      "cfres_id": "70969",
      "product_res_number": "Z-0515-2009",
      "event_date_initiated": "2007-06-01",
      "event_date_posted": "2009-01-17",
      "recall_status": "Terminated",
      "event_date_terminated": "2009-02-18",
      "res_event_number": "48190",
      "product_code": "JJY",
      "k_numbers": [
        "K970862"
      ],
      "product_description": "Quantimetrix SpinalScopics Spinal Fluid Cell Count Control.\n\nSpinalscopics is a plastic bottle liquid spinal fluid cell count control.  This product is intended for monitoring cell counts in patient cerebrospinal fluid samples performed manually using a hemocytometer.",
      "code_info": "Lot Numbers: 46130, 46131, 46132, 46140, 46141, 46142, 46150, 46151, 46152, 46160, 46161, 46162, 46170, 46171 and 46172.",
      "firm_fei_number": "2020715",
      "recalling_firm": "Quantimetrix Corporation",
      "address_1": "2005 Manhattan Beach Blvd",
      "city": "Redondo Beach",
      "state": "CA",
      "postal_code": "90278-1205",
      "additional_info_contact": "310-536-0006",
      "reason_for_recall": "This recall was initiated due to efficacy concerns with the stabilized erythrocytes, leukocytes and lymphocytes contained in this product group.",
      "root_cause_description": "Employee error",
      "action": "The recall communication was initiated when customer both direct and distributors who have purchased the lots were mailed a recall notification letter dated June 8, 2008 with instructions which was sent via first class mail.\n\nCustomers were requested to assist in the voluntary recall of the listed product and if any of the affected lots have been consumed please review their records to ensure that the obtained values conform to the stated specifications found in the lot specific product insert.",
      "product_quantity": "24,937 sets for all products.",
      "distribution_pattern": "Nationwide Distribution including states of AK, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NH, NM, NY, OR, PA, SC, SD, TN, TX, VA, WA and WY.",
      "openfda": {
        "k_number": [
          "K944452",
          "K993014",
          "K760818",
          "K990161",
          "K030482",
          "K051185",
          "K062668",
          "K122838",
          "K023250",
          "K833639",
          "K082067",
          "K971657",
          "K972235",
          "K960862",
          "K882728",
          "K833423",
          "K931373",
          "K102157",
          "K023669",
          "K945315",
          "K961359",
          "K102700",
          "K041361",
          "K071433",
          "K923663",
          "K935219",
          "K950326",
          "K134013",
          "K800558",
          "K972608",
          "K983166",
          "K955812",
          "K972904",
          "K090541",
          "K111063",
          "K101809",
          "K901212",
          "K874459",
          "K790264",
          "K082107",
          "K792024",
          "K863648",
          "K872777",
          "K904781",
          "K903933",
          "K023352",
          "K992349",
          "K022401",
          "K974384",
          "K984372",
          "K792375",
          "K941698",
          "K973629",
          "K812650",
          "K944728",
          "K882486",
          "K072246",
          "K821026",
          "K780351",
          "K092051",
          "K970979",
          "K872303",
          "K981532",
          "K070741",
          "K131993",
          "K813295",
          "K092320",
          "K890414",
          "K920940",
          "K041975",
          "K960824",
          "K912441",
          "K896246",
          "K992900",
          "K874810",
          "K931139",
          "K973469",
          "K982224",
          "K030942",
          "K972214",
          "K903407",
          "K110534",
          "K852594",
          "K933437",
          "K944486",
          "K874230",
          "K954323",
          "K040880",
          "K121794",
          "K834064",
          "K030088",
          "K881229",
          "K920636",
          "K012513",
          "K941876",
          "K800813",
          "K983492",
          "K990610",
          "K861448",
          "K801335",
          "K790371",
          "K822133",
          "K914255",
          "K060206",
          "K871058",
          "K792025",
          "K861567",
          "K991704",
          "K910440",
          "K965114",
          "K001632",
          "K770310",
          "K050125",
          "K813167",
          "K932076",
          "K881437",
          "K062379",
          "K771437",
          "K813294",
          "K883012",
          "K803098",
          "K013265",
          "K911984",
          "K791695",
          "K893580",
          "K861901",
          "K932458",
          "K992770",
          "K130761",
          "K021930",
          "K981474",
          "K920291",
          "K781893",
          "K800559",
          "K955690",
          "K895862",
          "K911169",
          "K972307",
          "K973790",
          "K123198",
          "K022867",
          "K992034",
          "K000875",
          "K972070",
          "K970147",
          "K990391",
          "K811466",
          "K920826",
          "K042625",
          "K051928",
          "K790330",
          "K821242",
          "K011731",
          "K121603",
          "K931366",
          "K955858",
          "K915191",
          "K811465",
          "K831263",
          "K002852",
          "K023731",
          "K843986",
          "K950469",
          "K020237",
          "K041227",
          "K854931",
          "K110904",
          "K974827",
          "K972048",
          "K872227",
          "K980514",
          "K012119",
          "K801538",
          "K945651",
          "K936184",
          "K923165",
          "K010111",
          "K951749",
          "K894852",
          "K022991",
          "K843611",
          "K810952",
          "K981401",
          "K861202",
          "K922046",
          "K140916",
          "K914619",
          "K112705",
          "K991182",
          "K083891",
          "K904099",
          "K012120",
          "K870683",
          "K950802",
          "K162530",
          "K042335",
          "K991447",
          "K883861",
          "K012855",
          "K883942",
          "K861185",
          "K984419",
          "K933844",
          "K903225",
          "K093492",
          "K010358",
          "K071980",
          "K032335",
          "K073080",
          "K903190",
          "K821982",
          "K852418",
          "K904829",
          "K935599",
          "K860942",
          "K800370",
          "K941328",
          "K944455",
          "K952877",
          "K943290",
          "K903310",
          "K923740",
          "K993546",
          "K861000",
          "K011494",
          "K873611",
          "K042318",
          "K964065",
          "K121468",
          "K123775",
          "K984477",
          "K022041",
          "K013332",
          "K855035",
          "K122291",
          "K861603",
          "K030452",
          "K033344",
          "K983513",
          "K896522",
          "K043108",
          "K020450",
          "K003716",
          "K070504",
          "K930534",
          "K813321",
          "K811100",
          "K790840",
          "K791696",
          "K970976",
          "K934357",
          "K102617",
          "K982473",
          "K911854",
          "K902616",
          "K962982",
          "K951170",
          "K143379",
          "K981455",
          "K971952",
          "K883124",
          "K030679",
          "K030705",
          "K142964",
          "K904784",
          "K983130",
          "K001758",
          "K993023",
          "K920926",
          "K902573",
          "K163629",
          "K901988",
          "K031364",
          "K983807",
          "K821560",
          "K130963",
          "K040822",
          "K040280",
          "K841452",
          "K940397",
          "K023661",
          "K970956",
          "K914890",
          "K973791",
          "K924915",
          "K883147",
          "K112742",
          "K881275",
          "K850017",
          "K962575",
          "K981915",
          "K000270",
          "K093032",
          "K990150",
          "K964640",
          "K821559",
          "K944525",
          "K143573",
          "K990016",
          "K863745",
          "K990888",
          "K040379",
          "K983241",
          "K022885",
          "K791313",
          "K012117",
          "K854967",
          "K891929",
          "K133676",
          "K891680",
          "K061182",
          "K011579",
          "K770312",
          "K823929",
          "K902673",
          "K103271",
          "K042389",
          "K861420",
          "K043208",
          "K813293",
          "K072892",
          "K112337",
          "K000810",
          "K946049",
          "K042815",
          "K080973",
          "K001373",
          "K061137",
          "K914958",
          "K980178",
          "K955252",
          "K812635",
          "K902740",
          "K943394",
          "K141073",
          "K070200",
          "K770380",
          "K933317",
          "K031575",
          "K894252",
          "K062593",
          "K924802",
          "K131595",
          "K970862",
          "K860998",
          "K854119",
          "K912725",
          "K020521",
          "K791958",
          "K984594",
          "K033937",
          "K974822",
          "K971405",
          "K933435",
          "K012431",
          "K962997",
          "K853943",
          "K111996",
          "K030566",
          "K890442",
          "K904768",
          "K944513",
          "K040535",
          "K030150",
          "K032453",
          "K993310",
          "K913224",
          "K934056",
          "K032237",
          "K050536",
          "K140393",
          "K862278",
          "K923795",
          "K872785",
          "K142978",
          "K820255",
          "K163015",
          "K132751",
          "K990531",
          "K872304",
          "K913133",
          "K863751",
          "K943754",
          "K952385",
          "K851942",
          "K860970",
          "K051517",
          "K882736",
          "K963346",
          "K103601",
          "K890576",
          "K040277",
          "K893142",
          "K860906",
          "K895905",
          "K062916",
          "K863849",
          "K013995",
          "K040459",
          "K111506",
          "K920350",
          "K902094",
          "K881113",
          "K030801",
          "K060429",
          "K140971",
          "K110880",
          "K091225",
          "K102016",
          "K002196",
          "K040352",
          "K844760",
          "K020817",
          "K103364",
          "K895943",
          "K780096",
          "K844125",
          "K901115",
          "K110738",
          "K924061",
          "K862565",
          "K030089",
          "K113720",
          "K060850",
          "K093384",
          "K954298",
          "K140764",
          "K936245",
          "K890708",
          "K082714",
          "K955260",
          "K880175",
          "K002460",
          "K902909",
          "K984074",
          "K860355",
          "K780552",
          "K002174",
          "K904968",
          "K833192",
          "K851820",
          "K844413",
          "K082036",
          "K031644",
          "K781593",
          "K100727",
          "K941912",
          "K840458",
          "K844848",
          "K821382",
          "K961941",
          "K880494",
          "K011867",
          "K123663",
          "K862185",
          "K981339",
          "K021381",
          "K941115",
          "K993631",
          "K992172",
          "K894078",
          "K923633",
          "K100718",
          "K992741",
          "K052056",
          "K961195",
          "K040925",
          "K070223",
          "K893141",
          "K131992",
          "K902536",
          "K982637",
          "K942458",
          "K925256",
          "K874280",
          "K912718",
          "K944059",
          "K972080",
          "K954247",
          "K802109",
          "K974613",
          "K923632",
          "K904293",
          "K904989",
          "K810906",
          "K862420",
          "K023867",
          "K871977",
          "K811695",
          "K082717",
          "K012243",
          "K823490",
          "K062674",
          "K110616",
          "K040091",
          "K153712",
          "K870684",
          "K050537",
          "K111964",
          "K830179",
          "K033162",
          "K011169",
          "K131479",
          "K823352",
          "K990424",
          "K881503",
          "K101863",
          "K042324",
          "K022591",
          "K854789",
          "K984325",
          "K844996",
          "K140522",
          "K972868",
          "K990282",
          "K973030",
          "K033379",
          "K874809",
          "K130337",
          "K791119",
          "K974452",
          "K791825",
          "K890470",
          "K973894",
          "K954655",
          "K022936",
          "K926046",
          "K862963",
          "K840457",
          "K013235",
          "K902235",
          "K882542",
          "K862586",
          "K022851",
          "K123725",
          "K841884",
          "K051543",
          "K811467",
          "K990149",
          "K881438",
          "K955024",
          "K833009",
          "K955691",
          "K942647",
          "K853906",
          "K911803",
          "K992710",
          "K896247",
          "K961912",
          "K821557",
          "K061196",
          "K791654",
          "K883247",
          "K944454",
          "K782038",
          "K950399",
          "K802962",
          "K993360",
          "K150300",
          "K031921",
          "K921286",
          "K904141",
          "K871342",
          "K023410",
          "K990394",
          "K955264",
          "K012118",
          "K813580",
          "K132091",
          "K974826",
          "K050504",
          "K971071",
          "K850575",
          "K041073",
          "K990017",
          "K881436",
          "K083554",
          "K833191",
          "K861068",
          "K892642",
          "K872376",
          "K032074",
          "K905389",
          "K992939",
          "K884977",
          "K925326",
          "K121759",
          "K071675",
          "K031577",
          "K971954",
          "K961293",
          "K101633",
          "K862179",
          "K032122",
          "K911755",
          "K090475",
          "K850574",
          "K031101",
          "K965048",
          "K101216",
          "K860999",
          "K812556",
          "K871537",
          "K920927",
          "K911171",
          "K990317",
          "K904075",
          "K041812",
          "K880447",
          "K862587",
          "K983507",
          "K032089",
          "K942295",
          "K903566",
          "K914204",
          "K962630",
          "K101226",
          "K971406",
          "K943729",
          "K921477",
          "K961750",
          "K903338",
          "K984225",
          "K902095",
          "K813322",
          "K003488",
          "K900717",
          "K870252",
          "K874772",
          "K965171",
          "K053247",
          "K963770",
          "K001249",
          "K844018",
          "K900993",
          "K072835",
          "K992550",
          "K124009",
          "K011853",
          "K853931",
          "K850202",
          "K801539",
          "K954656",
          "K930850",
          "K982290",
          "K802108",
          "K904872",
          "K842970",
          "K051687",
          "K132156",
          "K033387",
          "K050387",
          "K874731",
          "K133960",
          "K812583",
          "K822850",
          "K954074",
          "K852753",
          "K912455",
          "K012468",
          "K130762",
          "K811182",
          "K913687",
          "K011548",
          "K013734",
          "K973517",
          "K855036",
          "K132227",
          "K130162",
          "K000116",
          "K040273",
          "K924919",
          "K882727"
        ],
        "registration_number": [
          "3011325031",
          "2027969",
          "2432235",
          "1226774",
          "2182595",
          "3022977716",
          "9610806",
          "3002809144",
          "3007111389",
          "3006198300",
          "2915274",
          "3037000637",
          "1618982",
          "3011979858",
          "2085064",
          "1181121",
          "3009491259",
          "2517506",
          "2020715",
          "2020808",
          "3008723200",
          "8043792",
          "3000308930",
          "1319681",
          "2182501",
          "2050010",
          "9616673",
          "2032087",
          "3003795116",
          "2016706",
          "3003951180",
          "3024756665",
          "9610566",
          "1649661",
          "1319809",
          "2245285",
          "3007349751",
          "3005706626",
          "1319808",
          "3009984016",
          "2084025",
          "8023024",
          "3005935106",
          "3004205692",
          "1000403749",
          "3002637618",
          "1000256526",
          "3007490412",
          "3008156187",
          "3002806944",
          "9681753",
          "2245578",
          "3027273565",
          "3005146313",
          "1832216",
          "3021841051",
          "1220477",
          "3011096386",
          "3004102403",
          "3005333358",
          "8032314",
          "3012963943",
          "3004493545",
          "1220649",
          "8040374",
          "3008198807",
          "3004193489",
          "3013982035",
          "3003030793",
          "3007113174",
          "3006602209",
          "3010127294",
          "2026994",
          "1523456",
          "3014150341",
          "2020726",
          "2050012",
          "3012494290",
          "3012471076",
          "9613959",
          "3013660430",
          "3002642396",
          "3004973408",
          "3005643513",
          "1616487",
          "1222035",
          "9610126",
          "3007662974",
          "3010421384",
          "2250051",
          "9680745",
          "9613069",
          "3015525148",
          "3007088335",
          "3004130086",
          "1181055",
          "3003871639",
          "2022375",
          "2247288",
          "3009711478",
          "8030478",
          "1058931",
          "2435505",
          "3002800697",
          "3004152597",
          "9610529",
          "3001431095",
          "1319436",
          "9681518",
          "1219029",
          "1218959",
          "1220972",
          "2085085",
          "3002807968",
          "3010939897",
          "1827821",
          "3015171822",
          "2431530",
          "3014607865",
          "9680746",
          "3009092698",
          "3013513377",
          "1217183",
          "9610746",
          "9614373",
          "3000148879",
          "1717966",
          "8020316",
          "3000125561",
          "3039353646",
          "3032705"
        ],
        "fei_number": [
          "3011325031",
          "2432235",
          "1000120319",
          "2182595",
          "1000525241",
          "1000206808",
          "3003179344",
          "3022977716",
          "3002809144",
          "3007111389",
          "3006198300",
          "2915274",
          "3037000637",
          "3005774014",
          "3001576181",
          "1618982",
          "3011979858",
          "1000135102",
          "1181121",
          "1000119795",
          "3009491259",
          "1000142648",
          "2517506",
          "3002698769",
          "2020715",
          "2020808",
          "3008723200",
          "3000308930",
          "2182501",
          "2050010",
          "3003795116",
          "2016706",
          "3003951180",
          "3024756665",
          "2245285",
          "3007349751",
          "3005706626",
          "3009984016",
          "2084025",
          "3005935106",
          "6387",
          "3004205692",
          "1000403749",
          "3002637618",
          "1000256526",
          "3007490412",
          "3008156187",
          "3002806944",
          "5864",
          "2245578",
          "3027273565",
          "3005146313",
          "1832216",
          "3021841051",
          "1220477",
          "1000305840",
          "3011096386",
          "3004102403",
          "1000122044",
          "3005333358",
          "1000125596",
          "3012963943",
          "3004493545",
          "1220649",
          "3002806557",
          "3008198807",
          "3004193489",
          "3013982035",
          "3003030793",
          "3007113174",
          "3002808269",
          "3006602209",
          "3010127294",
          "2026994",
          "3002806559",
          "1523456",
          "3014150341",
          "2020726",
          "2050012",
          "3012494290",
          "3012471076",
          "3013660430",
          "3002642396",
          "3003355251",
          "3004973408",
          "3005643513",
          "3001451797",
          "3003045237",
          "3005695774",
          "1616487",
          "1222035",
          "3002701146",
          "3007662974",
          "3010421384",
          "2250051",
          "1220394",
          "3004130086",
          "1181055",
          "3003871639",
          "2022375",
          "3009711478",
          "3002808340",
          "3002808140",
          "3001644026",
          "3002800697",
          "3004152597",
          "3001431095",
          "3002808301",
          "3003407015",
          "1219029",
          "1218959",
          "1220972",
          "3002806809",
          "3002807968",
          "3000303338",
          "3010939897",
          "1827821",
          "3015171822",
          "2431530",
          "3014607865",
          "3002806560",
          "3009092698",
          "3013513377",
          "1000136573",
          "1000305841",
          "1217183",
          "3000148879",
          "1717966",
          "3000125561",
          "3039353646",
          "3002807876"
        ],
        "device_name": "Multi-Analyte Controls, All Kinds (Assayed)",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1660",
        "device_class": "1"
      }
    }
  ]
}