{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "85372",
      "product_res_number": "Z-0063-2010",
      "event_date_initiated": "2009-10-01",
      "event_date_posted": "2009-12-08",
      "recall_status": "Terminated",
      "event_date_terminated": "2013-09-27",
      "res_event_number": "53383",
      "product_code": "MKJ",
      "k_numbers": [
        "K051632"
      ],
      "product_description": "Heartstart FR2+ automated external defibrillator Models M3860A and M3840A are equipped with an ECG Module. M3860A is under the Philips label and M3840A is under the Laerdal label.",
      "code_info": "4 model (catalog) numbers are affected: M3860A, M3861A, M3840A, and M3841A. Specifics for each code: M3860A: Philips Heartstart FR2+ Defibrillator with ECG Module (various languages)  M3861A: Philips Heartstart FR2+ Defibrillator Text Display, No ECG Module (various languages).  M3840A: Laerdal Branded Heartstart FR2+ Defibrillator with ECG Module (various languages).  M3841A: Laerdal Branded Heartstart FR2+ Defibrillator Text Display, No ECG Module (various languages).  The following letter codes are used in conjunction with the catalog numbers to indicate language of Instructions for Use (IFU) and FR2+ voice prompts: ABA=U.S. English; ABU=U.K. English; ABD=German; ABE=Spanish; ABF=French; ABH=Dutch; ABI=Hong Kong; ABJ=Japanese; ABN=Norwegian; ABO=Taiwanese; ABS=Swedish; ABX=Finland; ABZ=Italian; AB2=China; and  AC4=Portugal.  Additional (non-language related) letter codes for product: A01=Product with TSO (Technical Standard Order) approved battery; and RF=Service Replacement.  Following serial numbers listed as follows: M3860A=U.S. Distribution M3861A=U.S. Distribution M3860A=Canadian Distribution M3861A=Canadian Distribution M3860A=International, not including Canada M3861A=Interantional, not including Canada Laerdal=M3840A and M3841A=all international M3860A U.S. DISTRIBUTION:   0108267575, 0807248065, 0807248798, 0807249110, 0807250145, 0907253534, 0907254125, 0907254417, 0907254602, 0108267741, 0807248084, 0807248802, 0807249112, 0807250147, 0907253540, 0907254127, 0907254421, 0907254603, 0307230176, 0807248100, 0807248804, 0807249113, 0807250166, 0907253544, 0907254129, 0907254423, 0907254604, 0507237301, 0807248160, 0807248806, 0807249116, 0807250177, 0907253546, 0907254132, 0907254425, 0907254605, 0507237302, 0807248166, 0807248815, 0807249118, 0807250179, 0907253548, 0907254133, 0907254427, 0907254606, 0507237303, 0807248168, 0807248827, 0807249122, 0807250198, 0907253552, 0907254136, 0907254429, 0907254607, 0507237304, 0807248172, 0807248830, 0807249126, 0807250211, 0907253554, 0907254137, 0907254431, 0907254608, 0507237305, 0807248176, 0807248840, 0807249128, 0807250212, 0907253556, 0907254139, 0907254433, 0907254609, 0507237321, 0807248199, 0807248841, 0807249132, 0807250219, 0907253558, 0907254140, 0907254437, 0907254611, 0507237323, 0807248200, 0807248844, 0807249133, 0807250223, 0907253560, 0907254142, 0907254439, 0907254612, 0507237353, 0807248220, 0807248845, 0807249134, 0807250224, 0907253566, 0907254143, 0907254441, 0907254613, 0507237426, 0807248222, 0807248846, 0807249137, 0807250225, 0907253574, 0907254144, 0907254443, 0907254615, 0507237438, 0807248248, 0807248848, 0807249141, 0807250227, 0907253587, 0907254150, 0907254447, 0907254616, 0507237446, 0807248252, 0807248851, 0807249142, 0807250229, 0907253600, 0907254165, 0907254449, 0907254617, 0507237462, 0807248254, 0807248853, 0807249144, 0807250231, 0907253641, 0907254166, 0907254451, 0907254619, 0507237511, 0807248262, 0807248855, 0807249145, 0807250236, 0907253643, 0907254167, 0907254455, 0907254620, 0507237512, 0807248278, 0807248856, 0807249149, 0807250240, 0907253664, 0907254169, 0907254459, 0907254621, 0507237541, 0807248288, 0807248857, 0807249150, 0807250243, 0907253676, 0907254170, 0907254461, 0907254623, 0507237555, 0807248290, 0807248862, 0807249155, 0807250244, 0907253677, 0907254171, 0907254462, 0907254625, 0507237557, 0807248292, 0807248891, 0807249158, 0807250249, 0907253678, 0907254172, 0907254465, 0907254626, 0507237608, 0807248294, 0807248894, 0807249159, 0807250252, 0907253696, 0907254173, 0907254482, 0907254627, 0507237611, 0807248300, 0807248895, 0807249160, 0807250258, 0907253698, 0907254174, 0907254484, 0907254628, 0507237649, 0807248303, 0807248896, 0807249165, 0807250262, 0907253703, 0907254175, 0907254486, 0907254630, 0507237651, 0807248309, 0807248897, 0807249169, 0807250276, 0907253719, 0907254176, 0907254488, 0907254632, 0507237657, 0807248315, 0807248900, 0807249171, 0807250342, 0907253791, 0907254179, 0907254492, 0907254634, 0507237679, 0807248319, 0807248901, 0807249175, 0807250344, 0907253792, 0907254183, 0907254494, 0907254638, 0507237682, 0807248323, 0807248902, 0807249177, 0807250349, 0907253806, 0907254187, 0907254498, 0907254696, 0507237686, 0807248327, 0807248905, 0807249181, 0807250354, 0907253832, 0907254189, 0907254500, 0907254804, 0507237699, 0807248337, 0807248907, 0807249183, 0807250362, 0907253836, 0907254191, 0907254505, 0907254806, 0507237700, 0807248391, 0807248912, 0807249185, 0807250367, 0907253844, 0907254195, 0907254507, 0907254810, 0507237704, 0807248412, 0807248917, 0807249192, 0807250377, 0907253882, 0907254215, 0907254509, 0907254826, 0507237765, 0807248456, 0807248922, 0807249194, 0807250378, 0907253885, 0907254223, 0907254511, 0907254836, 0507237816, 0807248468, 0807248923, 0807249196, 0807250385, 0907253886, 0907254225, 0907254512, 0907254901, 0507237846, 0807248472, 0807248924, 0807249200, 0807250388, 0907253887, 0907254227, 0907254513, 0907254903, 0507237848, 0807248474, 0807248926, 0807249205, 0807250390, 0907253889, 0907254229, 0907254517, 0907254905, 0607239532, 0807248476, 0807248930, 0807249207, 0807250392, 0907253893, 0907254231, 0907254518, 0907254907, 0607239882, 0807248480, 0807248932, 0807249225, 0807250398, 0907253895, 0907254233, 0907254520, 0907254909, 0607239884, 0807248482, 0807248933, 0807249226, 0807250446, 0907253899, 0907254239, 0907254522, 0907254911, 0607239886, 0807248490, 0807248935, 0807249227, 0807250915, 0907253901, 0907254241, 0907254524, 0907254913, 0607239888, 0807248492, 0807248938, 0807249231, 0907252699, 0907253906, 0907254243, 0907254526, 0907254915, 0607239910, 0807248496, 0807248940, 0807249237, 0907252731, 0907253921, 0907254245, 0907254528, 0907254917, 0607239911, 0807248509, 0807248944, 0807249239, 0907253237, 0907253923, 0907254247, 0907254530, 0907254919, 0607239953, 0807248512, 0807248948, 0807249241, 0907253241, 0907253926, 0907254257, 0907254532, 0907254921, 0607239994, 0807248517, 0807248958, 0807249243, 0907253329, 0907253928, 0907254259, 0907254538, 0907254923, 0607239996, 0807248519, 0807248964, 0807249245, 0907253337, 0907253930, 0907254261, 0907254540, 0907254925, 0607240000, 0807248524, 0807248967, 0807249247, 0907253398, 0907253932, 0907254263, 0907254542, 0907254927, 0607240002, 0807248535, 0807248969, 0807249249, 0907253400, 0907253934, 0907254281, 0907254544, 0907254929, 0607240089, 0807248537, 0807248971, 0807249255, 0907253401, 0907253936, 0907254283, 0907254546, 0907254931, 0607240114, 0807248589, 0807248972, 0807249257, 0907253402, 0907253941, 0907254285, 0907254552, 0907254933, 0607240130, 0807248622, 0807248974, 0807249258, 0907253407, 0907253942, 0907254289, 0907254554, 0907254935, 0607240134, 0807248626, 0807248975, 0807249262, 0907253409, 0907253946, 0907254301, 0907254556, 0907254937, 0607240315, 0807248628, 0807248978, 0807249270, 0907253411, 0907253947, 0907254305, 0907254558, 0907254939, 0607240317, 0807248630, 0807248980, 0807249272, 0907253413, 0907253952, 0907254315, 0907254560, 0907254941, 0607240319, 0807248632, 0807248982, 0807249280, 0907253420, 0907253954, 0907254328, 0907254562, 0907254943, 0607240321, 0807248636, 0807248984, 0807249284, 0907253425, 0907253956, 0907254335, 0907254564, 0907254945, 0607240325, 0807248638, 0807248986, 0807249305, 0907253426, 0907253962, 0907254345, 0907254566, 0907254947, 0607240327, 0807248642, 0807248988, 0807249307, 0907253428, 0907253965, 0907254347, 0907254568, 0907254949, 0607240331, 0807248644, 0807248989, 0807249309, 0907253430, 0907253966, 0907254349, 0907254569, 0907254951, 0607240974, 0807248646, 0807248990, 0807249311, 0907253436, 0907253968, 0907254353, 0907254571, 0907254953, 0607241467, 0807248648, 0807248992, 0807249313, 0907253441, 0907253969, 0907254361, 0907254572, 0907254955, 0607241477, 0807248650, 0807248994, 0807249315, 0907253442, 0907253973, 0907254363, 0907254573, 0907254957, 0607241479, 0807248658, 0807248996, 0807249317, 0907253452, 0907253981, 0907254365, 0907254575, 0907254959, 0607241481, 0807248667, 0807248998, 0807249321, 0907253454, 0907253983, 0907254371, 0907254576, 0907254961, 0607241874, 0807248669, 0807249000, 0807249323, 0907253456, 0907253989, 0907254383, 0907254577, 0907254963, 0607242131, 0807248672, 0807249013, 0807249325, 0907253463, 0907253999, 0907254385, 0907254579, 0907254965, 0607242143, 0807248675, 0807249017, 0807249692, 0907253476, 0907254004, 0907254387, 0907254580, 0907254967, 0607242145, 0807248676, 0807249019, 0807249745, 0907253478, 0907254029, 0907254389, 0907254581, 0907254973, 0607242147, 0807248681, 0807249023, 0807249760, 0907253484, 0907254031, 0907254391, 0907254582, 0907254975, 0707243308, 0807248683, 0807249025, 0807249877, 0907253492, 0907254051, 0907254392, 0907254583, 0907254977, 0707243632, 0807248691, 0807249027, 0807249879, 0907253493, 0907254053, 0907254393, 0907254585, 0907254979, 0707244329, 0807248693, 0807249029, 0807249973, 0907253497, 0907254087, 0907254394, 0907254586, 0907254981, 0707244388, 0807248695, 0807249031, 0807250042, 0907253499, 0907254104, 0907254395, 0907254590, 0907254983, 0707244494, 0807248699, 0807249033, 0807250059, 0907253502, 0907254108, 0907254396, 0907254591, 0907254985, 0707245904, 0807248732, 0807249035, 0807250079, 0907253503, 0907254111, 0907254397, 0907254592, 0907254987, 0707246078, 0807248734, 0807249037, 0807250097, 0907253504, 0907254114, 0907254399, 0907254593, 0907254989, 0707246276, 0807248736, 0807249039, 0807250101, 0907253507, 0907254115, 0907254401, 0907254594, 0907254991, 0807247654, 0807248750, 0807249041, 0807250103, 0907253512, 0907254116, 0907254403, 0907254595, 0907254993, 0807247692, 0807248754, 0807249043, 0807250109, 0907253520, 0907254117, 0907254405, 0907254596, 0907254995, 0807247809, 0807248756, 0807249045, 0807250113, 0907253524, 0907254119, 0907254407, 0907254597, 0907254997, 0807247811, 0807248758, 0807249101, 0807250117, 0907253526, 0907254120, 0907254409, 0907254598, 0908287926, 0807248037, 0807248764, 0807249106, 0807250129, 0907253528, 0907254121, 0907254411, 0907254599, 0908287992, 0807248059, 0807248766, 0807249107, 0807250135, 0907253530, 0907254122, 0907254413, 0907254600, 1007255537, 0807248063, 0807248768, 0807249109, 0807250144, 0907253532, 0907254123, 0907254415, AND 0907254601.==================================================  M3861A U.S. DISTRIBUTION:    0108267721, 0807248357, 0907253438, 0907254101, 0907254280, 0907254452, 0907254699, 0907254819, 0108267727, 0807248359, 0907253485, 0907254103, 0907254282, 0907254454, 0907254700, 0907254820, 0308273383, 0807248361, 0907253487, 0907254106, 0907254284, 0907254456, 0907254701, 0907254821, 0507237606, 0807248373, 0907253488, 0907254109, 0907254288, 0907254458, 0907254702, 0907254822, 0507237614, 0807248379, 0907253489, 0907254110, 0907254290, 0907254460, 0907254703, 0907254824, 0507237737, 0807248381, 0907253491, 0907254112, 0907254291, 0907254464, 0907254704, 0907254825, 0507237897, 0807248399, 0907253495, 0907254113, 0907254293, 0907254466, 0907254705, 0907254827, 0607239687, 0807248460, 0907253496, 0907254126, 0907254294, 0907254468, 0907254706, 0907254828, 0607239744, 0807248514, 0907253519, 0907254128, 0907254296, 0907254470, 0907254707, 0907254829, 0607239756, 0807248518, 0907253538, 0907254130, 0907254298, 0907254472, 0907254708, 0907254830, 0607239890, 0807248525, 0907253542, 0907254134, 0907254300, 0907254474, 0907254709, 0907254831, 0607239947, 0807248531, 0907253557, 0907254135, 0907254302, 0907254476, 0907254710, 0907254832, 0607239957, 0807248533, 0907253611, 0907254138, 0907254304, 0907254478, 0907254711, 0907254833, 0607240006, 0807248539, 0907253660, 0907254145, 0907254306, 0907254480, 0907254712, 0907254835, 0607240106, 0807248564, 0907253667, 0907254147, 0907254308, 0907254490, 0907254713, 0907254838, 0607240112, 0807248583, 0907253668, 0907254148, 0907254310, 0907254502, 0907254714, 0907254840, 0607240124, 0807248595, 0907253669, 0907254149, 0907254312, 0907254506, 0907254715, 0907254841, 0607240126, 0807248634, 0907253671, 0907254151, 0907254314, 0907254510, 0907254717, 0907254843, 0607240521, 0807248645, 0907253672, 0907254152, 0907254316, 0907254514, 0907254718, 0907254844, 0607241459, 0807248684, 0907253674, 0907254154, 0907254318, 0907254523, 0907254719, 0907254846, 0607241461, 0807248702, 0907253682, 0907254155, 0907254320, 0907254534, 0907254720, 0907254847, 0607241469, 0807248739, 0907253692, 0907254156, 0907254322, 0907254567, 0907254721, 0907254849, 0607241473, 0807248916, 0907253700, 0907254157, 0907254326, 0907254584, 0907254722, 0907254850, 0607241475, 0807248937, 0907253717, 0907254158, 0907254330, 0907254588, 0907254725, 0907254851, 0607241591, 0807248945, 0907253724, 0907254159, 0907254332, 0907254614, 0907254726, 0907254854, 0607241595, 0807248947, 0907253725, 0907254160, 0907254334, 0907254618, 0907254727, 0907254856, 0607241597, 0807248997, 0907253728, 0907254161, 0907254336, 0907254622, 0907254728, 0907254860, 0607241605, 0807250127, 0907253732, 0907254162, 0907254338, 0907254624, 0907254729, 0907254862, 0607241866, 0807250131, 0907253734, 0907254164, 0907254339, 0907254629, 0907254730, 0907254864, 0607241868, 0807250133, 0907253736, 0907254168, 0907254340, 0907254631, 0907254731, 0907254866, 0607241872, 0807250137, 0907253738, 0907254177, 0907254341, 0907254633, 0907254732, 0907254868, 0607241878, 0807250143, 0907253740, 0907254178, 0907254342, 0907254635, 0907254733, 0907254870, 0607241880, 0807250149, 0907253763, 0907254180, 0907254343, 0907254636, 0907254734, 0907254871, 0607241882, 0807250150, 0907253773, 0907254181, 0907254357, 0907254637, 0907254735, 0907254872, 0607241884, 0807250151, 0907253783, 0907254182, 0907254359, 0907254639, 0907254736, 0907254873, 0607241886, 0807250153, 0907253790, 0907254184, 0907254360, 0907254640, 0907254737, 0907254874, 0607241888, 0807250155, 0907253798, 0907254186, 0907254362, 0907254641, 0907254738, 0907254875, 0607241890, 0807250156, 0907253800, 0907254188, 0907254367, 0907254642, 0907254739, 0907254877, 0607242129, 0807250157, 0907253818, 0907254190, 0907254368, 0907254644, 0907254740, 0907254878, 0607242133, 0807250161, 0907253822, 0907254192, 0907254369, 0907254646, 0907254741, 0907254879, 0607242135, 0807250164, 0907253824, 0907254194, 0907254370, 0907254647, 0907254747, 0907254880, 0607242139, 0807250172, 0907253830, 0907254197, 0907254372, 0907254648, 0907254748, 0907254881, 0607242141, 0807250174, 0907253838, 0907254198, 0907254373, 0907254649, 0907254749, 0907254883, 0607242149, 0807250178, 0907253842, 0907254202, 0907254374, 0907254651, 0907254750, 0907254884, 0607242166, 0807250182, 0907253848, 0907254204, 0907254375, 0907254652, 0907254751, 0907254885, 0607242168, 0807250190, 0907253896, 0907254206, 0907254376, 0907254656, 0907254753, 0907254886, 0607242170, 0807250192, 0907253897, 0907254208, 0907254377, 0907254658, 0907254754, 0907254887, 0607242172, 0807250200, 0907253898, 0907254210, 0907254378, 0907254660, 0907254755, 0907254888, 0607242377, 0807250207, 0907253900, 0907254212, 0907254379, 0907254661, 0907254756, 0907254889, 0707245469, 0807250246, 0907253908, 0907254214, 0907254380, 0907254662, 0907254757, 0907254891, 0707245624, 0807250248, 0907253909, 0907254216, 0907254381, 0907254663, 0907254758, 0907254893, 0707246066, 0807250256, 0907253910, 0907254230, 0907254382, 0907254664, 0907254759, 0907254895, 0807247650, 0807250260, 0907253912, 0907254232, 0907254384, 0907254665, 0907254760, 0907254897, 0807248038, 0807250278, 0907253913, 0907254234, 0907254386, 0907254666, 0907254761, 0907254899, 0807248041, 0807250282, 0907253933, 0907254236, 0907254388, 0907254667, 0907254762, 0907254922, 0807248075, 0807250286, 0907253938, 0907254240, 0907254390, 0907254668, 0907254765, 0907254924, 0807248077, 0807250288, 0907253940, 0907254242, 0907254398, 0907254669, 0907254767, 1007255202, 0807248097, 0807250290, 0907253943, 0907254244, 0907254400, 0907254670, 0907254769, 1007256655, 0807248099, 0807250292, 0907253944, 0907254246, 0907254402, 0907254671, 0907254771\n\n 0807248103, 0807250294, 0907253948, 0907254248, 0907254404, 0907254672, 0907254773\n\n 0807248109, 0807250296, 0907253955, 0907254249, 0907254406, 0907254673, 0907254774\n\n 0807248121, 0807250302, 0907253961, 0907254250, 0907254408, 0907254674, 0907254775\n\n 0807248123, 0807250308, 0907253963, 0907254251, 0907254410, 0907254675, 0907254777\n\n 0807248125, 0807250316, 0907253977, 0907254252, 0907254412, 0907254676, 0907254779, 0807248129, 0807250318, 0907254001, 0907254253, 0907254414, 0907254677, 0907254781, 0807248131, 0807250322, 0907254003, 0907254254, 0907254416, 0907254678, 0907254783, 0807248133, 0807250326, 0907254006, 0907254256, 0907254418, 0907254679, 0907254787, 0807248135, 0807250346, 0907254009, 0907254258, 0907254420, 0907254680, 0907254789, 0807248137, 0807250366, 0907254017, 0907254260, 0907254422, 0907254681, 0907254791, 0807248139, 0807250387, 0907254027, 0907254262, 0907254424, 0907254683, 0907254793, 0807248141, 0807250395, 0907254045, 0907254264, 0907254426, 0907254684, 0907254795, 0807248143, 0807251069, 0907254047, 0907254265, 0907254428, 0907254685, 0907254797, 0807248145, 0907251325, 0907254049, 0907254266, 0907254430, 0907254686, 0907254799, 0807248147, 0907253257, 0907254055, 0907254267, 0907254432, 0907254687, 0907254801, 0807248162, 0907253259, 0907254059, 0907254268, 0907254434, 0907254688, 0907254803, 0807248164, 0907253265, 0907254061, 0907254269, 0907254436, 0907254689, 0907254805, 0807248170, 0907253275, 0907254065, 0907254270, 0907254438, 0907254690, 0907254807, 0807248214, 0907253323, 0907254067, 0907254271, 0907254440, 0907254691, 0907254809, 0807248317, 0907253404, 0907254073, 0907254272, 0907254442, 0907254693, 0907254811, 0807248325, 0907253418, 0907254089, 0907254273, 0907254444, 0907254694, 0907254813, 0807248339, 0907253421, 0907254091, 0907254274, 0907254446, 0907254695, 0907254814, 0807248351, 0907253432, 0907254093, 0907254276, 0907254448, 0907254697, 0907254815, 0807248355, 0907253434, 0907254100, 0907254278, 0907254450, 0907254698, AND 0907254816.  M3860A  CANADA DISTRIBUTION: 0607240550, 0707243439, 0707243892, 0807246607, 0807246609, 0807246615, 0807246637, 0807248842, 0807248908, 0807248968, 0807249278, 0807249282, 0807249296, 0807250176, 0807250811, 0907253718, 0907253796, 0907254275, 0907254279, 0907254364, 0907254366, 0907254496, 0907254548, 0907254570, AND 0907254610.  =============================================\t\t\t\t==============================================\t M3861A CANADA DISTRIBUTION: 0607239912, 0807250320, 0907253701, 0907253726, 0907254023, 0907254105, 0907254141, 0907254153, AND 0907254504.  M3860A INTERNATIONAL-NOT INCLUDING CANADA: 0108267591, 0108267701, 0108267705, 0108267715, 0108267723, 0507236763, 0507237318, 0507237325, 0507237333, 0507237369, 0507237377, 0507237391, 0507237393, 0507237409, 0507237410, 0507237460, 0507237504, 0507237506, 0507237527, 0507237547, 0507237551, 0507237668, 0507237676, 0507237778, 0507237803, 0507237805, 0507237808, 0507237824, 0507237826, 0508275018, 0607239952, 0607239954, 0607239956, 0607239958, 0607240008, 0607240013, 0607240021, 0607240029, 0607240037, 0607240039, 0607240043, 0607240045, 0607240047, 0607240049, 0607240051, 0607239952, 0607239954, 0607239956, 0607239958, 0607240008, 0607240013, 0607240021, 0607240029, 0607240037, 0607240039, 0607240043, 0607240045, 0607240047, 0607240049, 0607240051, 0607240105, 0607240167, 0607240171, 0607240205, 0607240259, 0607240272, 0607240306, 0607240316, 0607240318, 0607240320, 0607240322, 0607240323, 0607240332, 0607240334, 0607240346, 0607240356, 0607240360, 0607240365, 0607240370, 0607240372, 0607240374, 0607240376, 0607240380, 0607240809, 0607240811, 0607240864, 0607240878, 0607240880, 0607240882, 0607240886, 0607240904, 0607241005, 0607241471, 0607241876, 0607241894, 0607242365, 0607242466, 0707243291, 0707243340, 0707243399, 0707243444, 0707243838, 0707244007, 0707244013, 0707244020, 0707244024, 0707244039, 0707244041, 0707244044, 0707244399, 0707244405, 0707244408, 0707244483, 0707244485, 0707244493, 0707244503, 0707245561, 0707245706, 0707245707, 0707245708, 0707245712, 0707245714, 0707246252, 0807248019, 0807248023, 0807248053, 0807248055, 0807248061, 0807248072, 0807248078, 0807248119, 0807248127, 0807248204, 0807248216, 0807248224, 0807248228, 0807248232, 0807248236, 0807248244, 0807248255, 0807248259, 0807248261, 0807248263, 0807248265, 0807248269, 0807248271, 0807248273, 0807248275, 0807248277, 0807248283, 0807248285, 0807248287, 0807248289, 0807248291, 0807248293, 0807248297, 0807248302, 0807248307, 0807248345, 0807248347, 0807248367, 0807248369, 0807248420, 0807248422, 0807248424, 0807248428, 0807248430, 0807248434, 0807248436, 0807248438, 0807248440, 0807248446, 0807248452, 0807248462, 0807248464, 0807248478, 0807248488, 0807248668, 0807249047, 0807249063, 0807249204, 0807249276, 0807249303, 0807249601, 0807249605, 0807249609, 0807249611, 0807249614, 0807249615, 0807249616, 0807249617, 0807249619, 0807249620, 0807249621, 0807249623, 0807249624, 0807249626, 0807249627, 0807249628, 0807249630, 0807249632, 0807249633, 0807249634, 0807249635, 0807249636, 0807249637, 0807249638, 0807249639, 0807249640, 0807249641, 0807249642, 0807249643, 0807249645, 0807249646, 0807249647, 0807249648, 0807249649, 0807249652, 0807249655, 0807249659, 0807249672, 0807249674, 0807249676, 0807249681, 0807249685, 0807249688, 0807249689, 0807249699, 0807249749, 0807249802, 0807249804, 0807249807, 0807249816, 0807249818, 0807249822, 0807249828, 0807249830, 0807249834, 0807249836, 0807249844, 0807249866, 0807249868, 0807249870, 0807249872, 0807249874, 0807249876, 0807249882, 0807249886, 0807249888, 0807249890, 0807249892, 0807249894, 0807249901, 0807249903, 0807249904, 0807249905, 0807249907, 0807249911, 0807249912, 0807249913, 0807249915, 0807249917, 0807249919, 0807249921, 0807249925, 0807249933, 0807249935, 0807249937, 0807249957, 0807249977, 0807249979, 0807249981, 0807249995, 0807249999, 0807250077, 0807250116, 0807250168, 0807250170, 0807250187, 0807250218, 0807250560, 0807250562, 0807250590, 0807250647, 0807250651, 0807250913, 0807250927, 0807250943, 0807250945, 0807250947, 0807250955, 0907253219, 0907253245, 0907253251, 0907253262, 0907253267, 0907253269, 0907253271, 0907253272, 0907253278, 0907253280, 0907253281, 0907253283, 0907253319, 0907253321, 0907253325, 0907253327, 0907253394, 0907253794, 0907254025, 0907254131, 0907254146, 0907254235, 0907254237, 0907254287, 0907254419, 0907254654, 0907254742, 0907254746, 0907254766, 0907254782, 0907254788, 0907254794, 0907254896, 0907254906, 0907254908, 0907254912, 0907254914, 0907254916, 0907254960, 0907254964, 0907254966, 0907254968, 0907254970, 0907254971, 0907254999, 0907255003, 0907255007, 0907255013, 0907255028, 0907255045, 0907255057, 0907255061, 0907255065, 0907255067, 0907255069, 0907255071, 0907255073, 0907255077, 0907255081, 0907255085, 0907255087, 0907255089, 0907255091, 0907255095, 0907255108, 0907255110, 0907255113, 0907255116, 0907255118, 0907255120, 0907255121, 0907255122, 0907255136, 0907255141, 0907255163, 0907255165, 1007255210, 1007255218, 1007255231, 1007255233, 1007255237, 1007255241, 1007255243, 1007255249, 1007255267, 1007255271, 1007255301, 1007255303, 1007255305, 1007255307, 1007255313, 1007255317, 1007255319, 1007255323, 1007255371, 1007256663, 1007256675, 1007256677, 1007256679, 1007256681, 1007256683, 1007256704, 1007256767, 1007256773, 1007256777, 1007256781, 1007256787, 1007256793, 1007256795, 1007256799, 1007257012, 1007257014, 1007257016, 1007257063, 1007257071, 1007257073, 1007257006, 1207267451, 1207267455, AND 1207267459.================================================================= M3861A INTERNATIONAL-NOT INCLUDING CANADA: 0108267589, 0507236701, 0507236709, 0507236711, 0507236713, 0507236715, 0507236717, 0507236719, 0507236721, 0507236723, 0507236725, 0507236729, 0507236731, 0507236733, 0507236735, 0507236737, 0507236739, 0507236741, 0507236743, 0507236745, 0507236753, 0507236755.0507236757, 0507236777, 0507236785, 0507236791, 0507236795, 0507236799, 0507237051, 0507237057, 0507237063, 0507237067, 0507237069, 0507237073, 0507237075, 0507237077, 0507237079, 0507237081, 0507237083, 0507237089, 0507237093, 0507237095, 0507237097, 0507237099, 0507237119, 0507237122, 0507237123, 0507237125, 0507237127, 0507237129, 0507237131, 0507237133, 0507237135, 0507237167, 0507237169, 0507237179, 0507237181, 0507237193, 0507237224, 0507237226, 0507237230, 0507237232, 0507237234, 0507237236, 0507237238, 0507237244, 0507237248, 0507237260, 0507237295, 0507237297, 0507237308, 0507237310, 0507237327, 0507237331, 0507237339, 0507237361, 0507237367, 0507237371, 0507237385, 0507237387, 0507237397, 0507237401, 0507237411, 0507237417, 0507237419, 0507237420, 0507237421, 0507237423, 0507237424, 0507237428, 0507237429, 0507237452, 0507237457, 0507237507, 0507237509, 0507237517, 0507237519, 0507237521, 0507237525, 0507237539, 0507237543, 0507237549, 0507237553, 0507237607, 0507237615, 0507237619, 0507237621, 0507237624, 0507237625, 0507237626, 0507237628, 0507237630, 0507237631, 0507237632, 0507237634, 0507237636, 0507237637, 0507237638, 0507237639, 0507237640, 0507237645, 0507237647, 0507237648, 0507237650, 0507237654, 0507237658, 0507237664, 0507237666, 0507237672, 0507237675, 0507237677, 0507237680, 0507237681, 0507237685, 0507237693, 0507237697, 0507237707, 0507237708, 0507237709, 0507237710, 0507237712, 0507237716, 0507237718, 0507237720, 0507237724, 0507237730, 0507237739, 0507237742, 0507237743, 0507237744, 0507237745, 0507237749, 0507237762, 0507237763, 0507237767, 0507237772, 0507237773, 0507237775, 0507237779, 0507237780, 0507237782, 0507237783, 0507237785, 0507237789, 0507237791, 0507237792, 0507237794, 0507237796, 0507237800, 0507237809, 0507237813, 0507237815, 0507237817, 0507237818, 0507237821, 0507237822, 0507237827, 0507237828, 0507237829, 0507237836, 0507237842, 0507237844, 0507237850, 0507237851, 0507237852, 0507237853, 0507237855, 0507237856, 0507237859, 0507237861, 0507237862, 0507237866, 0507237867, 0507237869, 0507237871, 0507237873, 0507237877, 0507237879, 0507237881, 0507237882, 0507237885, 0507237889, 0507237894, 0507237899, 0507237900, 0507237901, 0507237902, 0507237907, 0507237911, 0507237913, 0507237915, 0507237919, 0507237920, 0507237921, 0507237926, 0507237928, 0507237931, 0507237932, 0507237934, 0507237936, 0507237940, 0507237942, 0507237943, 0507237944, 0507237946, 0507237949, 0507237954, 0507237960, 0507237962, 0507237965, 0507237966, 0507237971, 0507237994, 0507238002, 0508275018, 0607238010, 0607238013, 0607238024, 0607238026, 0607238027, 0607238029, 0607238030, 0607238031, 0607238043, 0607238045, 0607238047, 0607238049, 0607238063, 0607238065, 0607238074, 0607238078, 0607238080, 0607238086, 0607238088, 0607238090, 0607238091, 0607238092, 0607238098, 0607238099, 0607238100, 0607238101, 0607238102, 0607238104, 0607238248, 0607238250, 0607238252, 0607238264, 0607238321, 0607238337, 0607238339, 0607238367, 0607238381, 0607238403, 0607238407, 0607238409, 0607238411, 0607238415, 0607238419, 0607238425, 0607238433, 0607238655, 0507237962, 0507237965, 0507237966, 0507237971, 0507237994, 0507238002, 0607238010, 0607238013, 0607238024, 0607238026, 0607238027, 0607238029, 0607238030, 0607238031, 0607238043, 0607238045, 0607238047, 0607238049, 0607238063, 0607238065, 0607238074, 0607238078, 0607238080, 0607238086, 0607238088, 0607238090, 0607238091, 0607238092, 0607238098, 0607238099, 0607238100, 0607238101, 0607238102, 0607238104, 0607238248, 0607238250, 0607238252, 0607238264, 0607238321, 0607238337, 0607238339, 0607238367, 0607238381, 0607238403, 0607238407, 0607238409, 0607238411, 0607238415, 0607238419, 0607238425, 0607238433, 0607238655, 0607238685, 0607238688, 0607238692, 0607238694, 0607238696, 0607238705, 0607238997, 0607239199, 0607239308, 0607239310, 0607239312, 0607239314, 0607239316, 0607239318, 0607239320, 0607239322, 0607239324, 0607239326, 0607239328, 0607239330, 0607239334, 0607239336, 0607239338, 0607239340, 0607239342, 0607239387, 0607239389, 0607239391, 0607239393, 0607239395, 0607239397, 0607239399, 0607239401, 0607239403, 0607239405, 0607239408, 0607239410, 0607239416, 0607239418, 0607239420, 0607239422, 0607239426, 0607239428, 0607239430, 0607239432, 0607239434, 0607239436, 0607239440, 0607239442, 0607239448, 0607239452, 0607239454, 0607239500, 0607239504, 0607239506, 0607239510, 0607239514, 0607239516, 0607239518, 0607239520, 0607239522, 0607239524, 0607239526, 0607239528, 0607239530, 0607239534, 0607239536, 0607239538, 0607239540, 0607239542, 0607239544, 0607239546, 0607239673, 0607239679, 0607239681, 0607239683, 0607239685, 0607239689, 0607239691, 0607239693, 0607239697, 0607239699, 0607239701, 0607239703, 0607239705, 0607239706, 0607239708, 0607239710, 0607239712, 0607239714, 0607239716, 0607239718, 0607239720, 0607239722, 0607239724, 0607239726, 0607239730, 0607239732, 0607239734, 0607239736, 0607239738, 0607239740, 0607239742, 0607239746, 0607239748, 0607239750, 0607239752, 0607239754, 0607239758, 0607239760, 0607239762, 0607239764, 0607239766, 0607239768, 0607239770, 0607239772, 0607239774, 0607239776, 0607239778, 0607239892, 0607239894, 0607239898, 0607239900, 0607239901, 0607239902, 0607239904, 0607239907, 0607239908, 0607239914, 0607239939, 0607239944, 0607239946, 0607239948, 0607239949, 0607239950, 0607239951, 0607239955, 0607239966, 0607239998, 0607240015, 0607240025, 0607240027, 0607240031, 0607240035, 0607240041, 0607240053, 0607240055, 0607240101, 0607240118, 0607240120, 0607240122, 0607240132, 0607240143, 0607240145, 0607240163, 0607240173, 0607240175, 0607240177, 0607240179, 0607240183, 0607240185, 0607240187, 0607240189, 0607240191, 0607240193, 0607240199, 0607240203, 0607240207, 0607240209, 0607240211, 0607240213, 0607240215, 0607240217, 0607240221, 0607240251, 0607240255, 0607240261, 0607240263, 0607240265, 0607240267, 0607240270, 0607240271, 0607240273, 0607240274, 0607240275, 0607240276, 0607240277, 0607240278, 0607240279, 0607240280, 0607240281, 0607240282, 0607240283, 0607240284, 0607240285, 0607240286, 0607240288, 0607240289, 0607240290, 0607240291, 0607240292, 0607240293, 0607240295, 0607240296, 0607240297, 0607240298, 0607240300, 0607240301, 0607240302, 0607240303, 0607240304, 0607240305, 0607240308, 0607240314, 0607240324, 0607240326, 0607240336, 0607240338, 0607240339, 0607240341, 0607240342, 0607240343, 0607240344, 0607240345, 0607240347, 0607240348, 0607240349, 0607240351, 0607240352, 0607240353, 0607240354, 0607240355, 0607240357, 0607240358, 0607240362, 0607240363, 0607240364, 0607240366, 0607240368, 0607240369, 0607240371, 0607240373, 0607240375, 0607240377, 0607240379, 0607240381, 0607240383, 0607240385, 0607240386, 0607240387, 0607240397, 0607240399, 0607240401, 0607240403, 0607240405, 0607240406, 0607240407, 0607240408, 0607240409, 0607240410, 0607240411, 0607240412, 0607240413, 0607240414, 0607240415, 0607240416, 0607240417, 0607240418, 0607240420, 0607240421, 0607240422, 0607240423, 0607240424, 0607240425, 0607240427, 0607240428, 0607240429, 0607240430, 0607240431, 0607240432, 0607240433, 0607240434, 0607240435, 0607240436, 0607240437, 0607240438, 0607240439, 0607240440, 0607240453, 0607240455, 0607240457, 0607240466, 0607240476, 0607240478, 0607240482, 0607240484, 0607240486, 0607240488, 0607240490, 0607240492, 0607240494, 0607240496, 0607240498, 0607240500, 0607240504, 0607240506, 0607240507, 0607240508, 0607240509, 0607240510, 0607240511, 0607240512, 0607240513, 0607240514, 0607240516, 0607240517, 0607240518, 0607240519, 0607240520, 0607240523, 0607240524, 0607240525, 0607240526, 0607240527, 0607240528, 0607240529, 0607240530, 0607240532, 0607240533, 0607240535, 0607240537, 0607240538, 0607240539, 0607240540, 0607240541, 0607240542, 0607240543, 0607240544, 0607240545, 0607240546, 0607240547, 0607240548, 0607240549, 0607240551, 0607240552, 0607240553, 0607240554, 0607240555, 0607240556, 0607240557, 0607240558, 0607240559, 0607240560, 0607240561, 0607240562, 0607240563, 0607240564, 0607240565, 0607240566, 0607240567, 0607240568, 0607240569, 0607240570, 0607240571, 0607240572, 0607240573, 0607240574, 0607240575, 0607240576, 0607240577, 0607240578, 0607240579, 0607240580, 0607240582, 0607240583, 0607240584, 0607240585, 0607240587, 0607240588, 0607240589, 0607240590, 0607240591, 0607240592, 0607240593, 0607240595, 0607240596, 0607240597, 0607240598, 0607240600, 0607240601, 0607240602, 0607240603, 0607240604, 0607240605, 0607240609, 0607240611, 0607240613, 0607240615, 0607240617, 0607240619, 0607240621, 0607240623, 0607240625, 0607240629, 0607240633, 0607240635, 0607240639, 0607240641, 0607240643, 0607240645, 0607240647, 0607240649, 0607240653, 0607240655, 0607240657, 0607240661, 0607240663, 0607240665, 0607240669, 0607240671, 0607240673, 0607240677, 0607240689, 0607240691, 0607240693, 0607240695, 0607240697, 0607240699, 0607240701, 0607240703, 0607240705, 0607240772, 0607240791, 0607240793, 0607240795, 0607240805, 0607240806, 0607240808, 0607240812, 0607240814, 0607240815, 0607240816, 0607240818, 0607240819, 0607240820, 0607240821, 0607240822, 0607240824, 0607240826, 0607240828, 0607240830, 0607240832, 0607240834, 0607240836, 0607240838, 0607240842, 0607240846, 0607240850, 0607240851, 0607240852, 0607240854, 0607240856, 0607240858, 0607240859, 0607240860, 0607240868, 0607240883, 0607240885, 0607240887, 0607240889, 0607240891, 0607240892, 0607240893, 0607240894, 0607240895, 0607240906, 0607240908, 0607240914, 0607240916, 0607240918, 0607240920, 0607240922, 0607240954, 0607240964, 0607240969, 0607240971, 0607240972, 0607240973, 0607240975, 0607240976, 0607240977, 0607240979, 0607240980, 0607240981, 0607240983, 0607240985, 0607240987, 0607240989, 0607240991, 0607240993, 0607240995, 0607241003, 0607241031, 0607241033, 0607241042, 0607241044, 0607241062, 0607241138, 0607241140, 0607241144, 0607241455, 0607241465, 0607241593, 0607241624, 0607241700, 0607242137, 0608278916, 0608278920, 0608279889, 0608280706, 0707243297, 0707243343, 0707243379, 0707243397, 0707243405, 0707243406, 0707243408, 0707243410, 0707243414, 0707243416, 0707243424, 0707243426, 0707243428, 0707243434, 0707243435, 0707243440, 0707243441, 0707243443, 0707243447, 0707243448, 0707243450, 0707243452, 0707243454, 0707243458, 0707243462, 0707243466, 0707243468, 0707243476, 0707243477, 0707243478, 0707243481, 0707243482, 0707243484, 0707243486, 0707243487, 0707243488, 0707243489, 0707243490, 0707243491, 0707243492, 0707243495, 0707243499, 0707243501, 0707243503, 0707243505, 0707243508, 0707243510, 0707243512, 0707243520, 0707243522, 0707243528, 0707243530, 0707243532, 0707243534, 0707243536, 0707243540, 0707243543, 0707243544, 0707243545, 0707243546, 0707243547, 0707243549, 0707243553, 0707243554, 0707243555, 0707243557, 0707243559, 0707243561, 0707243563, 0707243567, 0707243569, 0707243575, 0707243583, 0707243591, 0707243593, 0707243597, 0707243599, 0707243601, 0707243606, 0707243607, 0707243611, 0707243612, 0707243613, 0707243614, 0707243615, 0707243616, 0707243619, 0707243620, 0707243622, 0707243624, 0707243626, 0707243628, 0707243630, 0707243634, 0707243637, 0707243638, 0707243639, 0707243643, 0707243644, 0707243645, 0707243649, 0707243651, 0707243653, 0707243655, 0707243657, 0707243661, 0707243667, 0707243727, 0707243731, 0707243741, 0707243807, 0707243814, 0707243815, 0707243816, 0707243817, 0707243818, 0707243821, 0707243824, 0707243828, 0707243829, 0707243831, 0707243835, 0707243837, 0707243842, 0707243843, 0707243844, 0707243845, 0707243846, 0707243847, 0707243849, 0707243850, 0707243851, 0707243852, 0707243853, 0707243857, 0707243858, 0707243859, 0707243861, 0707243862, 0707243863, 0707243866, 0707243868, 0707243870, 0707243871, 0707243872, 0707243873, 0707243874, 0707243875, 0707243876, 0707243878, 0707243880, 0707243882, 0707243884, 0707243888, 0707243894, 0707243900, 0707243902, 0707243903, 0707243904, 0707243906, 0707243908, 0707243909, 0707243911, 0707243912, 0707243914, 0707243916, 0707243917, 0707243918, 0707243920.0707243924, 0707243926, 0707243928, 0707243931, 0707243934, 0707243938, 0707243939, 0707243941, 0707243948, 0707243950, 0707243952, 0707243953, 0707243954, 0707243955, 0707243956, 0707243957, 0707243960, 0707243963, 0707243969, 0707243977, 0707243981, 0707243982, 0707243983, 0707243991, 0707243993, 0707243995, 0707243997, 0707243999, 0707244001, 0707244005, 0707244055, 0707244087, 0707244095, 0707244112, 0707244113, 0707244114, 0707244123, 0707244129, 0707244131, 0707244135, 0707244141, 0707244145, 0707244147, 0707244159, 0707244173, 0707244174, 0707244178, 0707244181, 0707244185, 0707244188, 0707244189, 0707244190, 0707244196, 0707244210, 0707244223, 0707244234, 0707244269, 0707244273, 0707244275, 0707244279, 0707244281, 0707244285, 0707244291, 0707244293, 0707244295, 0707244297, 0707244306, 0707244307, 0707244309, 0707244311, 0707244313, 0707244316, 0707244317, 0707244319, 0707244320, 0707244322, 0707244324, 0707244325, 0707244327, 0707244333, 0707244335, 0707244336, 0707244337, 0707244338, 0707244340, 0707244344, 0707244348, 0707244358, 0707244360, 0707244362, 0707244366, 0707244368, 0707244370, 0707244372, 0707244374, 0707244378, 0707244380, 0707244382, 0707244386, 0707244390, 0707244392, 0707244393, 0707244395, 0707244398, 0707244404, 0707244492, 0707244501, 0707244836, 0707244848, 0707244854, 0707244862, 0707244866, 0707244880, 0707244890, 0707244902, 0707245024, 0707245033, 0707245044, 0707245045, 0707245055, 0707245067, 0707245069, 0707245071, 0707245075, 0707245140, 0707245427, 0707245433, 0707245439, 0707245441, 0707245447, 0707245449, 0707245451, 0707245457, 0707245459, 0707245461, 0707245463, 0707245465, 0707245467, 0707245471, 0707245473, 0707245475, 0707245485, 0707245491, 0707245495, 0707245509, 0707245511, 0707245513, 0707245515, 0707245519, 0707245521, 0707245523, 0707245525, 0707245529, 0707245531, 0707245533, 0707245535, 0707245541, 0707245543, 0707245545, 0707245547, 0707245553, 0707245555, 0707245557, 0707245559, 0707245563, 0707245569, 0707245575, 0707245602, 0707245604, 0707245606, 0707245608, 0707245616, 0707245618, 0707245638, 0707245644, 0707245650, 0707245807, 0707245808, 0707245809, 0707245810, 0707245811, 0707245813, 0707245814, 0707245815, 0707245817, 0707245820, 0707245821, 0707245826, 0707245830, 0707245834, 0707245838, 0707245840, 0707245842, 0707245844, 0707245846, 0707245848, 0707245850, 0707245852, 0707245856, 0707245858, 0707245860, 0707245862, 0707245864, 0707245865, 0707245866, 0707245868, 0707245872, 0707245874, 0707245876, 0707245878, 0707245879, 0707245880, 0707245882, 0707245886, 0707245888, 0707245896, 0707245898, 0707245900, 0707245902, 0707246040, 0707246062, 0707246064, 0707246068, 0707246072, 0707246076, 0707246082, 0707246084, 0707246088, 0707246092, 0707246098, 0707246100, 0708282150, 0708282152, 0708282154, 0708282200, 0708282231, 0807246564, 0807246572, 0807246574, 0807246586, 0807246635, 0807246695, 0807246718, 0807246728, 0807246730, 0807246732, 0807246734, 0807246738, 0807247027, 0807247029, 0807247769, 0807248021, 0807248027, 0807248029, 0807248031, 0807248033, 0807248035, 0807248039, 0807248040, 0807248043, 0807248044, 0807248045, 0807248046, 0807248047, 0807248048, 0807248049, 0807248051, 0807248052, 0807248054, 0807248056, 0807248058, 0807248060, 0807248062, 0807248064, 0807248066, 0807248067, 0807248068, 0807248069, 0807248070, 0807248074, 0807248079, 0807248080, 0807248081, 0807248083, 0807248085, 0807248087, 0807248088, 0807248089, 0807248093, 0807248094, 0807248095, 0807248096, 0807248098, 0807248107, 0807248115, 0807248117, 0807248180, 0807248182, 0807248206, 0807248210, 0807248212, 0807248218, 0807248230, 0807248234, 0807248238, 0807248240, 0807248246, 0807248249, 0807248250, 0807248251, 0807248253, 0807248256, 0807248257, 0807248258, 0807248260, 0807248264, 0807248266, 0807248268, 0807248270, 0807248272, 0807248274, 0807248276, 0807248279, 0807248280, 0807248281, 0807248282, 0807248284, 0807248286, 0807248296, 0807248301, 0807248311, 0807248313, 0807248321, 0807248331, 0807248349, 0807248363, 0807248377, 0807248383, 0807248385, 0807248389, 0807248397, 0807248414, 0807248416, 0807248426, 0807248432, 0807248442, 0807248444, 0807248454, 0807248466, 0807248470, 0807248486, 0807248498, 0807248500, 0807248501, 0807248502, 0807248503, 0807248504, 0807248505, 0807248506, 0807248507, 0807248510, 0807248511, 0807248513, 0807248515, 0807248516, 0807248520, 0807248521, 0807248522, 0807248523, 0807248527, 0807248529, 0807248532, 0807248543, 0807248545, 0807248547, 0807248549, 0807248551, 0807248553, 0807248555, 0807248557, 0807248560, 0807248561, 0807248562, 0807248563, 0807248565, 0807248566, 0807248567, 0807248568, 0807248569, 0807248570, 0807248572, 0807248573, 0807248575, 0807248576, 0807248577, 0807248581, 0807248585, 0807248587, 0807248590, 0807248591, 0807248592, 0807248593, 0807248594, 0807248596, 0807248597, 0807248598, 0807248599, 0807248600, 0807248601, 0807248602, 0807248604, 0807248605, 0807248606, 0807248607, 0807248608, 0807248609, 0807248610, 0807248611, 0807248612, 0807248613, 0807248614, 0807248615, 0807248616, 0807248617, 0807248618, 0807248619, 0807248620, 0807248621, 0807248623, 0807248624, 0807248627, 0807248629, 0807248631, 0807248633, 0807248635, 0807248639, 0807248640, 0807248641, 0807248652, 0807248654, 0807248656, 0807248660, 0807248662, 0807248664, 0807248665, 0807248666, 0807248670, 0807248671, 0807248673, 0807248677, 0807248678, 0807248680, 0807248682, 0807248686, 0807248687, 0807248689, 0807248690, 0807248692, 0807248694, 0807248696, 0807248697, 0807248698, 0807248700, 0807248701, 0807248703, 0807248704, 0807248706, 0807248707, 0807248708, 0807248709, 0807248710, 0807248712, 0807248714, 0807248715, 0807248716, 0807248717, 0807248719, 0807248720, 0807248722, 0807248724, 0807248725, 0807248726, 0807248728, 0807248729, 0807248730, 0807248731, 0807248733, 0807248737, 0807248738, 0807248740, 0807248742, 0807248743, 0807248745, 0807248746, 0807248748, 0807248751, 0807248752, 0807248753, 0807248757, 0807248759, 0807248760, 0807248761, 0807248762, 0807248763, 0807248765, 0807248769, 0807248770, 0807248771, 0807248773, 0807248774, 0807248776, 0807248777, 0807248778, 0807248779, 0807248780, 0807248781, 0807248782, 0807248783, 0807248784, 0807248785, 0807248786, 0807248787, 0807248788, 0807248789, 0807248791, 0807248792, 0807248793, 0807248794, 0807248795, 0807248797, 0807248799, 0807248800, 0807248801, 0807248803, 0807248805, 0807248807, 0807248808, 0807248809, 0807248810, 0807248811, 0807248812, 0807248813, 0807248814, 0807248816, 0807248819, 0807248828, 0807248832, 0807248836, 0807248837, 0807248838, 0807248849, 0807248859, 0807248861, 0807248863, 0807248864, 0807248865, 0807248867, 0807248869, 0807248870, 0807248871, 0807248873, 0807248874, 0807248875, 0807248877, 0807248879, 0807248881, 0807248883, 0807248885, 0807248889, 0807248893, 0807248899, 0807248934, 0807248952, 0807248961, 0807248979, 0807249075, 0807249103, 0807249105, 0807249111, 0807249114, 0807249119, 0807249121, 0807249123, 0807249127, 0807249129, 0807249131, 0807249135, 0807249136, 0807249138, 0807249139, 0807249143, 0807249147, 0807249148, 0807249151, 0807249153, 0807249157, 0807249161, 0807249162, 0807249163, 0807249167, 0807249170, 0807249179, 0807249187, 0807249189, 0807249198, 0807249294, 0807249319, 0807249501, 0807249503, 0807249504, 0807249505, 0807249507, 0807249509, 0807249511, 0807249513, 0807249515, 0807249519, 0807249521, 0807249523, 0807249525, 0807249527, 0807249529, 0807249531, 0807249533, 0807249535, 0807249537, 0807249539, 0807249543, 0807249545, 0807249547, 0807249549, 0807249553, 0807249555, 0807249557, 0807249559, 0807249561, 0807249563, 0807249565, 0807249567, 0807249569, 0807249571, 0807249573, 0807249575, 0807249577, 0807249579, 0807249581, 0807249583, 0807249585, 0807249586, 0807249587, 0807249588, 0807249589, 0807249590, 0807249591, 0807249592, 0807249593, 0807249595, 0807249596, 0807249597, 0807249598, 0807249599, 0807249600, 0807249602, 0807249604, 0807249606, 0807249608, 0807249610, 0807249612, 0807249613, 0807249618, 0807249650, 0807249653, 0807249661, 0807249662, 0807249663, 0807249664, 0807249665, 0807249666, 0807249667, 0807249668, 0807249669, 0807249670, 0807249671, 0807249673, 0807249675, 0807249678, 0807249680, 0807249682, 0807249683, 0807249684, 0807249686, 0807249687, 0807249690, 0807249691, 0807249693, 0807249695, 0807249696, 0807249697, 0807249698, 0807249700, 0807249702, 0807249704, 0807249706, 0807249707, 0807249708, 0807249709, 0807249710, 0807249711, 0807249712, 0807249713, 0807249714, 0807249715, 0807249716, 0807249717, 0807249718, 0807249719, 0807249720, 0807249721, 0807249722, 0807249723, 0807249724, 0807249725, 0807249726, 0807249727, 0807249728, 0807249729, 0807249730, 0807249731, 0807249732, 0807249733, 0807249734, 0807249735, 0807249736, 0807249737, 0807249738, 0807249739, 0807249740, 0807249741, 0807249742, 0807249743, 0807249744, 0807249746, 0807249747, 0807249748, 0807249750, 0807249751, 0807249753, 0807249754, 0807249755, 0807249756, 0807249757, 0807249758, 0807249759, 0807249761, 0807249762, 0807249763, 0807249764, 0807249766, 0807249767, 0807249768, 0807249770, 0807249772, 0807249773, 0807249774, 0807249775, 0807249777, 0807249778, 0807249779, 0807249780, 0807249781, 0807249782, 0807249783, 0807249784, 0807249785, 0807249786, 0807249787, 0807249788, 0807249789, 0807249790, 0807249791, 0807249792, 0807249793, 0807249794, 0807249795, 0807249796, 0807249797, 0807249798, 0807249799, 0807249800, 0807249801, 0807249803, 0807249806, 0807249809, 0807249810, 0807249812, 0807249813, 0807249814, 0807249817, 0807249819, 0807249821, 0807249823, 0807249824, 0807249825, 0807249826, 0807249827, 0807249829, 0807249831, 0807249833, 0807249835, 0807249837, 0807249838, 0807249839, 0807249840, 0807249841, 0807249842, 0807249843, 0807249845, 0807249846, 0807249847, 0807249849, 0807249850, 0807249851, 0807249852, 0807249853, 0807249854, 0807249855, 0807249856, 0807249857, 0807249858, 0807249859, 0807249860, 0807249861, 0807249862, 0807249863, 0807249864, 0807249865, 0807249867, 0807249869, 0807249871, 0807249873, 0807249875, 0807249880, 0807249883, 0807249885, 0807249887, 0807249889, 0807249891, 0807249893, 0807249895, 0807249897, 0807249899, 0807249902, 0807249909, 0807249910, 0807249914, 0807249916, 0807249918, 0807249920, 0807249922, 0807249929, 0807249930, 0807249932, 0807249934, 0807249936, 0807249938, 0807249940, 0807249942, 0807249943, 0807249944, 0807249946, 0807249947, 0807249948, 0807249949, 0807249950, 0807249951, 0807249952, 0807249953, 0807249954, 0807249955, 0807249956, 0807249960, 0807249962, 0807249963, 0807249964, 0807249965, 0807249966, 0807249967, 0807249968, 0807249969, 0807249970, 0807249971, 0807249972, 0807249976, 0807249978, 0807249980, 0807249982, 0807249984, 0807249986, 0807249987, 0807249988, 0807249990, 0807249991, 0807249992, 0807249994, 0807249996, 0807249997, 0807249998, 0807250000, 0807250001, 0807250002, 0807250003, 0807250004, 0807250005, 0807250006, 0807250007, 0807250008, 0807250009, 0807250010, 0807250011, 0807250012, 0807250013, 0807250014, 0807250015, 0807250016, 0807250017, 0807250018, 0807250020, 0807250021, 0807250022, 0807250023, 0807250025, 0807250026, 0807250027, 0807250028, 0807250029, 0807250030, 0807250031, 0807250032, 0807250033, 0807250034, 0807250035, 0807250036, 0807250037, 0807250038, 0807250039, 0807250040, 0807250041, 0807250043, 0807250044, 0807250045, 0807250046, 0807250047, 0807250048, 0807250049, 0807250051, 0807250052, 0807250054, 0807250055, 0807250056, 0807250057, 0807250058, 0807250060, 0807250061, 0807250062, 0807250063, 0807250064, 0807250065, 0807250066, 0807250067, 0807250068, 0807250069, 0807250070, 0807250071, 0807250072, 0807250073, 0807250074, 0807250075, 0807250076, 0807250078, 0807250080, 0807250081, 0807250082, 0807250083, 0807250084, 0807250085, 0807250086, 0807250087, 0807250088, 0807250089, 0807250090, 0807250091, 0807250092, 0807250093, 0807250094, 0807250095, 0807250096, 0807250098, 0807250099, 0807250100, 0807250102.0807250104, 0807250105, 0807250106, 0807250107, 0807250108, 0807250110, 0807250111, 0807250112, 0807250114, 0807250118, 0807250119, 0807250120, 0807250121, 0807250122, 0807250123, 0807250125, 0807250126, 0807250128, 0807250130, 0807250132, 0807250134, 0807250136, 0807250138, 0807250139, 0807250140, 0807250141, 0807250142, 0807250146, 0807250148, 0807250152, 0807250154, 0807250158, 0807250159, 0807250160, 0807250162, 0807250165, 0807250167, 0807250169, 0807250171, 0807250175, 0807250180, 0807250181, 0807250183, 0807250184, 0807250185, 0807250186, 0807250188, 0807250189, 0807250191, 0807250193, 0807250194, 0807250195, 0807250197, 0807250199, 0807250201, 0807250202, 0807250203, 0807250204, 0807250205, 0807250206, 0807250208, 0807250210, 0807250213, 0807250214, 0807250215, 0807250217, 0807250220, 0807250221, 0807250222, 0807250228, 0807250230, 0807250232, 0807250233, 0807250234, 0807250237, 0807250238, 0807250239, 0807250241, 0807250242, 0807250247, 0807250250, 0807250251, 0807250253, 0807250254, 0807250255, 0807250257, 0807250259, 0807250261, 0807250263, 0807250264, 0807250265, 0807250266, 0807250267, 0807250268, 0807250269, 0807250270, 0807250271, 0807250272, 0807250273, 0807250274, 0807250275, 0807250277, 0807250281, 0807250283, 0807250284, 0807250285, 0807250287, 0807250289, 0807250291, 0807250293, 0807250295, 0807250297, 0807250298, 0807250299, 0807250300, 0807250301, 0807250303, 0807250304, 0807250305, 0807250306, 0807250307, 0807250309, 0807250310, 0807250311, 0807250312, 0807250313, 0807250314, 0807250315, 0807250317, 0807250319, 0807250321, 0807250323, 0807250324, 0807250325, 0807250327, 0807250329, 0807250330, 0807250331, 0807250332, 0807250333, 0807250334, 0807250335, 0807250336, 0807250337, 0807250338, 0807250339, 0807250340, 0807250341, 0807250343, 0807250345, 0807250347, 0807250348, 0807250350, 0807250351, 0807250352, 0807250353, 0807250355, 0807250356, 0807250357, 0807250358, 0807250359, 0807250360, 0807250361, 0807250363, 0807250364, 0807250365, 0807250368, 0807250369, 0807250370, 0807250371, 0807250372, 0807250373, 0807250374, 0807250375, 0807250376, 0807250379, 0807250380, 0807250381, 0807250382, 0807250383, 0807250384, 0807250386, 0807250391, 0807250393, 0807250394, 0807250396, 0807250399, 0807250404, 0807250406, 0807250408, 0807250410, 0807250412, 0807250414, 0807250416, 0807250418, 0807250420, 0807250422, 0807250424, 0807250426, 0807250428, 0807250430, 0807250434, 0807250436, 0807250438, 0807250440, 0807250442, 0807250444, 0807250448, 0807250450, 0807250452, 0807250454, 0807250501, 0807250502, 0807250503, 0807250504, 0807250505, 0807250506, 0807250507, 0807250508, 0807250509, 0807250510, 0807250511, 0807250512, 0807250513, 0807250514, 0807250516, 0807250517, 0807250518, 0807250519, 0807250520, 0807250521, 0807250522, 0807250523, 0807250524, 0807250525, 0807250526, 0807250527, 0807250529, 0807250530, 0807250531, 0807250532, 0807250533, 0807250535, 0807250537, 0807250539, 0807250541, 0807250542, 0807250543, 0807250544, 0807250545, 0807250547, 0807250548, 0807250549, 0807250550, 0807250551, 0807250552, 0807250553, 0807250554, 0807250555, 0807250556, 0807250557, 0807250559, 0807250561, 0807250563, 0807250565, 0807250566, 0807250567, 0807250568, 0807250569, 0807250570, 0807250571, 0807250573, 0807250575, 0807250577, 0807250578, 0807250579, 0807250581, 0807250583, 0807250584, 0807250585, 0807250586, 0807250587, 0807250589, 0807250591, 0807250592, 0807250593, 0807250595, 0807250596, 0807250597, 0807250598, 0807250599, 0807250600, 0807250602, 0807250606, 0807250608, 0807250612, 0807250614, 0807250616, 0807250618, 0807250620, 0807250621, 0807250622, 0807250623, 0807250624, 0807250627, 0807250631, 0807250633, 0807250635, 0807250639, 0807250641, 0807250643, 0807250645, 0807250649, 0807250653, 0807250655, 0807250657, 0807250661, 0807250665, 0807250667, 0807250669, 0807250671, 0807250675, 0807250677, 0807250679, 0807250681, 0807250683, 0807250685, 0807250687, 0807250689, 0807250691, 0807250693, 0807250695, 0807250697, 0807250699, 0807250766, 0807250768, 0807250770, 0807250772, 0807250774, 0807250776, 0807250778, 0807250800, 0807250801, 0807250803, 0807250805, 0807250809, 0807250813, 0807250815, 0807250817, 0807250842, 0807250903, 0807250905, 0807250909, 0807250911, 0807250919, 0807250921, 0807250925, 0807250929, 0807250931, 0807250935, 0807250937, 0807250939, 0807250949, 0807250951, 0807250953, 0807250959, 0807250963, 0807250965, 0807250967, 0807250969, 0807250971, 0807250973, 0807250975, 0807250977, 0807251041, 0807251043, 0807251045, 0807251047, 0807251049, 0807251055, 0807251067, 0807251071, 0807251073, 0807251075, 0807251077, 0807251079, 0807251081, 0807251083, 0807251085, 0807251087, 0807251089, 0807251091, 0807251093, 0807251095, 0907251307, 0907251309, 0907251311, 0907251313, 0907251315, 0907251317, 0907251319, 0907251321, 0907251327, 0907251329, 0907252089, 0907252516, 0907252520, 0907252523, 0907252529, 0907252531, 0907252540, 0907252542, 0907252544, 0907252545, 0907252548, 0907252549, 0907252551, 0907252554, 0907252557, 0907252558, 0907252565, 0907252571, 0907252573, 0907252575, 0907252583, 0907252585, 0907252599, 0907252601, 0907252603, 0907252605, 0907252607, 0907252611, 0907252613, 0907252615, 0907252623, 0907252634, 0907252643, 0907252647, 0907252657, 0907252661, 0907252667, 0907252669, 0907252673, 0907252677, 0907252681, 0907252687, 0907252695, 0907252701, 0907252707, 0907252717, 0907252723, 0907252727, 0907252746, 0907252748, 0907252784, 0907252788, 0907252790, 0907252792, 0907252796, 0907252798, 0907252802, 0907252808, 0907252843, 0907252845, 0907252847, 0907252849, 0907252856, 0907252860, 0907252868, 0907252878, 0907252920, 0907252924, 0907252926, 0907252938, 0907252942, 0907253048, 0907253056, 0907253066, 0907253088, 0907253090, 0907253096, 0907253100, 0907253123, 0907253125, 0907253148, 0907253215, 0907253221, 0907253223, 0907253225, 0907253227, 0907253229, 0907253233, 0907253235, 0907253284, 0907253290, 0907253510, 0907253562, 0907253655, 0907254043, 0907254107, 0907254199, 0907254255, 0907254277, 0907254286, 0907254292, 0907254324, 0907254351, 0907254508, 0907254516, 0907254536, 0907254550, 0907254643, 0907254645, 0907254744, 0907254745, 0907254763, 0907254772, 0907254778, 0907254780, 0907254817, 0907254842, 0907254848, 0907254852, 0907254900, 0907254910, 0907254918, 0907254920, 0907254926, 0907254928, 0907254930, 0907254969, 0907254986, 0907255083, 0907255097, 0907255099, 0907255106, 0907255117, 0907255126, 0907255167, 1007255204, 1007255206, 1007255208, 1007255212, 1007255214, 1007255216, 1007255220, 1007255222, 1007255224, 1007255235, 1007255239, 1007255245, 1007255247, 1007255251, 1007255253, 1007255255, 1007255257, 1007255259, 1007255261, 1007255263, 1007255265, 1007255269, 1007255273, 1007255275, 1007255277, 1007255279, 1007255281, 1007255283, 1007255285, 1007255287, 1007255289, 1007255291, 1007255293, 1007255295, 1007255297, 1007255299, 1007255302, 1007255304, 1007255306, 1007255308, 1007255309, 1007255310, 1007255311, 1007255312, 1007255314, 1007255315, 1007255316, 1007255318, 1007255320, 1007255321, 1007255322, 1007255324, 1007255325, 1007255326, 1007255327, 1007255328, 1007255329, 1007255333, 1007255335, 1007255336, 1007255337, 1007255338, 1007255339, 1007255343, 1007255345, 1007255347, 1007255349, 1007255351, 1007255353, 1007255355, 1007255357, 1007255359, 1007255361, 1007255363, 1007255365, 1007255367, 1007255369, 1007255373, 1007255377, 1007255379, 1007255381, 1007255383, 1007255385, 1007255387, 1007255389, 1007255391, 1007255393, 1007255395, 1007255397, 1007255399, 1007255458, 1007255460, 1007255494, 1007255496, 1007255498, 1007255502, 1007255513, 1007255515, 1007255519, 1007255521, 1007255523, 1007255525, 1007255527, 1007255529, 1007255531, 1007255533, 1007255535, 1007255539, 1007255545, 1007255547, 1007255549, 1007255551, 1007255553, 1007255555, 1007255557, 1007255559, 1007255561, 1007255563, 1007255565, 1007255567, 1007255569, 1007255571, 1007255573, 1007255575, 1007255577, 1007255579, 1007255581, 1007255583, 1007255585, 1007255587, 1007255589, 1007255591, 1007255595, 1007255597, 1007255599, 1007255703, 1007255712, 1007255714, 1007255716, 1007255717, 1007255718, 1007255719, 1007255720, 1007255721, 1007255723, 1007255975, 1007256263, 1007256448, 1007256657, 1007256659, 1007256661, 1007256665, 1007256667, 1007256687, 1007256702, 1007256757, 1007256759, 1007256761, 1007256775, 1007256779, 1007256783, 1007256785, 1007256791, 1007256797, 1007256934, 1007257008, 1007257010, 1007257018, 1007257344, 1007257346, 1007257348, 1007257356, 1007257358, AND 1007257360.======================================= LAERDAL M3840A AND M3841A:  0607240946, 0807246508, 0807249629, 0807249657, 0807249705, 0807249939, 0807249989, 0108267713, 0607240924, 0907253243, 0108267745, 0108267749, 0507237315, 0507237337, 0507237351, 0507237458, 0507237463, 0507237472, 0508275051, 0508275053, 0508275055, 0508275061, 0508275063, 0508275065, 0508275067, 0508275069, 0508275071, 0508275099, 0607240057, 0607240063, 0607240311, 0607240313, 0607240388, 0607240390, 0607240392, 0607240398, 0607240400, 0607240404, 0607240515, 0607240685, 0607240807, 0607240817, 0607240862, 0607240866, 0607240870, 0607240872, 0607240874, 0607240884, 0607240888, 0607240890, 0607240896, 0607240902, 0607240912, 0607240997, 0607240999, 0607241001, 0807248822, 0807248824, 0807248858, 0807248860, 0807248884, 0807248886, 0807248888, 0807248892, 0807248950, 0807248973, 0807248977, 0807248981, 0807248985, 0807248995, 0807248999, 0807249001, 0807249003, 0807249005, 0807249009, 0807249011, 0807249021, 0807249104, 0807249117, 0807249120, 0807249124, 0807249176, 0807249202, 0807249203, 0807249214, 0807249216, 0807249218, 0807249220, 0807249224, 0807249228, 0807249232, 0807249236, 0807249240, 0807249242, 0807249244, 0807249246, 0807249261, 0807249271, 0807249277, 0807249281, 0807249286, 0807249290, 0807249300, 0807249301, 0807249327, 0807250564, 0807250576, 0807250580, 0807250582, 0807250588, 0807250594, 0807250604, 0807250610, 0807250625, 0807250629, 0807250659, 0807250663, 0807250901, 0807250923, 0807250961, 0507236781, 0507236783, 0507237360, 0507237408, 0507237501, 0507237505, 0807248032, 0807248152, 0807248329, 0807248333, 0807248335, 0807248353, 0807248365, 0807248371, 0807248375, 0807248395, 0807248820, 0807248850, 0807248852, 0807248854, 0807248868, 0807248876, 0807248878, 0807248925, 0807248931, 0807248941, 0807248949, 0807248965, 0807248987, 0807249102, 0807249108, 0807249140, 0807249164, 0807249166, 0807249168, 0807249172, 0807249180, 0807249188, 0807249193, 0807249199, 0807249215, 0807249235, 0807249251, 0807249253, 0807249259, 0409300614, 0507237356, 0507237473, 0507237475, 0507237485, 0507237491, 0507237495, 0507237523, 0607240071, 0607241142, 0607241148, 0607241463, 0707243385, 0707243825, 0807248082, 0807248086, 0807248090, 0807248092, 0807248105, 0807248113, 0108267755, 0507237317, 0507237320, 0507237324, 0507237329, 0507237335, 0507237338, 0507237340, 0507237343, 0507237345, 0507237347, 0507237349, 0507237395, 0507237403, 0507237477, 0507237478, 0507237489, 0607238324, 0607239962, 0607239964, 0607240095, 0607240097, 0607240108, 0607240110, 0607240219, 0607240223, 0607240257, 0607240266, 0607240307, 0607240335, 0607240391, 0607240393, 0607240395, 0607240534, 0607240536, 0607241023, 0607241025, 0607241027, 0607241029, 0607241134, 0607241136, 0607241146, 0707243401, 0707243437, 0707243459, 0707243465, 0707243823, 0707243833, 0707243889, 0707243946, 0707243964, 0707243998, 0707244003, 0707244177, 0807248073, 0807248091, 0807248101, 0807248158, 0807248174, 0807248178, 0807248202, 0807248208, 0807248242, 0807248341, 0807248343, 0807248450, 0807248508, 0807248534, 0807248536, 0807248541, 0807248571, 0807248579, 0807248688, 0807248705, 0807248711, 0807248713, 0807248723, 0807248727, 0807248741, 0807248747, 0807248749, 0807248755, 0907253247, 0907253253, 0907253261, 1007257077, 0807250941, 0308273587, 0308273589, 0308273591, 0308273595, 0507237328, 0507237344, 0507237413, 0507237535, 0507237609, 0507237670, 0507237703, 0507237735, 0507237781, 0507237798, 0507237832, 0507237837, 0507237838, 0507237854, 0507237860, 0507237870, 0507237891, 0508275012, 0508275014, 0508275016, 0508275020, 0508275024, 0508275026, 0508275028, 0508275030, 0508275032, 0508275036, 0508275049, 0508275057, 0508275059, 0508275073, 0508275075, 0508275077, 0508275079, 0508275081, 0508275083, 0508275085, 0508275087, 0508275089, 0508275091, 0508275093, 0508275095, 0508275097, 0508275101, 0508275103, 0508275105, 0607238011, 0607238020, 0607238032, 0607238238, 0607238260, 0607238326, 0607238328, 0607238647, 0607238657, 0607238661, 0607238675, 0607238679, 0607238687, 0607238691, 0607238693, 0607238699, 0607239001, 0607239187, 0607239960, 0607240009, 0607240011, 0607240017, 0607240019, 0607240141, 0607240153, 0607240155, 0607240159, 0607240161, 0607240165, 0607240169, 0607240181, 0607240310, 0607240340, 0607240813, 0607240825, 0607240827, 0607240829, 0607240831, 0607240833, 0607240835, 0607240843, 0607240855, 0607240857, 0607240861, 0607240881, 0607240898, 0607240900, 0607240926, 0607240930, 0607240932, 0607240936, 0607240938, 0607240940, 0607240952, 0607240956, 0607240958, 0607240960, 0607240962, 0607240966, 0607240968, 0607240970, 0707244715, 0707244824, 0707244830, 0707244832, 0707244870, 0707244894, 0707244904, 0807249941, 0807249945, 0907252534, 0507237736, 0807248674, 0807248906, 0807248943, 0807249015, 0807249085, 0807249087, 0807249095, 0807249625, 0807249631, 0807249644, 0807249654, 0807249656, 0807249660, 0807249677, 0807249679, 0807249694, 0807249703, 0807249769, 0807249771, 0807249906, 0807249908, 0807249927, 0807249958, 0807249959, 0807249961, 0807249975, 0807249983, 0807249985, 1207267284, 0507237741, 0507237777, 0507237841, 0507237883, 0607238242, 0607238244, 0707244856, 0707244878, 0508275047, 0607240061, 0607240394, 0607240396, 0607241032, 0607241038, 0707243381, 0707243383, 0707243422, 0707243432, 0707243433, 0707243460, 0707243467, 0707243855, 0707243865, 0707243877, 0707243879, 0707243881, 0707243885, 0707243890, 0707243896, 0707244864, 0807250528, 0807250538, 0807250540, 0807250572, 0807250574, 0807250637, 0807250848, 0807250907, 0907254015, 0907254752, 0907254764, 0907254768, 0907254770, 0907254776, 0907254784, 0907254786, 0907254790, 0907254792, 0907254796, 0907254798, 0907254800, 0907254834, 0907254882, 0907254890, 0907254894, 0907254898, 0907254902, 0907254904, 0907254932, 0907254934, 0907254936, 0907254938, 0907254940, 0907254942, 0907254944, 0907254946, 0907254948, 0907254950, 0907254952, 0907254954, 0907254956, 0907254958, 0907254962, 0907254972, 0907254974, 0907254976, 0907254978, 0907254980, 0907254982, 0907254984, 0907254988, 0907254990, 0907254992, 0907254994, 0907254996, 0907254998, 0907255000, 0907255001, 0907255005, 0907255009, 0907255011, 0907255017, 0907255026, 0907255039, 0907255047, 0907255049, 0907255051, 0907255053, 0907255055, 0907255059, 0907255063, 0907255079, 0907255093, 0907255101, 0907255103, 0907255104, 0907255105, 0907255107, 0907255109, 0907255111, 0907255112, 0907255114, 0907255115, 0907255119, 0907255123, 0907255124, 0907255125, 0907255128, 0907255129, 0907255130, 0907255131, 0907255132, 0907255133, 0907255134, 0907255161, 0108269537, 0507237595, 0507237603, 0507237620, 0507237687, 0507237691, 0507237702, 0507237747, 0507237751, 0507237793, 0507237795, 0507237811, 0507237835, 0507237843, 0507237845, 0507237847, 0507237884, 0507237886, 0507237922, 0507237924, 0507237933, 0507237938, 0507237948, 0507238000, 0507238005, 0607238012, 0607238014, 0607238017, 0607238018, 0607238019, 0607238021, 0607238035, 0607238046, 0607238059, 0607238060, 0607238070, 0607238083, 0607238094, 0607238096, 0607238427, 0607239344, 0607239346, 0607239406, 0607239414, 0607239508, 0607242373, 0607242375, 0707243436, 0707243438, 0707243442, 0707243813, 0707243856, 0707243864, 0707243898, 0807247353, 0807247355, 0807248393, 0807248898, 0807248903, 0807248904, 0807248909, 0807248910, 0807248915, 0807248918, 0807248928, 0807248929, 0807248936, 0807248939, 0807248942, 0807248946, 0807248951, 0807248953, 0807248954, 0807248955, 0807248956, 0807248957, 0807248959, 0807248960, 0807248963, 0807248966, 0807248970, 0807248976, 0807248983, 0807248991, 0807248993, 0807249007, 0807249178, 0807249182, 0807249184, 0807249186, 0807249190, 0807249208, 0807249209, 0807249210, 0807249211, 0807249212, 0807249213, 0807249219, 0807249221, 0807249222, 0807249223, 0807249229, 0807249230, 0807249233, 0807249234, 0807249238, 0807249248, 0807249250, 0807249252, 0807249254, 0807249256, 0807249260, 0807249263, 0807249264, 0807249265, 0807249266, 0807249268, 0807249269, 0807249274, 0807249275, 0807249279, 0807249283, 0807249285, 0807249287, 0807249288, 0807249289, 0807249291, 0807249292, 0807249293, 0807249295, 0807249297, 0807249298, 0807249299, 0807250840, 0907253246, 0907253248, 0907253250, 0907253252, 0907253254, 0907253256, 0907253258, 0907253260, 0907253266, 0907253268, 0907253270, 0907253274, 0907253276, 0907253282, 0907253288, 0907253292, 0907253294, 0907253296, 0907253298, 0907253300, 0907253302, 0907253304, 0907253306, 0907253310, 0907253312, 0907253314, 0907253316, 0907253318, 0907253320, 0907253322, 0907253324, 0907253326, 0907253328, 0907253330, 0907253331, 0907253332, 0907253333, 0907253334, 0907253335, 0907253336, 0907253338, 0907253339, 0907253340, 0907253341, 0907253342, 0907253388, 0907253390, 0907253392, 0907253396, 0907253405, 0907253415, 0907253417, 0907253419, 0907253423, 0907253433, 0907253439, 0907253443, 0907253445, 0907253459, 0907253461, 0907253465, 0907253467, 0907253469, 0907253471, 0907253473, 0907253479, 0907253481, 0907253483, 0907253501, 0907253505, 0907253509, 0907253511, 0907253517, 0907253521, 0907253523, 0907253525, 0907253543, 0907253547, 0907253555, 0907253597, 0907253599, 0907254013, 0607239913, 0607240402, 0807246506, 0907253217, 0907253277, 1007257075, 0308273599, 0507237326, 0507237366, 0507237499, 0507237950, 0508275022, 0607238649, 0607238682, 0607239903, 0607239905, 0607240023, 0607240033, 0607240147, 0607240149, 0607240157, 0607240195, 0607240197, 0607240201, 0607240350, 0807248042, 0807248050, 0807248305, 0807249622, 0807249993, 0908288146, 0908288152, 0607240928, 0607240944, 0108267577, 0108267709, 0108267739, 0108267743, 0108267753, 0507237195, 0507237197, 0507237199, 0507237330, 0507237358, 0507237373, 0507237375, 0507237433, 0507237444, 0507237450, 0507237481, 0507237493, 0607240059, 0607240065, 0607240067, 0607240069, 0607240099, 0607240136, 0607240138, 0607240309, 0607240329, 0607240333, 0607240337, 0607240359, 0607240361, 0607240367, 0607240382, 0607240384, 0607240659, 0607240667, 0608278912, 0807246502, 0807246621, 0807246639, 0807246641, 0807246643, 0807248685, 0807248744, 0807248772, 0807248818, 0807248834, 0807248847, 0807248872, 0807248880, 0807248882, 0807248890, 0807248919, 0807249125, 0807249146, 0807249152, 0807249154, 0807249156, 0807249174, 1207267276, 1207267280, 1207267282, 1207267286, 1207267290, 1207267292, 1207267294, 1207267296, 1207267298, 1207267409, 1207267411, 1207267415, 1207267419, 1207267441, 1207267449, 1207267453, 1207267457, 1207267463, 1207267465, 1207267471, 0807248927, 0807249701, 0807249923, 0907253244, 0907253255, 0907253286, 0907253403, 0807249267, AND 0807249273.",
      "firm_fei_number": "1000524572",
      "recalling_firm": "Philips Medical Systems",
      "address_1": "2301 5th Ave Ste 200 2301 5th Ave",
      "address_2": "Ste 200",
      "city": "Seattle",
      "state": "WA",
      "postal_code": "98121",
      "additional_info_contact": "George Biggins\n206-664-5215",
      "reason_for_recall": "A potential exists for a component failure, which, if it occurs, could result in the inability to deliver therapy.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "On September 28, 2009 Philips issued a nationwide press release. On October 1, 2009 the firm issued notification letters to consignees arranging for the return and replacement of the affected units.  The firm also set up a page on their website with a serial number look-up tool.\n\nFor further information or support concerning this issue, customers are asked to contact Philips at 1.800.263.3342 or visit www.philips.com/FR2PlusAction for the serial number look-up list, answers to frequently asked questions, directions for the battery insertion test, and the replacement AED selection form.",
      "product_quantity": "5,808",
      "distribution_pattern": "Domestic and Worldwide distribution to:\nUnited States, Canada, Australia, Austria, Bulgaria, China, Finland, France, Germany, Gibraltar, Great Britain, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Jordan, Lebanon, Malaysia, Mexico, Netherlands, Norway, Portugal, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Arab Emirates.",
      "openfda": {
        "k_number": [
          "K970741",
          "K883344",
          "K142803",
          "K954504",
          "K955923",
          "K141231",
          "K072923",
          "K981737",
          "K032051",
          "K982710",
          "K860676",
          "K111693",
          "K003565",
          "K981802",
          "K972241",
          "K122839",
          "K110825",
          "K992543",
          "K142915",
          "K150198",
          "K014157",
          "K091265",
          "K062233",
          "K042007",
          "K091561",
          "K012197",
          "K081371",
          "K012766",
          "K110168",
          "K954609",
          "K990106",
          "K020203",
          "K002079",
          "K001725",
          "K002445",
          "K971148",
          "K072430",
          "K080232",
          "K031187",
          "K081828",
          "K051475",
          "K110742",
          "K010634",
          "K980857",
          "K052465",
          "K970391",
          "K142430",
          "K964991",
          "K983393",
          "K123881",
          "K130454",
          "K040637",
          "K010257",
          "K020984",
          "K993072",
          "K022732",
          "K033474",
          "K131494",
          "K002232",
          "K984286",
          "K143714",
          "K992977",
          "K040056",
          "K002806",
          "K073089",
          "K942850",
          "K072596",
          "K990762",
          "K011901",
          "K033144",
          "K120350",
          "K984582",
          "K082937",
          "K943698",
          "K070455",
          "K013896",
          "K052161",
          "K040404",
          "K102972",
          "K110526",
          "K940445",
          "K002280",
          "K041459",
          "K110361",
          "K063375",
          "K133659",
          "K870622",
          "K040438",
          "K052057",
          "K141774",
          "K112432",
          "K014209",
          "K962359",
          "K063510",
          "K103186",
          "K883341",
          "K955828",
          "K932931",
          "K051632",
          "K882067",
          "K122600",
          "K141795",
          "K973486",
          "K050004",
          "K040479",
          "K020288",
          "K060559",
          "K926172",
          "K965117",
          "K970719",
          "K042417",
          "K960329",
          "K052316",
          "K020715",
          "K062537",
          "K041883",
          "K983287",
          "K110978",
          "K133269",
          "K982025",
          "K073716",
          "K103567",
          "K982992",
          "K000638",
          "K021168",
          "K892065",
          "K972418",
          "K953204",
          "K955628",
          "K062970",
          "K080903",
          "K120406",
          "K991910",
          "K991007",
          "K030898",
          "K033771",
          "K993745",
          "K883584",
          "K131525",
          "K003548",
          "K092391",
          "K993533",
          "K991556",
          "K991871",
          "K071025",
          "K020743",
          "K853564",
          "K072812",
          "K150055",
          "K100565",
          "K112761",
          "K003228",
          "K012428",
          "K023854",
          "K011541",
          "K892005",
          "K941262",
          "K040775",
          "K091548",
          "K954957",
          "K952101",
          "K130153",
          "K013425",
          "K955882",
          "K942344",
          "K130684",
          "K041067",
          "K032439",
          "K051134",
          "K972970",
          "K033896",
          "K934692",
          "K932028",
          "K955890",
          "K061707",
          "K012030",
          "K031987",
          "K941970",
          "K971149",
          "K051076",
          "K121853",
          "K010214",
          "K993036",
          "K040256",
          "K032421",
          "K935922",
          "K091943",
          "K072233",
          "K082861",
          "K063119",
          "K120907",
          "K082090",
          "K113787",
          "K131550",
          "K010207",
          "K955854",
          "K060934",
          "K011144",
          "K041892",
          "K062177",
          "K982264",
          "K090989",
          "K081081",
          "K021453",
          "K022210",
          "K111594",
          "K042404",
          "K120536",
          "K925936",
          "K010918",
          "K033275",
          "K133484",
          "K142709",
          "K102496",
          "K983307",
          "K133239",
          "K092598",
          "K122758",
          "K081259",
          "K042088",
          "K133441",
          "K896919",
          "K011388",
          "K003819",
          "K033642",
          "K022929",
          "K080421",
          "K110154"
        ],
        "pma_number": [
          "P160032",
          "P190013",
          "P160034",
          "P160008",
          "P190007",
          "P180028",
          "P200007",
          "P160012",
          "P170018",
          "P210015",
          "P200005",
          "P160022",
          "P160033",
          "P200004",
          "P160028",
          "P160015",
          "P160026"
        ],
        "registration_number": [
          "",
          "1423537",
          "9681191",
          "3030733800",
          "3002807314",
          "3004035727",
          "3012494290",
          "3002714378",
          "3029871637",
          "3026007685",
          "3009077524",
          "3012502601",
          "3014579161",
          "3043226252",
          "3013517171",
          "1218058",
          "3003521780",
          "1921846",
          "3012972516",
          "3038259592",
          "3013826848",
          "3003537036",
          "3005621729",
          "3014572985",
          "1223925",
          "3006108746",
          "2112020",
          "1423662",
          "3008717264",
          "8020045",
          "3030446844",
          "3026976873",
          "3010317211",
          "3021559257",
          "9681860",
          "3004906872",
          "3010783192",
          "3020460367",
          "3010839093",
          "3030677",
          "3013596024",
          "3012104670",
          "3005906887",
          "3027815",
          "3015876",
          "3013557562",
          "3003800159",
          "3017422354",
          "3009886119",
          "3010030355",
          "3003867964",
          "1219103",
          "1317188",
          "1320894",
          "1720929",
          "1220908",
          "3013500228",
          "1930027",
          "3011823740",
          "3000126629",
          "3015997711",
          "2132111",
          "3012607778",
          "3010703925",
          "3004123209",
          "3015134086",
          "3007029079"
        ],
        "fei_number": [
          "",
          "3030733800",
          "3002807314",
          "3004035727",
          "3012494290",
          "3002714378",
          "3029871637",
          "3026007685",
          "3009077524",
          "3012502601",
          "1000524572",
          "3001451571",
          "3014579161",
          "1000305720",
          "3043226252",
          "3013517171",
          "1218058",
          "3001236905",
          "3003521780",
          "1000148701",
          "1921846",
          "3012972516",
          "3038259592",
          "3013826848",
          "3003537036",
          "3005621729",
          "3014572985",
          "3006108746",
          "2112020",
          "3008717264",
          "3008207730",
          "3030446844",
          "3026976873",
          "3010317211",
          "3021559257",
          "1000127185",
          "3004906872",
          "6736",
          "3010783192",
          "3020460367",
          "3010839093",
          "3000215273",
          "3012104670",
          "3013596024",
          "3005906887",
          "1000140026",
          "3015876",
          "3013557562",
          "3003800159",
          "3017422354",
          "3009886119",
          "3010030355",
          "3003867964",
          "1317188",
          "1720929",
          "1220908",
          "3013500228",
          "1930027",
          "3011823740",
          "3000126629",
          "3003491392",
          "3015997711",
          "3012607778",
          "3010703925",
          "3004123209",
          "3015134086",
          "3007029079"
        ],
        "device_name": "Automated External Defibrillators (Non-Wearable)",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.5310",
        "device_class": "3"
      }
    }
  ]
}