  "results": [
    {
      "cfres_id": "73274",
      "product_res_number": "Z-0004-2009",
      "event_date_initiated": "2008-09-19",
      "event_date_posted": "2008-11-21",
      "recall_status": "Terminated",
      "event_date_terminated": "2010-12-27",
      "res_event_number": "49413",
      "product_code": "NBW",
      "k_numbers": [
        "K062195"
      ],
      "product_description": "One Touch Ultra Test Strips, part number 020-994-02, manufactured by LifeScan Inc, Milpitas, CA, 25 count.\n\nIntended to be used for quantitative measurement of glucose in fresh capillary whole blood.  Intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control.  Intended for use on the finger, arm, or palm.",
      "code_info": "Lot number 2829235",
      "firm_fei_number": "2939301",
      "recalling_firm": "Lifescan Inc",
      "address_1": "1000 Gibraltar Dr",
      "city": "Milpitas",
      "state": "CA",
      "postal_code": "95035",
      "additional_info_contact": "408-942-5903",
      "reason_for_recall": "Products exceed inaccuracy threshold , with many complaints resulting from inaccurate low results.",
      "root_cause_description": "Process control",
      "action": "Urgent: Medical Device Recall letters, dated 09/19/2008, were sent to consumer end users, direct account distributors and wholesalers, health care providers, and pharmacists notifying them of the recall.  The letter states that if customers have strips from Lot 2829235, they should call LifeScan at 1-866-247-1281 and tell customer service that they have strips that need to be replaced.  Customer service and QA scripts were provided to LifeScan staff to provide to any callers who inquire about the recall.  The distributors and wholesalers were to locate and sequester any inventory from the affected Lot 2829235; request all recalled inventory from customers; call Carolina Logistics for an RGA; and when only the affected returned product is received, a credit will be issued.",
      "product_quantity": "35308 vials",
      "distribution_pattern": "Nationwide Distribution --- including Puerto Rico.",
      "openfda": {
        "k_number": [
          "K093930",
          "K143548",
          "K151265",
          "K120568",
          "K193475",
          "K053635",
          "K181273",
          "K122181",
          "K124040",
          "K072274",
          "K080283",
          "K080710",
          "K091898",
          "K070593",
          "K132821",
          "K193340",
          "K091547",
          "K102049",
          "K172723",
          "K182428",
          "K062377",
          "K162430",
          "K170143",
          "K182286",
          "K190564",
          "K130295",
          "K143169",
          "K240640",
          "K073118",
          "K071493",
          "K191038",
          "K171941",
          "K070239",
          "K133522",
          "K123090",
          "K123010",
          "K222126",
          "K070053",
          "K081715",
          "K090712",
          "K051147",
          "K150833",
          "K070017",
          "K150274",
          "K130886",
          "K153278",
          "K062538",
          "K040670",
          "K121087",
          "K083099",
          "K141867",
          "K110709",
          "K153330",
          "K231192",
          "K121133",
          "K200277",
          "K201880",
          "K223293",
          "K162382",
          "K042005",
          "K160179",
          "K090057",
          "K122307",
          "K111029",
          "K061118",
          "K072441",
          "K031843",
          "K090629",
          "K091168",
          "K160742",
          "K091157",
          "K162203",
          "K173638",
          "K123089",
          "K131013",
          "K110435",
          "K141036",
          "K190579",
          "K022724",
          "K043158",
          "K022581",
          "K231476",
          "K093764",
          "K011616",
          "K080014",
          "K020210",
          "K073699",
          "K160156",
          "K082020",
          "K123016",
          "K033845",
          "K063068",
          "K101457",
          "K130535",
          "K262116",
          "K101307",
          "K130696",
          "K191657",
          "K051514",
          "K122338",
          "K141862",
          "K120183",
          "K133790",
          "K081975",
          "K223722",
          "K241787",
          "K040814",
          "K093755",
          "K024234",
          "K032029",
          "K142785",
          "K120423",
          "K171822",
          "K130181",
          "K202739",
          "K093940",
          "K133647",
          "K101631",
          "K070524",
          "K231679",
          "K070941",
          "K243060",
          "K052469",
          "K051651",
          "K170079",
          "K051802",
          "K182057",
          "K082201",
          "K062187",
          "K070559",
          "K050985",
          "K060620",
          "K213131",
          "K240637",
          "K051936",
          "K131419",
          "K023256",
          "K141351",
          "K170264",
          "K093035",
          "K051839",
          "K122370",
          "K092887",
          "K132288",
          "K090495",
          "K150942",
          "K082683",
          "K122525",
          "K091815",
          "K082513",
          "K133346",
          "K043543",
          "K090405",
          "K161790",
          "K192859",
          "K052762",
          "K090609",
          "K073030",
          "K111890",
          "K181588",
          "K091765",
          "K021513",
          "K072039",
          "K062829",
          "K070585",
          "K170231",
          "K123559",
          "K152442",
          "K202534",
          "K072489",
          "K170267",
          "K173139",
          "K160944",
          "K083468",
          "K123136",
          "K093262",
          "K110074",
          "K051854",
          "K011240",
          "K180866",
          "K100398",
          "K233058",
          "K153561",
          "K012324",
          "K133045",
          "K132663",
          "K060793",
          "K010039",
          "K010553",
          "K180196",
          "K041107",
          "K150332",
          "K143467",
          "K133260",
          "K203562",
          "K120989",
          "K020208",
          "K163159",
          "K090389",
          "K092894",
          "K141656",
          "K212140",
          "K173511",
          "K020866",
          "K062195",
          "K080538",
          "K102751",
          "K150817",
          "K093712",
          "K170241",
          "K072369",
          "K131029",
          "K140210",
          "K072543",
          "K030703",
          "K113077",
          "K073231",
          "K161299",
          "K051213",
          "K103021",
          "K152493",
          "K151658",
          "K132406",
          "K081703",
          "K101371",
          "K160682",
          "K100322",
          "K010403",
          "K193406",
          "K210687",
          "K132966",
          "K091820",
          "K093789",
          "K181070",
          "K102678",
          "K113135",
          "K132862",
          "K090404",
          "K212248",
          "K102481",
          "K152534",
          "K190189",
          "K083628",
          "K073494",
          "K101597",
          "K241304",
          "K102816",
          "K111456",
          "K192220",
          "K123007",
          "K102063",
          "K062770",
          "K150531",
          "K153286",
          "K023279",
          "K120064",
          "K073137",
          "K152599",
          "K140068",
          "K110017",
          "K082618",
          "K153201",
          "K070472",
          "K061234",
          "K132086",
          "K012205",
          "K071507",
          "K110571",
          "K191286",
          "K143387",
          "K113098",
          "K093475",
          "K202885",
          "K012975",
          "K201396",
          "K014099",
          "K061328",
          "K083265",
          "K193407",
          "K040603",
          "K130265",
          "K052818",
          "K192175",
          "K093635",
          "K091481",
          "K112275",
          "K132511",
          "K141292",
          "K092099",
          "K090188",
          "K102383",
          "K112638",
          "K161867",
          "K130094",
          "K142035",
          "K072784",
          "K101926",
          "K133862",
          "K170614",
          "K171480",
          "K091296",
          "K072854",
          "K130284",
          "K062800",
          "K112963",
          "K121679",
          "K014167",
          "K160038",
          "K092638",
          "K083263",
          "K080501",
          "K230625",
          "K063771",
          "K063026",
          "K062235",
          "K173658",
          "K022171",
          "K161834",
          "K113614",
          "K182046",
          "K110737",
          "K220421",
          "K021819",
          "K103115",
          "K050500",
          "K102345",
          "K132929",
          "K181316",
          "K060470",
          "K070255",
          "K090653",
          "K100701",
          "K022545",
          "K140141",
          "K150214",
          "K062555",
          "K181131",
          "K170464",
          "K163087",
          "K150299",
          "K151611",
          "K011738",
          "K120708",
          "K071393",
          "K072413",
          "K140100",
          "K133741",
          "K032657",
          "K201037",
          "K160365",
          "K031260",
          "K201258",
          "K151595",
          "K090185",
          "K090332",
          "K111932",
          "K031501",
          "K042795",
          "K073119",
          "K072905",
          "K103044",
          "K073492",
          "K131363",
          "K023584",
          "K093745",
          "K091555",
          "K113192",
          "K131173",
          "K021827",
          "K081318",
          "K043311",
          "K141829",
          "K120351",
          "K140371",
          "K032985",
          "K130138",
          "K122142",
          "K103396",
          "K203711",
          "K090536",
          "K092602",
          "K173505",
          "K120649",
          "K182047",
          "K092902",
          "K021448",
          "K070641",
          "K072054",
          "K213887",
          "K023832",
          "K150052",
          "K081796",
          "K082521",
          "K141914",
          "K101509",
          "K060467",
          "K222234",
          "K171785",
          "K160430",
          "K112738",
          "K100937",
          "K183306",
          "K081957",
          "K132180",
          "K081726",
          "K063409",
          "K100437",
          "K080960",
          "K101543",
          "K123935",
          "K150396",
          "K100560",
          "K040678",
          "K083273",
          "K161738",
          "K052343",
          "K052108",
          "K080641",
          "K081451",
          "K062058",
          "K080853",
          "K162564",
          "K012210",
          "K182992",
          "K120558",
          "K081450",
          "K250085",
          "K062395",
          "K073573",
          "K130657",
          "K093819",
          "K173140",
          "K070088",
          "K082486",
          "K181527",
          "K053529",
          "K092875",
          "K123008",
          "K094005",
          "K132338",
          "K040848",
          "K132272",
          "K193590",
          "K131469",
          "K113343",
          "K081144",
          "K093592",
          "K103542",
          "K131230",
          "K133389",
          "K082169",
          "K201394",
          "K160282",
          "K042678",
          "K120866",
          "K112916",
          "K133509",
          "K100806",
          "K091229",
          "K022182",
          "K140325",
          "K123087",
          "K111509",
          "K040628",
          "K132633",
          "K122340",
          "K123517",
          "K113137",
          "K050224",
          "K103544",
          "K170064",
          "K052208",
          "K062567",
          "K080923",
          "K070850",
          "K032552",
          "K103329",
          "K092052",
          "K140150",
          "K062434",
          "K090628",
          "K093005",
          "K162336",
          "K023219",
          "K142928",
          "K152365",
          "K101635",
          "K120813",
          "K151164",
          "K053312",
          "K041750",
          "K024194",
          "K130621",
          "K012014",
          "K062347",
          "K132072",
          "K061181",
          "K013203",
          "K110948",
          "K213061",
          "K043340",
          "K151100",
          "K081389",
          "K090187",
          "K100732",
          "K053593",
          "K082328",
          "K063212",
          "K130244"
        ],
        "registration_number": [
          "3022346573",
          "3017500903",
          "3001058982",
          "3014353045",
          "3032948391",
          "2030633",
          "3021286079",
          "3015967131",
          "3022147580",
          "3042026207",
          "3016525500",
          "3011213605",
          "3005395953",
          "3020976354",
          "3006733253",
          "3014585508",
          "3033394445",
          "3015058854",
          "9610126",
          "3017520431",
          "3013223546",
          "3020885868",
          "3006532271",
          "9710595",
          "3003674698",
          "3043513123",
          "3004145393",
          "3003917514",
          "3015347724",
          "3008810357",
          "3017086896",
          "3045492872",
          "1000115322",
          "3013896610",
          "3007384659",
          "3011121740",
          "3005621222",
          "3005798905",
          "3022418645",
          "3004193489",
          "3003741676",
          "3011061417",
          "3006371704",
          "3034603279",
          "1810909",
          "3006755541",
          "3024379416",
          "3013957847",
          "3001452053",
          "3015525148",
          "3012104670",
          "3012014399",
          "3014302067",
          "3007031103",
          "3043013796",
          "3030978471",
          "3011065399",
          "3008959428",
          "3009204541",
          "3004130086",
          "3015997711",
          "3008780388",
          "3011536730",
          "9710492",
          "3038419886",
          "3013472743",
          "3015489159",
          "1000113657",
          "3009129531",
          "1649661",
          "3013160357",
          "3012188053",
          "3002736372",
          "9616936",
          "3033530347",
          "3011420269",
          "3003600763",
          "3007594734",
          "1220477",
          "3017550662",
          "3015188281",
          "1616487",
          "3005627859",
          "3042364147",
          "3010131137",
          "3030311186",
          "3003132490",
          "9611444",
          "1832816",
          "3004183148",
          "3011393376",
          "3004176168",
          "3007421145",
          "3013298431",
          "3005706626",
          "3012725363",
          "3026260647",
          "3007363438",
          "3024512705",
          "3004146366",
          "3008606429",
          "3008824097",
          "3005393828",
          "3004101980",
          "3007499349",
          "3016863723",
          "1037998",
          "3012421607",
          "3005644014",
          "9615507",
          "3007710663",
          "3009901017",
          "3030979885",
          "3045752480",
          "3043025312",
          "3010830834",
          "3004167884",
          "3006537560",
          "3012118659",
          "3008514395",
          "3003366417",
          "3031789039",
          "3010852191",
          "3004637226",
          "3003902721",
          "3015109923",
          "3031259044",
          "3007382160",
          "3008382007",
          "3008700497",
          "3010564249",
          "3042301847",
          "1832216",
          "3009456923",
          "3005862821",
          "3020460367",
          "3017548799",
          "3007134734",
          "2954323",
          "3014353121",
          "3013729860",
          "3002807314",
          "3010622184",
          "3004009431",
          "1000134386",
          "3005360469",
          "3009171251",
          "3045951983",
          "1219029",
          "9612841",
          "1220649",
          "3008573045",
          "3018940143",
          "1181121",
          "3008789114",
          "3021611414",
          "3011191255",
          "3004539335",
          "1450743",
          "2531491",
          "1066444",
          "3011979858",
          "1450662",
          "2953359",
          "3010713600",
          "2029275",
          "9616530",
          "3030186688",
          "3012133212",
          "3003422726",
          "3004043187"
        ],
        "fei_number": [
          "3022346573",
          "3017500903",
          "3001058982",
          "3014353045",
          "3032948391",
          "3021286079",
          "3015967131",
          "3022147580",
          "3042026207",
          "3016525500",
          "3004631140",
          "3011213605",
          "3005395953",
          "3020976354",
          "3006733253",
          "3014585508",
          "3033394445",
          "3015058854",
          "3017520431",
          "3003516723",
          "3013223546",
          "3020885868",
          "3006532271",
          "3003674698",
          "3043513123",
          "3004145393",
          "3003917514",
          "3015347724",
          "3008810357",
          "3017086896",
          "3045492872",
          "1000115322",
          "1000306431",
          "3013896610",
          "3007384659",
          "3003500908",
          "3005621222",
          "3005798905",
          "3022418645",
          "3004193489",
          "3003741676",
          "3011061417",
          "3006371704",
          "3034603279",
          "3001451797",
          "3003116995",
          "1810909",
          "3006755541",
          "3002698769",
          "3024379416",
          "3013957847",
          "3001452053",
          "3012104670",
          "3012014399",
          "3014302067",
          "3007031103",
          "3003405786",
          "3043013796",
          "3030978471",
          "3011065399",
          "3008959428",
          "3009204541",
          "3004130086",
          "3015997711",
          "3008780388",
          "3011536730",
          "3038419886",
          "3013472743",
          "3015489159",
          "1000113657",
          "3009129531",
          "3013160357",
          "3012188053",
          "3033530347",
          "3003600763",
          "3011420269",
          "3007594734",
          "1220477",
          "3017550662",
          "3015188281",
          "1616487",
          "3002829675",
          "3005627859",
          "3042364147",
          "3010131137",
          "3030311186",
          "3003132490",
          "1832816",
          "3004183148",
          "3011393376",
          "3011774813",
          "3004176168",
          "3013298431",
          "3005706626",
          "3012725363",
          "3026260647",
          "3003400058",
          "3000215205",
          "3024512705",
          "3004146366",
          "3002807423",
          "3008606429",
          "3008824097",
          "3005393828",
          "3004101980",
          "3016863723",
          "1037998",
          "3005664575",
          "3012421607",
          "3005644014",
          "3009901017",
          "3030979885",
          "3006759904",
          "3045752480",
          "3043025312",
          "3005670738",
          "3004167884",
          "3006537560",
          "3012118659",
          "3008514395",
          "3003366417",
          "3031789039",
          "3010852191",
          "3004637226",
          "3003902721",
          "3015109923",
          "3002806559",
          "3002671736",
          "3031259044",
          "3004525092",
          "3008382007",
          "3008700497",
          "3007382160",
          "3010564249",
          "3042301847",
          "1832216",
          "3002803444",
          "3002806387",
          "3009456923",
          "3005862821",
          "3020460367",
          "3017548799",
          "3007134734",
          "3014353121",
          "3013729860",
          "3002807314",
          "3010622184",
          "3004009431",
          "1000134386",
          "3045951983",
          "1219029",
          "3010700165",
          "1220649",
          "3008573045",
          "3018940143",
          "1181121",
          "3008789114",
          "3021611414",
          "3011191255",
          "3004539335",
          "1450743",
          "3004173703",
          "3011979858",
          "1450662",
          "3010713600",
          "3030186688",
          "3000210620",
          "3012133212",
          "3003422726",
          "3004043187"
        ],
        "device_name": "System, Test, Blood Glucose, Over The Counter",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1345",
        "device_class": "2"
      }
    }
  ]
}