  "results": [
    {
      "cfres_id": "85537",
      "product_res_number": "Z-0256-2010",
      "event_date_initiated": "2009-10-08",
      "event_date_posted": "2009-11-13",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-05-16",
      "res_event_number": "53399",
      "product_code": "JWH",
      "k_numbers": [
        "K070214"
      ],
      "product_description": "Zimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, ultracongruent, Size 3,4,5, right, 13 MM height, for use with baseplate size 3,4 or 5, sterile, Zimmer, Warsaw, IN; REF 00-5428-023-13.\n\nThe device is a semiconstrained total knee prosthesis components indicated for uncemented or cemented use in skeletally mature individuals.",
      "code_info": "Lots 60933716, 60933717, 60964509, 60981897, 60987856, 60988253, 60993073, 60993074, 60993075, 61001578, 61005473, 61023710, 61027950, 61027951, 61045118, 61051513, 61058641, 61058642, 61075277, 61075278, 61078791, 61085909, 61091205, 61091229, 61096401, 61106597, 61106598, 61114214, 61122297 and 61136942.",
      "firm_fei_number": "1000220733",
      "recalling_firm": "Zimmer Inc.",
      "address_1": "345 E Main St",
      "city": "Warsaw",
      "state": "IN",
      "postal_code": "46580-2746",
      "additional_info_contact": "800-613-6131",
      "reason_for_recall": "The surgeon may have difficulty inserting the polyethylene into the tibial baseplate, which may result in intraoperative damage to the device.",
      "root_cause_description": "Other",
      "action": "An \"Urgent: Field Correction\" dated October 8, 2009 was issued to Consignees.  The notification letter described the affected product, issue, health risks and actions for customers to take.  The letter notified consignees of a changed to the surgical technique, to include the use of a new peripheral rib impactor tool.\n\nNotifications of this field notification were sent to all affected direct accounts of Zimmer,  Inc.  Contact your Zimmer Sales Representative if\nyou have any questions regarding this action.  \n\nFor shipping assistance, questions or assistance in notifying your accounts about the field correction please contact Zimmer, Inc. at 1-800-613-6131.",
      "product_quantity": "26,404 of all products",
      "distribution_pattern": "Worldwide Distribution -- United States, Austria, Belgium, France, Germany, Japan, Spain and Switzerland.",
      "openfda": {
        "k_number": [
          "K810328",
          "K000113",
          "K153595",
          "K000666",
          "K100981",
          "K040281",
          "K193057",
          "K253314",
          "K143226",
          "K934695",
          "K864671",
          "K121416",
          "K914565",
          "K855166",
          "K171269",
          "K935262",
          "K023724",
          "K170573",
          "K971550",
          "K150496",
          "K934381",
          "K131019",
          "K131116",
          "K881982",
          "K231233",
          "K110837",
          "K012095",
          "K171389",
          "K121149",
          "K180446",
          "K182125",
          "K954138",
          "K963148",
          "K984054",
          "K920080",
          "K041825",
          "K161668",
          "K982585",
          "K132161",
          "K133256",
          "K261563",
          "K123339",
          "K933958",
          "K183583",
          "K143192",
          "K153217",
          "K122239",
          "K872134",
          "K910990",
          "K921640",
          "K210252",
          "K191625",
          "K972863",
          "K041106",
          "K190911",
          "K242410",
          "K973595",
          "K913492",
          "K180159",
          "K123721",
          "K894334",
          "K053579",
          "K925837",
          "K131467",
          "K190577",
          "K963796",
          "K183017",
          "K940235",
          "K924340",
          "K230537",
          "K910992",
          "K243991",
          "K123692",
          "K926228",
          "K193540",
          "K964008",
          "K170106",
          "K080528",
          "K973524",
          "K210829",
          "K911197",
          "K131398",
          "K932776",
          "K971652",
          "K910235",
          "K101981",
          "K170240",
          "K953274",
          "K914169",
          "K052687",
          "K991428",
          "K113378",
          "K934166",
          "K093360",
          "K100370",
          "K982903",
          "K920426",
          "K210483",
          "K936079",
          "K242377",
          "K211004",
          "K230653",
          "K944285",
          "K962137",
          "K242046",
          "K863805",
          "K131337",
          "K172336",
          "K023546",
          "K112285",
          "K052879",
          "K925372",
          "K163332",
          "K760200",
          "K150783",
          "K111785",
          "K010101",
          "K231947",
          "K860654",
          "K934620",
          "K892366",
          "K936274",
          "K122883",
          "K182872",
          "K020295",
          "K993111",
          "K033883",
          "K960535",
          "K980632",
          "K951185",
          "K050210",
          "K202022",
          "K022586",
          "K871146",
          "K051756",
          "K033260",
          "K911972",
          "K994128",
          "K021559",
          "K093552",
          "K021578",
          "K160771",
          "K180750",
          "K960976",
          "K152704",
          "K892503",
          "K120068",
          "K253144",
          "K921182",
          "K113550",
          "K042343",
          "K120698",
          "K031901",
          "K242831",
          "K830121",
          "K110397",
          "K973672",
          "K913188",
          "K211609",
          "K213381",
          "K933785",
          "K933867",
          "K011219",
          "K102380",
          "K162222",
          "K100391",
          "K945589",
          "K210809",
          "K090411",
          "K043101",
          "K121879",
          "K183533",
          "K161366",
          "K980781",
          "K912663",
          "K951655",
          "K103837",
          "K160025",
          "K911572",
          "K003026",
          "K093567",
          "K020830",
          "K242984",
          "K202740",
          "K063732",
          "K833411",
          "K002678",
          "K190280",
          "K895639",
          "K170149",
          "K143047",
          "K951956",
          "K970498",
          "K962196",
          "K032606",
          "K905613",
          "K833363",
          "K060192",
          "K863667",
          "K003910",
          "K202684",
          "K955884",
          "K972270",
          "K843241",
          "K874547",
          "K954208",
          "K936008",
          "K790219",
          "K111062",
          "K040630",
          "K972864",
          "K142404",
          "K991581",
          "K181146",
          "K994076",
          "K810326",
          "K060370",
          "K973406",
          "K140830",
          "K946088",
          "K926257",
          "K081458",
          "K200075",
          "K232414",
          "K901457",
          "K920396",
          "K191765",
          "K081400",
          "K951987",
          "K001456",
          "K893040",
          "K870273",
          "K931054",
          "K212742",
          "K936037",
          "K011904",
          "K943462",
          "K983004",
          "K244039",
          "K874923",
          "K222431",
          "K212034",
          "K032479",
          "K023788",
          "K925242",
          "K082358",
          "K112780",
          "K111283",
          "K082826",
          "K012776",
          "K010027",
          "K141585",
          "K211602",
          "K094013",
          "K021740",
          "K200653",
          "K140452",
          "K003296",
          "K210551",
          "K040770",
          "K936159",
          "K983835",
          "K173884",
          "K932210",
          "K041046",
          "K093293",
          "K023416",
          "K170808",
          "K220705",
          "K152430",
          "K140826",
          "K933539",
          "K220896",
          "K981560",
          "K901992",
          "K932690",
          "K910500",
          "K952943",
          "K023211",
          "K003182",
          "K092803",
          "K973121",
          "K063406",
          "K132752",
          "K150241",
          "K091263",
          "K963242",
          "K142161",
          "K261032",
          "K132046",
          "K101994",
          "K173057",
          "K200765",
          "K153075",
          "K170534",
          "K933610",
          "K243724",
          "K871776",
          "K853991",
          "K230846",
          "K060569",
          "K872266",
          "K851231",
          "K131481",
          "K011643",
          "K063686",
          "K991753",
          "K984158",
          "K970031",
          "K932858",
          "K131368",
          "K792089",
          "K994370",
          "K922841",
          "K071220",
          "K980545",
          "K240683",
          "K073602",
          "K180930",
          "K012866",
          "K071738",
          "K241570",
          "K994214",
          "K161360",
          "K984623",
          "K261734",
          "K122975",
          "K170648",
          "K894504",
          "K231975",
          "K223650",
          "K142069",
          "K930228",
          "K875201",
          "K002281",
          "K051976",
          "K240327",
          "K150829",
          "K213673",
          "K864518",
          "K181530",
          "K170806",
          "K222278",
          "K121037",
          "K012251",
          "K150105",
          "K031201",
          "K961685",
          "K051229",
          "K180334",
          "K181338",
          "K071499",
          "K171045",
          "K241000",
          "K133919",
          "K961379",
          "K982869",
          "K963117",
          "K933229",
          "K152170",
          "K160515",
          "K953443",
          "K102580",
          "K080204",
          "K072281",
          "K120313",
          "K853661",
          "K162026",
          "K122500",
          "K123687",
          "K201611",
          "K935080",
          "K113122",
          "K980925",
          "K965173",
          "K180906",
          "K071071",
          "K940861",
          "K040964",
          "K912245",
          "K991052",
          "K031962",
          "K172326",
          "K073529",
          "K131864",
          "K030367",
          "K130084",
          "K971189",
          "K012778",
          "K133630",
          "K081023",
          "K925072",
          "K933304",
          "K013991",
          "K072336",
          "K172936",
          "K173875",
          "K173890",
          "K234011",
          "K011976",
          "K240324",
          "K221939",
          "K152382",
          "K972643",
          "K202924",
          "K904300",
          "K930188",
          "K060722",
          "K213793",
          "K020114",
          "K252974",
          "K062274",
          "K111184",
          "K964951",
          "K904205",
          "K160159",
          "K945087",
          "K910563",
          "K081127",
          "K133560",
          "K212129",
          "K221337",
          "K971770",
          "K143635",
          "K182370",
          "K922859",
          "K232175",
          "K183225",
          "K140073",
          "K112941",
          "K935789",
          "K954818",
          "K163167",
          "K123166",
          "K020214",
          "K121640",
          "K220737",
          "K993692",
          "K030301",
          "K082022",
          "K133449",
          "K122991",
          "K060515",
          "K192148",
          "K211938",
          "K941755",
          "K872379",
          "K982302",
          "K892800",
          "K923451",
          "K143443",
          "K213389",
          "K240863",
          "K871769",
          "K920698",
          "K952271",
          "K041466",
          "K141934",
          "K190355",
          "K151266",
          "K250382",
          "K251554",
          "K250677",
          "K873602",
          "K113369",
          "K253793",
          "K870702",
          "K090705",
          "K191192",
          "K030577",
          "K896397",
          "K946150",
          "K093746",
          "K032522",
          "K010434",
          "K140423",
          "K883122",
          "K926334",
          "K243817",
          "K944321",
          "K941306",
          "K153310",
          "K173723",
          "K911198",
          "K051977",
          "K252725",
          "K202248",
          "K901936",
          "K122985",
          "K915186",
          "K933843",
          "K140902",
          "K090988",
          "K904448",
          "K001147",
          "K863668",
          "K935773",
          "K990030",
          "K974556",
          "K093806",
          "K200227",
          "K981996",
          "K912266",
          "K923573",
          "K954909",
          "K112906",
          "K133854",
          "K970758",
          "K142388",
          "K170613",
          "K935726",
          "K202429",
          "K230952",
          "K230078",
          "K925900",
          "K032295",
          "K905076",
          "K011288",
          "K222816",
          "K884796",
          "K143725",
          "K050539",
          "K210191",
          "K183025",
          "K150928",
          "K830927",
          "K884338",
          "K063211",
          "K082424",
          "K861266",
          "K932425",
          "K230169",
          "K183344",
          "K894937",
          "K063633",
          "K991106",
          "K896081",
          "K173331",
          "K932911",
          "K883215",
          "K843242",
          "K933289",
          "K972967",
          "K171054",
          "K021313",
          "K150544",
          "K914072",
          "K901085",
          "K223062",
          "K063423",
          "K142649",
          "K022672",
          "K201851",
          "K923277",
          "K180743",
          "K091280",
          "K024037",
          "K092201",
          "K931466",
          "K895962",
          "K953439",
          "K021908",
          "K032079",
          "K962557",
          "K221059",
          "K130673",
          "K002735",
          "K941769",
          "K024161",
          "K892394",
          "K053390",
          "K960279",
          "K002740",
          "K121835",
          "K221404",
          "K013906",
          "K230441",
          "K962152",
          "K022010",
          "K022204",
          "K121443",
          "K000091",
          "K974328",
          "K871349",
          "K904170",
          "K030686",
          "K141471",
          "K190562",
          "K140027",
          "K160700",
          "K864632",
          "K073286",
          "K111492",
          "K243024",
          "K100900",
          "K980713",
          "K895145",
          "K914230",
          "K170753",
          "K173445",
          "K201343",
          "K901406",
          "K931651",
          "K122033",
          "K091014",
          "K961483",
          "K181947",
          "K033959",
          "K050723",
          "K182085",
          "K160157",
          "K121307",
          "K241180",
          "K983733",
          "K922048",
          "K070095",
          "K033890",
          "K915798",
          "K083772",
          "K250889",
          "K991461",
          "K952830",
          "K843907",
          "K222570",
          "K001307",
          "K935523",
          "K012172",
          "K122765",
          "K200395",
          "K181635",
          "K922285",
          "K152008",
          "K033272",
          "K072221",
          "K080199",
          "K963028",
          "K113829",
          "K932837",
          "K102795",
          "K110404",
          "K143242",
          "K143106",
          "K183010",
          "K162775",
          "K932246",
          "K061340",
          "K211877",
          "K233093",
          "K001010",
          "K902246",
          "K021657",
          "K094017",
          "K902335",
          "K972626",
          "K190122",
          "K182301",
          "K191755",
          "K133162",
          "K130143",
          "K972770",
          "K993371",
          "K161945",
          "K112891",
          "K153721",
          "K181732",
          "K211671",
          "K040267",
          "K002772",
          "K102578",
          "K160342",
          "K251717",
          "K120955",
          "K230471",
          "K190402",
          "K111400",
          "K032829",
          "K142550",
          "K922869",
          "K834432",
          "K251776",
          "K243574",
          "K240875",
          "K094050",
          "K222380",
          "K961482",
          "K915192",
          "K243277",
          "K901338",
          "K951027",
          "K100542",
          "K072531",
          "K031729",
          "K140942",
          "K032950",
          "K193223",
          "K192084",
          "K102437",
          "K954684",
          "K102367",
          "K103170",
          "K210695",
          "K011258",
          "K893762",
          "K871585",
          "K871772",
          "K023528",
          "K932070",
          "K243295",
          "K963255",
          "K905810",
          "K950010",
          "K230120",
          "K973164",
          "K922620",
          "K923443",
          "K893252",
          "K193105",
          "K852059",
          "K993159",
          "K943025",
          "K771598",
          "K230321",
          "K061882",
          "K772225",
          "K980276",
          "K041085",
          "K203447",
          "K042515",
          "K222700",
          "K131679",
          "K932677",
          "K980999",
          "K915512",
          "K961021",
          "K183596",
          "K251241",
          "K212411",
          "K904208",
          "K004013",
          "K042883",
          "K896542",
          "K032215",
          "K910940",
          "K253197",
          "K120507",
          "K033342",
          "K984291",
          "K864627",
          "K922734",
          "K922104",
          "K123457",
          "K162422",
          "K110547",
          "K211664",
          "K854883",
          "K860594",
          "K934354",
          "K934589",
          "K230844",
          "K242401",
          "K051640",
          "K933290",
          "K884398",
          "K120790",
          "K941128",
          "K940813",
          "K103733",
          "K062768",
          "K101281",
          "K161414",
          "K131310",
          "K840100",
          "K920034",
          "K182574",
          "K082019",
          "K212839",
          "K140555",
          "K082500",
          "K943299",
          "K221479",
          "K991143",
          "K924482",
          "K980637",
          "K912894",
          "K021451",
          "K974843",
          "K955421",
          "K240669",
          "K231037",
          "K833921",
          "K020703",
          "K113060",
          "K983153",
          "K991882",
          "K142933",
          "K182251",
          "K781294",
          "K960856",
          "K200209",
          "K223252",
          "K896452",
          "K122232",
          "K043440",
          "K201023",
          "K101499",
          "K041591",
          "K120316",
          "K153437",
          "K972752",
          "K102125",
          "K930831",
          "K160484",
          "K894403",
          "K973412",
          "K890132",
          "K915584",
          "K873197",
          "K123342",
          "K122870",
          "K042757",
          "K190068",
          "K802922",
          "K133342",
          "K870730",
          "K261458",
          "K152169",
          "K162035",
          "K040166",
          "K031183",
          "K210961",
          "K152494",
          "K933494",
          "K103117",
          "K896045",
          "K924940",
          "K192507",
          "K122183",
          "K924698",
          "K234044",
          "K930189",
          "K113325",
          "K140222",
          "K915525",
          "K062037",
          "K000590",
          "K032151",
          "K170282",
          "K181569",
          "K101032",
          "K142787",
          "K142807",
          "K200912",
          "K113482",
          "K053514",
          "K193284",
          "K230555",
          "K000978",
          "K972501",
          "K944845",
          "K060525",
          "K082469",
          "K133737",
          "K180493",
          "K000232",
          "K200407",
          "K002335",
          "K923807",
          "K071991",
          "K012702",
          "K930190",
          "K262251",
          "K072858",
          "K012348",
          "K020383",
          "K885036",
          "K925901",
          "K931960",
          "K113571",
          "K971789",
          "K042271",
          "K110362",
          "K150890",
          "K813598",
          "K882234",
          "K894658",
          "K131472",
          "K963709",
          "K910989",
          "K884410",
          "K130708",
          "K140833",
          "K123159",
          "K920524",
          "K915132",
          "K200826",
          "K161705",
          "K162273",
          "K131409",
          "K051146",
          "K882322",
          "K930554",
          "K152726",
          "K023186",
          "K874635",
          "K930595",
          "K042993",
          "K080936",
          "K972774",
          "K030978",
          "K940713",
          "K992167",
          "K060742",
          "K912147",
          "K111711",
          "K070981"
        ],
        "registration_number": [
          "",
          "1221053",
          "3009760038",
          "3009546990",
          "3014763043",
          "9614209",
          "3008572101",
          "3006395932",
          "3009732568",
          "1000547811",
          "3004641308",
          "3014833750",
          "3006356043",
          "3004196040",
          "1833920",
          "3010288357",
          "3014273644",
          "1827096",
          "3007366790",
          "3004450489",
          "2249697",
          "1649518",
          "9617840",
          "3010392243",
          "9610612",
          "2134947",
          "3008812251",
          "3010375065",
          "3010287737",
          "3007963827",
          "3010097171",
          "3008102042",
          "1226544",
          "1000517406",
          "3015916578",
          "3009973336",
          "3022142628",
          "3004638532",
          "2183744",
          "3010314800",
          "3009973505",
          "3035366890",
          "3008824097",
          "3005551626",
          "3008791302",
          "3007539489",
          "3010132111",
          "2244478",
          "3010178296",
          "1828288",
          "1219602",
          "3005596372",
          "3000236920",
          "3006849754",
          "3005718816",
          "3005852519",
          "1047843",
          "3027484613",
          "3002807314",
          "3008985661",
          "3013417938",
          "3015542154",
          "1531174",
          "3016851379",
          "3013055499",
          "3010388970",
          "3007334927",
          "3033796901",
          "3010039400",
          "3019767615",
          "3010462278",
          "3011061242",
          "3017397663",
          "3004513872",
          "2245304",
          "3002807315",
          "2528981",
          "1722511",
          "3003998208",
          "3005581234",
          "3010363503",
          "3005652161",
          "3004371426",
          "1057425",
          "3021566676",
          "3015806723",
          "3004153896",
          "1422549",
          "1218882",
          "3008110533",
          "3010400367",
          "3013176080",
          "1835251",
          "3015167917",
          "3012130008",
          "3009241418",
          "3015398319",
          "3009848551",
          "3009887475",
          "3019006180",
          "9616680",
          "3005061536",
          "3010220595",
          "3038503932",
          "3013450937",
          "3007675554",
          "3008421730",
          "2648920",
          "3005180920",
          "1220477",
          "3006128100",
          "1000286794",
          "2183946",
          "3020703621",
          "3004976965",
          "1833986",
          "3003387384",
          "1530390",
          "3010468876",
          "3014162916",
          "1221763",
          "9616671",
          "3004153240",
          "2531477",
          "3014662844",
          "8043792",
          "9681258",
          "1835572",
          "1832228",
          "1000200989",
          "1219655",
          "3004748528",
          "3010123256",
          "3002558847",
          "3016470314",
          "3018052045",
          "3010173425",
          "3017410889",
          "3009813841",
          "3009971621",
          "3020745236",
          "1529009",
          "3003916006",
          "3010163695",
          "3011724582",
          "3012725451",
          "1423662",
          "3042508254",
          "1018470",
          "3034676720",
          "3005562917",
          "3009923578",
          "1834331",
          "3005360851",
          "2242737",
          "3016438694",
          "3010536692",
          "3010407203",
          "8010379",
          "3016090213",
          "1828464",
          "3011170410",
          "3003477135",
          "9613350",
          "1822565",
          "9681642",
          "3006801265",
          "3006946276",
          "3011683674",
          "3005874553",
          "1054811",
          "3012319330",
          "1721676",
          "3007311878",
          "3016096064",
          "3020155054",
          "3026771806",
          "1064858",
          "9610726",
          "3026776",
          "2134285",
          "3015259876",
          "1643817",
          "3036756245",
          "3009555447",
          "1061927",
          "3007913880",
          "3002806470",
          "3012523063",
          "3011187779",
          "9614841",
          "3011461101",
          "3009882462",
          "3002808022",
          "3009293341",
          "3010856250",
          "3013014058",
          "1724955",
          "1646498",
          "3002807310",
          "1835444",
          "3021902344",
          "3008262872",
          "3025603301",
          "3002806603",
          "3005751028",
          "3011295718",
          "3006406161",
          "2249615",
          "3010120104",
          "1000550978",
          "3014252644",
          "3021008900",
          "3010898856",
          "1645151",
          "3008285983",
          "3007923096",
          "3013302242",
          "3002807295",
          "3009173317",
          "1038671",
          "3012275107",
          "9681465",
          "3011015572",
          "3000264985",
          "3006694268",
          "3007189546",
          "3006791286",
          "3008868758",
          "3006563559",
          "1223434",
          "2133928",
          "3012329926",
          "3009656988",
          "3010162973",
          "3042172882",
          "3014315560",
          "1319660",
          "1833506",
          "1043653",
          "3009475821",
          "3023852420",
          "3009998573",
          "3009871135",
          "1450662",
          "1226188",
          "3008021110",
          "3008395353",
          "1644408",
          "3015207155",
          "3002807112",
          "3013843055",
          "3015231789",
          "3003541440",
          "3010009693",
          "1526534",
          "2032521",
          "1720929",
          "3008658204",
          "3010576282",
          "3008544874",
          "3006487372",
          "3010041693",
          "3015516266",
          "3009590742",
          "3013186738",
          "3013791180",
          "1424263",
          "3010386387",
          "9680215",
          "1643264",
          "8010916",
          "3005144609",
          "1818910",
          "3006015326",
          "3009106092",
          "3009555440",
          "3005279746",
          "3010370880",
          "9617083",
          "2134470",
          "3009888740",
          "3013194153",
          "3010155661",
          "2246552",
          "3010331645",
          "3010531069",
          "3033509898",
          "2030624",
          "3040391045",
          "3012491394",
          "3014302784",
          "3007420745",
          "3010390570",
          "3006017180",
          "3007516837",
          "1825034",
          "9613369",
          "3010047454",
          "3009582259",
          "1020279",
          "3010537287",
          "3030446955",
          "3002907620",
          "3011274144",
          "3003639920",
          "3007353762",
          "3030412764",
          "3014644608",
          "3010846112",
          "3004142400",
          "3012392319",
          "9611390",
          "2025182",
          "3007200272",
          "3010057495",
          "1220246",
          "1836357"
        ],
        "fei_number": [
          "",
          "1221053",
          "1000118610",
          "1000132037",
          "3009760038",
          "3009546990",
          "3014763043",
          "3005985723",
          "3008572101",
          "3006395932",
          "1000143502",
          "3009732568",
          "1000547811",
          "3004641308",
          "1000137563",
          "3014833750",
          "1000114185",
          "3006356043",
          "3004196040",
          "3003569173",
          "3002993297",
          "3010288357",
          "3014273644",
          "1827096",
          "3004450489",
          "3010392243",
          "3002806725",
          "3010375065",
          "3002806773",
          "3001452082",
          "3010287737",
          "3007963827",
          "3003015552",
          "3010097171",
          "1000139987",
          "3002462840",
          "1000517406",
          "3015916578",
          "3009973336",
          "3022142628",
          "3004638532",
          "2183744",
          "3010314800",
          "3009973505",
          "3035366890",
          "3008824097",
          "3005551626",
          "3008791302",
          "3007539489",
          "3000205591",
          "3005601732",
          "3010132111",
          "1828288",
          "1219602",
          "3005596372",
          "3006849754",
          "3005718816",
          "3005852519",
          "1047843",
          "3027484613",
          "3002807314",
          "3002807172",
          "3003128516",
          "3008985661",
          "3013417938",
          "3002459190",
          "3015542154",
          "3016851379",
          "3013055499",
          "3010388970",
          "3007334927",
          "3033796901",
          "3010039400",
          "3019767615",
          "3002808402",
          "3010462278",
          "3011061242",
          "3000203000",
          "3017397663",
          "3004513872",
          "2245304",
          "3002807315",
          "2528981",
          "1000395711",
          "3003998208",
          "3005581234",
          "3003574398",
          "1000149660",
          "3010363503",
          "3005652161",
          "3021566676",
          "3015806723",
          "3004153896",
          "1218882",
          "3008110533",
          "3010400367",
          "3013176080",
          "3015167917",
          "3012130008",
          "3009241418",
          "3002924544",
          "3009848551",
          "3009887475",
          "3019006180",
          "3005061536",
          "3010220595",
          "3038503932",
          "3013450937",
          "3007675554",
          "3008421730",
          "3010938185",
          "3002806315",
          "3005180920",
          "1220477",
          "3006128100",
          "1000286794",
          "3020703621",
          "3001084824",
          "3004976965",
          "3003387384",
          "3003070421",
          "3002807893",
          "3008342029",
          "1000116116",
          "3007282753",
          "3014162916",
          "1832656",
          "1221763",
          "3004153240",
          "3014662844",
          "1832228",
          "1000200989",
          "1219655",
          "3004748528",
          "3010123256",
          "3002558847",
          "3016470314",
          "3002854251",
          "3018052045",
          "3009813841",
          "1000115656",
          "3009971621",
          "3020745236",
          "3000717687",
          "3003916006",
          "3010163695",
          "3011724582",
          "3012725451",
          "3002808293",
          "3042508254",
          "1018470",
          "3034676720",
          "3005562917",
          "3009923578",
          "3005360851",
          "3001236277",
          "3016438694",
          "1000086912",
          "3010536692",
          "3010407203",
          "3016090213",
          "1828464",
          "3011170410",
          "3003477135",
          "3006801265",
          "3006946276",
          "3011683674",
          "3005874553",
          "3000206435",
          "3012319330",
          "1721676",
          "3007311878",
          "3016096064",
          "3020155054",
          "3026771806",
          "3015259876",
          "1643817",
          "3036756245",
          "3009555447",
          "3007913880",
          "3002806470",
          "3012523063",
          "3011187779",
          "3011461101",
          "3009882462",
          "3002808022",
          "3009293341",
          "3010856250",
          "3013014058",
          "3003714383",
          "3000215103",
          "3002807310",
          "3021902344",
          "3008262872",
          "1000306340",
          "3025603301",
          "3002806603",
          "3005751028",
          "3002807113",
          "3011295718",
          "3006406161",
          "1000221916",
          "3010120104",
          "1000550978",
          "3014252644",
          "1000136269",
          "3021008900",
          "3008397795",
          "3010898856",
          "3008285983",
          "3007923096",
          "1000138247",
          "3013302242",
          "3002807295",
          "3009173317",
          "1038671",
          "3012275107",
          "1000159999",
          "3011015572",
          "3000264985",
          "3002287318",
          "3007932279",
          "3006694268",
          "3007189546",
          "3006791286",
          "3008868758",
          "3006563559",
          "3012329926",
          "3009656988",
          "3003386935",
          "3042172882",
          "3014315560",
          "3002808269",
          "1043653",
          "3009475821",
          "3023852420",
          "1000116912",
          "3009998573",
          "3009871135",
          "1450662",
          "1000136478",
          "3008021110",
          "3008395353",
          "3015207155",
          "3002807112",
          "3013843055",
          "3015231789",
          "3003541440",
          "3010009693",
          "1526534",
          "1000199976",
          "1720929",
          "3008658204",
          "3003142829",
          "3002806534",
          "3010576282",
          "3008544874",
          "3000215273",
          "3001354499",
          "3006487372",
          "3010041693",
          "3015516266",
          "3001049282",
          "3009590742",
          "3013186738",
          "1000220733",
          "3013791180",
          "3010386387",
          "1643264",
          "3005144609",
          "1818910",
          "3006015326",
          "3009106092",
          "3009555440",
          "3005279746",
          "3010370880",
          "3009888740",
          "3013194153",
          "3010155661",
          "2246552",
          "3010331645",
          "3010531069",
          "3033509898",
          "3040391045",
          "3001642380",
          "3000717902",
          "3012491394",
          "3002807041",
          "3014302784",
          "3007420745",
          "3010390570",
          "3001548076",
          "3006017180",
          "3007516837",
          "1825034",
          "3003110073",
          "3010047454",
          "3009582259",
          "1020279",
          "3010537287",
          "3002807036",
          "3030446955",
          "3002907620",
          "3011274144",
          "3003639920",
          "3005233524",
          "3007353762",
          "3030412764",
          "3014644608",
          "3010846112",
          "3004142400",
          "3012392319",
          "2025182",
          "3007200272",
          "3010057495",
          "1220246",
          "1000135253",
          "3003064967"
        ],
        "device_name": "Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3560",
        "device_class": "2"
      }
    }
  ]
}