{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-13",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "85112",
      "product_res_number": "Z-0147-2010",
      "event_date_initiated": "2009-08-11",
      "event_date_posted": "2009-11-09",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-04-04",
      "res_event_number": "53257",
      "product_code": "dxg",
      "k_numbers": [
        "K932414"
      ],
      "product_description": "Power cord for QVue Continuous Cardiac Output (CCO)/SO2 computer, list #s 52230-04-03, 52230-04-05, 52230-04-07, 52230-04-09, 52230-04-11, 52230-04-13, 52230-04-15, 52230-04-17, 52230-04-25, 52230-04-31, 52230-04-33, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.",
      "code_info": "AC power cords, all units of  Part Numbers: 826-85559-001,  826-85559-003,  826-85559-008,  826-85559-010,  826-96057-003",
      "firm_fei_number": "2921482",
      "recalling_firm": "Hospira Inc",
      "reason_for_recall": "Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard.",
      "root_cause_description": "Component design/selection",
      "action": "Hospira initiated its recall on 08/14/2009.  A nationwide press release was issued to AP, and a letter to consignees was sent via UPS.  The firm will replace affected cords and effectiveness checks will be conducted through tracking of servicing.",
      "product_quantity": "204,999 power cords distributed for all devices",
      "distribution_pattern": "Worldwide distribution.",
      "openfda": {}
    }
  ]
}