      "cfres_id": "71403",
      "product_res_number": "Z-0097-2009",
      "event_date_initiated": "2007-11-06",
      "event_date_posted": "2008-10-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2010-12-18",
      "res_event_number": "48432",
      "k_numbers": [
        "K060226"
      ],
      "product_description": "PRIMEVIEW 3i System 2.0 equipped with a Coherence Therapist R2.0 or R2.1, and PRIMEVIEW 3iR2.0 or R2.1, part number 8139847\n\nMedical charged-particle radiation therapy system.",
      "code_info": "Serial numbers associated with RTT2.0 and RTT2.1",
      "firm_fei_number": "2937457",
      "recalling_firm": "Siemens Medical Solutions USA,  Inc",
      "address_1": "4040 Nelson Ave",
      "city": "Concord",
      "state": "CA",
      "postal_code": "94520-1200",
      "additional_info_contact": "925-246-8407",
      "reason_for_recall": "Flat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.",
      "root_cause_description": "Software design",
      "action": "Service calibration inspection/correction Update instructions TH038/07/S was issued on 11/06/2007 to all customers.  The instructions identified the affected products, stated the reason for the update, and listed the steps to update the software.  Customer notification was also issued.  Software patch planned for release in June 2008.",
      "product_quantity": "50 units",
      "distribution_pattern": "Worldwide Distribution",