 4200153, 4200154, 4200155, 4200156, 4200157, 4200158, 4200159, 4200160, 4200161, 4200162, 4200163, 4200164, 4200165, 4200166, 4200167, 4200168, 4200169, 4200170, 4200171, 4200172, 4200173, 4200174, 4200175, 4200176, 4200177, 4200178, 4200179, 4200180, 4200181, 4200182, 4200183, 4200184, 4200185, 4200186, 4200187, 4200188, 4200189, 4200190, 4200191, 4200192, 4200193, 4200194, 4200195, 4200196, 4200197, 4200198, 4200199, 4200200, 4200201, 4200202, 4200203, 4200204, 4200205, 4200206, 4200207, 4200208, 4200209, 4200210, 4200211, 4200212, 4200213, 4200214, 4200215, 4200216, 4200217, 4200218, 4200219, 4200220, 4200221, 4200222, 4200223, 4200224, 4200225, 4200226, 4200227, 4200228, 4200229, 4200230, 4200231, 4200232, 4200233, 4200234, 4200235, 4200236, 4200237, 4200238, 4200239, 4200240, 4200241, 4200242, 4200243, 4200244, 4200245, 4200246, 4200247, 4200248, 4200249, 4200250, 4200251, 4200252, 4200253, 4200254, 4200255, 4200256, 4200257, 4200258, 4200259, 4200260, 4200261, 4200262, 4200263, 4200264, 4200265, 4200266, 4200267, 4200268, 4200269, 4200270, 4200271, 4200272, 4200273, 4200274, 4200275, 4200276, 4200277, 4200278, 4200279, 4200280, 4200281, 4200282, 4200283, 4200284, 4200285, 4200286, 4200287, 4200288, 4200289, 4200290, 4200291, 4200292, 4200293, 4200294, 4200295, 4200296, 4200297, 4200298, 4200299, 4200300, 4200301, 4200302, 4200303, 4200304, 4200305, 4200306, 4200307, 4200308, 4200309, 4200310, 4200311, 4200312, 4200313, 4200314, 4200315, 4200316, 4200317, 4200318, 4200319, 4200320, 4200321, 4200322, 4200323, 4200324, 4200325, 4200326, 4200327, 4201328, 4201329, 4201330, 4201331, 4201332, 4201333, 4201334, 4201335, 4201336, 4201337, 4201338, 4201339, 4201340, 4201341, 4201342, 4201343, 4201344, 4201345, 4201346, 4201347, 4201348, 4201349, 4201350, 4201351, 4201352, 4201353, 4201354, 4201355, 4201356, 4201357, 4201358, 4201359, 4201360, 4201361, 4201362, 4201363, 4201364, 4201365, 4201367, 4201368, 4201369, 4201370, 4201372, 4201373, 4201374, 4201375, 4201376, 4201377, 4201378, 4201379, 4201380, 4201381, 4201382, 4201383, 4201384, 4201385, 4201386, 4201387, 4201388, 4201389, 4201390, 4201391, 4201392, 4201393, 4201394, 4201395, 4201396, 4201397, 4201398, 4201399, 4201400, 4201401, 4201402, 4201403, 4201404, 4201405, 4201406, 4201407, 4201408, 4201409, 4201410, 4201411, 4201412, 4201413, 4201414, 4201415, 4201416, 4201417, 4201418, 4201419, 4201420, 4201421, 4201422, 4201423, 4201424, 4201425, 4201426, 4201427, 4210905, 4210906, 4210907, 4210908, 4210909, 4210910, 4210911, 4210912, 4210913, 4210914, 4210915, 4210916, 4210917, 4210918, 4210919, 4210920, 4210921, 4210922, 4210923, 4210924, 4210925, 4210926, 4210927, 4210928, 4210929, 4210930, 4210931, 4210932, 4210933, 4210934, 4210935, 4210936, 4210937, 4210938, 4210939, 4210940, 4210941, 4210942, 4210943, 4210944, 4210945, 4210946, 4210947, 4210948, 4210949, 4210950, 4210951, 4210952, 4210954, 4210956, 4210957, 4210958, 4210959, 4210960, 4210961, 4210962, 4210963, 4210964, 4210965, 4210966, 4210967, 4210968, 4210969, 4210970, 4210971, 4210972, 4210973, 4210974, 4210975, 4210976, 4210977, 4210978, 4210979, 4210980, 4210981, 4210982, 4210983, 4210984, 4210985, 4210986, 4210987, 4210988, 4210989, 4210990, 4210991, 4210992, 4210993, 4210994, 4210995, 4210996, 4210997, 4210998, 4210999, 4211000, 4211001, 4211002, 4211003, 4211004, 4211005, 4211006, 4211007, 4211008, 4211009, 4211010, 4211011, 4211012, 4211013, 4211014, 4211015, 4211016, 4211017, 4211018, 4211019, 4211020, 4211021, 4211022, 4211023, 4211024, 4211025, 4211026, 4211027, 4211028, 4211029, 4211030, 4211031, 4211032, 4211033, 4211034, 4211035, 4211036, 4211037, 4211038, 4211039, 4211040, 4211041, 4211042, 4211043, 4211044, 4211045, 4211046, 4211047, 4211048, 4211049, 4211050, 4211051, 4211052, 4211053, 4211054, 4211055, 4211056, 4211057, 4211058, 4211059, 4211060, 4211061, 4211062, 4211063, 4211064, 4211065, 4211066, 4211067, 4211068, 4211069, 4211070, 4211071, 4211072, 4211073, 4211074, 4211075, 4211076, 4211077, 4211078, 4211079, 4211080, 4211081, 4211082, 4211083, 4211084, 4211085, 4211086, 4211087, 4211088, 4211089, 4211090, 4211091, 4211092, 4211093, 4211094, 4211095, 4211096, 4211097, 4211098, 4211099, 4211100, 4211101, 4211102, 4211103, 4211104, 4211105, 4211106, 4211107, 4211108, 4211109, 4211110, 4211111, 4211112, 4211113, 4211114, 4211115, 4211116, 4211117, 4211118, 4211119, 4211120, 4211121, 4211122, 4211123, 4211124, 4211125, 4211126, 4211127, 4211128, 4211129, 4211130, 4211131, 4211132, 4211133, 4211134, 4211135, 4211136, 4211137, 4211138, 4211139, 4211140, 4211141, 4211142, 4211143, 4211144, 4211146, 4211147, 4211148, 4211149, 4211150, 4211151, 4211152, 4211153, 4211154, 4226054, 4226055, 4226056, 4226057, 4226058, 4226059, 4226060, 4226061, 4226062, 4226063, 4226064, 4226065, 4226066, 4226067, 4226068, 4226069, 4226070, 4226071, 4226072, 4226073, 4226074, 4226075, 4226076, 4226077, 4226078, 4226079, 4226080, 4226081, 4226082, 4226083, 4226084, 4226085, 4226086, 4226087, 4226088, 4226089, 4226090, 4226091, 4226092, 4226093, 4226094, 4226095, 4226096, 4226097, 4226098, 4226099, 4226100, 4226101, 4226102, 4226103, 4226104, 4226105, 4226106, 4226107, 4226108, 4226109, 4226110, 4226111, 4226112, 4226113, 4226114, 4226115, 4226116, 4226117, 4226118, 4226119, 4226120, 4226121, 4226122, 4226123, 4226124, 4226125, 4226126, 4226127, 4226128, 4226129, 4226130, 4226131, 4226132, 4226133, 4226134, 4226135, 4226136, 4226137, 4226138, 4226139, 4226140, 4226141, 4226142, 4226143, 4226144, 4226145, 4226146, 4226147, 4226148, 4226149, 4226150, 4226151, 4226152, 4226153, 4226154, 4226155, 4226156, 4226157, 4226158, 4226159, 4226160, 4226161, 4226162, 4226163, 4226164, 4226165, 4226166, 4226167, 4226168, 4226169, 4226170, 4226171, 4226172, 4226173, 4226174, 4226175, 4226176, 4226177, 4226178, 4226179, 4226180, 4226181, 4226182, 4226183, 4226184, 4226185, 4226186, 4226187, 4226188, 4226189, 4226190, 4226191, 4226192, 4226193, 4226194, 4226195, 4226196, 4226197, 4226198, 4226199, 4226200, 4226201, 4226202, 4226203, 4226204, 4226205, 4226206, 4226207, 4226208, 4226209, 4226210, 4226211, 4226212, 4226213, 4226214, 4226215, 4226216, 4226217, 4226218, 4226219, 4226220, 4226221, 4226222, 4226223, 4226224, 4226225, 4226226, 4226227, 4226228, 4226229, 4226230, 4226231, 4226232, 4226233, 4226234, 4226235, 4226236, 4226237, 4226238, 4226239, 4226240, 4226241, 4226242, 4226243, 4226244, 4226245, 4226246, 4226247, 4226248, 4226249, 4226250, 4226251, 4226252, 4226253, 4226254, 4226255, 4226256, 4226257, 4226258, 4226259, 4226260, 4226261, 4226262, 4226263, 4226264, 4226265, 4226266, 4226267, 4226268, 4226269, 4226270, 4226271, 4226272, 4226273, 4226274, 4226275, 4226276, 4226277, 4226278, 4226279, 4226280, 4226281, 4226282, 4226283, 4226284, 4226285, 4226286, 4226287, 4226288, 4226289, 4226290, 4226291, 4226292, 4226293, 4226294, 4226295, 4226296, 4226297, 4226298, 4226299, 4226300, 4226301, 4226302, 4226303, 4226304, 4226305, 4226306, 4226307, 4226308, 4226309, 4226310, 4226311, 4226312, 4226313, 4226314, 4226315, 4226316, 4226317, 4226318, 4226319, 4226320, 4226321, 4226322, 4226323, 4226324, 4226325, 4226326, 4226327, 4226328, 4226329, 4226330, 4226331, 4226332, 4226333, 4226334, 4226335, 4226336, 4226337, 4226338, 4226339, 4226340, 4226341, 4226342, 4226343, 4226344, 4226345, 4226346, 4226347, 4226348, 4226349, 4226350, 4226351, 4226352, 4226353, 4232504, 4232505, 4232506, 4232507, 4232508, 4232509, 4232510, 4232511, 4232512, 4232513, 4232514, 4232515, 4232516, 4232517, 4232518, 4232519, 4232520, 4232521, 4232522, 4232523, 4232524, 4232525, 4232526, 4232527, 4232528, 4232529, 4232530, 4232531, 4232532, 4232533, 4232534, 4232535, 4232536, 4232537, 4232538, 4232539, 4232540, 4232541, 4232542, 4232543, 4232544, 4232545, 4232546, 4232547, 4232548, 4232549, 4232550, 4232551, 4232552, 4232553, 4232554, 4232555, 4232556, 4232557, 4232558, 4232559, 4232560, 4232561, 4232562, 4232563, 4232564, 4232565, 4232566, 4232567, 4232568, 4232569, 4232570, 4232571, 4232572, 4232573, 4232574, 4232575, 4232576, 4232577, 4232578, 4232579, 4232580, 4232581, 4232582, 4232583, 4232584, 4232585, 4232586, 4232587, 4232588, 4232589, 4232590, 4232591, 4232592, 4232593, 4232594, 4232595, 4232596, 4232597, 4232598, 4232599, 4232600, 4232601, 4232602, 4232603, 4232604, 4232605, 4232606, 4232607, 4232608, 4232609, 4232610, 4232611, 4232612, 4232613, 4232614, 4232615, 4232616, 4232617, 4232618, 4232619, 4232620, 4232621, 4232622, 4232623, 4232624, 4232625, 4232626, 4232627, 4232628, 4232629, 4232630, 4232631, 4232632, 4232633, 4232634, 4232635, 4232636, 4232637, 4232638, 4232639, 4232640, 4232641, 4232642, 4232643, 4232644, 4232645, 4232646, 4232647, 4232648, 4232649, 4232650, 4232651, 4232652, 4232653, 4232654, 4232655, 4232656, 4232657, 4232658, 4232659, 4232660, 4232661, 4232662, 4232663, 4232664, 4232665, 4232666, 4232667, 4232668, 4232669, 4232670, 4232671, 4232672, 4232673, 4232674, 4232675, 4232676, 4232677, 4232678, 4232679, 4232680, 4232681, 4232682, 4232683, 4232684, 4232685, 4232686, 4232687, 4232688, 4232689, 4232690, 4232691, 4232692, 4232693, 4232694, 4232695, 4232696, 4232697, 4232698, 4232699, 4232700, 4232701, 4232702, 4232703, 4232704, 4232705, 4232706, 4232707, 4232708, 4232709, 4232710, 4232711, 4232712, 4232713, 4232714, 4232715, 4232716, 4232717, 4232718, 4232719, 4232720, 4232721, 4232722, 4232723, 4232724, 4232725, 4232726, 4232727, 4232728, 4232729, 4232730, 4232731, 4232732, 4232733, 4232734, 4232735, 4232736, 4232737, 4232738, 4232739, 4232740, 4232741, 4232742, 4232743, 4232744, 4232745, 4232746, 4232747, 4232748, 4232749, 4232750, 4232751, 4232752, 4232753, 4234754, 4234755, 4234756, 4234757, 4234758, 4234759, 4234760, 4234761, 4234762, 4234763, 4234764, 4234765, 4234766, 4234767, 4234768, 4234769, 4234770, 4234771, 4234772, 4234773, 4234774, 4234775, 4234776, 4234777, 4234778, 4234779, 4234780, 4234781, 4234782, 4234783, 4234784, 4234785, 4234786, 4234787, 4234788, 4234789, 4234790, 4234791, 4234792, 4234793, 4234794, 4234795, 4234796, 4234797, 4234798, 4234799, 4234800, 4234801, 4234802, 4234803, 4234804, 4234805, 4234806, 4234807, 4234808, 4234809, 4234810, 4234811, 4234812, 4234813, 4234814, 4234815, 4234816, 4234817, 4234818, 4234819, 4234820, 4234821, 4234822, 4234823, 4234824, 4234825, 4234826, 4234827, 4234828, 4234829, 4234830, 4234831, 4234832, 4234833, 4234834, 4234835, 4234836, 4234837, 4234838, 4234839, 4234840, 4234841, 4234842, 4234843, 4234844, 4234845, 4234846, 4234847, 4234848, 4234849, 4234850, 4234851, 4234852, 4234853, 4238304, 4238305, 4238306, 4238307, 4238308, 4238309, 4238310, 4238312, 4238313, 4238314, 4238315, 4238316, 4238319, 4238322, 4238324, 4238325, 4238326, 4238331, 4238336, 4238337, 4238338, 4238339, 4238340, 4238341, 4238342, 4238343, 4238344, 4238345, 4238346, 4238347, 4238348, 4238349, 4238350, 4238351, 4238352, 4238353, 4238354, 4238355, 4238356, 4238357, 4238358, 4238359, 4238360, 4238361, 4238362, 4238363, 4238364, 4238365, 4238366, 4238367, 4238368, 4238369, 4238370, 4238371, 4238372, 4238373, 4238374, 4238375, 4238376, 4238377, 4238378, 4238379, 4238380, 4238381, 4238382, 4238383, 4238384, 4238385, 4238386, 4238387, 4238388, 4238389, 4238390, 4238391, 4238392, 4238393, 4238394, 4238395, 4238396, 4238397, 4238398, 4238399, 4238400, 4238401, 4238402, 4238403, 4238404, 4238405, 4238406, 4238407, 4238408, 4238409, 4238410, 4238411, 4238412, 4238413, 4238415, 4238416, 4238417, 4238418, 4238419, 4238420, 4238421, 4238422, 4238423, 4238424, 4238425, 4238426, 4238427, 4238428, 4238429, 4238430, 4238431, 4238432, 4238433, 4238434, 4238435, 4238436, 4238437, 4238438, 4238439, 4238442, 4238443, 4238444, 4238445, 4238446, 4238447, 4238448, 4238449, 4238450, 4238451, 4238452, 4238453, 4238454, 4238455, 4238456, 4238457, 4238458, 4238459, 4238460, 4238461, 4238462, 4238463, 4238464, 4238465, 4238466, 4238467, 4238468, 4238469, 4238470, 4238471, 4238472, 4238473, 4238474, 4238475, 4238476, 4238477, 4238478, 4238479, 4238481, 4238482, 4238483, 4238484, 4238485, 4238486, 4238487, 4238488, 4238489, 4238490, 4238491, 4238492, 4238493, 4238494, 4238495, 4238496, 4238497, 4238498, 4238499, 4238500, 4238501, 4238502, 4238506, 4238507, 4238509, 4238510, 4238511, 4238512, 4238513, 4238514, 4238515, 4238516, 4238517, 4238518, 4238519, 4238520, 4238521, 4238522, 4238523, 4238524, 4238525, 4238526, 4238527, 4238528, 4238529, 4238530, 4238531, 4238532, 4238533, 4238534, 4238535, 4238536, 4238537, 4238538, 4238539, 4238540, 4238541, 4238542, 4238543, 4238544, 4238545, 4238546, 4238547, 4238548, 4238549, 4238550, 4238551, 4238552, 4238553, 4244254, 4244255, 4244256, 4244257, 4244258, 4244259, 4244260, 4244261, 4244262, 4244263, 4244264, 4244265, 4244266, 4244267, 4244268, 4244269, 4244270, 4244271, 4244272, 4244273, 4244274, 4244275, 4244276, 4244277, 4244278, 4244279, 4244280, 4244281, 4244282, 4244283, 4244284, 4244285, 4244286, 4244287, 4244288, 4244289, 4244290, 4244291, 4244292, 4244293, 4244294, 4244295, 4244296, 4244297, 4244298, 4244299, 4244300, 4244301, 4244302, 4244303, 4244304, 4244305, 4244306, 4244307, 4244308, 4244309, 4244310, 4244311, 4244312, 4244313, 4244314, 4244315, 4244316, 4244317, 4244318, 4244319, 4244320, 4244321, 4244322, 4244323, 4244324, 4244325, 4244326, 4244327, 4244328, 4244329, 4244330, 4244331, 4244332, 4244333, 4244334, 4244335, 4244336, 4244337, 4244338, 4244339, 4244340, 4244341, 4244342, 4244343, 4244344, 4244345, 4244346, 4244347, 4244348, 4244349, 4244350, 4244351, 4244352, 4244353, 4244354, 4244355, 4244356, 4244357, 4244358, 4244359, 4244360, 4244361, 4244362, 4244363, 4244364, 4244365, 4244366, 4244367, 4244368, 4244369, 4244370, 4244371, 4244372, 4244373, 4244374, 4244375, 4244376, 4244377, 4244378, 4244379, 4244380, 4244381, 4244382, 4244383, 4244384, 4244385, 4244386, 4244387, 4244388, 4244389, 4244390, 4244391, 4244392, 4244393, 4244394, 4244395, 4244396, 4244397, 4244398, 4244399, 4244400, 4244402, 4244403, 4244404, 4244405, 4244406, 4244407, 4244408, 4244409, 4244410, 4244411, 4244412, 4244413, 4244414, 4244415, 4244416, 4244417, 4244418, 4244419, 4244420, 4244421, 4244422, 4244423, 4244424, 4244425, 4244426, 4244427, 4244428, 4244429, 4244430, 4244431, 4244432, 4244433, 4244434, 4244435, 4244436, 4244437, 4244438, 4244439, 4244440, 4244441, 4244442, 4244443, 4244444, 4244445, 4244446, 4244447, 4244448, 4244449, 4244450, 4244451, 4244452, 4244453, 4244454, 4244455, 4244456, 4244457, 4244458, 4244459, 4244460, 4244461, 4244462, 4244463, 4244464, 4244465, 4244466, 4244467, 4244468, 4244469, 4244470, 4244471, 4244472, 4244473, 4244474, 4244475, 4244476, 4244477, 4244478, 4244479, 4244480, 4244481, 4244482, 4244483, 4244484, 4244485, 4244487, 4244488, 4244489, 4244490, 4244491, 4244492, 4244493, 4244494, 4244495, 4244496, 4244497, 4244498, 4244499, 4244500, 4244501, 4244502, 4244503, 4249504, 4249505, 4249506, 4249507, 4249508, 4249509, 4249510, 4249511, 4249512, 4249513, 4249514, 4249515, 4249516, 4249517, 4249518, 4249519, 4249520, 4249521, 4249522, 4249523, 4249524, 4249525, 4249526, 4249527, 4249528, 4249529, 4249530, 4249531, 4249532, 4249533, 4249534, 4249535, 4249536, 4249537, 4249538, 4249539, 4249540, 4249541, 4249542, 4249543, 4249544, 4249545, 4249546, 4249547, 4249548, 4249549, 4249550, 4249551, 4249552, 4249553, 4249554, 4249555, 4249556, 4249557, 4249558, 4249559, 4249560, 4249561, 4249562, 4249563, 4249564, 4249565, 4249566, 4249567, 4249568, 4249569, 4249570, 4249571, 4249572, 4249573, 4249574, 4249575, 4249576, 4249577, 4249578, 4249579, 4249580, 4249582, 4249583, 4249584, 4249585, 4249586, 4249587, 4249588, 4249589, 4249590, 4249591, 4249592, 4249593, 4249594, 4249595, 4249596, 4249597, 4249598, 4249599, 4249600, 4249601, 4249602, 4249603, 4249604, 4249605, 4249606, 4249607, 4249608, 4249609, 4249610, 4249611, 4249612, 4249613, 4249614, 4249615, 4249616, 4249617, 4249618, 4249619, 4249620, 4249621, 4249622, 4249623, 4249624, 4249625, 4249626, 4249627, 4249628, 4249629, 4249630, 4249631, 4249632, 4249633, 4249634, 4249635, 4249636, 4249637, 4249638, 4249639, 4249640, 4249641, 4249642, 4249643, 4249644, 4249645, 4249646, 4249647, 4249648, 4249649, 4249650, 4249651, 4249652, 4249653, 4249654, 4249655, 4249656, 4249657, 4249658, 4249659, 4249660, 4249661, 4249662, 4249663, 4249664, 4249665, 4249666, 4249667, 4249668, 4249669, 4249670, 4249671, 4249672, 4249673, 4249674, 4249675, 4249676, 4249677, 4249678, 4249679, 4249680, 4249681, 4249682, 4249683, 4249684, 4249685, 4249686, 4249687, 4249688, 4249689, 4249690, 4249691, 4249692, 4249693, 4249694, 4249695, 4249696, 4249697, 4249698, 4249699, 4249700, 4249701, 4249702, 4249703, 4249704, 4249705, 4249706, 4249707, 4249708, 4249709, 4249710, 4249711, 4249712, 4249713, 4249714, 4249715, 4249716, 4249717, 4249718, 4249719, 4249720, 4249721, 4249722, 4249723, 4249725, 4249726, 4249727, 4249728, 4249729, 4249730, 4249731, 4249732, 4249733, 4249734, 4249735, 4249736, 4249737, 4249738, 4249739, 4249740, 4249741, 4249742, 4249743, 4249744, 4249745, 4249746, 4249747, 4249748, 4249749, 4249750, 4249751, 4249752, 4249754, 4249755, 4249756, 4249759, 4249760, 4249761, 4249762, 4249763, 4249765, 4249767, 4249769, 4249771, 4249772, 4249774, 4249775, 4249776, 4249777, 4249778, 4249780, 4249781, 4249782, 4249783, 4249784, 4249785, 4249786, 4249787, 4249788, 4249789, 4249790, 4249791, 4249792, 4249793, 4249794, 4249795, 4249796, 4249797, 4249798, 4249799, 4249800, 4249801, 4249802, 4249803, 4259710, 4259711, 4259712, 4259713, 4259714, 4259715, 4259716, 4259717, 4259718, 4259719, 4259720, 4259721, 4259722, 4259723, 4259724, 4259725, 4259726, 4259727, 4259728, 4259729, 4259730, 4259731, 4259732, 4259733, 4259734, 4259735, 4259736, 4259737, 4259738, 4259739, 4259740, 4259741, 4259742, 4259743, 4259744, 4259745, 4259746, 4259747, 4259748, 4259749, 4259750, 4259751, 4259752, 4259753, 4259754, 4259755, 4259756, 4259757, 4259758, 4259759, 4259760, 4259761, 4259762, 4259763, 4259764, 4259765, 4259766, 4259767, 4259768, 4259769, 4259770, 4259771, 4259772, 4259773, 4259774, 4259775, 4259776, 4259777, 4259778, 4259779, 4259780, 4259781, 4259782, 4259783, 4259784, 4259785, 4259786, 4259787, 4259788, 4259789, 4259790, 4259791, 4259792, 4259793, 4259794, 4259795, 4259796, 4259797, 4259798, 4259799, 4259800, 4259801, 4259802, 4259803, 4259804, 4259805, 4259806, 4259807, 4259808, 4259809, 4262140, 4262141, 4262142, 4262143, 4262144, 4262145, 4262146, 4262147, 4262148, 4262149, 4262150, 4262151, 4262152, 4262153, 4262154, 4262155, 4262156, 4262157, 4262158, 4262159, 4262160, 4262161, 4262162, 4262163, 4262164, 4262165, 4262166, 4262167, 4262168, 4262169, 4262170, 4262171, 4262172, 4262173, 4262174, 4262175, 4262176, 4262177, 4262178, 4262179, 4262180, 4262181, 4262182, 4262183, 4262184, 4262185, 4262186, 4262187, 4262188, 4262189, 4262190, 4262191, 4262192, 4262193, 4262194, 4262195, 4262196, 4262197, 4262198, 4262199, 4262200, 4262201, 4262202, 4262203, 4262204, 4262205, 4262206, 4262207, 4262208, 4262209, 4262210, 4262211, 4262212, 4262213, 4262214, 4262215, 4262216, 4262217, 4262218, 4262219, 4262220, 4262221, 4262222, 4262223, 4262224, 4262225, 4262226, 4262227, 4262228, 4262229, 4262230, 4262231, 4262232, 4262233, 4262234, 4262235, 4262236, 4262237, 4262238, 4262239, 4262240, 4262241, 4262242, 4262243, 4262244, 4262245, 4262246, 4262247, 4262248, 4262249, 4262250, 4262251, 4262252, 4262253, 4262254, 4262255, 4262256, 4262257, 4262258, 4262259, 4262260, 4262261, 4262262, 4262263, 4262264, 4262265, 4262266, 4262267, 4262268, 4262269, 4262270, 4262271, 4262272, 4262273, 4262274, 4262275, 4262276, 4262277, 4262278, 4262279, 4262280, 4262281, 4262282, 4262283, 4262284, 4262285, 4262286, 4262287, 4262288, 4262289, 4262290, 4262291, 4262292, 4262293, 4262294, 4262295, 4262296, 4262297, 4262298, 4262299, 4262300, 4262301, 4262302, 4262303, 4262304, 4262305, 4262306, 4262307, 4262308, 4262309, 4262310, 4262311, 4262312, 4262313, 4262314, 4262315, 4262316, 4262317, 4262318, 4262319, 4262320, 4262321, 4262322, 4262323, 4262324, 4262325, 4262326, 4262327, 4262328, 4262329, 4262330, 4262331, 4262332, 4262333, 4262334, 4262335, 4262336, 4262337, 4262338, 4262339, 4263440, 4263441, 4263442, 4263443, 4263444, 4263445, 4263446, 4263447, 4263448, 4263449, 4263450, 4263451, 4263452, 4263453, 4263454, 4263455, 4263456, 4263457, 4263458, 4263459, 4263460, 4263461, 4263462, 4263463, 4263464, 4263465, 4263466, 4263467, 4263468, 4263469, 4263470, 4263471, 4263472, 4263473, 4263474, 4263475, 4263476, 4263477, 4263478, 4263479, 4263480, 4263481, 4263482, 4263483, 4263484, 4263485, 4263486, 4263487, 4263488, 4263489, 4263490, 4263491, 4263492, 4263493, 4263494, 4263495, 4263496, 4263497, 4263498, 4263499, 4263500, 4263501, 4263502, 4263503, 4263504, 4263505, 4263506, 4263507, 4263508, 4263509, 4263510, 4263511, 4263512, 4263513, 4263514, 4263515, 4263516, 4263517, 4263518, 4263519, 4263520, 4263521, 4263522, 4263523, 4263524, 4263525, 4263526, 4263527, 4263528, 4263529, 4263530, 4263531, 4263532, 4263533, 4263534, 4263535, 4263536, 4263537, 4263538, 4263539, 4265991, 4265992, 4265993, 4265994, 4265995, 4266596, 4266597, 4266598, 4266599, 4266600, 4266601, 4266602, 4266603, 4266604, 4266605, 4266606, 4266607, 4266608, 4266609, 4266610, 4266611, 4266612, 4266613, 4266614, 4266615, 4266616, 4266617, 4266618, 4266619, 4266620, 4266621, 4266622, 4266623, 4266624, 4266625, 4266626, 4266627, 4266628, 4266629, 4266630, 4266631, 4266632, 4266633, 4266634, 4266635, 4266636, 4266637, 4266638, 4266639, 4266640, 4266641, 4266642, 4266643, 4266644, 4266645, 4266796, 4266797, 4266798, 4266799, 4266800, 4266801, 4266802, 4266803, 4266804, 4266805, 4266806, 4266807, 4266808, 4266809, 4266810, 4266811, 4266812, 4266813, 4266814, 4266815, 4266816, 4266817, 4266818, 4266819, 4266820, 4266821, 4266822, 4266823, 4266824, 4266825, 4266826, 4266827, 4266828, 4266832, 4266834, 4266835, 4266836, 4266837, 4266838, 4266839, 4266840, 4266841, 4266842, 4266843, 4266844, 4266845, 4266846, 4266847, 4266848, 4266849, 4266850, 4266851, 4266852, 4266853, 4266854, 4266855, 4266856, 4266857, 4266858, 4266859, 4266860, 4266861, 4266862, 4266863, 4266864, 4266865, 4266866, 4266867, 4266868, 4266869, 4266870, 4266871, 4266872, 4266873, 4266874, 4266875, 4266876, 4266877, 4266878, 4266879, 4266880, 4266881, 4266882, 4266883, 4266884, 4266885, 4266886, 4266887, 4266888, 4266889, 4266890, 4266891, 4266892, 4266893, 4266894, 4266895, 4266896, 4266897, 4266898, 4266899, 4266900, 4266901, 4266902, 4266903, 4266904, 4266905, 4269907, 4269916, 4269917, 4269918, 4269919, 4269920, 4269921, 4269922, 4269923, 4269924, 4269925, 4269926, 4269927, 4269928, 4269929, 4269930, 4269931, 4270010 and 9015899.",
      "firm_fei_number": "2112020",
      "recalling_firm": "Cardiac Science Corporation",
      "address_1": "3303 Monte Villa Pkwy",
      "city": "Bothell",
      "state": "WA",
      "postal_code": "98021-8969",
      "additional_info_contact": "Michael Matysik\n425-402-2206",
      "reason_for_recall": "Potential for devices not to deliver therapy",
      "root_cause_description": "Process control",
      "action": "Cardiac Science issued an \"Urgent - Voluntary Medical Device Correction AED Resistor Issue\" letter dated November 13, 2009.  Consignees were informed of the affected product and instructed on preventive measures.\n\nFor further information, contact Cardiac Science at 1-877-901-1788 in the US and +44-161-926-0011 for outside the US.",
      "product_quantity": "8488",
      "distribution_pattern": "Worldwide Distribution -- UNITED STATES, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, CZECHOSLOVAKIA, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GRECE, GUAM, GUATEMALA, HOLLAND, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KENYA, KOREA, KUWAIT, LEBANON, LIBYA, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONACO, MORROCCO, NAMIBIA, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SCOTLAND, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UKRAINE, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM, VIRGIN ISLANDS AND YEMEN.",
      "openfda": {
        "k_number": [
          "K982710",
          "K042404",
          "K051632",
          "K883584",
          "K962359",
          "K040438",
          "K981802",
          "K123881",
          "K012766",
          "K032439",
          "K955882",
          "K033896",
          "K052161",
          "K033642",
          "K120406",
          "K860676",
          "K993533",
          "K122839",
          "K955923",
          "K954957",
          "K133441",
          "K990762",
          "K063510",
          "K052465",
          "K040479",
          "K892005",
          "K992977",
          "K982992",
          "K041459",
          "K131494",
          "K013425",
          "K110526",
          "K142803",
          "K984582",
          "K110154",
          "K030898",
          "K072233",
          "K040256",
          "K882067",
          "K023854",
          "K981737",
          "K100565",
          "K061707",
          "K972418",
          "K990106",
          "K062233",
          "K133484",
          "K081828",
          "K013896",
          "K983393",
          "K150198",
          "K892065",
          "K932931",
          "K020288",
          "K130153",
          "K052057",
          "K014157",
          "K110361",
          "K112761",
          "K022210",
          "K060559",
          "K014209",
          "K870622",
          "K935922",
          "K103567",
          "K965117",
          "K934692",
          "K041892",
          "K003819",
          "K942344",
          "K955854",
          "K040637",
          "K080421",
          "K133239",
          "K143714",
          "K992543",
          "K131525",
          "K003565",
          "K012428",
          "K051076",
          "K133659",
          "K041883",
          "K111594",
          "K051475",
          "K052316",
          "K092598",
          "K133269",
          "K003228",
          "K020743",
          "K971149",
          "K993072",
          "K002445",
          "K991007",
          "K130454",
          "K020715",
          "K082090",
          "K110978",
          "K953204",
          "K142915",
          "K926172",
          "K010918",
          "K142709",
          "K042088",
          "K955628",
          "K970719",
          "K121853",
          "K896919",
          "K033771",
          "K080903",
          "K970741",
          "K941262",
          "K031187",
          "K050004",
          "K012030",
          "K002280",
          "K112432",
          "K120536",
          "K040775",
          "K973486",
          "K972241",
          "K091943",
          "K113787",
          "K041067",
          "K020203",
          "K954504",
          "K984286",
          "K141231",
          "K952101",
          "K991871",
          "K980857",
          "K110825",
          "K120350",
          "K092391",
          "K102972",
          "K081371",
          "K000638",
          "K011901",
          "K063119",
          "K954609",
          "K072430",
          "K042417",
          "K993745",
          "K925936",
          "K081259",
          "K970391",
          "K103186",
          "K072812",
          "K062177",
          "K073716",
          "K991910",
          "K080232",
          "K063375",
          "K022929",
          "K942850",
          "K082937",
          "K022732",
          "K110742",
          "K003548",
          "K122600",
          "K010207",
          "K032421",
          "K091548",
          "K021168",
          "K011541",
          "K020984",
          "K982264",
          "K073089",
          "K943698",
          "K131550",
          "K940445",
          "K130684",
          "K122758",
          "K991556",
          "K141774",
          "K993036",
          "K110168",
          "K012197",
          "K090989",
          "K040056",
          "K081081",
          "K002079",
          "K941970",
          "K011388",
          "K091265",
          "K032051",
          "K955890",
          "K031987",
          "K072923",
          "K883341",
          "K150055",
          "K983307",
          "K033474",
          "K883344",
          "K972970",
          "K982025",
          "K033144",
          "K072596",
          "K010214",
          "K960329",
          "K001725",
          "K033275",
          "K062970",
          "K002806",
          "K011144",
          "K070455",
          "K964991",
          "K060934",
          "K141795",
          "K102496",
          "K091561",
          "K932028",
          "K120907",
          "K111693",
          "K051134",
          "K971148",
          "K983287",
          "K071025",
          "K082861",
          "K002232",
          "K142430",
          "K955828",
          "K042007",
          "K010634",
          "K853564",
          "K040404",
          "K010257",
          "K021453",
          "K062537"
        ],
        "pma_number": [
          "P200005",
          "P160034",
          "P170018",
          "P160033",
          "P160028",
          "P160015",
          "P160022",
          "P180028",
          "P190013",
          "P160032",
          "P210015",
          "P160008",
          "P160026",
          "P200004",
          "P200007",
          "P160012",
          "P190007"
        ],
        "registration_number": [
          "",
          "1720929",
          "3004906872",
          "3010839093",
          "3038259592",
          "3008717264",
          "3020460367",
          "3013500228",
          "3030446844",
          "3004123209",
          "3003521780",
          "1930027",
          "3014572985",
          "1320894",
          "3010030355",
          "3010317211",
          "3000126629",
          "1218058",
          "3002714378",
          "1423537",
          "8020045",
          "3009886119",
          "1317188",
          "3030733800",
          "3043226252",
          "3030677",
          "3013557562",
          "3010783192",
          "1223925",
          "3015876",
          "1921846",
          "3013826848",
          "3003537036",
          "1220908",
          "3027815",
          "3009077524",
          "3003867964",
          "3029871637",
          "3012972516",
          "9681860",
          "3005621729",
          "3002807314",
          "3026976873",
          "3003800159",
          "3015997711",
          "3012104670",
          "3017422354",
          "3011823740",
          "3014579161",
          "3005906887",
          "3007029079",
          "3026007685",
          "3004035727",
          "3015134086",
          "3013517171",
          "3021559257",
          "3012502601",
          "2132111",
          "3012607778",
          "1219103",
          "3012494290",
          "3013596024",
          "9681191",
          "3010703925",
          "1423662",
          "2112020",
          "3006108746"
        ],
        "fei_number": [
          "",
          "1720929",
          "3004906872",
          "3010839093",
          "3038259592",
          "3008717264",
          "3020460367",
          "3013500228",
          "3030446844",
          "3004123209",
          "3003521780",
          "1930027",
          "3014572985",
          "3010030355",
          "3010317211",
          "3000126629",
          "1218058",
          "3002714378",
          "3009886119",
          "1317188",
          "3030733800",
          "3043226252",
          "3013557562",
          "3000215273",
          "3010783192",
          "1000148701",
          "3015876",
          "1921846",
          "3013826848",
          "3003537036",
          "1220908",
          "1000305720",
          "3009077524",
          "3003867964",
          "3003491392",
          "3029871637",
          "3012972516",
          "3001236905",
          "3005621729",
          "3002807314",
          "3026976873",
          "3001451571",
          "3003800159",
          "1000140026",
          "3015997711",
          "3012104670",
          "3017422354",
          "3011823740",
          "3014579161",
          "3005906887",
          "3007029079",
          "3026007685",
          "3004035727",
          "3015134086",
          "3013517171",
          "3021559257",
          "3012502601",
          "1000127185",
          "3012607778",
          "3012494290",
          "3013596024",
          "1000524572",
          "3010703925",
          "2112020",
          "3006108746",
          "6736",
          "3008207730"
        ],
        "device_name": "Automated External Defibrillators (Non-Wearable)",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.5310",
        "device_class": "3"
      }
    }
  ]
}