      "cfres_id": "75641",
      "product_res_number": "Z-0737-2009",
      "event_date_initiated": "2008-08-01",
      "event_date_posted": "2009-02-03",
      "recall_status": "Terminated",
      "event_date_terminated": "2009-11-12",
      "res_event_number": "48991",
      "product_code": "CAI",
      "product_description": "King Systems Universal Flex2 Breathing Circuit, Latex Free, 50 per case, King Systems Corporation, Noblesville, IN; REF # DF470-6127Z.\n\nThis breathing circuit is a device intended to administer medical gases and/or anesthetic gases to a patient during anesthesia for inhalation, or medical gases to a patient during respiratory care inhalation.",
      "code_info": "Lots IKG75, IKP73, IKXI3, IL0Z7 and IL4L6.",
      "firm_fei_number": "1824226",
      "recalling_firm": "King Systems Corp.",
      "address_1": "15011 Herriman Blvd",
      "city": "Noblesville",
      "state": "IN",
      "postal_code": "46060",
      "additional_info_contact": "317-776-6823",
      "reason_for_recall": "The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels.",
      "root_cause_description": "Pending",
      "action": "Consignees were notified by an Urgent Product Recall letter on or about 7/29/2008 to cease sales, notify their customers to the hospital level, and to return product on hand.",
      "product_quantity": "Approx. 500,000 units of all products.",
      "distribution_pattern": "Worldwide Distribution - including USA and countries of Australia, Colombia, Germany, Netherlands, and Peru.",
      "openfda": {
        "k_number": [
          "K934462",
          "K852313",
          "K953787",
          "K801191",
          "K904538",
          "K800499",
          "K770549",
          "K932575",
          "K864588",
          "K852405",
          "K932589",
          "K952020",
          "K912504",
          "K823714",
          "K932590",
          "K811858",
          "K922789",
          "K932576",
          "K932588",
          "K942123",
          "K954480",
          "K940889",
          "K932190",
          "K852312",
          "K952009",
          "K932497",
          "K944295",
          "K932781",
          "K932486",
          "K925232",
          "K821087",
          "K862003",
          "K925543",
          "K821311",
          "K852311",
          "K900594",
          "K942131",
          "K871669",
          "K881141",
          "K880683",
          "K890499",
          "K942133",
          "K840850",
          "K932485",
          "K840056",
          "K771017",
          "K861438",
          "K812624",
          "K760565",
          "K852316",
          "K932578",
          "K833088",
          "K932583",
          "K926454",
          "K852315",
          "K932192",
          "K833478",
          "K853342",
          "K822769",
          "K932587",
          "K942570",
          "K942130",
          "K912065",
          "K813382",
          "K851530",
          "K891479",
          "K905148",
          "K790115",
          "K932562",
          "K870859",
          "K820352",
          "K942135",
          "K925231",
          "K911880",
          "K841264",
          "K911263",
          "K940890",
          "K822379",
          "K952014",
          "K930713",
          "K932196",
          "K932780",
          "K925893",
          "K811015",
          "K842488",
          "K932586",
          "K832453",
          "K924627",
          "K932195",
          "K932579",
          "K932558",
          "K771034",
          "K930252",
          "K821086",
          "K771518",
          "K884072",
          "K840740",
          "K932559",
          "K871386",
          "K901739",
          "K770286",
          "K851529",
          "K893919",
          "K932489",
          "K760149",
          "K873359",
          "K896897",
          "K932582",
          "K945984",
          "K840431",
          "K932488",
          "K841455",
          "K910896",
          "K851531",
          "K771307",
          "K853343",
          "K830470",
          "K932585",
          "K935044",
          "K942122",
          "K952016",
          "K932197",
          "K780044",
          "K800498",
          "K915452",
          "K770838",
          "K830475",
          "K760819",
          "K942568",
          "K932194",
          "K900428",
          "K801111",
          "K932708",
          "K771134",
          "K790214",
          "K833871",
          "K942127",
          "K811700",
          "K905450",
          "K942126",
          "K932707",
          "K852314",
          "K945314",
          "K952017",
          "K944533",
          "K911736",
          "K952019",
          "K792499",
          "K811413",
          "K932193",
          "K931487",
          "K942128",
          "K853065",
          "K851193",
          "K932487",
          "K942124",
          "K852632",
          "K932580",
          "K932288",
          "K895357",
          "K924889",
          "K942134",
          "K953788",
          "K932490",
          "K933641",
          "K853344",
          "K944429",
          "K854573",
          "K935168",
          "K945565",
          "K852866",
          "K942136",
          "K942132",
          "K932710",
          "K913753",
          "K952018",
          "K781908",
          "K896365",
          "K942125",
          "K844149",
          "K842492",
          "K932577",
          "K942129",
          "K833196"
        ],
        "registration_number": [
          "",
          "3004111573",
          "3010397428",
          "3014208300",
          "3018762668",
          "3006546082",
          "3008693658",
          "9617566",
          "3018094310",
          "3005347827",
          "3003862188",
          "9617604",
          "8030673",
          "3006984712",
          "3017828726",
          "3021947238",
          "3008294339",
          "1423395",
          "3029706818",
          "9610849",
          "3006534321",
          "3043619925",
          "3016634999",
          "3038195661",
          "3006446479",
          "3008806809",
          "9616088",
          "3006210673",
          "3004116032",
          "1824226",
          "3001650535",
          "2523320",
          "8010042",
          "3006666710",
          "1282497",
          "3003902679",
          "9710567",
          "3021039857",
          "3016195240",
          "9610861",
          "1314417",
          "3013550220",
          "3005595283",
          "2320762",
          "3011987967",
          "1924066",
          "3018783526",
          "1319447",
          "3012834717",
          "1220908",
          "3006606984",
          "8021995",
          "3036979319",
          "8043316",
          "3010131137",
          "3010957755",
          "3010580321",
          "2112667",
          "3045058049",
          "3017178358",
          "3021244152",
          "3038293079",
          "3015168843",
          "9681622",
          "3006157842",
          "1043214",
          "3013507143",
          "3012050423",
          "3022631687",
          "2028807",
          "3008258694",
          "3016764932",
          "2648727",
          "8010939",
          "3017171849",
          "3004365956",
          "2020550",
          "3010817335",
          "3018366470",
          "3001415205",
          "3011780089",
          "3009302812",
          "3005998497",
          "3011587055",
          "3003903031",
          "3011986317",
          "3005012805",
          "1722746",
          "1423537",
          "3006897516",
          "3016452400",
          "1061124",
          "3000126629",
          "3005767446",
          "1836161",
          "3010202439",
          "2085602",
          "3010515776",
          "3030446844",
          "3007997596",
          "1417592",
          "3005543241",
          "3008867836",
          "1832562",
          "3032012299",
          "3005273623",
          "3010621964",
          "3015515504",
          "3013576541",
          "8044169",
          "9610105",
          "3031369651",
          "9616354",
          "3010683054",
          "3037918935",
          "3004122598",
          "2518157",
          "3016458361",
          "3007092328",
          "2022648",
          "3010834934",
          "3045816902",
          "3009171251",
          "1420751",
          "3006547005",
          "3020777588",
          "9616991",
          "8044004",
          "3019793765",
          "3014437893",
          "3013783488",
          "9611500",
          "1000604079",
          "2025816",
          "9710391",
          "3006783791",
          "2938401",
          "3009107690",
          "3002807267",
          "3011185837",
          "3010187802",
          "8022265",
          "9610617",
          "3015907115",
          "9616679",
          "3014916466",
          "3005587132",
          "2518422",
          "3013298431",
          "3016052406",
          "3044320625",
          "3011137372",
          "3038562397",
          "2518436",
          "3017202422",
          "3015495690",
          "3034629649",
          "3015537026",
          "3044994756",
          "3014342096",
          "3006897996",
          "3015134079",
          "3010439239",
          "8044099",
          "9616445",
          "3013530405",
          "3023272766",
          "9613910",
          "3010400865",
          "3010707607",
          "3015173212",
          "3010889742",
          "1024404",
          "3001313714",
          "3015058854",
          "1928237",
          "3010838917",
          "2024500",
          "3013187142",
          "3015191173",
          "3021010418",
          "9680661",
          "3007127332",
          "3006061749",
          "1218704",
          "2936999",
          "3016734145",
          "9611451",
          "3012494290",
          "9710644",
          "3002905553",
          "3010402691",
          "3007289408",
          "9681384",
          "3014615028",
          "3008157119",
          "3013557562",
          "3012187635",
          "3015142721",
          "3003449582",
          "3013412170",
          "3003898360",
          "3020460367",
          "3014579161",
          "3004050971"
        ],
        "fei_number": [
          "",
          "3004111573",
          "3010397428",
          "3014208300",
          "3018762668",
          "3006546082",
          "3008693658",
          "3018094310",
          "3005347827",
          "3005747797",
          "3003862188",
          "3003205539",
          "3006984712",
          "3017828726",
          "3021947238",
          "3008294339",
          "3003244771",
          "3029706818",
          "3006534321",
          "3043619925",
          "3002808128",
          "3002806988",
          "3016634999",
          "3038195661",
          "3004574120",
          "3008806809",
          "3006446479",
          "3000210616",
          "3006210673",
          "3004116032",
          "1000306216",
          "1824226",
          "3001650535",
          "2523320",
          "3006666710",
          "1282497",
          "3003902679",
          "3021039857",
          "3016195240",
          "1314417",
          "3013550220",
          "3005595283",
          "3011987967",
          "1924066",
          "3003202124",
          "3018783526",
          "3012834717",
          "3003256355",
          "1220908",
          "3006606984",
          "3036979319",
          "3010131137",
          "2112667",
          "3010957755",
          "3010580321",
          "3045058049",
          "3017178358",
          "3021244152",
          "3038293079",
          "3002808696",
          "3015168843",
          "3006157842",
          "3002861325",
          "3001104093",
          "1043214",
          "3013507143",
          "3006759904",
          "3012050423",
          "3003319165",
          "3022631687",
          "3008258694",
          "3016764932",
          "2648727",
          "3017171849",
          "3004365956",
          "2020550",
          "3002808268",
          "3010817335",
          "3004213967",
          "3018366470",
          "3001415205",
          "3011780089",
          "3009302812",
          "3005998497",
          "3011587055",
          "3003903031",
          "3011986317",
          "3005012805",
          "3002807391",
          "3006897516",
          "3016452400",
          "3000126629",
          "3005767446",
          "3010515776",
          "3030446844",
          "3002807293",
          "3003418980",
          "3007997596",
          "1417592",
          "3005543241",
          "3008867836",
          "1832562",
          "3032012299",
          "3005273623",
          "3010621964",
          "3015515504",
          "3002806783",
          "3013576541",
          "3002807836",
          "3031369651",
          "3010683054",
          "3037918935",
          "3004122598",
          "2518157",
          "3016458361",
          "3007092328",
          "2022648",
          "1000374126",
          "3010834934",
          "3045816902",
          "1420751",
          "3006547005",
          "3004458261",
          "3020777588",
          "1000512168",
          "3014437893",
          "3019793765",
          "3013783488",
          "1000604079",
          "2025816",
          "3001236905",
          "3002558692",
          "3006783791",
          "2938401",
          "3009107690",
          "3002807267",
          "3003057578",
          "3011185837",
          "3010187802",
          "3015907115",
          "3014916466",
          "3005587132",
          "2518422",
          "3013298431",
          "3016052406",
          "3044320625",
          "3001227743",
          "3038562397",
          "2518436",
          "3002807626",
          "3017202422",
          "3015495690",
          "3034629649",
          "3015537026",
          "3044994756",
          "3014342096",
          "3003949221",
          "3006897996",
          "3015134079",
          "3010439239",
          "3000717925",
          "3013530405",
          "3023272766",
          "3002902735",
          "1000283452",
          "3010400865",
          "3010707607",
          "3015173212",
          "3010889742",
          "1024404",
          "3015058854",
          "1928237",
          "3010838917",
          "3003040625",
          "2024500",
          "3013187142",
          "3015191173",
          "3021010418",
          "3005749238",
          "3007127332",
          "3006061749",
          "3003952420",
          "1000120301",
          "2936999",
          "3016734145",
          "3002808440",
          "3012494290",
          "3002905553",
          "3010402691",
          "3007289408",
          "3014615028",
          "3008157119",
          "3007475226",
          "3012187635",
          "3013557562",
          "3015142721",
          "3013412170",
          "3003449582",
          "1000508235",
          "3006363476",
          "3003898360",
          "3002750084",
          "3020460367",
          "3014579161",
          "3004050971"
        ],
        "device_name": "Circuit, Breathing (W Connector, Adaptor, Y Piece)",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.5240",
        "device_class": "1"
      }
    }
  ]
}