{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-08-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "73737",
      "product_res_number": "Z-0271-2009",
      "event_date_initiated": "2008-09-22",
      "event_date_posted": "2008-10-31",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-01-24",
      "res_event_number": "49525",
      "product_code": "LGW",
      "pma_numbers": [
        "P030017"
      ],
      "product_description": "Precision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System labeled as BIONICS SCS-CHARGING KIT  CONTENTS Charger (1)  Base Station (1)   Belt   Power Supply (1)  Adhesive Kit  ADVANCED BIONICS CORP  12740 San Fernando Rd.   Sylmar, CA  91342   USA.    The Precision Spinal Cord Stimulator System (Precision System) is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs, including unilateral or bilateral pain associated with following: failed back surgery syndrome, intractable low back pain and leg pain. The Precision System includes an implantable 16-output, multi-channel stimulator (IPG - Implantable Pulse Generator) with a rechargeable battery power source. The IPG, commonly implanted in the abdomen or buttock area (less frequently in the subclavicular area), can be configured to accept one or two leads. The leads are implanted in the epidural space, adjacent to the spinal column. The stimulation delivered at the distal end of the lead is intended to mask the pain signals. An externally used Charger is utilized to transcutaneously recharge the IPG (Implantable Pulse Generator) battery. Currently, Boston Scientific distributes Charger 2.0 (Model Number SC-5312). For improved heat management, Charger 2.0 is equipped with thermistor (temperature sensing component), designed to activate/deactivate charging at a set temperature.",
      "code_info": "Model Number SC-5300; Serial Numbers: 1020 1022 1035 1048 1060 1062 1098 1105 1120 1130 1131 1153 1158 1167 1173 1175 1177 1178 1181 1182 1183 1184 1187 1188 1191 1192 1193 1196 1197 1199 1201 1203 1204 1205 1206 1209 1210 1211 1214 1218 1219 1220 1221 1222 1223 1227 1228 1230 1235 1236 1239 1241 1244 1249 1250 1253 1258 1261 1266 1267 1268 1269 1270 1271 1272 1274 1275 1276 1277 1279 1281 1282 1283 1284 1287 1288 1289 1292 1297 1300 1303 1305 1306 1309 1310 1311 1312 1314 1316 1317 1319 1323 1333 1335 1336 1337 1338 1340 1343 1344 1347 1348 1349 1351 1352 1354 1356 1357 1360 1361 1362 1363 1364 1366 1368 1369 1370 1371 1372 1373 1374 1376 1377 1378 1380 1381 1382 1384 1385 1386 1387 1389 1390 1391 1392 1393 1394 1396 1397 1400 1401 1404 1405 1407 1411 1412 1414 1415 1422 1424 1425 1426 1428 1431 1434 1435 1439 1440 1444 1447 1450 1451 1454 1455 1456 1457 1458 1459 1460 1461 1462 1466 1468 1469 1471 1472 1473 1475 1476 1477 1479 1481 1482 1484 1485 1486 1487 1489 1490 1491 1492 1493 1494 1496 1497 1498 1500 1501 1505 1506 1508 1509 1510 1511 1512 1513 1515 1516 1517 1518 1519 1521 1525 1526 1528 1530 1532 1535 1537 1541 1542 1543 1547 1549 1550 1553 1554 1556 1557 1560 1561 1565 1566 1569 1572 1574 1576 1579 1580 1581 1582 1584 1585 1586 1588 1590 1592 1593 1596 1597 1600 1601 1602 1604 1607 1608 1609 1611 1613 1614 1615 1618 1619 1620 1624 1625 1627 1628 1630 1633 1634 1637 1639 1640 1642 1644 1645 1646 1649 1650 1651 1653 1654 1657 1658 1660 1661 1662 1663 1664 1665 1666 1669 1675 1678 1679 1691 1692 1694 1697 3000 3001 3007 3008 3009 3010 3011 3013 3014 3015 3018 3019 3022 3023 3026 3028 3029 3030 3032 3033 3034 3039 3040 3041 3043 3044 3045 3046 3047 3052 3053 3054 3055 3057 3058 3061 3064 3065 3066 3068 3072 3074 3077 3078 3079 3080 3083 3084 3085 3087 3088 3089 3090 3091 3092 3093 3094 3095 3097 3098 3102 3104 3105 3106 3107 3108 3109 3110 3112 3114 3119 3152 3153 3158 3163 3168 3172 3176 3177 3178 3182 3184 3185 3187 3189 3191 3193 3194 3195 3196 3197 3198 3199 3200 3201 3203 3204 3207 3209 3215 3216 3217 3219 3220 3222 3226 3227 3228 3230 3231 3232 3233 3237 3238 3239 3241 3244 3248 3249 3250 3251 3252 3255 3257 3259 3260 3262 3264 3265 3266 3268 3269 3270 3272 3276 3277 3280 3282 3283 3286 3287 3289 3291 3296 3297 3298 3300 3302 3305 3306 3307 3308 3310 3312 3315 3317 3318 3319 3320 3321 3323 3325 3326 3328 3329 3330 3332 3335 3336 3337 3340 3341 3342 3343 3344 3345 3346 3349 3350 3351 3352 3354 3356 3358 3362 3363 3364 3365 3366 3367 3368 3369 3371 3379 3380 3384 3385 3387 3388 3389 3390 3392 3393 3395 3396 3397 3398 3399 3401 3402 3403 3404 3406 3408 3410 3411 3414 3415 3416 3417 3419 3420 3421 3422 3424 3425 3426 3428 3429 3439 3441 3442 3443 3445 3446 3448 3450 3451 3452 3455 3456 3457 3458 3460 3461 3462 3463 3464 3466 3467 3468 3470 3473 3477 3478 3482 3485 3489 3491 3492 3493 3495 3498 3500 3501 3505 3506 3515 3517 3528 3530 3534 3540 3541 3543 3544 3545 3546 3547 3549 3550 3551 3552 3553 3555 3556 3557 3558 3559 3560 3562 3563 3564 3568 3569 3570 3571 3572 3573 3574 3577 3578 3579 3580 3581 3582 3583 3584 3586 3587 3588 3591 3592 3593 3595 3599 3600 3601 3604 3607 3610 3611 3612 3613 3614 3616 3617 3618 3619 3620 3621 3622 3624 3625 3628 3629 3630 3632 3634 3636 3637 3638 3640 3642 3644 3646 3647 3648 3650 3651 3652 3653 3655 3656 3657 3660 3663 3664 3665 3666 3667 3668 3670 3671 3673 3674 3679 3684 3685 3686 3688 3690 3692 3693 3694 3695 3697 3698 3700 3701 3702 3703 3705 3706 3707 3708 3713 3714 3715 3717 3718 3719 3720 3721 3722 3723 3727 3728 3729 3731 3732 3733 3734 3736 3738 3741 3742 3744 3746 3748 3749 3752 3756 3757 3763 3764 3766 3769 3770 3773 3774 3776 3777 3779 3781 3782 3783 3785 3786 3788 3789 3791 3792 3793 3794 3796 3798 3799 3800 3804 3805 3806 3808 3809 3810 3811 3813 3815 3823 3825 3826 3828 3829 3831 3832 3833 3834 3835 3838 3839 3841 3843 3844 3845 3847 3848 3850 3851 3855 3856 3859 3860 3861 3863 3871 3873 3874 3875 3876 3878 3880 3881 3882 3885 3886 3887 3890 3892 3894 3895 3897 3898 3899 3900 3901 3903 3904 3906 3907 3910 3912 3913 3914 3915 3917 3918 3919 3921 3922 3925 3928 3929 3930 3931 3932 3933 3935 3936 3937 3938 3939 3940 3941 3942 3943 3945 3946 3947 3948 3949 3950 3951 3952 3953 3955 3956 3958 3959 3960 3961 3962 3964 3965 3966 3967 3968 3971 3972 3973 3974 3975 3976 3979 3980 3982 3983 3984 3985 3986 3987 3988 3989 3991 3993 3995 3996 3997 3998 4000 4001 4003 4004 4006 4007 4008 4010 4011 4013 4014 4015 4016 4017 4018 4019 4020 4021 4022 4027 4028 4029 4030 4031 4033 4035 4037 4039 4040 4041 4042 4043 4044 4045 4047 4049 4054 4055 4056 4057 4060 4061 4063 4064 4066 4067 4069 4070 4071 4072 4077 4078 4079 4080 4082 4083 4084 4085 4086 4087 4091 4093 4094 4095 4096 4097 4098 4099 4100 4103 4104 4107 4108 4109 4110 4112 4115 4116 4117 4118 4121 4123 4124 4125 4126 4127 4128 4129 4130 4136 4138 4139 4142 4143 4144 4145 4146 4148 4150 4154 4155 4156 4157 4160 4162 4163 4164 4165 4166 4167 4168 4169 4171 4173 4174 4175 4176 4177 4178 4179 4181 4182 4184 4185 4186 4188 4189 4190 4191 4193 4194 4196 4197 4198 4199 4201 4202 4203 4204 4205 4206 4207 4210 4211 4213 4214 4216 4217 4218 4219 4220 4223 4224 4225 4226 4227 4230 4231 4232 4233 4234 4236 4237 4238 4240 4241 4243 4244 4246 4247 4249 4250 4251 4252 4253 4255 4256 4257 4258 4259 4260 4261 4262 4264 4265 4268 4269 4271 4272 4273 4275 4277 4279 4281 4282 4283 4284 4285 4286 4287 4289 4290 4291 4295 4297 4299 4300 4302 4303 4304 4305 4308 4309 4310 4311 4312 4313 4315 4316 4317 4318 4321 4322 4323 4325 4326 4327 4329 4330 4332 4334 4336 4337 4338 4339 4340 4346 4348 4350 4351 4352 4353 4354 4355 4359 4361 4364 4365 4370 4371 4373 4374 4375 4376 4377 4378 4381 4382 4385 4386 4388 4390 4395 4396 4397 4398 4402 4403 4404 4405 4406 4407 4408 4409 4411 4412 4414 4416 4417 4418 4420 4421 4422 4423 4424 4432 4433 4434 4436 4437 4438 4440 4441 4442 4443 4444 4445 4446 4449 4450 4451 4452 4456 4457 4459 4460 4462 4464 4465 4467 4468 4469 4472 4473 4474 4475 4476 4477 4478 4479 4482 4483 4484 4485 4487 4488 4489 4492 4493 4494 4496 4497 4501 4502 4504 4505 4506 4507 4509 4510 4511 4513 4514 4515 4516 4517 4518 4519 4520 4521 4522 4523 4524 4526 4532 4533 4535 4536 4539 4540 4542 4545 4547 4549 4550 4551 4552 4553 4557 4558 4559 4560 4561 4564 4565 4566 4571 4572 4576 4577 4578 4580 4583 4584 4587 4589 4591 4592 4593 4594 4597 4599 4600 4601 4602 4603 4604 4605 4606 4607 4608 4612 4613 4615 4616 4617 4618 4619 4621 4622 4623 4624 4625 4630 4631 4632 4634 4635 4636 4637 4638 4639 4640 4642 4643 4645 4646 4647 4651 4652 4653 4654 4655 4656 4658 4659 4660 4662 4663 4664 4665 4667 4669 4670 4671 4672 4673 4675 4677 4678 4679 4680 4681 4682 4685 4686 4688 4689 4690 4691 4692 4694 4696 4697 4698 4699 4700 4701 4702 4703 4704 4705 4706 4707 4708 4710 4711 4712 4713 4714 4715 4718 4719 4720 4721 4723 4726 4727 4728 4729 4730 4731 4732 4733 4734 4735 4736 4737 4739 4740 4741 4742 4743 4744 4746 4747 4752 4753 4754 4755 4756 4757 4758 4759 4760 4761 4762 4763 4764 4765 4766 4767 4769 4770 4772 4773 4774 4776 4777 4778 4779 4781 4782 4784 4785 4786 4787 4788 4789 4790 4791 4792 4793 4794 4795 4796 4797 4798 4799 4801 4802 4804 4805 4808 4809 4810 4811 4812 4813 4815 4818 4819 4822 4823 4824 4825 4827 4829 4830 4831 4833 4837 4838 4841 4842 4843 4844 4849 4850 4851 4852 4853 4854 4855 4858 4859 4860 4861 4864 4865 4867 4869 4870 4872 4873 4874 4875 4877 4879 4882 4885 4888 4889 4890 4891 4892 4894 4895 4896 4897 4898 4899 4900 4902 4903 4904 4905 4906 4907 4908 4909 4910 4911 4912 4913 4914 4915 4916 4917 4918 4920 4921 4922 4923 4924 4925 4927 4929 4932 4933 4934 4935 4936 4938 4942 4943 4944 4945 4946 4948 4949 4950 4951 4954 4956 4957 4960 4964 4965 4966 4967 4968 4970 4971 4972 4973 4974 4976 4977 4978 4981 4982 4983 4984 4985 4986 4987 4988 4989 4992 4993 4994 4996 4997 4999 5000 5001 5002 5003 5004 5006 5007 5009 5010 5011 5012 5013 5015 5016 5017 5018 5019 5020 5022 5024 5026 5027 5028 5031 5033 5036 5039 5040 5041 5042 5043 5045 5046 5047 5049 5051 5052 5053 5054 5055 5058 5059 5061 5063 5064 5065 5067 5068 5069 5070 5071 5072 5075 5076 5078 5080 5081 5082 5083 5085 5086 5088 5089 5090 5092 5096 5097 5098 5099 5101 5102 5105 5107 5108 5109 5110 5111 5112 5113 5114 5118 5120 5122 5123 5125 5128 5129 5131 5132 5133 5135 5141 5142 5143 5144 5145 5148 5149 5150 5151 5152 5153 5154 5156 5158 5159 5161 5162 5166 5167 5168 5169 5170 5175 5176 5177 5178 5179 5184 5185 5186 5187 5188 5189 5190 5191 5192 5193 5195 5197 5200 5201 5202 5203 5204 5205 5207 5208 5210 5212 5213 5214 5221 5222 5223 5226 5228 5229 5230 5231 5232 5233 5234 5235 5236 5238 5239 5240 5241 5242 5243 5244 5245 5247 5248 5249 5253 5254 5255 5258 5260 5261 5262 5264 5265 5266 5267 5268 5269 5270 5272 5273 5276 5277 5279 5280 5282 5283 5285 5286 5287 5288 5292 5293 5294 5295 5297 5298 5300 5301 5302 5304 5307 5309 5310 5311 5313 5315 5317 5319 5320 5324 5326 5327 5328 5329 5330 5331 5332 5333 5334 5336 5337 5338 5339 5340 5342 5343 5344 5345 5346 5347 5349 5350 5351 5352 5353 5354 5357 5358 5359 5361 5362 5364 5365 5366 5367 5369 5370 5373 5374 5375 5376 5377 5378 5379 5380 5384 5385 5387 5388 5389 5390 5391 5393 5396 5397 5399 5400 5403 5405 5406 5407 5409 5410 5411 5412 5413 5414 5415 5418 5419 5420 5422 5423 5424 5426 5428 5429 5430 5431 5432 5433 5434 5436 5439 5440 5441 5442 5443 5444 5445 5446 5448 5449 5451 5453 5454 5456 5457 5458 5459 5460 5461 5462 5463 5465 5466 5467 5468 5470 5471 5474 5475 5477 5479 5480 5481 5482 5483 5485 5486 5487 5488 5489 5491 5494 5496 5497 5498 5500 5501 5502 5504 5508 5509 5510 5511 5512 5513 5514 5515 5516 5517 5518 5520 5521 5522 5523 5524 5526 5529 5530 5532 5535 5537 5538 5540 5541 5542 5543 5544 5547 5548 5549 5550 5552 5553 5555 5556 5559 5560 5562 5563 5565 5566 5567 5568 5569 5571 5573 5574 5576 5578 5579 5580 5581 5584 5585 5586 5587 5588 5590 5592 5593 5595 5597 5598 5604 5606 5607 5610 5612 5613 5614 5616 5617 5619 5620 5622 5623 5624 5625 5626 5627 5628 5629 5630 5631 5632 5633 5634 5635 5636 5637 5638 5641 5642 5643 5644 5645 5648 5649 5652 5653 5654 5655 5657 5658 5660 5661 5662 5663 5665 5667 5669 5670 5671 5672 5673 5674 5675 5677 5678 5679 5680 5682 5683 5686 5687 5688 5691 5693 5695 5696 5697 5699 5702 5703 5704 5705 5707 5708 5710 5712 5714 5715 5716 5717 5718 5720 5722 5723 5724 5726 5727 5728 5730 5731 5732 5735 5736 5737 5738 5739 5740 5741 5742 5744 5745 5746 5749 5751 5752 5753 5754 5756 5757 5758 5759 5760 5761 5763 5764 5765 5766 5769 5770 5771 5772 5773 5775 5777 5779 5780 5781 5782 5784 5786 5787 5788 5789 5790 5791 5792 5794 5797 5800 5802 5803 5805 5809 5810 5812 5814 5816 5817 5818 5819 5822 5823 5825 5828 5830 5832 5833 5834 5835 5836 5837 5838 5840 5841 5842 5843 5844 5847 5848 5849 5851 5852 5853 5854 5855 5856 5857 5858 5860 5861 5862 5866 5868 5870 5871 5872 5873 5874 5875 5876 5877 5879 5880 5881 5882 5884 5885 5886 5887 5888 5889 5890 5891 5892 5893 5894 5895 5896 5897 5898 5899 5900 5901 5902 5903 5905 5906 5907 5908 5909 5912 5913 5914 5915 5917 5918 5919 5921 5922 5923 5924 5926 5927 5931 5932 5933 5934 5935 5937 5938 5939 5940 5942 5943 5946 5947 5948 5949 5950 5951 5952 5954 5955 5956 5958 5959 5960 5961 5963 5964 5968 5969 5970 5971 5972 5973 5975 5977 5979 5980 5982 5983 5984 5986 5987 5990 5992 5993 5994 5995 5996 5997 5998 5999 6000 6002 6003 6008 6009 6010 6012 6015 6016 6018 6019 6021 6022 6023 6024 6025 6031 6033 6036 6037 6038 6039 6041 6042 6043 6045 6046 6047 6048 6049 6050 6051 6055 6056 6057 6058 6059 6060 6062 6064 6067 6070 6071 6072 6073 6074 6076 6077 6079 6082 6083 6084 6087 6089 6090 6091 6092 6094 6095 6096 6097 6099 6101 6108 6110 6114 6115 6119 6122 6123 6124 6126 6128 6129 6130 6132 6133 6134 6136 6138 6139 6140 6142 6143 6144 6146 6150 6153 6154 6157 6159 6160 6161 6162 6163 6164 6165 6166 6167 6168 6169 6170 6171 6173 6175 6176 6177 6179 6181 6183 6184 6186 6188 6189 6190 6191 6196 6197 6203 6204 6207 6208 6209 6211 6212 6214 6215 6216 6217 6218 6219 6222 6224 6225 6226 6229 6230 6232 6233 6235 6236 6237 6240 6242 6243 6245 6246 6247 6248 6252 6253 6254 6257 6258 6259 6260 6261 6262 6263 6265 6267 6268 6269 6270 6271 6272 6273 6274 6276 6278 6280 6282 6283 6286 6287 6289 6290 6291 6292 6296 6297 6298 6299 6300 6301 6304 6307 6308 6309 6310 6311 6312 6314 6315 6316 6317 6318 6319 6322 6324 6325 6326 6327 6328 6329 6330 6332 6333 6334 6336 6337 6338 6341 6342 6345 6346 6347 6349 6350 6352 6353 6354 6356 6358 6359 6360 6363 6364 6365 6368 6371 6372 6375 6378 6379 6380 6381 6382 6383 6384 6386 6387 6389 6390 6391 6393 6394 6395 6396 6397 6398 6399 6401 6402 6403 6404 6405 6409 6410 6411 6413 6414 6415 6416 6417 6418 6421 6423 6424 6425 6426 6427 6429 6430 6431 6432 6433 6434 6435 6436 6437 6439 6441 6443 6444 6445 6447 6448 6449 6450 6452 6453 6454 6457 6459 6460 6461 6462 6463 6465 6466 6467 6468 6471 6472 6473 6474 6475 6476 6477 6478 6479 6480 6482 6483 6486 6487 6488 6489 6492 6494 6496 6498 6499 6503 6504 6506 6507 6508 6511 6512 6514 6516 6517 6520 6521 6522 6523 6524 6525 6528 6531 6535 6536 6537 6539 6540 6543 6544 6546 6550 6551 6552 6553 6554 6558 6559 6560 6561 6568 6570 6573 6575 6576 6579 6581 6584 6585 6586 6588 6589 6591 6592 6594 6595 6601 6602 6605 6610 6612 6613 6614 6621 6624 6626 6627 6628 6629 6630 6631 6633 6634 6635 6636 6637 6638 6639 6640 6641 6642 6644 6645 6646 6647 6648 6649 6650 6651 6652 6653 6654 6658 6660 6661 6665 6668 6669 6671 6672 6676 6678 6680 6682 6683 6684 6685 6686 6688 6689 6690 6691 6693 6694 6695 6697 6698 6699 6700 6701 6703 6704 6705 6706 6707 6708 6709 6710 6712 6716 6717 6719 6720 6722 6723 6724 6725 6726 6727 6728 6729 6732 6733 6734 6735 6736 6737 6739 6741 6743 6745 6747 6751 6752 6753 6754 6755 6756 6757 6758 6759 6761 6763 6764 6765 6766 6768 6769 6770 6772 6773 6774 6776 6777 6778 6779 6780 6781 6782 6783 6786 6787 6788 6789 6790 6793 6794 6795 6796 6797 6799 6800 6801 6803 6804 6805 6806 6808 6811 6812 6814 6817 6818 6819 6821 6822 6825 6826 6827 6831 6832 6835 6836 6837 6838 6843 6844 6846 6847 6849 6851 6852 6856 6859 6863 6865 6866 6867 6870 6872 6874 6875 6876 6877 6878 6880 6881 6882 6885 6886 6889 6890 6891 6892 6893 6894 6895 6896 6898 6899 6900 6902 6903 6904 6905 6907 6908 6909 6910 6913 6915 6920 6921 6922 6923 6924 6925 6926 6928 6929 6930 6932 6933 6935 6936 6937 6938 6940 6941 6942 6943 6945 6946 6947 6948 6949 6954 6958 6962 6963 6964 6965 6966 6967 6968 6969 6970 6972 6973 6974 6975 6976 6977 6978 6979 6980 6981 6982 6983 6985 6986 6987 6989 6990 6993 6994 6995 6996 6997 6999 7001 7002 7003 7004 7005 7006 7008 7010 7011 7013 7015 7016 7019 7020 7023 7024 7027 7028 7032 7033 7035 7036 7037 7039 7040 7043 7046 7047 7048 7050 7051 7052 7053 7054 7055 7056 7059 7060 7063 7066 7068 7069 7071 7074 7077 7078 7079 7082 7083 7085 7086 7087 7091 7094 7096 7097 7098 7099 7101 7102 7103 7105 7107 7109 7112 7115 7117 7118 7120 7121 7122 7123 7124 7125 7127 7129 7130 7131 7134 7135 7139 7141 7142 7143 7144 7145 7146 7147 7148 7149 7151 7152 7153 7154 7155 7156 7157 7158 7159 7162 7163 7164 7165 7169 7171 7172 7173 7174 7175 7176 7177 7178 7179 7180 7182 7183 7184 7185 7186 7187 7189 7190 7191 7192 7193 7194 7195 7197 7199 7201 7203 7204 7205 7206 7207 7208 7209 7211 7214 7215 7219 7221 7222 7223 7224 7225 7226 7227 7228 7229 7230 7231 7232 7234 7239 7240 7244 7245 7247 7248 7249 7250 7251 7252 7253 7257 7259 7260 7261 7264 7265 7266 7267 7268 7269 7270 7271 7273 7274 7275 7276 7277 7280 7282 7283 7284 7285 7287 7288 7290 7294 7302 7305 7306 7307 7308 7309 7310 7311 7313 7314 7316 7318 7320 7321 7322 7323 7326 7327 7328 7329 7330 7331 7332 7333 7334 7335 7337 7338 7340 7341 7342 7343 7344 7345 7350 7352 7353 7354 7357 7358 7359 7361 7363 7364 7365 7366 7367 7368 7370 7371 7372 7373 7374 7375 7377 7378 7380 7381 7383 7386 7387 7388 7389 7390 7391 7393 7394 7395 7397 7398 7399 7400 7402 7403 7404 7405 7406 7407 7408 7409 7410 7412 7413 7414 7416 7417 7419 7420 7421 7423 7424 7426 7427 7428 7429 7430 7431 7432 7433 7434 7435 7436 7437 7438 7439 7440 7441 7442 7443 7444 7445 7446 7448 7449 7451 7452 7456 7458 7459 7460 7461 7463 7464 7465 7466 7467 7468 7469 7474 7479 7480 7483 7484 7485 7486 7487 7488 7490 7497 7498 7499 7501 7502 7503 7504 7505 7506 7507 7509 7510 7512 7514 7515 7516 7518 7519 7520 7522 7526 7528 7530 7531 7532 7533 7535 7539 7540 7543 7546 7547 7548 7550 7552 7553 7554 7556 7557 7560 7563 7564 7567 7568 7571 7573 7574 7577 7579 7581 7583 7584 7585 7588 7589 7590 7591 7592 7593 7594 7595 7596 7597 7598 7600 7602 7603 7604 7606 7608 7610 7612 7613 7614 7615 7616 7617 7618 7620 7621 7622 7623 7624 7626 7627 7628 7630 7631 7632 7633 7635 7636 7638 7639 7640 7641 7642 7643 7644 7645 7646 7647 7649 7651 7652 7654 7656 7657 7658 7659 7660 7661 7662 7663 7664 7666 7667 7668 7669 7670 7671 7672 7673 7674 7675 7678 7679 7680 7681 7682 7683 7684 7685 7686 7687 7690 7691 7692 7693 7695 7696 7697 7698 7699 7700 7702 7703 7704 7705 7706 7707 7708 7710 7713 7714 7716 7717 7718 7720 7721 7723 7724 7725 7726 7727 7729 7732 7735 7736 7739 7740 7741 7744 7745 7746 7748 7749 7750 7753 7754 7756 7757 7758 7762 7763 7764 7765 7766 7767 7768 7770 7771 7772 7775 7776 7777 7778 7779 7781 7783 7784 7787 7790 7791 7792 7795 7796 7797 7800 7802 7803 7804 7805 7810 7813 7816 7817 7818 7819 7820 7821 7823 7824 7825 7827 7828 7829 7831 7832 7836 7842 7844 7846 7847 7849 7850 7852 7853 7854 7856 7858 7859 7861 7862 7863 7865 7866 7867 7870 7871 7873 7874 7875 7876 7877 7878 7879 7881 7883 7885 7886 7887 7888 7889 7890 7891 7892 7893 7895 7896 7897 7898 7900 7901 7902 7904 7905 7906 7907 7908 7909 7910 7912 7913 7914 7915 7916 7917 7919 7920 7921 7922 7923 7926 7928 7929 7930 7932 7933 7934 7936 7940 7941 7942 7943 7944 7945 7948 7949 7950 7951 7954 7955 7956 7957 7958 7959 7963 7964 7965 7967 7968 7970 7971 7972 7976 7977 7979 7980 7981 7982 7986 7989 7991 7992 7994 7995 7997 7998 7999 8000 8001 8002 8003 8004 8005 8006 8010 8012 8013 8016 8017 8019 8020 8021 8022 8024 8025 8026 8028 8029 8030 8032 8033 8035 8036 8039 8040 8041 8045 8047 8048 8050 8051 8052 8053 8054 8056 8057 8058 8063 8064 8065 8066 8067 8068 8069 8070 8071 8072 8076 8078 8079 8080 8082 8084 8087 8091 8092 8093 8094 8095 8096 8097 8099 8100 8101 8102 8103 8104 8105 8106 8107 8108 8109 8110 8113 8114 8115 8116 8118 8119 8120 8121 8122 8123 8125 8126 8127 8129 8130 8131 8132 8133 8134 8135 8136 8137 8139 8140 8142 8143 8145 8146 8147 8148 8149 8150 8151 8152 8155 8158 8159 8160 8161 8162 8165 8167 8170 8171 8174 8175 8178 8180 8181 8183 8184 8185 8186 8188 8189 8191 8192 8193 8194 8197 8199 8200 8204 8205 8206 8209 8210 8212 8213 8214 8215 8216 8217 8218 8220 8221 8224 8226 8227 8229 8233 8234 8235 8236 8238 8240 8242 8243 8244 8245 8248 8249 8251 8253 8254 8256 8257 8258 8259 8262 8264 8266 8267 8268 8269 8270 8271 8272 8274 8275 8276 8277 8278 8279 8280 8281 8282 8283 8285 8287 8288 8292 8293 8296 8298 8299 8300 8301 8302 8303 8304 8306 8307 8308 8313 8314 8315 8316 8317 8318 8320 8322 8324 8325 8327 8328 8329 8330 8331 8332 8333 8334 8335 8337 8338 8339 8340 8341 8342 8344 8345 8348 8349 8350 8351 8353 8354 8355 8357 8358 8359 8360 8363 8364 8366 8367 8368 8369 8370 8371 8372 8373 8374 8375 8377 8378 8379 8380 8381 8382 8383 8384 8385 8386 8387 8390 8391 8392 8393 8394 8395 8396 8397 8398 8399 8400 8401 8402 8403 8404 8405 8406 8407 8409 8411 8413 8414 8415 8416 8417 8418 8420 8421 8422 8423 8424 8425 8426 8427 8429 8430 8431 8435 8437 8439 8440 8441 8442 8443 8445 8447 8448 8450 8452 8454 8455 8458 8459 8460 8461 8462 8463 8464 8465 8466 8467 8469 8470 8471 8473 8474 8475 8477 8478 8479 8480 8481 8482 8483 8484 8485 8486 8489 8491 8492 8493 8495 8496 8497 8499 8500 8501 8502 8503 8505 8506 8507 8508 8509 8513 8514 8517 8518 8520 8521 8523 8524 8525 8526 8527 8528 8531 8532 8534 8535 8536 8538 8539 8540 8543 8545 8546 8548 8549 8550 8551 8557 8560 8561 8563 8564 8566 8567 8569 8571 8573 8576 8577 8579 8580 8581 8583 8584 8585 8586 8587 8588 8590 8591 8593 8594 8596 8597 8599 8600 8601 8602 8603 8605 8606 8607 8608 8610 8611 8612 8614 8615 8616 8617 8618 8619 8620 8621 8622 8623 8624 8627 8628 8629 8631 8634 8635 8636 8638 8641 8642 8643 8644 8645 8648 8649 8651 8654 8657 8658 8660 8662 8667 8669 8670 8671 8673 8675 8676 8677 8679 8681 8682 8683 8687 8689 8690 8691 8692 8695 8698 8699 8700 8701 8704 8705 8706 8707 8710 8711 8714 8716 8717 8718 8719 8720 8724 8727 8729 8732 8737 8738 8740 8742 8743 8744 8745 8746 8747 8749 8750 8751 8752 8753 8754 8755 8756 8757 8759 8760 8763 8767 8769 8771 8778 8780 8782 8783 8785 8786 8787 8789 8791 8792 8793 8794 8798 8799 8802 8804 8805 8806 8807 8811 8812 8815 8819 8820 8821 8823 8824 8825 8826 8827 8829 8831 8832 8833 8834 8836 8841 8842 8843 8844 8845 8846 8847 8849 8850 8851 8852 8853 8854 8855 8856 8857 8858 8861 8862 8863 8865 8866 8869 8870 8871 8872 8873 8876 8877 8878 8880 8881 8882 8883 8884 8887 8890 8891 8892 8893 8894 8895 8896 8897 8898 8899 8900 8901 8903 8904 8906 8907 8909 8912 8914 8915 8920 8922 8926 8927 8929 8931 8932 8934 8935 8937 8939 8940 8941 8943 8944 8946 8947 8948 8949 8950 8952 8953 8957 8959 8960 8961 8962 8963 8964 8965 8968 8969 8970 8972 8974 8975 8976 8978 8979 8980 8982 8983 8984 8985 8986 8987 8988 8990 8991 8992 8993 8994 8995 8996 8997 8998 8999 9001 9003 9004 9005 9006 9008 9009 9010 9011 9012 9013 9014 9016 9017 9018 9019 9022 9026 9027 9029 9030 9031 9033 9035 9036 9037 9040 9041 9043 9044 9045 9046 9047 9050 9052 9053 9054 9055 9056 9057 9058 9059 9060 9061 9062 9063 9064 9065 9066 9067 9068 9069 9070 9072 9073 9075 9077 9081 9083 9084 9085 9086 9088 9089 9090 9111 9112 9113 9114 9115 9120 9121 9124 9125 9126 9127 9128 9129 9130 9133 9134 9135 9136 9137 9139 9142 9143 9145 9148 9149 9150 9152 9153 9154 9158 9159 9160 9161 9163 9164 9165 9166 9168 9170 9171 9172 9174 9175 9176 9177 9178 9179 9181 9182 9183 9184 9187 9188 9189 9190 9259 9260 9264 9265 9268 9269 9270 9271 9273 9276 9277 9278 9280 9281 9282 9283 9284 9289 9290 9292 9294 9295 9296 9297 9298 9302 9303 9304 9305 9308 9309 9310 9313 9314 9315 9318 9320 9322 9324 9325 9330 9331 9337 9338 9339 9341 9342 9343 9344 9345 9348 9349 9351 9353 9356 9358 9359 9363 9365 9368 9369 9370 9377 9378 9380 9383 9385 9386 9387 9388 9391 9393 9395 9396 9397 9398 9401 9403 9404 9405 9407 9408 9409 9412 9413 9415 9417 9421 9422 9426 9427 9428 9429 9431 9433 9434 9438 9439 9440 9441 9442 9444 9446 9447 9448 9451 9452 9454 9455 9458 9459 9461 9462 9463 9464 9466 9468 9470 9472 9473 9475 9476 9477 9479 9482 9483 9486 9489 9491 9492 9493 9494 9495 9496 9502 9506 9512 9513 9514 9515 9516 9517 9518 9519 9520 9524 9525 9526 9528 9532 9533 9536 9538 9539 9540 9542 9544 9546 9547 9548 9549 9550 9552 9553 9554 9555 9557 9559 9560 9564 9566 9568 9569 9570 9571 9573 9574 9578 9579 9580 9581 9582 9583 9584 9585 9588 9589 9591 9593 9594 9597 9598 9599 9600 9601 9603 9605 9612 9614 9616 9618 9620 9621 9622 9623 9626 9628 9631 9635 9638 9639 9643 9644 9646 9647 9648 9650 9653 9656 9658 9660 9661 9664 9665 9666 9667 9670 9671 9672 9673 9677 9678 9679 9680 9682 9688 9689 9692 9697 9698 9699 9700 9702 9703 9704 9705 9706 9707 9710 9711 9712 9716 9718 9719 9720 9721 9722 9723 9724 9728 9729 9731 9732 9733 9736 9737 9738 9739 9742 9744 9745 9747 9750 9751 9753 9754 9755 9756 9759 9760 9762 9763 9764 9765 9767 9771 9772 9773 9774 9775 9776 9777 9778 9780 9781 9782 9784 9785 9786 9791 9793 9795 9798 9799 9805 9806 9809 9810 9812 9813 9814 9816 9818 9819 9820 9825 9828 9829 9830 9831 9832 9835 9837 9842 9844 9847 9849 9850 9852 9853 9855 9856 9857 9858 9860 9862 9865 9867 9868 9872 9873 9877 9878 9880 9881 9883 9885 9889 9891 9893 9894 9895 9896 9897 9899 9901 9902 9903 9905 9912 9913 9914 9917 9918 9919 9921 9923 9924 9925 9926 9927 9928 9929 9930 9932 9934 9935 9936 9937 9938 9939 9940 9941 9942 9943 9944 9945 9946 9947 9949 9950 9953 9955 9956 9960 9962 9963 9964 9965 9966 9967 9969 9970 9971 9974 9976 9977 9978 9980 9981 9982 9984 9985 9986 9987 9988 9989 9990 9991 9992 9996 9998 9999 10000 10002 10003 10004 10005 10006 10007 10008 10009 10010 10011 10012 10016 10017 10018 10019 10020 10023 10024 10025 10027 10028 10029 10031 10032 10034 10035 10037 10038 10039 10041 10042 10045 10047 10048 10049 10050 10052 10054 10055 10057 10058 10062 10063 10064 10065 10067 10068 10075 10079 10080 10083 10086 10089 10091 10092 10097 10099 10101 10103 10104 10105 10106 10109 10110 10111 10112 10113 10114 10115 10116 10117 10119 10120 10121 10123 10125 10128 10129 10132 10133 10134 10135 10136 10137 10138 10139 10141 10142 10145 10147 10150 10152 10153 10157 10158 10159 10162 10164 10171 10172 10174 10176 10177 10178 10179 10180 10181 10182 10183 10184 10185 10187 10188 10189 10190 10191 10192 10193 10194 10196 10198 10199 10200 10201 10202 10204 10205 10206 10208 10210 10211 10212 10215 10216 10217 10218 10219 10221 10222 10223 10224 10226 10227 10228 10229 10230 10231 10233 10234 10236 10237 10239 10240 10241 10242 10244 10245 10247 10248 10249 10250 10254 10256 10257 10258 10259 10260 10263 10264 10266 10268 10269 10270 10273 10276 10277 10278 10279 10280 10281 10282 10283 10284 10286 10287 10288 10289 10290 10292 10294 10297 10298 10299 10300 10301 10302 10303 10304 10305 10306 10309 10310 10311 10312 10313 10314 10318 10319 10320 10321 10322 10329 10330 10331 10332 10333 10334 10335 10336 10337 10338 10339 10340 10344 10345 10346 10347 10349 10350 10351 10352 10353 10354 10355 10356 10357 10358 10360 10363 10368 10369 10372 10373 10377 10384 10386 10387 10388 10392 10393 10394 10397 10398 10399 10402 10403 10404 10406 10408 10410 10411 10412 10413 10414 10418 10420 10423 10426 10427 10431 10432 10433 10438 10441 10442 10443 10445 10448 10450 10451 10452 10457 10458 10460 10461 10462 10463 10468 10471 10473 10475 10476 10479 10480 10481 10482 10487 10490 10491 10492 10493 10494 10496 10497 10498 10500 10501 10502 10503 10508 10509 10511 10512 10514 10521 10524 10527 10529 10533 10534 10536 10537 10540 10555 10556 10557 10558 10559 10565 10566 10567 10570 10572 10573 10577 10578 10580 10582 10583 10584 10587 10589 10591 10592 10595 10596 10597 10598 10600 10606 10610 10612 10613 10614 10616 10617 10618 10619 10621 10622 10630 10631 10641 10644 10645 10646 10653 10654 10655 10657 10660 10665 10667 10669 10670 10672 10673 10676 10678 10679 10680 10681 10682 10683 10684 10685 10687 10689 10690 10692 10693 10694 10695 10696 10697 10700 10701 10702 10703 10704 10705 10706 10708 10709 10712 10713 10714 10717 10718 10719 10720 10721 10722 10724 10725 10726 10727 10728 10729 10730 10731 10735 10737 10738 10740 10742 10743 10744 10745 10746 10747 10752 10754 10755 10756 10757 10759 10762 10768 10769 10770 10771 10772 10775 10776 10777 10779 10780 10782 10785 10786 10787 10788 10789 10790 10792 10794 10795 10797 10799 10801 10802 10804 10806 10807 10809 10810 10811 10813 10814 10815 10816 10818 10819 10820 10823 10829 10830 10831 10832 10833 10834 10835 10836 10839 10841 10844 10845 10846 10849 10850 10854 10855 10856 10858 10860 10863 10864 10866 10868 10869 10870 10871 10872 10875 10876 10878 10880 10881 10882 10883 10884 10886 10887 10888 10889 10891 10892 10893 10894 10896 10898 10901 10904 10906 10908 10909 10911 10912 10915 10917 10920 10921 10924 10925 10926 10929 10930 10932 10933 10935 10936 10938 10939 10941 10945 10947 10948 10950 10952 10953 10957 10958 10963 10965 10966 10969 10974 10977 10980 10983 10984 10985 10991 10993 10995 10999 11001 11006 11007 11009 11010 11014 11016 11017 11019 11020 11022 11024 11027 11030 11034 11036 11037 11041 11042 11043 11044 11045 11054 11055 11059 11060 11061 11062 11067 11068 11072 11073 11076 11077 11078 11082 11087 11091 11094 11096 11097 11098 11100 11102 11104 11105 11107 11110 11111 11113 11114 11116 11117 11118 11119 11121 11126 11129 11130 11134 11135 11136 11137 11139 11147 11148 11150 11154 11156 11160 11163 11164 11167 11169 11171 11172 11173 11174 11177 11178 11180 11183 11184 11185 11186 11192 11194 11195 11197 11198 11199 11201 11203 11204 11205 11208 11210 11211 11214 11216 11217 11219 11222 11223 11224 11226 11227 11230 11233 11238 11239 11240 11243 11248 11250 11252 11253 11255 11257 11261 11262 11264 11266 11267 11268 11273 11274 11277 11279 11282 11285 11287 11288 11290 11292 11293 11294 11296 11297 11299 11301 11302 11308 11309 11310 11311 11312 11313 11314 11315 11316 11317 11319 11320 11321 11322 11323 11330 11331 11332 11334 11337 11338 11339 11340 11341 11344 11346 11347 11348 11349 11350 11351 11352 11353 11354 11356 11357 11358 11359 11360 11362 11363 11364 11365 11366 11368 11369 11371 11373 11374 11375 11380 11383 11384 11386 11387 11388 11389 11391 11392 11393 11394 11395 11398 11399 11400 11401 11402 11403 11408 11409 11413 11416 11420 11421 11424 11428 11431 11438 11440 11441 11449 11450 11454 11457 11461 11464 11465 11469 11472 11480 11485 11498 11508 11513 11517 11519 11521 11523 11525 11534 11537 11539 11540 11542 11549 11551 11552 11553 11554 11555 11560 11563 11566 11817 11821 11822 11829 11837 11838 11842 11843 11845 11847 11848 11852 11860 11862 11863 11864 11868 11873 11875 11876 11878 11880 11891 11899 11905 11908 11911 11914 11917 11919 11920 11926 11927 11928 11929 11930 11931 11935 11938 11942 11947 11949 11951 11955 11956 11959 11968 11969 11970 11976 11978 11979 11980 11984 12001 12052 12053 12055 12057 12059 12066 12069 12070 12071 12072 12076 12077 12081 12093 12097 12119 12127 12138 12140 12146 12152 12172 12174 12178 12185 12189 12190 12191 12196 12200 12201 12202 12206 12227 12232 12235 12237 12238 12239 12240 12241 12243 12253 12263 12264 12265 12266 12268 12270 12271 12272 12276 12280 12282 12285 12287 12288 12289 12292 12293 12294 12295 12296 12298 12302 12304 12312 12314 1032M 1040M 1043M 1045M 1046M 1047M 1050M 1051M 1053M 1056M 1063M 1064M 1067M 1068M 1071M 1072M 1073M 1074M 1078M 1079M 1085M 1087M 1091M 1092M 1093M 1094M 1096M 1097M 1099M 1100M 1101M 1102M 1105M 1106M 1107M 1109M 1112M 1114M 1115M 1116M 1118M 1121M 1123M 1124M 1127M 1129M 1133M 1134M 1135M 1136M 1139M 1140M 1141M 1143M 1145M 1146M 1147M 1149M 1150M 1151M 1152M 1154M 1155M 1157M 1158M 1159M 1160M 1162M 1163M 1164M 1167M 1172M 1173M 1174M 1175M 1176M 1178M 1181M 1182M 1183M 1185M 1186M 1187M 1188M 1191M 1192M 1194M 1196M 1197M 1198M 1200M 1203M 1204M 1205M 1208M 1209M 1212M 1216M 1218M 1219M 1220M 1221M 1223M 1224M 1226M 1228M 1230M 1233M 1235M 1236M 1238M 1239M 1240M 1242M 1246M 1250M 1252M 1254M 1256M 1258M 1259M 1261M 1263M 1264M 1266M 1268M 1269M 1270M 1271M 1273M 1274M 1275M 1277M 1278M 1281M 1282M 1284M 1285M 1288M 1291M 1293M 1294M 1295M 1298M 1299M 1300M 1306M 1310M 1312M 1314M 1315M 1319M 1321M 1322M 1324M 1330M 1332M 1333M 1335M 1336M 1337M 1338M 1339M 1340M 1341M 1343M 1344M 1345M 1346M 1348M 1350M 1351M 1355M 1356M 1357M 1358M 1359M 1360M 1361M 1362M 1363M 1367M 1368M 1370M 1372M 1375M 1376M 1677M 1680M 1685M 1699M 1702M 1705M 1707M 1713M 1068 1096 1174 1589 3005 3120 3274 3309 3554 3724 3778 3837 3884 4215 4400 4500 4876 4928 4955 4991 4995 5023 5382 5469 5519 5528 5768 5774 5799 5936 5989 6527 6565 6571 6574 6582 6603 6666 6951 7022 7061 7062 7084 7297 7336 7471 7482 7496 7508 7811 7872 8081 8112 8263 8412 8530 8547 8574 8632 8702 8908 8917 9042 9049 9155 9275 9312 9402 9410 9727 9815 9997 10093 10102 10127 10235 10238 10370 10401 10585 11307 11448 11468 11504 11511 11561 11869 11887 11973 11983 11994 12084 12134 12137 12262 12275 12278 1039M 1080M 1130M 1165M 1241M 1301M 1378M 1714M 6484 10417 10419 10421 10428 10430 10440 10446 10513 10515 10516 10517 10519 10523 10525 10526 10532 10535 10542 10543 10549 10551 10552 10569 10581 10599 10608 10633 10634 10638 10662 10663 11618 11715 11812 12135 12136 12165 12167 12369 12370 12389 12406 12407 12409 12410 12413 12414 12423 12424 12440 12453 12466 12468 12470 12471 12479 12493 12499 12504 12506 4293 8770 3969 5038 5106 5218 5533 5564 5681 5967 6020 6385 6400 6440 6548 6569 6616 6662 6664 6715 6718 6738 6816 6855 6871 6917 7138 7188 7198 7235 7289 7299 7360 7376 7415 7470 7521 7538 7773 7809 7857 7864 7869 7882 7931 7938 8037 8089 8144 8166 8361 8451 8488 8542 8598 8723 8888 9076 9080 9157 9263 9286 9334 9362 9371 9411 9562 9572 9577 9630 9659 9675 9741 9792 9826 9890 10538 10571 12125 11025 4751 8153 5115 7292 7392 7396 7453 7455 7689 7894 7899 7903 7925 8015 8187 8190 8260 8261 8297 8570 8731 8733 8765 8768 8879 8918 8971 9274 9437 9511 9521 9523 9586 9633 9636 9652 9685 9686 9690 9693 9740 9748 9757 9779 9796 9823 9908 9954 9959 10014 10043 10046 10051 10056 10060 10081 10082 10088 10232 10243 10252 10253 10453 10507 10686 10763 10767 10865 10907 10949 10955 10982 11004 11015 11433 11436 11442 11446 11460 11478 11486 11535 11567 11569 11572 11573 11578 11582 11583 11584 11585 11586 11587 11588 11592 11593 11594 11598 11599 11601 11607 11609 11614 11615 11617 11619 11620 11622 11623 11625 11626 11627 11631 11632 11636 11644 11645 11649 11650 11651 11652 11655 11656 11659 11661 11662 11664 11665 11666 11667 11668 11669 11671 11673 11675 11676 11677 11681 11682 11683 11685 11686 11687 11688 11691 11692 11696 11697 11699 11700 11706 11712 11713 11714 11716 11717 11718 11719 11720 11721 11722 11723 11724 11725 11726 11727 11730 11731 11733 11734 11735 11736 11737 11738 11739 11740 11741 11742 11743 11744 11745 11746 11748 11749 11750 11751 11752 11753 11755 11756 11757 11758 11764 11765 11769 11770 11772 11781 11784 11794 11796 11805 11811 11813 11814 11816 11818 11819 11828 11832 11836 11839 11841 11849 11857 11858 11859 11861 11866 11870 11874 11877 11879 11881 11882 11884 11885 11886 11888 11889 11890 11900 11906 11921 11939 11941 11945 11946 11948 11952 11958 11961 11972 11988 11997 12010 12030 12047 12063 12067 12073 12078 12079 12080 12083 12087 12090 12091 12095 12101 12102 12103 12104 12105 12108 12111 12114 12115 12118 12121 12123 12128 12129 12130 12143 12144 12147 12150 12151 12153 12155 12156 12157 12158 12159 12163 12166 12179 12180 12181 12183 12184 12192 12195 12198 12204 12205 12210 12211 12215 12219 12221 12222 12223 12226 12228 12230 12233 12236 12244 12247 12248 12250 12254 12255 12258 12321 12322 12324 12331 12335 12340 12342 12344 12345 12346 12347 12348 12349 12351 12352 12356 12357 12366 12367 4114 4962 5008 5321 5355 5503 5557 5734 5811 5883 5911 6178 6195 6234 6303 6458 6714 6792 6810 6823 7788 7993 8762 8772 8859 9300 9558 9717 9766 9794 9804 9824 3908 4379 5506 6357 6407 6456 6490 6505 6657 6721 6749 6873 7752 8074 8077 8433 8625 9147 9424 9541 9808 9875 9904 and 9916",
      "firm_fei_number": "1000220815",
      "recalling_firm": "Advanced Bionics Corp",
      "address_1": "25129 Rye Canyon Loop",
      "city": "Valencia",
      "state": "CA",
      "postal_code": "91355",
      "additional_info_contact": "866-566-8913",
      "reason_for_recall": "Since April 2004, 27 of 8,769 (0.31%) patients have reported receiving second degree burns in the area of charging and 3 of 8,769 (0.03%) patients have reported receiving third degree burns in the area of charging while using Charger 1 .0. Some of these events were the direct result of the patient disregarding the instructions for use, such as sleeping with the Charger on or placing it directly on",
      "root_cause_description": "Other",
      "action": "In the United States, notification will be sent to patients with the affected devices. Additionally, their physicians on record will be provided with the notification.  The Company is currently planning to initiate this field action on September 22, 2008. The U.S. notification will be sent via Fed Ex or U.S. Postal Service certified mail.  The letter advises that Boston Scientific Neuromodulation is voluntarily recalling its first generation charging device, Precision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System.  The Charger 1.0 devices are being recalled and will be replaced with the new second generation Precision Charger 2.0 (SC-5312, currently in distribution) that features a temperature limiter for improved temperature management. The Charger 1.0 can be used until replacement with Charger 2.0 by following the instructions on Page 2 of the letter.  Customers are asked to complete the acknowledgement form at the end of the letter and return it to Boston Scientific in the return envelope provided. A representative from Boston Scientific will contact them upon receiving the completed form to schedule an in-person visit at their physician's office to exchange the device.  Contact Boston Scientific Technical Serives at (866) 566-8913 for assistance.",
      "product_quantity": "6549 units",
      "distribution_pattern": "Nationwide and Internationally - Ireland, Korea, Mexico, The Netherlands, United Kingdom, Canada, Australia and Israel",
      "openfda": {
        "pma_number": [
          "P130022",
          "P130028",
          "P800040",
          "P030017",
          "P010032",
          "P190002",
          "P210037",
          "P840001"
        ],
        "registration_number": [
          "3004593495",
          "3019807891",
          "3008514029",
          "9681477",
          "3010052483",
          "2024311",
          "9610139",
          "2032521",
          "2030624",
          "1035166",
          "3018094310",
          "2182291",
          "2649622",
          "3014507028",
          "3007882731",
          "2183744",
          "3003120897",
          "3010086866",
          "3006630150",
          "3008403546",
          "3013973539",
          "2648729",
          "3003477135",
          "1627487",
          "3009380063",
          "3008988055",
          "1928237",
          "2133810",
          "3015173212",
          "3013023969",
          "3020969413",
          "2133641",
          "3015924373",
          "3004209178",
          "3010326622",
          "3007616088",
          "3038429246",
          "2134604",
          "2182207",
          "3003335547",
          "3016618265",
          "3005210579",
          "3021836309",
          "3006705815",
          "9616693",
          "3013368220",
          "2124215",
          "3027063190",
          "2135338",
          "3012693387",
          "3004519921",
          "3015200795",
          "3010972758",
          "2133772",
          "3010531069",
          "3010966701",
          "3005160949"
        ],
        "fei_number": [
          "3004593495",
          "3019807891",
          "3008514029",
          "3010052483",
          "3004369318",
          "2024311",
          "1035166",
          "3018094310",
          "2182291",
          "2649622",
          "3014507028",
          "3007882731",
          "2183744",
          "3003120897",
          "3010086866",
          "3005846326",
          "3006630150",
          "3008403546",
          "3013973539",
          "2648729",
          "3003477135",
          "1627487",
          "3009380063",
          "3008988055",
          "1928237",
          "3003574398",
          "3015173212",
          "3013023969",
          "3020969413",
          "3000203466",
          "3015924373",
          "3000203066",
          "3010326622",
          "3002807439",
          "3038429246",
          "3007616088",
          "2182207",
          "3003335547",
          "3016618265",
          "3005210579",
          "3021836309",
          "3006705815",
          "3005700659",
          "3013368220",
          "3002623849",
          "2124215",
          "3002618605",
          "3027063190",
          "3012693387",
          "3000209794",
          "3015200795",
          "3002806500",
          "3000206435",
          "3010972758",
          "3010531069",
          "3010966701",
          "3005160949"
        ],
        "device_name": "Stimulator, Spinal-Cord, Totally Implanted For Pain Relief",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3"
      }
    }
  ]
}