{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "75534",
      "product_res_number": "Z-0643-2009",
      "event_date_initiated": "2008-11-10",
      "event_date_posted": "2009-01-08",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-05-07",
      "res_event_number": "50501",
      "product_code": "DXT",
      "k_numbers": [
        "K010390",
        "K991103"
      ],
      "product_description": "ACIST CVi Angiographic Injection System with Interface Cable for Toshiba Infinix-i cardiovascular X-ray system, Part Number 700318-002.  The ACIST injection system is an angiographic injection system used in interventional cardiology, radiology, and vascular surgical procedures.  The ACIST injection system supplies radiopaque contrast media to a catheter at a user-determined variable flow rate and volume which can be instantaneously and continuously varied.   Manufactured by:  ACIST Medical Systems, Inc., 7905 Fuller Road, Eden Prairie, MN  55344, USA.    The ACIST CVi injection system is able to synchronize with certain X-ray imaging systems from Siemens, Toshiba, GE and Philips.  \n\nIntended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.",
      "code_info": "Not associated with specific serial or lot numbers. Model: CVi Toshiba interface Cable, part number 700318-002.",
      "firm_fei_number": "3001726808",
      "recalling_firm": "Acist Medical Systems",
      "address_1": "7905 Fuller Rd",
      "city": "Eden Prairie",
      "state": "MN",
      "postal_code": "55344",
      "additional_info_contact": "952-995-9300",
      "reason_for_recall": "Multiple Injections may result.  It has been determined that when using the ACIST CVi Angiographic Injection system to synchronize with the Toshiba Infinix -i cardiovascular x-ray system there is an injection scenario that may result in multiple injections of contract media when one injection was intended.",
      "root_cause_description": "Software design",
      "action": "ACIST is distributing an Urgent Medical Device Correction letter, dated November 10, 2008, to affected customers.  The letter describes the issue and the scenario when the issue to occur,  instructs them to discontinue use of the synchronized mode between the ACIST Angiographic Injection System model CVi and the Toshiba Infinix-i X-ray fluoroscopic system by disconnecting the interface cable immediately.  The letter states that ACIST is working with Toshiba to resolve this issue.  A response form is asked to be returned either by fax, e-mail or mail.  If there are any questions, they should be directed to either ACIST at 952-995-9300 or local Sales Reps or Clinical Applications Specialist.",
      "product_quantity": "18 cables",
      "distribution_pattern": "Worldwide Distribution --- USA including states of AR, AZ, FL, and NY and countries of France, Germany, Italy, Japan, and Spain.",
      "openfda": {
        "k_number": [
          "K020892",
          "K984418",
          "K830754",
          "K823253",
          "K942381",
          "K823256",
          "K140425",
          "K152906",
          "K823920",
          "K863596",
          "K904212",
          "K171646",
          "K162088",
          "K823254",
          "K912944",
          "K042744",
          "K092896",
          "K133600",
          "K910594",
          "K873173",
          "K072696",
          "K991557",
          "K011991",
          "K090487",
          "K955684",
          "K881521",
          "K800613",
          "K173450",
          "K113133",
          "K844045",
          "K862547",
          "K873955",
          "K984088",
          "K161471",
          "K192657",
          "K193028",
          "K841092",
          "K993728",
          "K863537",
          "K890208",
          "K984231",
          "K962141",
          "K934086",
          "K791782",
          "K905819",
          "K142081",
          "K960430",
          "K150902",
          "K173773",
          "K013815",
          "K983826",
          "K090621",
          "K791793",
          "K120892",
          "K181936",
          "K230668",
          "K902221",
          "K023183",
          "K152218",
          "K955685",
          "K840281",
          "K151300",
          "K872981",
          "K790835",
          "K924116",
          "K061976",
          "K172694",
          "K182289",
          "K875196",
          "K062168",
          "K903494",
          "K160190",
          "K010390",
          "K781334",
          "K993774",
          "K841493",
          "K011811",
          "K232388",
          "K920135",
          "K033247",
          "K822536",
          "K823192",
          "K935668",
          "K082905",
          "K883769",
          "K913624",
          "K160107",
          "K911549",
          "K001663",
          "K800823",
          "K955381",
          "K823191",
          "K242212",
          "K133189",
          "K183267",
          "K170279",
          "K251788",
          "K903390",
          "K854310",
          "K910873",
          "K033881",
          "K093830",
          "K991103",
          "K880310",
          "K903495",
          "K932083",
          "K902475",
          "K142745",
          "K863037",
          "K973334",
          "K242143",
          "K982814",
          "K153141",
          "K883452",
          "K050456",
          "K190102",
          "K073592",
          "K252891",
          "K871129",
          "K880877",
          "K083523",
          "K823340",
          "K823238",
          "K163054",
          "K973230",
          "K896415",
          "K171217",
          "K252652",
          "K131478",
          "K923117",
          "K040298",
          "K012950",
          "K911166",
          "K082212",
          "K963982",
          "K052633",
          "K031483",
          "K875324",
          "K182276",
          "K880987",
          "K933846",
          "K873410",
          "K211564",
          "K203004",
          "K850861",
          "K152978",
          "K170431",
          "K895318",
          "K192370",
          "K000013",
          "K051799",
          "K143538",
          "K821401",
          "K834107",
          "K253048",
          "K150485",
          "K960674",
          "K944875",
          "K960965",
          "K971712",
          "K163084",
          "K823356",
          "K191060",
          "K082439",
          "K890898",
          "K902377",
          "K022116",
          "K252653",
          "K201395",
          "K252638",
          "K851402",
          "K951802",
          "K882329",
          "K020333",
          "K873597",
          "K873687",
          "K042784",
          "K830609",
          "K161505",
          "K963071",
          "K900034",
          "K031339",
          "K964642",
          "K071691",
          "K032920",
          "K833138",
          "K872472",
          "K933908",
          "K171698",
          "K880493",
          "K140469",
          "K151746",
          "K904994",
          "K112086",
          "K860204",
          "K072886",
          "K884913",
          "K890292",
          "K241109",
          "K070798",
          "K113198",
          "K132928",
          "K052744",
          "K151960",
          "K833645",
          "K182273",
          "K860842",
          "K901382",
          "K823190",
          "K244038"
        ],
        "registration_number": [
          "2134243",
          "3035858922",
          "3004050971",
          "1060680",
          "3020460367",
          "1058584",
          "2029275",
          "1417592",
          "3003120897",
          "9616990",
          "3015183635",
          "3012085143",
          "3007007790",
          "9610612",
          "3012044688",
          "1047843",
          "2027111",
          "3042190142",
          "3004056159",
          "3014498725",
          "3006950086",
          "3003829571",
          "3009763019",
          "3007048453",
          "9610849",
          "3007138831",
          "3013413516",
          "1038185",
          "3045750327",
          "3009882464",
          "1625425",
          "3008850249",
          "1526711",
          "3004068971",
          "3010160527",
          "2011171",
          "3009746061",
          "3008692839",
          "3014246639",
          "2520313",
          "1721686",
          "1720929",
          "3018094310",
          "3031503890",
          "3003786696",
          "2032112",
          "3013820501",
          "3013529745",
          "3003898360",
          "3004111573",
          "2936999",
          "3003561450",
          "3011642792",
          "3015309643",
          "3011604145",
          "8043817",
          "2523676",
          "3002808086",
          "3015009200",
          "2521402",
          "3016096098",
          "3015276088",
          "3010934635",
          "1018470",
          "1724474",
          "3010041511",
          "3012127758",
          "1518293",
          "3009051888",
          "8044099",
          "1526534",
          "1422634",
          "1721504",
          "1043214",
          "3004727093",
          "3003627840",
          "3011334555",
          "1721676",
          "3010131137",
          "3009196578",
          "3010665433",
          "2020394",
          "8022210",
          "1000604079",
          "3004753774",
          "3015173212",
          "1061124",
          "1054713",
          "3007361214",
          "3010009632",
          "2243252",
          "3007289408",
          "1055236",
          "3004605321",
          "3019807891",
          "3006791331",
          "3014615028",
          "2032098",
          "3033589330",
          "3012536737",
          "3012395857",
          "1643817",
          "3002949731",
          "3015910259",
          "1526863",
          "9616567",
          "3020298301",
          "3013764800",
          "3034544377",
          "3015225571",
          "3001650535",
          "3004080920",
          "2022435",
          "3003542872",
          "2249018",
          "1928237",
          "1319639",
          "1043729",
          "3003915875",
          "1054241",
          "1123137",
          "3009504230",
          "3011137372",
          "2320762",
          "3007084575",
          "3003768277"
        ],
        "fei_number": [
          "1000138054",
          "3004050971",
          "3035858922",
          "1000219637",
          "3005747797",
          "3020460367",
          "1000208624",
          "1417592",
          "3003120897",
          "3015183635",
          "3012085143",
          "3007007790",
          "3003040625",
          "3012044688",
          "1047843",
          "3002807391",
          "2027111",
          "1000514656",
          "3042190142",
          "3004056159",
          "3014498725",
          "3006950086",
          "3003829571",
          "3002806315",
          "3009763019",
          "3007048453",
          "3007138831",
          "3013413516",
          "1038185",
          "3045750327",
          "3009882464",
          "1625425",
          "3008850249",
          "1526711",
          "3004068971",
          "1000114111",
          "2011171",
          "3003076831",
          "3009746061",
          "3008692839",
          "3002807040",
          "3014246639",
          "2520313",
          "1721686",
          "1720929",
          "3018094310",
          "3031503890",
          "3003786696",
          "3013820501",
          "3013529745",
          "3003898360",
          "1000306431",
          "3004111573",
          "2936999",
          "3003561450",
          "3011642792",
          "3015309643",
          "3011604145",
          "2523676",
          "3002808086",
          "3001726808",
          "3015009200",
          "2521402",
          "3010934635",
          "1018470",
          "3010041511",
          "3015276088",
          "3016096098",
          "3003058448",
          "3012127758",
          "1518293",
          "3000718866",
          "3009051888",
          "1526534",
          "1721504",
          "1043214",
          "1000112968",
          "3002963732",
          "3004727093",
          "3003627840",
          "3011334555",
          "1721676",
          "3010131137",
          "3009196578",
          "3010665433",
          "1000512168",
          "2020394",
          "1000604079",
          "3004753774",
          "3015173212",
          "3007361214",
          "3010009632",
          "2243252",
          "3007289408",
          "3004605321",
          "3019807891",
          "3006791331",
          "3014615028",
          "1000112877",
          "3033589330",
          "3012536737",
          "3012395857",
          "1643817",
          "3002949731",
          "3015910259",
          "1526863",
          "3020298301",
          "3003094575",
          "3013764800",
          "3002807880",
          "3034544377",
          "3015225571",
          "3001650535",
          "3004080920",
          "2022435",
          "3003542872",
          "3932",
          "3003094851",
          "1928237",
          "1043729",
          "3003915875",
          "1000123180",
          "3009504230",
          "3008918179",
          "3007084575",
          "3004458261"
        ],
        "device_name": "Injector And Syringe, Angiographic",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1650",
        "device_class": "2"
      }
    }
  ]
}