      "cfres_id": "85111",
      "product_res_number": "Z-0146-2010",
      "event_date_initiated": "2009-08-11",
      "event_date_posted": "2009-11-09",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-04-04",
      "res_event_number": "53257",
      "product_code": "dxg",
      "k_numbers": [
        "K932414"
      ],
      "product_description": "Power cord for Q2Plus/Continuous Cardiac Output (CCO)/SO2 Computer, list #s 56711-04-01, 56711-04-03, 5671-04-05, 56711-04-51, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.",
      "code_info": "AC power cords, all units of  Part Numbers: 826-85559-001,  826-85559-003,  826-85559-008,  826-85559-010,  826-96057-003",
      "firm_fei_number": "2921482",
      "recalling_firm": "Hospira Inc",
      "reason_for_recall": "Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard.",
      "root_cause_description": "Component design/selection",
      "action": "Hospira initiated its recall on 08/14/2009.  A nationwide press release was issued to AP, and a letter to consignees was sent via UPS.  The firm will replace affected cords and effectiveness checks will be conducted through tracking of servicing.",
      "product_quantity": "204,999 power cords distributed for all devices",
      "distribution_pattern": "Worldwide distribution.",