  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "73911",
      "product_res_number": "Z-0296-2009",
      "event_date_initiated": "2008-05-23",
      "event_date_posted": "2008-11-07",
      "recall_status": "Terminated",
      "event_date_terminated": "2009-08-04",
      "res_event_number": "49604",
      "product_code": "DTQ",
      "k_numbers": [
        "K871131"
      ],
      "product_description": "Terumo Sarns Level Sensor II (red) (Part # 195274) alarm level sensor transducer for use with Sarns Advanced Perfusion System I, Perfusion System 800 and Perfusion System 900; Terumo Cardiovascular Systems Corp., Ann Arbor, MI.; REF 195274.  Level sensors are distributed as a finished device, as part of a kit, with a perfusion system 8000 safety monitor or with an Advanced Perfusion System 1 base.\n\nLevel sensors are attached to the color coded connectors on the module and then to a hard-shell reservoir.  Level detection is used to monitor blood levels in hard-shell reservoirs.",
      "code_info": "Serial Numbers:  2264, 2281, 3077 through 3082, 3084 through 3088, 3092, 3094 through 3110, 3112 through 3115, 3117, 1932x, 1939x through 1943x, 1950x , 1951x, 1954x, 1973x, 1979x, 1980x, 1982x, 1990x, 1994x, 2000x, 2009x, 2014x through 2064x, 2066x through 2163x, 2194x through 2222x, 2225x through 2232x, 2280x, 2282x, 2283x, 2290x, 2294x, 2295x, 2298x through 2301x, 2303x, 2305x, 2306x, 2309x through 2311x, 2313x through 2315x, 2318x, 2319x, 2322x through 2324x, 2330x through 2332x, 2334x through 2347x, 2350x through 2467x, 2469x through 2471x, 2473x t hrough 2503x, 2505x, 2507x through 2529x, 2533x, 2536x, 2539x through 2541x, 2543x through 2551x, 2553x through 2572x, 2574x through 2583x, 2585x through 2609x and 2611x through 2624x.",
      "firm_fei_number": "1828100",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Rd",
      "city": "Ann Arbor",
      "state": "MI",
      "postal_code": "48103",
      "additional_info_contact": "800-262-3304",
      "reason_for_recall": "The level sensor may not properly couple to the reservoir, resulting in a sensor \"not attached\" message, an alert or alarm condition, or a failure to detect a low level condition.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "U.S. consignees were notified by (Urgent Medical Device Removal) letter dated 5/23/08 via FedEx and international affiliates were notified via e-mail on 5/27/08.  Consignees were advised of the problem, and to continue using the affected sensors until replacements are available in late 2008.\n\nFor questions or additional information, contact the Terumo CVS Customer Service at 1-800-521-2818.",
      "product_quantity": "547",
      "distribution_pattern": "Worldwide including USA and countries of Australia, Belgium, Bolivia, Canada, Chile, Ecuador, Egypt, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Latin America, Malaysia, Panama, Philippines, Saudi Arabia, Singapore, South Africa, South Korea, Sri Lanka, Taiwan, Thailand, United Arab Emirates, Venezuela and Vietnam.",
      "openfda": {
        "k_number": [
          "K842138",
          "K011088",
          "K052555",
          "K813034",
          "K863960",
          "K090006",
          "K822068",
          "K133598",
          "K021694",
          "K210130",
          "K861388",
          "K002116",
          "K071318",
          "K173834",
          "K943803",
          "K871131",
          "K221373",
          "K862973",
          "K851615",
          "K080832",
          "K870976",
          "K801097",
          "K915183",
          "K151349",
          "K060053",
          "K092463",
          "K190282",
          "K030141",
          "K070213",
          "K091007",
          "K020571",
          "K163531",
          "K772001",
          "K062396",
          "K103762",
          "K002118",
          "K863541",
          "K982014",
          "K102726",
          "K022947",
          "K942582",
          "K770002",
          "K232291",
          "K153376",
          "K221895",
          "K964017",
          "K202125",
          "K950990",
          "K172220",
          "K131041",
          "K851614",
          "K093882"
        ],
        "registration_number": [
          "2916596",
          "9680841",
          "3007427881",
          "1828100",
          "3010163695",
          "3042990265",
          "2184009",
          "1218444",
          "8010762",
          "9611109",
          "3019807891",
          "3006073153",
          "3002807090",
          "3010703925",
          "3009507273",
          "3005941719",
          "3011554160",
          "3001589655"
        ],
        "fei_number": [
          "2916596",
          "1000116158",
          "3007427881",
          "1828100",
          "3010163695",
          "3042990265",
          "1218444",
          "3002808267",
          "3019807891",
          "3006073153",
          "3010703925",
          "3009649876",
          "3006003617",
          "3002807111",
          "3009507273",
          "3011554160",
          "3001589655",
          "3005426550"
        ],
        "device_name": "Console, Heart-Lung Machine, Cardiopulmonary Bypass",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.4220",
        "device_class": "2"
      }
    }
  ]
}