  "results": [
    {
      "cfres_id": "73582",
      "product_res_number": "Z-0080-2009",
      "event_date_initiated": "2008-08-29",
      "event_date_posted": "2008-10-10",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-04-06",
      "res_event_number": "49405",
      "product_code": "FGE",
      "k_numbers": [
        "K053404"
      ],
      "product_description": "Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards Lifesciences LLC  Irvine, CA 92614, Model EX100601CD.  The LifeStent FlexStar self-expanding biliary stent system is intended for use in the palliation of malignant strictures (neoplasms) in the biliary tree.",
      "code_info": "Lot Numbers: R58285865, R58312972, 58337682, 58415748, FR6J0229, and RFR6K0303",
      "firm_fei_number": "2020394",
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "address_1": "1625 W 3rd St",
      "city": "Tempe",
      "state": "AZ",
      "postal_code": "85281-2438",
      "additional_info_contact": "480-894-9515",
      "reason_for_recall": "Some LifeStent FlexStar Systems may exhibit a gap between the tip of\nthe delivery system and the primary sheath such that the guidewire lumen is visible.",
      "root_cause_description": "Process control",
      "action": "The customer letters for the tip to sheath gap were sent on 8/29/2008 with return receipt.  The letter advises that C.R. Bard has determined that some LifeStent FlexStar Systems may exhibit a gap between the tip of\nthe delivery system and the primary sheath such that the guidewire lumen is visible.  Per the letter, the firm's analysis has determined that a gap between the tip of the delivery system and the delivery system sheath may be associated with difficult advancement to the treatment site.  Users are instructed to inspect all devices to verify that there is not a gap between the tip and sheath (orange colored guidewire lumen is visible), prior to use and to not use any devices that exhibit a tip to sheath gap condition. If devices that exhibit a tip to sheath gap are identified, customers are advised to call Customer Service  at 1-800-321-4254 to return the device and a replacement will be provided at no cost.  The letter also advises that the Instructions for Use for the LifeStent FlexStar System will be updated to include the inspection step which is provided in the letter.",
      "product_quantity": "281 units",
      "distribution_pattern": "Nationwide",
      "openfda": {
        "k_number": [
          "K851956",
          "K112543",
          "K980191",
          "K913932",
          "K990592",
          "K993588",
          "K821817",
          "K003450",
          "K912122",
          "K033497",
          "K024303",
          "K844384",
          "K020682",
          "K980372",
          "K014007",
          "K900904",
          "K920236",
          "K910493",
          "K072301",
          "K002650",
          "K000564",
          "K210419",
          "K963397",
          "K032530",
          "K003186",
          "K823428",
          "K974815",
          "K910671",
          "K000550",
          "K041796",
          "K033134",
          "K924608",
          "K083627",
          "K992318",
          "K990854",
          "K896504",
          "K042518",
          "K021630",
          "K152524",
          "K894871",
          "K033834",
          "K032457",
          "K965147",
          "K892383",
          "K033212",
          "K251123",
          "K014205",
          "K963691",
          "K033394",
          "K003753",
          "K172044",
          "K011881",
          "K030409",
          "K992755",
          "K000744",
          "K020809",
          "K061231",
          "K053459",
          "K103478",
          "K992816",
          "K896518",
          "K860990",
          "K163169",
          "K112522",
          "K130591",
          "K973114",
          "K023666",
          "K083374",
          "K954043",
          "K014215",
          "K223469",
          "K955728",
          "K851954",
          "K945498",
          "K903420",
          "K130981",
          "K050725",
          "K050627",
          "K053245",
          "K111149",
          "K012347",
          "K914495",
          "K032137",
          "K924107",
          "K011039",
          "K024048",
          "K834574",
          "K020788",
          "K032768",
          "K032436",
          "K251229",
          "K012056",
          "K012090",
          "K241801",
          "K102519",
          "K140630",
          "K963767",
          "K924451",
          "K940370",
          "K851960",
          "K030839",
          "K900923",
          "K834413",
          "K003929",
          "K020528",
          "K251658",
          "K052539",
          "K905738",
          "K040151",
          "K982005",
          "K050502",
          "K905229",
          "K873954",
          "K001648",
          "K232596",
          "K021898",
          "K853666",
          "K122072",
          "K141335",
          "K884714",
          "K061235",
          "K112994",
          "K011261",
          "K023243",
          "K925406",
          "K073667",
          "K050534",
          "K023870",
          "K202973",
          "K040800",
          "K170740",
          "K923096",
          "K994156",
          "K970052",
          "K000308",
          "K961216",
          "K012087",
          "K961262",
          "K060624",
          "K861220",
          "K896069",
          "K010684",
          "K931025",
          "K042204",
          "K253832",
          "K993892",
          "K913073",
          "K872509",
          "K990808",
          "K093392",
          "K101314",
          "K040930",
          "K152069",
          "K242950",
          "K170743",
          "K962899",
          "K180868",
          "K221071",
          "K002101",
          "K060057",
          "K032025",
          "K171623",
          "K122546",
          "K033862",
          "K863437",
          "K032360",
          "K031186",
          "K990130",
          "K896503",
          "K060487",
          "K915813",
          "K901431",
          "K101095",
          "K050777",
          "K253987",
          "K060817",
          "K862059",
          "K920361",
          "K904669",
          "K143303",
          "K051379",
          "K894867",
          "K914882",
          "K014136",
          "K260835",
          "K230127",
          "K902107",
          "K061566",
          "K902734",
          "K010549",
          "K896376",
          "K040413",
          "K023308",
          "K983008",
          "K991028",
          "K914699",
          "K914881",
          "K011806",
          "K973171",
          "K993091",
          "K011184",
          "K102082",
          "K002235",
          "K003365",
          "K020052",
          "K012066",
          "K980823",
          "K072708",
          "K991533",
          "K953827",
          "K010601",
          "K910768",
          "K060877",
          "K003040",
          "K030053",
          "K011506",
          "K063481",
          "K111627",
          "K993232",
          "K842147",
          "K905707",
          "K910006",
          "K060023",
          "K040653",
          "K955413",
          "K912288",
          "K133700",
          "K172057",
          "K242845",
          "K851647",
          "K993145",
          "K121832",
          "K983154",
          "K941350",
          "K081733",
          "K896323",
          "K991032",
          "K200103",
          "K023121",
          "K033909",
          "K062798",
          "K900205",
          "K050698",
          "K011909",
          "K030645",
          "K121125",
          "K181669",
          "K203162",
          "K972825",
          "K954713",
          "K861291",
          "K072720",
          "K944912",
          "K062264",
          "K834468",
          "K031821",
          "K983075",
          "K052114",
          "K111315",
          "K253132",
          "K101786",
          "K173035",
          "K211021",
          "K980290",
          "K172520",
          "K993303",
          "K982508",
          "K963254",
          "K924387",
          "K041370",
          "K010242",
          "K011817",
          "K001222",
          "K855142",
          "K103353",
          "K912971",
          "K152607",
          "K880343",
          "K223592",
          "K052132",
          "K012822",
          "K001843",
          "K971295",
          "K003997",
          "K896163",
          "K041606",
          "K911292",
          "K010445",
          "K002143",
          "K232245",
          "K003793",
          "K012752",
          "K980653",
          "K896423",
          "K885180",
          "K973113",
          "K152853",
          "K222663",
          "K962925",
          "K000864",
          "K935094",
          "K033314",
          "K162404",
          "K853543",
          "K913841",
          "K943143",
          "K014184",
          "K872458",
          "K251291",
          "K093003",
          "K040550",
          "K213946",
          "K182980",
          "K021824",
          "K130484",
          "K955110",
          "K833937",
          "K041032",
          "K071646",
          "K202477",
          "K991674",
          "K023217",
          "K130792",
          "K131201",
          "K991793",
          "K030633",
          "K993646",
          "K222627",
          "K021672",
          "K141408",
          "K903858",
          "K032206",
          "K050501",
          "K081996",
          "K023248",
          "K896506",
          "K002133",
          "K253393",
          "K905714",
          "K863957",
          "K133110",
          "K913986",
          "K042985",
          "K112701",
          "K901582",
          "K992569",
          "K901432",
          "K063733",
          "K833920",
          "K060557",
          "K093893",
          "K063532",
          "K151671",
          "K031777",
          "K033537",
          "K010107",
          "K031360",
          "K040505",
          "K030638",
          "K851242",
          "K944143",
          "K120452",
          "K910931",
          "K250409",
          "K021994",
          "K161203",
          "K901433",
          "K964119",
          "K001258",
          "K921654",
          "K053404",
          "K110122",
          "K021729",
          "K972583",
          "K990867",
          "K111766",
          "K040027",
          "K163018",
          "K141597",
          "K033333",
          "K941973",
          "K013737",
          "K171223",
          "K022026",
          "K081184",
          "K251019",
          "K090133",
          "K993392",
          "K032580",
          "K012993",
          "K012216",
          "K990631",
          "K910219",
          "K863571",
          "K994068",
          "K901427",
          "K991929",
          "K031197",
          "K012590",
          "K081956",
          "K113582",
          "K030107",
          "K010714",
          "K821819",
          "K151925",
          "K093947",
          "K973008",
          "K043481",
          "K042969",
          "K923993",
          "K830483",
          "K001338",
          "K834351",
          "K852293",
          "K021345",
          "K031331",
          "K953332",
          "K193600",
          "K080037",
          "K011080",
          "K934377",
          "K000793",
          "K911581",
          "K955126",
          "K032929",
          "K003336",
          "K180418",
          "K171548",
          "K964688",
          "K896391",
          "K033348",
          "K082093",
          "K882610",
          "K053454",
          "K940965",
          "K150487",
          "K892384",
          "K070610",
          "K972582",
          "K933200",
          "K930623",
          "K041962",
          "K162226",
          "K140760",
          "K992319",
          "K001224",
          "K040667",
          "K233079",
          "K121430",
          "K982346",
          "K092352",
          "K043243",
          "K921191",
          "K050103",
          "K223354",
          "K911630",
          "K010411",
          "K002130",
          "K963782",
          "K033393",
          "K180086",
          "K914277",
          "K223626",
          "K171809",
          "K073672",
          "K041423",
          "K240464",
          "K993904",
          "K143609",
          "K241888",
          "K183396",
          "K051124",
          "K851962",
          "K243412",
          "K051785",
          "K873606"
        ],
        "registration_number": [
          "",
          "3021009035",
          "3004837686",
          "3009018440",
          "3005099803",
          "1058584",
          "3027260597",
          "3010419931",
          "3003902943",
          "2124215",
          "3013152301",
          "3013247477",
          "2027111",
          "3011050570",
          "3014328942",
          "3006950086",
          "3009763019",
          "3013460719",
          "1828132",
          "3014222368",
          "3016678045",
          "3013440400",
          "9681442",
          "3002807314",
          "3014769442",
          "3004207966",
          "3006260740",
          "1625425",
          "3011347300",
          "1319211",
          "2183870",
          "3003910212",
          "2011171",
          "1000121056",
          "3014246639",
          "3010155661",
          "1225687",
          "1450395",
          "1644312",
          "3003898360",
          "3013337852",
          "3004111573",
          "9616684",
          "3011642792",
          "3015309643",
          "9614641",
          "3009526575",
          "3040505963",
          "3000247987",
          "3005089785",
          "3012480535",
          "3004091615",
          "3003790304",
          "3040096051",
          "3010288205",
          "2024024",
          "9680809",
          "3004859241",
          "1220477",
          "1724474",
          "3001845648",
          "3009906660",
          "1320894",
          "3003775027",
          "2134265",
          "1219930",
          "3005941719",
          "3008040402",
          "3011458193",
          "3030733800",
          "3009051888",
          "3006891655",
          "2648045",
          "3021632375",
          "1313046",
          "9710641",
          "1000523114",
          "3011146996",
          "1528319",
          "3032916632",
          "1721504",
          "3031276312",
          "1016427",
          "3007695959",
          "1721676",
          "2024168",
          "2020394",
          "3004129994",
          "2126666",
          "3011163668",
          "1061124",
          "3013557562",
          "1000393132",
          "1223925",
          "1222168",
          "1820334",
          "1531050",
          "3030447506",
          "1018233",
          "3008146331",
          "3005987240",
          "3013526170",
          "2032098",
          "3033589330",
          "1643817",
          "3012536737",
          "3035451914",
          "1317547",
          "3011689956",
          "3007146453",
          "1037905",
          "3005580113",
          "3010220595",
          "2029386",
          "2246552",
          "1417485",
          "3006082230",
          "9616099",
          "3015225571",
          "3013162291",
          "1928237",
          "3006131984",
          "2183744",
          "1319639",
          "9681477",
          "3010173425",
          "3002807090",
          "2939561",
          "3009121749",
          "1220592",
          "3008492676",
          "3010273872",
          "3030446844",
          "3016463905",
          "2030624",
          "3031230150",
          "9681260",
          "1123137",
          "9680794",
          "3011137372",
          "8010047",
          "2528981",
          "3014579161",
          "3010532612",
          "9710284",
          "3014279513",
          "3002504068"
        ],
        "fei_number": [
          "",
          "3002807795",
          "1000306647",
          "3021009035",
          "1000138054",
          "1000219637",
          "3005747797",
          "3004837686",
          "3003565456",
          "3009018440",
          "3005099803",
          "3001451463",
          "3027260597",
          "3003528016",
          "3010419931",
          "3003902943",
          "2124215",
          "3013152301",
          "3013247477",
          "2027111",
          "3011050570",
          "3014328942",
          "3006950086",
          "3009763019",
          "3013460719",
          "1828132",
          "3014222368",
          "3016678045",
          "3000206435",
          "3013440400",
          "3002807314",
          "3014769442",
          "3002807036",
          "3004207966",
          "3006260740",
          "3002807111",
          "1625425",
          "3011347300",
          "1319211",
          "2183870",
          "3003910212",
          "2011171",
          "1000121056",
          "3014246639",
          "3010155661",
          "1450395",
          "3003898360",
          "3013337852",
          "3004111573",
          "3011642792",
          "3015309643",
          "3009526575",
          "3040505963",
          "3002806565",
          "3000247987",
          "1000305720",
          "3012480535",
          "3005089785",
          "3004091615",
          "3003790304",
          "3040096051",
          "3010288205",
          "2024024",
          "3004859241",
          "3002755653",
          "1220477",
          "3001845648",
          "3004635528",
          "3003058448",
          "3009906660",
          "3003775027",
          "1219930",
          "3008040402",
          "3011458193",
          "3030733800",
          "3009051888",
          "3006891655",
          "2648045",
          "3021632375",
          "1313046",
          "1000523114",
          "3011146996",
          "3002806368",
          "3032916632",
          "3002095335",
          "1721504",
          "3031276312",
          "1016427",
          "3002913962",
          "3007695959",
          "1721676",
          "2024168",
          "3003737899",
          "3006003617",
          "1000512168",
          "3010532612",
          "2020394",
          "3004129994",
          "2126666",
          "3011163668",
          "3013557562",
          "1000393132",
          "1222168",
          "1820334",
          "3030447506",
          "3003995201",
          "1000117748",
          "1018233",
          "3008146331",
          "3005987240",
          "3013526170",
          "3003084171",
          "3033589330",
          "1643817",
          "3012536737",
          "3035451914",
          "1317547",
          "3011689956",
          "3007146453",
          "1037905",
          "3005580113",
          "3010220595",
          "2246552",
          "1417485",
          "3006082230",
          "3002807439",
          "1000122656",
          "3015225571",
          "3013162291",
          "3003094851",
          "1928237",
          "2183744",
          "3006131984",
          "3009121749",
          "1220592",
          "3008492676",
          "3010273872",
          "3030446844",
          "1000123180",
          "3002995060",
          "3016463905",
          "3031230150",
          "2528981",
          "3014579161",
          "3001451571",
          "3014279513",
          "3002504068"
        ],
        "device_name": "Stents, Drains And Dilators For The Biliary Ducts",
        "medical_specialty_description": "Gastroenterology, Urology",
        "regulation_number": "876.5010",
        "device_class": "2"
      }
    }
  ]
}