      "cfres_id": "72466",
      "product_res_number": "Z-0304-2009",
      "event_date_initiated": "2008-06-11",
      "event_date_posted": "2008-11-21",
      "recall_status": "Terminated",
      "event_date_terminated": "2009-11-10",
      "res_event_number": "48712",
      "product_code": "LZO",
      "k_numbers": [
        "K030724"
      ],
      "product_description": "Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 32 mm diameter, neck length +3.5 mm, sterile, Zimmer, Warsaw, IN; REF 00-6418-032-03 and 00-6428-032-03.\n\nThe Alumina Ceramic Femoral Heads are modular components used in total hip arthroplasty and indicated for patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur; patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis; patients suffering from disability due to previous fusion; patients with previously failed endoprostheses and/or total hip components in the operative extremity; and patients with acute neck fractures.",
      "code_info": "All lots.",
      "firm_fei_number": "1000220733",
      "recalling_firm": "Zimmer Inc.",
      "address_1": "345 E Main St",
      "city": "Warsaw",
      "state": "IN",
      "postal_code": "46580-2746",
      "additional_info_contact": "574-267-6131",
      "reason_for_recall": "When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.",
      "root_cause_description": "Device Design",
      "action": "The firm issued Product Compatibility Update # 13, dated 6/11/08, to consignees informing them which products are not compatible, and thus should not be used in combination.",
      "product_quantity": "49,474 of all products",
      "distribution_pattern": "Worldwide Distribution --- including USA and countries of Austria, Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Dubai, England, Finland, France, Germany, Greece, Hungary, India, Israel, Italy, Korea, Lebanon, Mexico, Netherlands, New Zealand, Poland, Romania, Singapore, Slovakia, Spain, Switzerland and Ukraine.",
      "openfda": {
        "k_number": [
          "K131884",
          "K223828",
          "K944601",
          "K122305",
          "K930788",
          "K170473",
          "K964395",
          "K061066",
          "K170089",
          "K181693",
          "K931617",
          "K131647",
          "K192827",
          "K010560",
          "K192126",
          "K920219",
          "K051858",
          "K250444",
          "K183532",
          "K112463",
          "K972637",
          "K933499",
          "K992144",
          "K261009",
          "K952418",
          "K213770",
          "K022985",
          "K122911",
          "K934457",
          "K892644",
          "K896669",
          "K173267",
          "K170339",
          "K131518",
          "K993043",
          "K933108",
          "K973681",
          "K930712",
          "K123392",
          "K203175",
          "K121439",
          "K083392",
          "K013935",
          "K201673",
          "K070554",
          "K953169",
          "K073337",
          "K093549",
          "K211548",
          "K211891",
          "K971752",
          "K163441",
          "K931189",
          "K042249",
          "K150576",
          "K961589",
          "K073567",
          "K041443",
          "K192416",
          "K190704",
          "K964249",
          "K150302",
          "K935913",
          "K040644",
          "K953690",
          "K032348",
          "K201047",
          "K051737",
          "K020713",
          "K922247",
          "K971426",
          "K043356",
          "K921181",
          "K061312",
          "K202730",
          "K161192",
          "K933942",
          "K183136",
          "K210263",
          "K203420",
          "K000432",
          "K935921",
          "K933567",
          "K053246",
          "K173898",
          "K223103",
          "K983404",
          "K992359",
          "K151739",
          "K192189",
          "K970846",
          "K945091",
          "K904028",
          "K152903",
          "K011836",
          "K250834",
          "K130128",
          "K920968",
          "K042992",
          "K240381",
          "K972361",
          "K912295",
          "K925682",
          "K140915",
          "K970200",
          "K121221",
          "K160895",
          "K242249",
          "K023901",
          "K122185",
          "K945090",
          "K132455",
          "K181744",
          "K212327",
          "K953925",
          "K945094",
          "K133381",
          "K140676",
          "K190890",
          "K980239",
          "K990261",
          "K130728",
          "K163577",
          "K002441",
          "K972228",
          "K123705",
          "K061564",
          "K173776",
          "K051738",
          "K221149",
          "K912641",
          "K112115",
          "K013106",
          "K923734",
          "K183141",
          "K042035",
          "K920577",
          "K203041",
          "K131458",
          "K955350",
          "K062997",
          "K193433",
          "K132435",
          "K143407",
          "K241767",
          "K152893",
          "K212069",
          "K251906",
          "K143096",
          "K052718",
          "K211567",
          "K932222",
          "K936256",
          "K141235",
          "K233476",
          "K923583",
          "K961727",
          "K032380",
          "K051411",
          "K171181",
          "K913420",
          "K213675",
          "K192024",
          "K941847",
          "K082525",
          "K150133",
          "K143509",
          "K935193",
          "K902712",
          "K082398",
          "K161840",
          "K090299",
          "K253171",
          "K161226",
          "K894904",
          "K050398",
          "K063279",
          "K191816",
          "K896850",
          "K890472",
          "K050556",
          "K980774",
          "K913633",
          "K151307",
          "K990183",
          "K220483",
          "K161830",
          "K981847",
          "K251292",
          "K142187",
          "K051741",
          "K945095",
          "K240452",
          "K931571",
          "K003413",
          "K121665",
          "K921224",
          "K141327",
          "K890060",
          "K931929",
          "K944209",
          "K051682",
          "K915310",
          "K082996",
          "K234025",
          "K070479",
          "K173380",
          "K202768",
          "K231925",
          "K201657",
          "K971721",
          "K052275",
          "K101451",
          "K170845",
          "K970490",
          "K993889",
          "K152919",
          "K022120",
          "K140975",
          "K973307",
          "K960588",
          "K042842",
          "K123924",
          "K181598",
          "K111046",
          "K123619",
          "K915641",
          "K063616",
          "K200007",
          "K093944",
          "K111193",
          "K885102",
          "K081973",
          "K011533",
          "K062684",
          "K060107",
          "K190344",
          "K192883",
          "K885069",
          "K040803",
          "K200112",
          "K891083",
          "K042091",
          "K082844",
          "K012091",
          "K905260",
          "K945089",
          "K973837",
          "K860275",
          "K932519",
          "K955877",
          "K141821",
          "K051588",
          "K151531",
          "K921111",
          "K953689",
          "K053587",
          "K964150",
          "K101472",
          "K902073",
          "K915396",
          "K183684",
          "K980473",
          "K243927",
          "K111936",
          "K180929",
          "K170359",
          "K260182",
          "K173794",
          "K162942",
          "K063816",
          "K200959",
          "K082991",
          "K191145",
          "K071082",
          "K943586",
          "K971414",
          "K192559",
          "K033944",
          "K070928",
          "K231109",
          "K921225",
          "K181061",
          "K991708",
          "K003923",
          "K010839",
          "K240528",
          "K180263",
          "K111572",
          "K213839",
          "K071830",
          "K211657",
          "K182705",
          "K241241",
          "K022555",
          "K072852",
          "K920145",
          "K260831",
          "K191632",
          "K944689",
          "K102080",
          "K071535",
          "K112682",
          "K243444",
          "K192352",
          "K000392",
          "K925345",
          "K953792",
          "K130634",
          "K181254",
          "K093846",
          "K183312",
          "K955386",
          "K972893",
          "K122296",
          "K894694",
          "K915184",
          "K102487",
          "K201471",
          "K181532",
          "K100008",
          "K905687",
          "K943774",
          "K030151",
          "K121392",
          "K120211",
          "K030373",
          "K991226",
          "K131952",
          "K991162",
          "K964357",
          "K161635",
          "K945093",
          "K123688",
          "K171156",
          "K962248",
          "K110101",
          "K243809",
          "K994184",
          "K952686",
          "K903976",
          "K243043",
          "K200196",
          "K964704",
          "K141653",
          "K953045",
          "K041940",
          "K914676",
          "K914237",
          "K111981",
          "K003379",
          "K172235",
          "K181171",
          "K120362",
          "K183365",
          "K083762",
          "K962101",
          "K052787",
          "K062748",
          "K992235",
          "K072656",
          "K991952",
          "K932935",
          "K013989",
          "K993874",
          "K240834",
          "K980081",
          "K970675",
          "K151902",
          "K182462",
          "K965002",
          "K123434",
          "K160289",
          "K172467",
          "K946026",
          "K182321",
          "K232371",
          "K941942",
          "K143739",
          "K001449",
          "K915222",
          "K203167",
          "K110087",
          "K243029",
          "K981201",
          "K971409",
          "K250450",
          "K920593",
          "K173960",
          "K170690",
          "K891082",
          "K210678",
          "K964431",
          "K884952",
          "K213129",
          "K900628",
          "K241483",
          "K070885",
          "K220405",
          "K971435",
          "K030724",
          "K221323",
          "K153381",
          "K193050",
          "K955117",
          "K101296",
          "K910548",
          "K033664",
          "K191374",
          "K992403",
          "K920290",
          "K103189",
          "K973087",
          "K150066",
          "K113556",
          "K193030",
          "K140043",
          "K935185",
          "K233724",
          "K082371",
          "K022403",
          "K083558",
          "K915309",
          "K240391",
          "K971497",
          "K953111",
          "K022549",
          "K021673",
          "K926395",
          "K952088",
          "K051223",
          "K191781",
          "K213580",
          "K152530",
          "K223745",
          "K123782",
          "K233233",
          "K191399",
          "K930963",
          "K192946",
          "K141377",
          "K921836",
          "K893685",
          "K002320",
          "K190535",
          "K073570",
          "K003391",
          "K041170",
          "K163585",
          "K945092",
          "K970166",
          "K122641",
          "K083495",
          "K181761",
          "K221462",
          "K103352",
          "K142675",
          "K935452",
          "K052542",
          "K090845",
          "K131986",
          "K945516",
          "K954768",
          "K972690",
          "K060918",
          "K896204",
          "K031803",
          "K914574",
          "K182048",
          "K100412",
          "K203532",
          "K100866",
          "K103012",
          "K861400",
          "K925695",
          "K031217",
          "K130984",
          "K103120",
          "K953688",
          "K902074",
          "K844906",
          "K130376",
          "K093736",
          "K003274",
          "K193545",
          "K250571",
          "K081860",
          "K003316",
          "K131684",
          "K192660",
          "K190834",
          "K192683",
          "K895399",
          "K093991",
          "K012918",
          "K191831",
          "K970346",
          "K933275",
          "K942330",
          "K933203",
          "K222296",
          "K120531",
          "K201951",
          "K130899",
          "K961921",
          "K073102",
          "K912985",
          "K931572",
          "K103233",
          "K130343",
          "K130167",
          "K243466",
          "K973261",
          "K052237",
          "K923808",
          "K083312",
          "K252401",
          "K060577",
          "K905739",
          "K040734",
          "K910547",
          "K920756",
          "K260037",
          "K121011",
          "K181491",
          "K934694",
          "K150062",
          "K945096",
          "K925512",
          "K243021",
          "K140674",
          "K981302"
        ],
        "registration_number": [
          "",
          "1020279",
          "3003791208",
          "3002806535",
          "3005551626",
          "3012392319",
          "2134285",
          "3006395932",
          "3010041693",
          "1000577893",
          "3002807315",
          "3012119238",
          "3011189525",
          "1531174",
          "3002807112",
          "1822565",
          "3030412764",
          "3010097171",
          "9610612",
          "1828288",
          "3009973505",
          "1218882",
          "3013055499",
          "1423662",
          "1047843",
          "3014833750",
          "9613350",
          "3007120646",
          "1644408",
          "3010314800",
          "3002498892",
          "3012425419",
          "3004153240",
          "3015542154",
          "2249697",
          "3003639920",
          "1828464",
          "3006946276",
          "3010375065",
          "1833920",
          "3010178296",
          "3009959868",
          "3010386387",
          "1825034",
          "2133928",
          "3011461101",
          "3009882462",
          "3010065126",
          "3012916784",
          "3009760038",
          "1835572",
          "3002807314",
          "8043792",
          "1835251",
          "3012329926",
          "3009475821",
          "3007923096",
          "3007913880",
          "9681642",
          "3000264985",
          "3014315560",
          "3010287737",
          "3021386586",
          "3002807408",
          "3017397663",
          "3011274144",
          "1526711",
          "3005718816",
          "1643264",
          "1827096",
          "3009888740",
          "3010155661",
          "3010206549",
          "1000547811",
          "2648920",
          "9610726",
          "3008421730",
          "1720929",
          "3030338875",
          "3008824097",
          "3004142400",
          "3005360851",
          "3005562917",
          "3018052045",
          "3036795922",
          "3007051929",
          "1043653",
          "9681465",
          "3007200272",
          "3000236920",
          "3015259876",
          "3007334927",
          "3009813841",
          "3014170143",
          "3008544874",
          "3008812251",
          "3010047454",
          "1645151",
          "3014272734",
          "3036773273",
          "3008556682",
          "3012523063",
          "3012439458",
          "3008572101",
          "1649518",
          "1064858",
          "3008868758",
          "1219655",
          "1833986",
          "3008102042",
          "3013413393",
          "3004638532",
          "3004748528",
          "3003595343",
          "3014302784",
          "1220477",
          "3005061536",
          "1018470",
          "3009871135",
          "3009241418",
          "3010468876",
          "3010576282",
          "3006946279",
          "3011277306",
          "1061927",
          "3005333358",
          "1450662",
          "3012725451",
          "3004513872",
          "2032521",
          "3008262872",
          "3009998573",
          "3042508254",
          "3003387384",
          "1835831",
          "3010009693",
          "3026771806",
          "3020155054",
          "3006128100",
          "3009973336",
          "3012130008",
          "3007311878",
          "9617840",
          "1038671",
          "3009546990",
          "3002808022",
          "3008793310",
          "3011015572",
          "3006809628",
          "1526534",
          "8010379",
          "1226544",
          "3009756327",
          "3013014058",
          "3010531069",
          "3008021110",
          "3008281151",
          "3016438694",
          "3023852420",
          "3038503932",
          "3010363503",
          "3003998208",
          "3012319330",
          "3026776",
          "3000258338",
          "3008668801",
          "3004188066",
          "3011295718",
          "1834331",
          "3002907620",
          "1835444",
          "3004641308",
          "3011683674",
          "9616944",
          "3020967480",
          "3013302242",
          "3008395353",
          "3006801265",
          "1057425",
          "3002806470",
          "1721676",
          "9614209",
          "3003541440",
          "3007391088",
          "3011170410",
          "1424263",
          "3004371426",
          "3006536454",
          "1834379",
          "3009732568",
          "3013194153",
          "3027484613",
          "9616671",
          "1530390",
          "3002807295",
          "1221763",
          "3007353762",
          "3014644608",
          "3033509898",
          "3016778562",
          "3005180920",
          "3003761012",
          "3021008900",
          "3022142628",
          "3005323511",
          "3007366790",
          "2245304",
          "1836357",
          "3014273644",
          "3005913274",
          "1818910",
          "1529009",
          "3010120135",
          "3019807891",
          "9616680",
          "2249615",
          "3010057495",
          "3012275107",
          "3006356043",
          "3010163695",
          "3015167917",
          "3009173317",
          "1643817",
          "1000200989",
          "3010846112",
          "1221053",
          "9613369",
          "9615800",
          "3010536692",
          "3006001176",
          "3010039400",
          "3004153896",
          "3010220595",
          "3004635447",
          "3001284227",
          "3010132111",
          "9615674",
          "2246552",
          "3007740680",
          "3020282716",
          "3014252644",
          "1000294921",
          "3008110533",
          "3015398319",
          "3036756245",
          "3005751028",
          "2244478",
          "3003477135",
          "2183744",
          "3000196378",
          "3025603301",
          "3010288357",
          "1226188",
          "3010173425",
          "3002807090",
          "3013176080",
          "3016438761",
          "3034676720",
          "3035366890",
          "2242737",
          "3002806603",
          "3009106092",
          "3014763043",
          "8044172",
          "2030624",
          "3001278222",
          "3043543260",
          "3006017180",
          "1722511",
          "2528981",
          "3002807310",
          "3015207155",
          "3015516266",
          "3004493876",
          "3013791180",
          "3013450937",
          "1833506"
        ],
        "fei_number": [
          "",
          "1020279",
          "3003128516",
          "3003791208",
          "3002806535",
          "1000118610",
          "3005551626",
          "3012392319",
          "3006395932",
          "3010041693",
          "3001548076",
          "3005233524",
          "1000577893",
          "3002807315",
          "3012119238",
          "3011189525",
          "3000717902",
          "3002807112",
          "3001452082",
          "3030412764",
          "3010097171",
          "1828288",
          "3009973505",
          "1000221916",
          "1000135253",
          "1218882",
          "3002808293",
          "3013055499",
          "1047843",
          "3014833750",
          "3007120646",
          "3010314800",
          "3002498892",
          "3007932279",
          "3005601732",
          "1000199976",
          "3004153240",
          "3015542154",
          "3006946276",
          "3003639920",
          "1828464",
          "3002806315",
          "3010375065",
          "3009959868",
          "3010386387",
          "1825034",
          "3011461101",
          "3009882462",
          "3010065126",
          "3012916784",
          "3000206435",
          "3009760038",
          "3002854251",
          "1000149660",
          "3003064967",
          "3002807314",
          "3012329926",
          "3009475821",
          "3007923096",
          "3007913880",
          "3002807036",
          "3000264985",
          "3014315560",
          "3010287737",
          "1000143502",
          "3021386586",
          "3002807408",
          "3017397663",
          "1000306340",
          "3011274144",
          "1526711",
          "3000205591",
          "3005718816",
          "3002807172",
          "1000395711",
          "1643264",
          "3007282753",
          "1827096",
          "3009888740",
          "3010155661",
          "3010206549",
          "3001236277",
          "1000547811",
          "3008421730",
          "1720929",
          "3030338875",
          "3008824097",
          "3004142400",
          "3010566635",
          "3005360851",
          "3005562917",
          "3003110073",
          "3018052045",
          "3036795922",
          "3007051929",
          "1000220733",
          "3008397795",
          "1043653",
          "3007200272",
          "3015259876",
          "3007334927",
          "1000136269",
          "3009813841",
          "3014170143",
          "1000115656",
          "3008544874",
          "3010047454",
          "3014272734",
          "3036773273",
          "3008556682",
          "3012523063",
          "3012439458",
          "3008572101",
          "3008868758",
          "1219655",
          "3013413393",
          "3000215103",
          "3004748528",
          "1000306303",
          "3003595343",
          "3014302784",
          "1220477",
          "3005061536",
          "1018470",
          "3009871135",
          "3009241418",
          "1450662",
          "3010576282",
          "3006946279",
          "3011277306",
          "3005333358",
          "3012725451",
          "3003414678",
          "3004513872",
          "3008262872",
          "3001049282",
          "3009998573",
          "3042508254",
          "3003387384",
          "3010009693",
          "3026771806",
          "3020155054",
          "3006128100",
          "3009973336",
          "3012130008",
          "3007311878",
          "1038671",
          "3009546990",
          "3002808022",
          "3008793310",
          "3011015572",
          "3002808269",
          "1000116912",
          "3006809628",
          "3003574398",
          "1526534",
          "3002993297",
          "3003714383",
          "3013014058",
          "3010531069",
          "3008021110",
          "3008281151",
          "3016438694",
          "3023852420",
          "3038503932",
          "3010363503",
          "3003998208",
          "3012319330",
          "3000258338",
          "3008668801",
          "3004188066",
          "3011295718",
          "1832656",
          "3002907620",
          "3004641308",
          "3001814066",
          "3011683674",
          "3003386935",
          "3020967480",
          "3013302242",
          "3008395353",
          "3006801265",
          "3002287318",
          "3002806470",
          "1721676",
          "3003541440",
          "3007391088",
          "3008342029",
          "3011170410",
          "3006003617",
          "3000215273",
          "3006536454",
          "3009732568",
          "3013194153",
          "3027484613",
          "3002807295",
          "1000132037",
          "1221763",
          "3007353762",
          "3014644608",
          "3033509898",
          "3016778562",
          "3005180920",
          "3003761012",
          "3021008900",
          "3000203000",
          "3005323511",
          "3022142628",
          "2245304",
          "3014273644",
          "3002806653",
          "3005913274",
          "1818910",
          "1000086912",
          "3010120135",
          "3002807893",
          "3019807891",
          "1000139987",
          "3010057495",
          "3012275107",
          "3006356043",
          "3010163695",
          "3015167917",
          "3002924544",
          "3009173317",
          "3003051987",
          "1000200989",
          "1643817",
          "3010846112",
          "1221053",
          "3010536692",
          "3006001176",
          "3010039400",
          "3004153896",
          "3010220595",
          "3004635447",
          "3001284227",
          "3010132111",
          "2246552",
          "3020282716",
          "3007740680",
          "3014252644",
          "1000294921",
          "3008110533",
          "3002807479",
          "3036756245",
          "3005751028",
          "3003477135",
          "3003015552",
          "2183744",
          "3000196378",
          "3025603301",
          "3010288357",
          "3002807113",
          "3013176080",
          "3016438761",
          "3034676720",
          "3035366890",
          "3002807038",
          "3003070421",
          "3002806603",
          "3009106092",
          "3002806725",
          "3014763043",
          "3001278222",
          "3043543260",
          "3002459190",
          "3006017180",
          "2528981",
          "3002807310",
          "3015207155",
          "3015516266",
          "3004493876",
          "3000717687",
          "3013791180",
          "3013450937",
          "3004638532"
        ],
        "device_name": "Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3353",
        "device_class": "2"
      }
    }
  ]
}